Connecting the steam sterilizer to a network
Manually configured
Automatically configured
network (DHCP)
Device
Steam sterilizer itself
Computer
Log printer
Label printer
Gateway
Subnetwork mask
Table 1: IP addresses saved in the steam sterilizer
Warning!
12
If the network is manually configured, then the subnetwork mask of the
network
medical practice network must be known. If a subnetwork mask is used other
than the one preset in the steam sterilizer, a software expert should adapt the
IP addresses in the steam sterilizer. If the masks conform and the above-
mentioned IP addresses are not assigned, then a log printer or label printer
can be connected directly to the steam sterilizer.
The network is automatically configured in a dynamic network.
In this case, in the Settings Logging menu set the IP address of
▪
the steam sterilizer on DHCP.
If logging takes place over a computer via FTP, then a fixed IP address must
be determined for the computer, which is then registered in the steam
sterilizer.
NOTICE!
If a DHCP server cannot be located in the practice network, the steam
sterilizer will automatically recieve the pre-set static IP address.
The IP addresses for the following devices are saved in the steam sterilizer
as parameters:
IP address
192.168.40.40
192.168.40.140
192.168.40.240
192.168.40.160
192.168.40.244
255.255.255.0
■
The input of false IP addresses can lead to disturbance of the
practice network.
Verifying and changing IP addresses
The IP addresses are changed in the Settings Logging menu.
▪
Follow the logging assistance until you reach the window in which the IP
addresses for the specific devices are listed.
▪
Select, for example, the steam sterilizer. It opens the next window.
▪
Select the numeric keypad which you wish to change directly by
choosing each respective number.
▪
Delete the digits with the C button, then enter a new number and
confirm with SAVE.
▪
Proceed in the same way with the other devices that are to be
connected to the network.