Download Print this page

Salter AIRE Elite 8353-1 Instruction Manual page 2

Advertisement

WARNING
To prevent possible risk of infection from
contaminated medication, cleaning of the nebulizer is
NOTE
The nebulizer kit is for single patient use only.
recommended after each treatment.
NOTE
Please follow national requirements to dispose the unit
properly.
6. Storage
Keep the unit and accessories dry. Avoid direct sunshine. See
specifications in Section 10 for appropriate environmental storage
conditions.
7. Maintenance
1.
This unit is oil-less. Do Not Lubricate.
7.1 General Information
2.
Risk of electric shock. Do not disassemble the main unit.
NOTE
Inspect the filter once every month and
7.2 Filter Change
replace as necessary or when filter turns gray.
Please follow the below instructions as right
figures.
1.
Open the filter cap.
2.
Inspect the filter and if dirty, remove filter
with a small, pointed object. Discard the
filter.
3.
Replace with a clean filter. Additional filters
should be purchased from your provider.
4.
Put the filter cap back.
CAUTION
Do not use cotton or any other material as a filter. Do not
wash or reuse the filter. Do not operate the unit without a filter.
Except for the above instructed user maintenance, all device
7.3 Service
servicing must be performed by a Salter Labs
authorized service
®
representative. There are no serviceable parts inside the unit.
Contact your Salter Labs dealer or authorized service center if your
unit needs repair.
WARNING
Do not tamper with or attempt to repair the device.
Refer servicing to qualified service personnel.
8. Expected Service Life
The products are intended to offer safe and reliable operation when
used or installed according to the instructions provided by Salter
Labs. Salter Labs recommends that the system be inspected and
serviced by authorized technicians if there are any signs of wear or
concerns with device function and indication on products.
Otherwise, service and inspection of the devices generally should
not be required.
9. Troubleshooting
If your Salter AIRE Elite Compressor fails to function, consult the
Troubleshooting guide below. If the problem persists, consult your
equipment provider.
‧Check if plug is properly fit into an appropriate
Problem
Cause and Solution
Device doesn't
electrical outlet.
operate.
‧When device has been run continuously for over
30 minutes right before using, an auto shut down
may activate by built-in thermal protector, cool
down device for 30 minutes before next usage.
‧Check for proper electrical voltage.
Weak
‧Check tubing for blockage or air leakage at
Nebulization
connection to Salter AIRE Elite Compressor or
nebulizer cup, replace as needed.
‧Check the nebulizer cup if it is properly assembled
and not damaged. If there is any damage, replace
as needed.
‧Check if filter is too dirty, replace as needed.
800987 Rev D (Side 2)
(All specifications are subject to change without notice.)
10. Specifications
120VAC,60Hz,1.2A
230VAC,50Hz,0.6A
Electrical Rating
( For 120V System)
( For 230V System)
(Note: Refer to the rating
≧30 psi
label on the product)
Maximum Compressor
≧5.5 lpm
≧5.0 lpm
Pressure
Class II.
Nebulizer Flow Rate
BF equipment.
IP2X
No AP/APG protection.
Classification
Mouthpiece or Nasal mask
14.6 × 20.3 × 9.5 cm /
Applied part
5.7" × 7.9" × 3.7"
Dimensions (W x D x H)
1.8 kg / 4.0 lb
Weight (approx.)
Fuse (non-user serviceable) F5AL 250V
T1.6AL 250V
5 Years
Warranty
Operation: 10℃ to 40℃/ 50℉ to 104℉
Storage: -15℃ to 50℃ / 5℉ to 122℉
Temperature
Transport: -15℃ to 70℃/ 5℉ to 158℉
Operation: 10% to 90%RH non-condensing
Environment
Storage: 10% to 90%RH non-condensing
Humidity
Transport: 10% to 90% RH non-condensing
Operation: 700-1060 hPa
Atmospheric
pressure
11. Accessories
8501-1-2
Salter AIRE Elite Replacement filters
Model
Description
8258-0-1
Compressor Carrying bag
8660
NebuTech® Nebulizer – Reusable
8960
NebuTech Nebulizer – Disposable
8967
NebuTech Nebulizer with Pediatric Mask
8984
NebuTech Nebulizer with Adult Mask
8900
Nebulizer with Tee Adapter
8906
Nebulizer with Pediatric Mask
8924
Nebulizer with Adult Mask
12. Appendix A: EMC Information
This device is intended for use in the electromagnetic environment
Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
specified below. The user of this device should make sure it is used
in such an environment.
Emissions
Compliance
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal
RF emissions
Group1
Test
function. Therefore, its RF emissions are very
CISPR 11
low and are not likely to cause any
interference in nearby electronic equipment
RF emissions
Class B
CISPR 11
Harmonic
emissions
The device is suitable for use in all
Class A
establishments, including domestic
IEC61000-3-2
establishments and those directly connected
Voltage
to the public low-voltage power supply
fluctuations /
network.
Flicker
Complies
emissions
IEC61000-3-3
This device is intended for use in the electromagnetic environment
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
specified below. The user of this device should make sure it is used
in such an environment.
IEC60601
Electromagnetic
Immunity Test
Compliance
Floors should be wood,
test level
Environment-Guidance
concrete or ceramic tile. If
Electrostatic
±6kV contact
±6kV contact
floors are covered with
Discharge (ESD)
±8kV air
±8kV air
synthetic material, the
IEC61000-4-2
relative humidity should be
at least 30%.
±2kV
for
±2kV for power
Mains power quality
Electrical fast
power supply
supply line
should be that of a typical
transient/ burst
line
±1kV
for
commercial or hospital
IEC61000-4-4
±1kV
for
input/out line
environment.
input/out line
± 1 kV line(s) to
± 1 kV line(s)
Mains power quality
Surge
line(s)
to
should be that of a typical
IEC61000-4-5
± 2 kV line(s) to
line(s)
commercial or hospital
earth
environment.
<5 % U
(>95
<5 % U
(>95
Mains power quality
% dip in U
T
)for
% dip in U
T
)
should be that of a typical
Voltage dips,
0,5 cycle
T
for 0,5 cycle
T
commercial or hospital
short
40 % U
(60 %
40 % U
(60
environment. If the user of
interruptions
dip in U
T
)
% dip in U
T
)
this device requires
and voltage
for 5 cycles
T
for 5 cycles
T
continued operation
variations on
70 % U
(30 %
70 % U
(30
during power mains
power supply
dip in U
T
)
% dip in U
T
)
interruptions, it is
input lines
for 25 cycles
for 25 cycles
recommended that the
T
T
IEC61000-4-11
<5 % U
(>95
<5 % U
(>95
device be powered from
% dip in U
T
) for
% dip in U
T
)
an uninterruptible power
5 sec
for 5 sec
supply or a battery.
T
T
Power frequency magnetic
Power
fields should be at levels
frequency
characteristic of a typical
(50/60Hz)
3 A/m
3 A/m
location in a typical
magnetic field
commercial or hospital
IEC61000-4-8
environment.
NOTE: U
is the AC mains voltage prior to the application of the test level
T
Recommended separation distances between portable and
This device is intended for use in an electromagnetic environment
mobile RF communications equipment and this device:
in which radiated RF disturbances are controlled. The customer or
the user of this device can help prevent electromagnetic
interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and this
device as recommended below, according to the maximum output
power of the communications equipment
Separation distance according to frequency of transmitter (m)
Rated
maximum
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
output
power
0.01
0.12
0.12
0.23
=
=
=
d
1
2 .
P
d
1
2 .
P
d
2
3 .
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
This device is intended for use in the electromagnetic environment
Guidance and Manufacturer's Declaration - Electromagnetic Immunity:
specified below. The user of this device should make sure it is used
in such an environment.
Immunity
IEC60601 test
Compliance
Electromagnetic
Portable and mobile RF
Test
level
Environment-Guidance
communications equipment
should be used no closer to any
part of this device, including
cables, than there commended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted
3Vrms150 kHz
3 Vrms
RF
to 80 MHz
d
outside ISM
IEC
d
bands
61000-4-6
a
Where P is the maximum output
d
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
3 V/m
the recommended separation
3 V/m 80
Radiated
distance in meters (m).
MHz to 2.5 GHz
RF
Field strengths from fixed RF
IEC
transmitters, as determined by an
61000-4-3
electromagnetic site survey
should be less than the compliance
level in each frequency range
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a/ The ISM (industrial, scientific and medical) bands between 150 kHz and
80 MHz are 6,765 MHz to 6,795 MHz;13,553 MHz to 13,567 MHz; 26,957
MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
b/ The compliance levels in the ISM frequency bands between 150 kHz and
80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to
decrease the likelihood that mobile/portable communications equipment
could cause interference if it is inadvertently brought into patient areas. For
this reason, an additional factor of 10/3 is used in calculating the
recommended separation distance for transmitters in these frequency
ranges.
c/ Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
P
measures may be necessary, such as reorienting or relocating the device.
d/ Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3 V/m.
100 W. Sycamore Rd.
Arvin, CA 93203 USA
Tel: 661-854-3166
800-421-0024 (in CA)
800-235-4203 (outside CA)
Fax: 661-854-3850
800-628-4690 (US & Canada)
Internet:
www.salterlabs.com
Authorized Representative in the E.U.:
MT Promedt Consulting GmbH
Altenhofstrasse 80
D-66386 St. Ingbert, Germany
NOTE
Consult distributor(s) or EU representative(s) for
additional technical documents.
150kHz to 80MHz
=
150kHz to 80MHz
1
2 .
P
=
80 MHz to 2.5G MHz
1
2 .
P
=
2
3 .
P
b
,
c
.
d
0086
666001-6720 V1.0

Advertisement

loading
Need help?

Need help?

Do you have a question about the AIRE Elite 8353-1 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Aire elite 8352-1Aire elite 8350-18350-1Aire elite8353-18352-1