Download Print this page

Troubleshooting - Terumo Elemano Quick Start User Manual

Advertisement

Troubleshooting

A
"LOW" INDICATION
"Low" indication displays when pressurization
is not enough. It is recommended to pressurize
30–40 mmHg higher than expected systolic.
If you see the LOW sign, keep pumping to
higher number. Make sure the tubes are
connected properly using the site window.
B
DIFFICULT PRESSURIZATION
Elemano may have a clogged air lter in
in ation bulb or there may be deterioration in
the bulb. In case of a clogged air lter, it
needs to be cleaned (refer to the Instructions
Manual, page E-18). If the in ation bulb has
deteriorated, replace it with a new bulb.
C
FLASHING PANEL
The Elemano display panel ashes to
warn abnormal conditions such as irregular
beats or weak pulse. If this happens, perform
the measurement in the Auscultation mode to
verify the condition.
©2008 Elemano is a trademark of Terumo Corporation. TERUMO is a registered trademark of Terumo Corporation.
D
Error indications
Errors
Symptoms
Err 1, Err 2,
Out of order
Err 5
Err 3
The cuff connector, the air tube
and/or the in ation bulb are
incorrectly connected to the
main body
The cuff is not applied properly
The air tube is not connected
The patient moved during
measurement
The main body was shaken or
vibrated during measurement
Err 4
Abnormal blood pressure/pulse
was detected
The cuff is not applied properly
The patient moved during
measurement
The main body was shaken or
vibrated during measurement
Please refer to the Instruction Manual or visit our website for detailed troubleshooting information.
Cuff Sizes
Technical Service Hotline:
866-301-8723
www.elemano.com
Measures
Contact Terumo technical service
Check the connection between the main
body and the cuff/in ation bulb
Reapply the cuff around the patient's arm
Check the connection between the main
body and the cuff
Measure again while keeping the patient at
rest with the arm rested on a rm surface
Do not shake or vibrate the main body during
measurement
Use Auscultation or palpation for
measurement
Reapply the cuff around the patient's arm
Measure again while keeping the patient at
rest with the arm rested on a rm surface
Do not shake or vibrate the main body during
measurement
Arm circumference sizing:
Size
Centimeters Inches
SS
13-20 cm
5.1"-7.9"
S
17-26 cm
6.7"-10.2"
M
24-32 cm
9.4"-12.6"
L
32-42 cm
12.6"-16.5"
TERUMO MEDICAL PRODUCTS