DOMINA PLUS B Water Steam Sterilizer User Manual Rev. 8 OM1004U Date: March 2017...
Page 2
DentalX. Information in this manual is subject to change without any warning or prior notice by Dental X and does not represent a commitment for the vendor.
8.5 CLASS B ADDITIONAL ALARMS ...................... 27 CONNECTIONS ............................28 9.1 CONNECTION TO AN EXTERNAL PRINTER ................... 28 9.2 INTEGRATED PRINTER (OPTION) ....................28 9.3 CONNECTION TO A COMPUTER (ONLY FOR SERVICE) ............... 29 OPERATOR’S NOTES ..........................30 APPENDIX: SERVICE BOOK Domina Plus B...
FORWARDER, DENTAL X AND ITS AREA AGENT. CONFORMITY TO EUROPEAN DIRECTIVES Domina Plus B is produced by Dental X and conforms to electromagnetic compatibility standards in ac- cordance with the Medical Devices Directive 93/42/EEC, with European Standard EN 13060 and with the Pressure Equipment Directive (PED) 97/23/CE.
57 Kg USABLE SPACE IN THE CHAMBER Diameter: 240 mm Depth: 384 mm Capacity: 17,5 liters Useful dimensions per tray: 315 x 214 mm (x 2), 315 x 168 mm (x 2) Useful tray volume : 10 liters Domina Plus B...
Elapsed 30 minutes from the end of the cycle without opening the door or without activating a front panel button, the unit will switch off automatically. This function is not operating if no sterilization cycle has been carried out and completed. Domina Plus B...
The unit is not to be used to sterilize liquids and flammable materials. The unit is designed for indoor use only. Do not use in presence of anesthetic or flammable gas. To avoid an excessive level of humidity, properly air the room where the unit is installed. Domina Plus B...
Page 9
the device. INTENDED USE Domina Plus B is a small table steam generator designed to decontaminate dental and medical products in accordance with ISO EN 13060 requirements. Its intended use is to perform: • Class B sterilisation The sterilisation of all wrapped or non-wrapped, solid, type A hollow-load products and porous products as represented by the test loads.
20% to 85%, atmospheric pressure from 750 mBar to 1050 mBar and alti- tude between 0 and +2000 meters. Storage conditions: temperature -10°C 50°C, humidity without condensation 10-95%, atmospheric pressure 500-1060 mBar. NEVER USE IN PRESENCE OF INFLAMMABILE ANESTHETIC GAS Domina Plus B...
Pollution degree: 2 Transient overvoltage protection: II Monitoring of the water conductivity Maximum volume available on the trays: 10 l Maximum chamber temperature: 135°C (-0/+2°C) Safety valve working pressure : 2,55 bar Pressurised container conforming to Directive 97/23/EC (PED) Domina Plus B...
(see next page). Then push the key to confirm the displayed value and start the automatic initialization procedure with the entry of water in the hydraulic circuit and chamber. Domina Plus B...
AL8 or AL5 alarm should be signaled. CONVERSION NOTE: to obtain meters multiply feet by 0.3048. These operations can be carried out by qualified personnel. Wrong settings can affect the quality of the sterilization. Domina Plus B...
Test : allows to carry out the Bowie & Dick Test with sterilizer normally turned on, or the Vacuum Test with sterilizer in stand-by and chamber temperature lower than 35°C. Domina Plus B...
TEST Not a sterilization COMPLY drying 10’ (6’ with vacuum + 4’ 2,04-2,25 bar or equivalent cycle cod. 1300) ventilation) Vacuum test Temperature under 35°C TEST < 35°C Empty chamber Not a sterilization cycle (*) Liquids excluded Domina Plus B...
The display Time shows the countdown of this phase. Follows the forced ventilation phase through the bacterial filter – the display Time shows also the countdown of this phase. Domina Plus B...
Remove the load and check for the presence of water into the chamber. In case of wrapped instruments, we suggest to replace with new bags. Before loading the chamber again, dry it carefully and wait 10 minutes to allow the water to evaporate and be drained completely. Domina Plus B...
CAUTION! The water contained in the used water tank is to be regarded as biologically con- taminating, therefore when this tank is emptied, suitable precautions should be taken. The dis- posal of recovery water needs to be done in accordance with the national or local regulation. Domina Plus B...
Press key measurement unit SET UNIT PSI Push again the key to exit the programming mode The equipment is normally preset in the factory with “°C” and “bar” measurement units; the default lan- guage is English (L2). Domina Plus B...
At the end of the special cycle S5, only the Led READY will switch on - not STERILIZE - to sig-nal that the efficiency of the cycle selected by the operator has not been validated by the manu-facturer. Domina Plus B...
In order to extend the sterilizer life, we recommend to carry out an accurate cleaning of instruments; one of the main causes of an early equipment wear is the settlement and accumulation of debris and fragments for inadequately in- strument cleaning, and consequent stains, fouling and progressive clogging of filters, electro-valves and tubing. Domina Plus B...
The frequency of these checks is regulated at local country level; check the regula- tions in force. For any explanations or information, contact your authorized reseller or directly Dental X. 6.5.1 Bowie & Dick test The test can be performed at any time with the sterilizer turned on and operating.
Push the key Electro-valve 4 energized (open) Power Push the key Electro-valve 5 energized (closed) Test Push the key Condenser’s fan powered Exit the diagnosis and return to the normal mode Push the key Domina Plus B...
< 0,5 mg/l < 0,1 mg/l Conductivity (at 20°C) < 15 uS < 3 uS from 5 to 7,5 from 5 to 7 appearance colorless, clean, without sediment colorless, clean, without sediment hardness < 0,02 mmol/l < 0,02 mmol/l Domina Plus B...
Need for the installation procedure. Chapter 3.2) NEED TEST Detected a pre-warning alarm See table B Clean the door gasket and repeat the test. TEST FAIL Negative result of the Vacuum Test Call for a technical service Domina Plus B...
(see § 6.1), and the sterilization phase the Vacuum Test. AL 10 Pressure too high Call for a technical service. Check the door tightness. Perform, if needed, AL 11 Pressure too low the automatic cleaning cycle (see § 6.1). Domina Plus B...
Solution AL 18 Drying phase Drying interrupted Dry the load AL 31 Drying phase Vacuum not sufficient Excess of load NOTE: The Class B alarms may occur only for the programs 1, 2, S2, S3 and S4. Domina Plus B...
The printer cable cannot exceed 3 meters. The serial port of the equipment can be connected to a printer with serial interface RS232. Please contact Dental X for further information. Switch-on the printer, Switch-on the sterilizer. The report is automatically printed on the course of the cycle and with the following data content: date and time of the process –...
Do not remove it if the device is not in use. Do not connect devices not supported by the manufacturer. THE RJ45 CONNECTOR IS NOT DESIGNED FOR A LAN CONNECTION, THIS OPER- ATION WILL DESTROY THE MOST OF THE ELECTRONIC PART OF THE AUTO- CLAVE. Domina Plus B...
“ NEED SERVICE”. This feature is in compliance with the norm EN13060 and answers to spe- cific safety requirements. The maintenance activities must be performed by authorized technicians (provided with card or certificate released by Dental X) and reported on this Service Book. Model ..............
Page 35
APPENDIX SERVICE BOOK □ □ □ □ Sterilizer OK Sterilizer OK Date ................Date ................No. of cycle ....Cleaning cycles ......No. of cycle ....Cleaning cycles ......Aborted cycles ... Alarm code ........Aborted cycles ..Alarm code ........Technician name: .............
Page 36
APPENDIX SERVICE BOOK □ □ □ □ Sterilizer OK Sterilizer OK Date ................Date ................No. of cycle ....Cleaning cycles ......No. of cycle ....Cleaning cycles ......Aborted cycles ..Alarm code ........Aborted cycles ... Alarm code ........Technician name: .............
Page 37
APPENDIX SERVICE BOOK □ □ □ □ Sterilizer OK Sterilizer OK Date ................Date ................No. of cycle ....Cleaning cycles ......No. of cycle ....Cleaning cycles ......Aborted cycles ... Alarm code ........Aborted cycles ..Alarm code ........Technician name: .............
Page 40
DENTAL X S.p.A. Via Marzotto 11 36031 Dueville (VI) Italy Tel. +39 0444 367400 Fax +39 0444 367436 e-mail: dentalx@dentalx.it http://www.dentalx.it Dental X 2017 Printed in Italy Sous réserve de modifi- Ci riserviamo il diritto di Reservados los derechos de Änderungen im Zuge techni-...