Electromagnetic Emissions And Immunity; Emc Guidance - Philips AVENT SCH740 Manual

Hide thumbs Also See for AVENT SCH740:

Advertisement

Problem
Possible cause and solution
Ensure that the thermometer is
pointed correctly towards the
eardrum (Fig. 10). An incorrect
measurement position (Fig. 11)
might lead to incorrect
measurement results. Make
sure that you have read the
user manual and know how to
use the device properly (see
'Measuring temperature').
Your room temperature is
too low or too high. Use
your thermometer at room
temperatures between 10.0°C
and 40.0°C (50.0°F and 104.0°F).
You are using the device
outdoors. The device is only
intended for indoor use.
You have held the device in your
hand too long. Put the device on
the table in the room where the
measurement is taking place and
let it cool down first.
The battery
The batteries are empty. Replace
symbol flashes
the empty batteries with new
on the display.
ones.
The display
Self-test error. Remove and
shows Err0
replace the batteries. If the error
persists, contact Philips Consumer
Care Center in your country.
The display
Stabilization error. Wait
shows Err2
30 minutes and measure again.
The display
Battery level low. Replace the
shows Err3
empty batteries with new ones.
Explanation of symbols
The warning signs and symbols are essential
to ensure that you use this product safely and
correctly and to protect you and others from injury.
Below you find the meaning of the warning signs
and symbols on the label and in the user manual.
Symbol for 'follow instructions for use'.
This symbol means that the part of the
device that comes into physical contact
with the user (also known as the applied
part, probe tip) is of type BF (Body
Floating) according to IEC 60601-1.
Symbol for 'the device complies with
European Medical Device Directive
93/42/EEC requirements'. 0344 refers to
the notified body.
Symbol for WEEE, waste electrical and
electronical equipment. Electrical waste
products should not be disposed of with
household waste. Please recycle where
facilities exist. Check with your local
authority or retailer for recycling advice
and see chapter 'Battery recycling'.
This symbol means that this product
contains batteries which shall not be
disposed of with normal household waste
(2006/66/EC).
Indicates the manufacturer, as defined in
EU Directives 93/42/EEC.
Symbol for the 'Bluetooth Combination
mark'. The device uses Bluetooth for
communication.
Indicates the manufacturer's serial number
so that a specific medical device can be
identified.
Symbol for indoor use only.
Indicates caution.The user should consult
the instructions for use for important
cautionary information such as warnings
and precautions that cannot, for a variety
of reasons, be presented on the medical
device itself.
IP22: The first number 2: Protected against
solid foreign objects of 12,5 mm Ф and
greater. The second number: Protected
against vertically falling water drops when
enclosure is tilted up to 15°. Vertically
falling drops shall have no harmful effects
when the enclosure is tilted at any angle
up to 15° on either side of the vertical.
Indicates the storage and transportation
60ºC
temperature limits to which the medical
–20ºC
device can be safely exposed: -20°C to
60°C.
Indicates the storage and transportation
relative humidity limits to which the
medical device can be safely exposed:
15% - 85%
Symbol for the 2 year Philips guarantee.
The Green Dot ('Der Grüne Punkt' in
German) is the license symbol of a
European network of industry-funded
systems for recycling the packaging
materials of consumer goods.
Electromagnetic emissions and
immunity
The device is approved according to EMC safety
standard EN 60601-1-2. It is designed to be used in
typical domestic environments.
Hereby, Philips Consumer Lifestyle BV declares
that the radio equipment type Bluetooth LE is in
compliance with Directive 2014/53/EU.
The full text of the EU declaration of conformity is
available at the following internet address:
www.philips.com/support

EMC Guidance

-
The ear thermometer needs special precautions
regarding EMC and needs to be installed
and put into service according to the EMC
information provided in this document.
-
Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations,
walkie-talkies can affect this equipment and
should be kept at least a distance d = 3.3 m
away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME
equipment, a typical cell phone with a maximum
output power of 2 W yields d = 3.3 m at an
immunity level of 3V/m.
Guidance and manufacturer's declaration
– electromagnetic emissions - for all ME
equipment and ME systems
Guidance and manufacturer's declaration –
electromagnetic emissions
The device is intended for use in the
electromagnetic environment specified below.
The customer or the user of the device should
assure that it is used in such an environment.
Emissions
Com pli­
Electromagnetic
test
ance
environment ­
guidance
RF emissions
Group 1
The SCH740 uses
CISPR 11
RF energy only for
its internal function.
Therefore, its RF
emissions are very
low and are not
likely to cause any
interference in nearby
electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic
Not
emissions IEC
applica-
61000-3-2
ble
Voltage fluc-
Not
tu a tions/flick-
applica-
er emissions
ble
IEC 61000-
3-3
Guidance and manufacturer's declaration
– electromagnetic immunity – for all ME
equipment and ME systems
Guidance and manufacturer's declaration –
electromagnetic immunity
The device is intended for use in the
electromagnetic environment specified below.
The customer or the user of the device should
assure that it is used in such an environment.
Immu­
IEC
Com­
Electromagnetic
ni ty
60601
pli ance
environment ­
test
test
level
guidance
level
Elec-
±6 kV
±6 kV
Floors should be
tro sta-
cont-
cont-
wood, concrete or
t ic dis-
act±8 kV
act±8
ceramic tile. If floors
charge
air
kV air
are covered with
(ESD)
synthetic material,
IEC
the relative
61000-
humidity should be
4-2
at least 30%.
Power
3A/m
3A/m
Power frequency
fre-
magnetic fields
quen cy
should be at levels
(50/60-
characteristic of
Hz)
a typical location
mag-
in a typical
net ic
commercial
field IEC
or hospital
61000-
environment.
4-8
Table 4 Guidance and manufacturer's
declaration – electromagnetic immunity
–for ME equipment and ME systems that
are not life supporting
Guidance and manufacturer's declaration –
electromagnetic immunity .The device is intended
for use in the electromagnetic environment
specified below. The customer or the user of the
device should assure that it is used in such an
environment.
IMMUNITY
IEC 60601
Compliance
test
TEST LEVEL
level
Conducted
3 Vrms150 kHz
3 Vrms
RFIEC 61000-
to80 MHz
4-6
Radiated
3 V/m80 MHz to
3 V/m
RFIEC 61000-
2.5 GHz
4-3
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part of
the device, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.

Advertisement

loading

This manual is also suitable for:

Avent sch740/01