Page 2
Avoid prolonged, direct exposure to the vapor generated by Solo 365. Safety Data Sheets for the Solo 370 battery pack and Solo ES6 Smoke Cartridge are available. It is recommended to review the safety data sheets for Solo 370 Battery Pack and Solo ES6 Smoke Cartridge before use (see www.sdifire.com/support).
Page 3
Information • Read this User Manual completely before using your • Do not use your Solo 365 if it is not operating correctly (see fault nding guides) and consult Solo 365. Troubleshooting (Section 9) of this manual. If required seek technical advice (Section 10).
Kit Contents Solo 365 Electronic Smoke Detector Tester Kit containing: Solo 371 Smoke Generator x 1 Solo 370 Lithium Ion Battery Pack x 1 Solo ES6 Smoke Cartridge x 1 Solo 365 Head Unit x 1 www.sdifire.com...
Page 5
Solo 365 Universal USB Charger x1 Quick Start Guide x1 USB Lead x1 Product Registration Card x1 www.sdifire.com...
6.6 Replacing the Battery 4 Using Solo 365 Page 18 7 Consumables & Accessories Page 11 4.1 Attaching Solo 365 to Solo Access Poles Page 19 8 Spares 4.2 Powering on Solo 365 8.1 Removing the Clear Cup 4.3 Carrying Out a Simple Test 8.2 Replacing the Clear Cup...
Page 8
Thank you for purchasing the Solo 365 Smoke Detector Tester. This manual is designed to assist you to get the best and most efcient use of the Solo 365 and provides all the information required to perform routine service and maintenance tasks with ease.
• When replacing a generator, any dust or debris within the powered USB socket using the supplied mains adaptor. the Solo 365 housing can be removed using an air • Ensure the battery is fully charged before use. duster. Any condensation can be removed by wiping •...
3.3 Inserting the Smoke Cartridge 3.4 Inserting the Battery • Remove cartridge from bag. • Once the battery is charged clip the battery pack into • Do not insert smoke cartridge until generator has the battery compartment. (Fig. 3). been fully fitted into the main unit. See installing the •...
4.1 Attaching Solo 365 to Solo Access Poles When working at height it is recommended Solo 365 is designed for use with the Solo range of access that a competent person carries out a suitable poles. (purchased separately). The product is not risk assessment.
Flashing Orange: Delayed start NOTE: For the best results the detector should be fully within the cup. Solo 365 may be used in ceiling voids and Slow ashing Red: Consumable low oor voids provided the detectors can be accessed safely.
On occasion it may be necessary to test detectors that do In low light levels an LED torch will automatically illuminate not easily t into the Solo 365 smoke cup or are obstructed from underneath the cup. This makes for easy alignment in some way.
5.1 Testing an ASD System Figure 8 An ASD System can be tested by using the ‘Delayed Start Mode’ and replacing the cup with the Solo 365 ASD Adaptor (Product Code: Solo 372-001). (Fig. 6) Users should note that using the Solo365 in this conguration is outside the scope of the...
5.3 Testing with the ASD Adaptor Power Solo 365 on Select the ‘Delayed Start Mode’ by pressing the ‘function button’ for one second – the status LEDs will ash orange Within 20 seconds ensure Solo365 is locat- ed over the sampling hole on the aspirating pipe to carry out the test (Fig.
Removing & Replacing Consumables 6.2 Replacing the Smoke Cartridge All critical parts of Solo 365 can be replaced in the eld Remove the cartridge from the bag without having to return the unit for service. Ensure Slide cartridge completely into generator housing that the unit is switched off during the replacement of following the guide rails (Fiig.
6.3 Removing the Smoke Generator 6.5 Removing the Battery for Charging Ensure the unit is switched off and the Smoke Ensure the unit is switched off and open the back Cartridge is removed (Section 6.1) cover. Disengage clip ‘1’ and lift. Repeat for clip ‘2’ to Unclip and remove the battery remove the used generator (Fig.
Consumables & Accessories Consumables ES6-6 PACK-024 Solo 365 Smoke Cartridges x 6 SOLO370- Solo 365 Lithium Ion Battery 1 PACK-024 Pack x1 SOLO371-024 Solo 365 Smoke Generator SERVICE360-024 Solo 365 Service Pack containing: Spare 1058 x 1, Solo 370 x 1 &...
Spares 8.1 Removing the Clear Cup 8.3 Removing & Replacing the Membrane Align the notch on the left hand side arm and Ease the membrane off the clear cup gently disengage by squeezing the cup and gently Take the replacement membrane and stretch it pulling the arm outwards.
10.1 Maintenance The Solo 365 unit contains no user serviceable parts. Do not disassemble. In the case of a build-up of dust or dirt on the surface of the product, wipe with a slightly damp lint-free cloth. The area around the smoke duct and within the cup should be cleaned periodically to remove any dirt or uid.
Page 22
SDi Fire www.sdifire.com 3535 Route 66, Building 6, Neptune, NJ 07753 (732) 751-9266 service@sdifire.com As our policy is one of continuous improvement, details of products described within this publication are subject to change without notice. All information provided here is believed to be correct at the time of going to press. Every effort has been made to ensure the accuracy of information which is provided in good faith but nothing contained herein is intended to incorporate any representation or warranty, either express or implied or to form the basis of any legal relations between the parties hereto, additional to or in lieu of such as may be applicable to a contract of sale or purchase.
Page 31
8) Shut the control unit off and then back on. The sensor is now reset. Try running the machine now. If it still fails it MUST be returned for Calibration Diagnostic Mode. The Trutest Request Return Materials Authorization (RMA) can be requested at: http://www.sdire.com/site/support/support_rma_request_form/ www.sdire.com...
Page 32
User Manual OPERATING, SAFETY AND MAINTENANCE INSTRUCTIONS Issue 2.1e Read and understand this manual before using Trutest. Failure to do so will increase the risk of injury. TRUT771...
Page 34
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ SAFETY INSTRUCTIONS Before operating Trutest, read this User Manual and become familiar with it and the equipment. Safe and efficient operation can be achieved only if the equipment is properly operated and maintained. Accidents are caused by failure to follow fundamental rules and precautions.
Page 35
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ ___________________________________________________________________ WARNING WORKING WITH TOOLS AT HEIGHT CAN BE DANGEROUS. Falling objects can cause serious injury or death. Trutest is designed to be used at height. It may be mounted solely on the top of SOLO telescopic operating poles.
Page 36
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ __________________________________________________________________ CAUTION AEROSOL CANISTERS CAN BE HAZARDOUS. Pressurized containers can explode if correct method of storage and use is not observed. Contents can be ignited under certain conditions. When using/handling/storing the Trutest aerosol: •...
Page 37
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ This page is intentionally blank ___________________________________________________________________ PRIMARY INFORMATION...
Page 38
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ TABLE OF CONTENTS Safety Instructions Table of Contents iii. Preface viii Important Notes for the User Statement Of Guarantee And Liability TECHNICAL DESCRIPTION 1.1. Why Test Sensitivity? 1.2. Why Test In-Situ? 1.3. Why Trutest? 1.4.
Page 39
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ OPERATING INSTRUCTIONS 4.1. Preparing to Test 4.2. Selecting the Detector Type 4.3. Selecting the Test Mode 4.3.1. Max/Min Test 4.3.2. Fast Ramp Test 4.3.3. Slow Ramp Test 4.4. Selecting a Detector Setup for Max/Min Mode 4.5.
Page 40
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 6.5. Troubleshooting Guide FREQUENTLY ASKED QUESTIONS 7.1. Questions on Usage 7.2. Questions on Sensitivity & Calibration 7.3. Questions on effects of Temperature & Humidity 7.4. Questions on Trutest Aerosol 7.5. Questions on Batteries and Power 7.6.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ iii. PREFACE This manual is: designed to outline the correct methods of operating and maintaining your Trutest system. written in a logical order for first time users. designed to be detailed. It is not, however, possible to cover all situations, possible problems or questions.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ IMPORTANT NOTES FOR THE USER Safety Read this manual and the Safety Instructions before using Trutest. The user must be thoroughly acquainted with the procedures in the manual to ensure maximum safety in use. User Registration Card Copy the serial number onto your registration card and return in order to register the product with the manufacturer.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ STATEMENT OF GUARANTEE AND LIABILITY No Climb Products Ltd. warrants this Trutest system described herein to be free from defect in material and factory workmanship for a period of one year from the date of shipment and agrees to repair such products which, under normal usage and service, disclose the defect to be the fault of No Climb Products.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ TECHNICAL DESCRIPTION 1.1. WHY TEST SENSITIVITY? Over-sensitive detectors are more prone to false alarms. Under-sensitive detectors tend to delay the alarm signal (if they alarm at all). sensitivity of modern smoke detectors are well controlled during manufacture but detectors in the field are subject to airborne contaminants, aging and outside interference - factors which can all significantly affect sensitivity.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 1.4. TRUTEST PRINCIPLES Trutest is fully automatic. The user merely determines which type of detector is under test and which type of test he wishes to perform. This information is input via a simple menu prompt system into the control unit carried on a shoulder strap.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ INTRODUCTION TO TRUTEST 2.1. INTRODUCTION Trutest is a technologically advanced, fully field-portable device for in-situ testing of smoke detector sensitivity. The equipment is suitable for testing photoelectric and ionization detectors; conventional, addressable and analogue addressable.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 2.2. COMPONENTS The complete set of equipment is comprised of the following main components (not all of which are necessarily supplied): • Main unit • Control unit • Synthetic test aerosols • Two interconnecting cables (differing lengths) •...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 2.2.3. Test Aerosols Trutest uses a form of synthetic test smoke stored in a pressurized aerosol container. This produces a fine mist of aerosol particles to simulate the smoke from a real fire. The aerosol formula and canister are specific to Trutest, and cannot be substituted with any other test aerosol.
Page 49
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ This page is intentionally blank ___________________________________________________________________ INTRODUCTION TO TRUTEST...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ ASSEMBLY INSTRUCTIONS 3.1. GENERAL Trutest is used in the following assembly: Fig. 1 __________________________________________________________________ ASSEMBLY INSTRUCTIONS...
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 3.2. COMPONENTS OF TRUTEST 3.2.1 The Control Unit Fig. 2 Carried on a shoulder strap by the operator, the control unit incorporates a microprocessor, LCD display, keyboard and rechargeable battery pack. Commonly used test limits may be pre-stored within the microprocessor for instant recall when required (see Section 4.5., Entering New Test Parameters for Max/Min Mode).
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 3.2.1.4 The rechargeable battery pack within the control unit will provide sufficient power for a day of testing. It may then be recharged overnight via the battery charger provided. Note: The control unit features an auto power down. If left on, it will automatically switch off after fifteen minutes, thereby saving battery power.
Page 53
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ Fig. 3 Components of the Main Unit: 3.2.2.1 The Aerosol Canister (A). The aerosol canister contains an "environment friendly", pressurized formulation which produces the test aerosol cloud. The aerosol particle sizes simulate the spectrum of particles generated by the smoke of a real fire.
Page 54
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ __________________________________________________________________ CAUTION AEROSOL CANISTERS CAN BE HAZARDOUS. Pressurized containers can explode if correct method of storage and use is not observed. Contents can be ignited under certain conditions. When using/handling/storing the Trutest aerosol: •...
Page 55
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ Note: Only aerosol cans specifically marked for use within Trutest can be used within this device. The use of any other aerosols will invalidate results and may cause irreparable damage to Trutest. Always re-order the same aerosol part number.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ Specifications). Before testing a detector, the user should always check that the correct size seal is being used in order that a good seal is maintained around the detector. The diaphragm helps to seal the detector within the cup, reducing leaks, and together with the slight positive pressure within the sensing loop, preventing the ingress of clean air into the sensing loop after the aerosol concentration has been...
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ WARNING Trutest is used above head height. • Read the Safety Instructions at the start of this User Manual. Failure to do so will increase the risk of injury. • When using Trutest at height, extreme care should be taken.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ CAUTION Trutest uses a pressurized aerosol canister. • Read the Safety Instructions at the start of this User Manual. Failure to do so will increase the risk of injury. • Always follow the instructions on the canister.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 3.4. FITTING THE AEROSOL CANISTER The aerosol canister is inserted into the main unit from the base of the main unit as follows: Locate the main unit at a convenient working height, not above the head.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 3.5. PREPARATORY CHECKS Before assembling Trutest for use, check the following: Ensure that the battery in the unit is fully charged. Ensure that you have the correct poles and extensions required to access the height of the detector.
Page 61
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ WARNING • Always ensure that Trutest aerosol canister is firmly located and secured in Trutest, before raising the unit above head height. This will mean that the canister will not be displaced during operation. Attach the interconnecting cable between Trutest main unit and control unit of appropriate length for the height you will be working.
Page 62
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ WARNING • Do not use Trutest unless it is mounted correct securely manner on the specified operating poles. • Always ensure that the spring buttons on the operating poles and the stem at the base of Trutest’s main unit are fully engaged, before raising the unit above head height.
Page 63
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ This page is intentionally blank ___________________________________________________________________ ASSEMBLY INSTRUCTIONS...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ OPERATING INSTRUCTIONS 4.1. PREPARING TO TEST Check that the unit has been correctly assembled as described in the assembly instructions (Chapter 3) before turning on the power. In particular, ensure that the battery has been charged, and the aerosol canister has been correctly installed (see section 3.4., Fitting the Aerosol Canister).
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ Fast ramp photo hi profile Press Start to test Slow ramp hi profile Press Start to test The user must determine which type and model of detector is to be tested, and which type of test is to be performed. The control unit requires up to three pieces of information to be set prior to performing the test.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 4.3. SELECTING THE TEST MODE Use the "Test mode" key on the keypad to toggle between these three types of test: 1. Max/min (see section 4.3.1.) 2. Fast ramp (see section 4.3.2.) 3.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ Max. is the maximum reading of obscuration in %/ft required to activate the detector. This figure can be determined from the higher sensitivity rating of the detector (printed on the label on the base of most UL listed detectors or from the manufacturers’...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 4.3.3. Slow Ramp Test This is the most accurate measurement of sensitivity within Trutest, and is used to carry out normal quantitative measurements. Because of delays built into certain detectors the ramp rate needs to be fairly slow to give an accurate reading.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ Select option 1 by pressing the numeric key "1". A screen similar to the following will appear: 01: NO_NAME_01 photo Which one 01-18? The display shows a detector setup that has been entered previously.
Page 70
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ From any of the default screens, (see Section 4.1., Preparing to Test) press the 'Set-up' key on the keypad. The following menu screen will appear: 1.Recall Setup 2.Change Setup 3.Number of Tests Select option 2 by pressing the number "2"...
Page 71
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ and “up” respectively. Press “Enter” to modify or enter data in that field. With the cursor positioned adjacent to the detector name field, pressing the “Enter” key reveals the following screen: 06: NO_NAME_06 ABCDEFGHIJKLM01234 - NOPQRSTUVWXYZ56789 A new cursor, “^”, shows beneath the first character of the detector...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ For the ramp speed field (slow ramp or fast ramp), the “Enter” key toggles between “slow ramp” and “fast ramp”. When all parameters have been entered/modified as required, move the cursor to the “exit” field and press “Enter”. Trutest saves all the changes and returns to the screen, which was displayed prior to entering setup.
Page 73
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ Fast ramp photo hi profile Resetting sensor 7 NO_NAME_01 min=1.00 max=2.00 Resetting sensor 7 This ‘initialization stage’ can be carried out without the Trutest cup positioned over the detector e.g. when moving from one test to another, but must be in place for the test to begin (see below).
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ NO_NAME_01 min=1.00 max=2.00 Reading=0.62 %/ft 4.6.4 At this point the user is simply a witness as Trutest performs the test. He/she should watch the detector to ensure that the diaphragm is sealed around the top and wait for the LED of the detector to activate when the detector reaches its alarm level.
Page 75
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ shown on the result screen. When the "Enter" key is pressed, the display returns to the default screen, still showing the amount of smoke cleared from the internal sensing loop. For example (max/min screen): NO_NAME_01 min=1.00 max=2.00...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ RESULTS 5.1. GENERAL OBJECTIVE OF RESULTS Trutest has been designed to perform on-site sensitivity tests on all makes of smoke detector. The sensitivity of a detector may be defined as the density (concentration) of smoke required to trigger an alarm. Testing with Trutest should enable the user to identify: The ability of the detector to respond to the presence of a pre- defined external smoke stimulus.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 5.2.3 If the obscuration level reaches the maximum level attainable (see section 8., Technical Specifications) and the detector has not alarmed, the test will stop automatically and the message “Full scale” will replace the “Result”...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ Result: Detector passed Reading=2.30 %/ft Press Enter 5.3.3 If the detector LED does not illuminate (the detector does not activate) by the “max” level, and the key is pushed to stop the test, the results display will read (for an alarm level of 3.1%/ft in this example): Result: Failed insensitive...
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ Should more than one of these messages be relevant at the time of a result screen display, only the highest priority warning will be shown. A two-second warning beep will accompany any of these warning messages to alert the user.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 5.6. INTERPRETATION OF RESULTS IMPORTANT - READ CAREFULLY It should be noted when using Trutest, that there are a large number of factors affecting the sensitivity readings, which should only be judged within the conditions and limits quoted in the Technical Specifications.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 5.7. ACTION IF A DETECTOR FAILS A TEST WITH TRUTEST If a detector fails a sensitivity test, there are steps you should take prior to condemning it. Ensure that the aerosol canister in Trutest is not nearly empty. This can easily be determined by shaking the canister and listening for any remaining liquid.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ SERVICE AND MAINTENANCE Trutest is a high quality electronic test instrument and as such, regular calibration and servicing as well as care in use are essential. Regular calibration ensures reliability, and continuing accuracy in use. It is recommended that this product be returned to an authorized service center annually.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 6.1.3. Replacing the Fuse The fuse holder is located on the side of the control unit and is clearly labeled. Never replace the fuse with any other type / rating than that recommended in the specifications.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 6.2. ANNUAL SERVICING In order to maintain the accuracy and performance from this instrument, it is recommended that Trutest be returned to an authorized service center for an annual service and re-calibration. This should be performed annually or after 5,000 tests, whichever is the sooner.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 6.4. WARNING MESSAGES The following warning messages are generally preceded by a two-second continuous beep from the control unit. 6.4.1. Low Battery 6.4.1.1 When the NiCd Battery is drained almost completely, the Low Battery Warning message is displayed.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 6.4.2.1 A warning message will occur when the number of tests remaining before service goes below 250. The message appears on the result screen, for example: Service due soon! Detector passed Reading=1.97 %/ft Press Enter The user can determine exactly how far away he/she is from an obligatory service by interrogating the control unit (see section 6.3.,...
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 6.4.4. Sensor Drift If the following message appears on a result screen: Sensor drifted! Detector passed Reading=2.64 %/ft Press Enter this means that the sensor may be dirty, blocked or misaligned. The results may be unreliable and care should be taken with their interpretation.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 6.4.7 Min/Max Error Should the setup parameters for a min/max test be entered in such a manner that the full-scale obscuration value of Trutest is exceeded for that type of detector, Trutest alerts the user prior to saving setup information.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 6.5. TROUBLESHOOTING GUIDE The following problems do not require professional servicing. problem cannot be remedied using this guide, contact your authorized service center for assistance. Problem Possible Causes Remedy Blank screen, no Blown fuse.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ FREQUENTLY ASKED QUESTIONS 7.1. QUESTIONS ON USAGE Does Trutest suit all makes of smoke detector? Yes, Trutest has been developed specifically to this end. It is used normally to test against the maximum and minimum values printed on the base of the detector, or in the manufacturers’...
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 7.2. QUESTIONS ON SENSITIVITY & CALIBRATION How do I know that Trutest will give accurate results? Trutest has been tested at Underwriters Laboratories in the United States, where it has been given a generic listing as an accessory for automatic fire detection systems.
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 7.3. QUESTIONS ON EFFECTS OF TEMPERATURE & HUMIDITY What is the effect of temperature on performance: Extensive tests have been carried out to determine the effect of temperature on the performance of Trutest. As Trutest uses a liquefied gas aerosol canister, the critical operating temperatures are determined by the gas characteristics.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ 7.4. QUESTIONS ON TRUTEST AEROSOL consistent repeatable aerosol characteristics? Very. This is a function of the aerosol canister, chemical formulation, valve and spray nozzle. All are subject to strict quality control (as is the rest of Trutest). What effect does "aging of smoke"...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ Can any other smoke aerosol canister be used in the device? No. The tool is designed to use only the special Trutest aerosol canisters, and a different product would not work or give correct results. Other formulations could cause serious damage to both the detector under test and Trutest.
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ How does the user know what stage of the test cycle is running? Either by looking at the message on the display or by hearing different beeps from the control unit. What memory has Trutest? The microprocessor has an internal memory for storing detector settings.
Page 96
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ How do we cope with detectors fitted in situations where Trutest can not reach or fit? The detectors which cannot be accessed can either be removed from their bases and placed in more accessible bases (exchanging the detectors) or powered up separately and tested.
Page 97
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ This page is intentionally blank ___________________________________________________________________ FREQUENTLY ASKED QUESTIONS...
Page 99
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ OPERATING DATA Trutest uncertainties in sensitivity measurements: (Note: For correct interpretation of results, the tolerance of the detector under test and any other measurement uncertainties must also be taken into account). Specified at 68°F +/- 5°F <...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ 8.2. AEROSOL DATA Description: Part Number: AERO400 General Description: An environmentally friendly aerosol canister for use with Trutest Smoke Detector Sensitivity Tester. Specifications: Dimensions: 2.1ins (53mm) dia x 6.8ins (173mm) Pressure: 18 bar maximum Material: Aluminum monobloc Contents:...
Page 101
Issue: 2.1e Trutest User Manual Date: 01/12/08 ___________________________________________________________________ UL Listing: 77TL Smoke-Automatic Fire Detector Accessory. Generic listing. Environmental Information: Tetrafluoroethane currently regulated under the Montreal Protocol and may be used as a replacement for regulated CFC's. ___________________________________________________________________ CAUTION AEROSOL CANISTERS CAN BE HAZARDOUS. Pressurized containers can explode if correct method of storage and use is not observed.
Issue: 2.1e Trutest User Manual Date: 01/12/2008 APPENDIX B - RESULTS PAGE Serial Number of Trutest Used_________________ Date of Tests_________ Sheet _____ of ______ Name of Test Engineer_____________ Name of Client __________________ Location (Site, Town, City, Country) _____________ Detector Tested Location Type of Test Result...
Issue: 2.1e Trutest User Manual Date: 01/12/08 __________________________________________________________________ APPENDIX C. - DETECTOR PERFORMANCE TABLES C.1. Method for Producing Detector Performance Tables If you cannot find the details for the particular type and model of detector you wish to test, the following procedure should be followed: Test a representative sample of the same model and type (e.g.
Issue: 2.1e Trutest User Manual Date: 01/12/08 Detector Performance Sheet Site:__________________________________ Date: ___________________ Model System Type Detector Type Max. Min. Diaphragm Manufacturer Ref. No. Conventional / Photoelectric / Pink or Gray Addressable Ionization 1 Model The manufacturer’s reference or model number for this detector. 2 Max.
Page 109
Contents Introduction and Key Features What does the Portalevel ® The Portalevel ® Operating Instructions Introduction Preparation 2.21 Function Test 2.22 Dip Test 2.23 Cylinder Preparations 2.24 Temperature Calibration Procedure Identification (CPI) 2.31 Test Procedure 1 2.32 Test Procedure 2 Troubleshooting Frequently Asked Questions Accessories...
1. Introduction and key features What does the Portalevel ® The Portalevel ® Max is designed to be a portable way to non-invasively locate the liquid level inside any single skinned cylinder. Portalevel s are capable of detecting the ® presence of any liquid externally;...
Page 111
Portalevel ® The Portalevel ® Max also includes; sensor, and ultrasonic gel (couplant). SPA: The SPA capability Gromit: The gromit at enables an increased the top of the unit is strength of signal output to the ultra secure simple achieve better readings for fastening for the sensor poor condition cylinders, more challenging...
2. Operating Instructions POINTS TO CONSIDER BEFORE PROCEEDING: Do not proceed before reading section 1. Do not proceed before familiarising yourself with the Training section on page 27. Introduction There are three basic procedures which must be carried out when using the Portalevel ®...
1. Preparation Preparation for testing is simple, quick and ensures you get the most reliable and accurate results from your testing. The FUNCTION Test and DIP Test both ensure the main unit and sensor are working correctly. CYLINDER PREPARATIONS ensure that you test on the best possible area of the cylinder wall.
CYLINDER PREPARATIONS: It is important to prepare the cylinders well to attain accurate and reliable readings. To perform CYLINDER PREPARATIONS: Step 1: Find on the cylinder the side which has the least damage, rust or chipped paint. For accurate testing you must find a vertical strip down the side of the cylinder which is as smooth as possible.
2. Calibration Procedure Identification (CPI) T he purpose of this test is to identify whether method 1 or 2 should be used for the cylinder being tested. The details and differences between methods 1 and 2 are explained in the next section. To perform CPI: Step 1: Place a strip of gel or water down the vertical strip on the cylinder you chose during CYLINDER PREPARATIONS.
Page 116
3. Testing PROCEDURE 1: TO BE USED WHERE HIGHER READINGS ARE FOUND IN THE GAS PHASE COMPARED TO THE LIQUID PHASE. Step 1: Place a strip of gel or water down the vertical strip on the cylinder you chose during CYLINDER PREPARATIONS. Step 2: Place the sensor on the area covered with gel or water and on the UPPER portion of the cylinder, where you are certain the sensor will be above the liquid level.
Page 117
Below liquid level Note: Some cylinders are prone to giving ‘false levels’;that is the bar graph may disappear after moving the sensor even though the Liquid Level has not been passed. If you think you may have found a false Liquid Level, simply move the sensor slightly to the left or right to check if the bar graph returns, as false levels can usually be caused by irregularities in a specific part of the steel wall.
Page 118
PROCEDURE 2: TO BE USED WHERE HIGHER READINGS ARE FOUND IN THE LIQUID PHASE COMPARED TO THE GAS PHASE. Step 1: Place a strip of gel or water down the vertical strip on the cylinder you chose during CYLINDER PREPARATIONS. Step 2: Place the sensor on the area covered with gel or water and on the LOWER portion of the cylinder, where you are certain the sensor will be below the Liquid Level.
Page 119
Above liquid level Note: Some cylinders are prone to giving ‘false levels’, that is the bar graph may disappear after moving the sensor even though the Liquid Level has not been passed. If you think you may have found a false Liquid Level, simply move the sensor slightly to the left or right to check if the bar graph returns, as false levels can usually be caused by irregularities in a specific part of the steel wall.
Page 120
Troubleshooting ISSUE 1: “The readings on the main unit did not change when the sensor was placed anywhere on the cylinder” SOLUTION: Firstly, ensure you are using an adequate quantity of water or gel to couple the sensor to the cylinder and also ensure that the rubber pad of the sensor is clean and undamaged.
Page 121
ISSUE 3: “Readings are rarely/never greater than 1000 on the side of the level where the highest readings are found.” SOLUTION: The Portalevel ® Max should, in most scenarios, be able to attain readings of more than 1000 on a cylinder. Achieving readings less than 1000 DOES NOT always prevent measurement, but in some cases may cause problems.
Page 122
ISSUE 4: “The bar graph does not fill up, even with CAL engaged” SOLUTION: The bar graph will only fill up if very high readings are found without CAL engaged, or once CAL is engaged and the current reading is significantly larger than the value displayed when CAL was pressed.
Page 123
ISSUE 5: “Debris/gel in sensor housing has caused the sensor mechanism to stick making it hard to use.” SOLUTION: The sensor can be removed from its housing and cleaned following the below steps. Step 1: Hold sensor in left hand with dot/top facing up. With right hand, unscrew rear cap completely by twisting anticlockwise.
Page 124
ISSUE 6: “There is not a significant difference between above and below level readings” Step 1: Carry out a ‘Calibration Procedure Identification’ (CPI) explained in Section 2.3, page 9. Step 2: Upon completing Step 1 press CAL on the exact location where the highest readings are found.
Frequently Asked Questions Why do I have to use water or gel with the sensor? The use of water or ultrasonic couplant gel is essential to the operation of a Portalevel ® unit. When the sensor both emits a high energy pulse and listens for the retuning echoes, excellent mechanical contact must be maintained between the senor and the container in order for the ultrasonic signals to pass into and out of the container efficiently and without interference.
Page 126
Why can’t I test CO and FE-13 in high temperatures? When testing some liquids, it is vital that testing is done under atmospheric temperatures lower than their critical temperature. At the critical temperature of a liquid, it transforms into a vapourous state in which a liquid level no longer exists inside the container to be measured.
3. Accessories Extension Rod (Marine Model) Sensor ® Ultrasonic Gel Portalevel ® Portatherm ® Portalevel ® Max in carry case with accessories Extension rod (Extra) The Portalevel ® extension rods are primarily developed for use in the marine industry and as such come as standard with the Portalevel® Max Marine (not with the Portalevel ®...
Page 128
The extension rod (Figure 3) sections simply screw together, with the sensor at one end (F) and the handle at the other (D). The cable connecting the sensor to the main unit (B) threads through the hollow sections and out of the handle to avoid tangling. Using the adjustment nut and an Alan key, the sensor (F) can be loosened and slid horizontally along the adjustment rail (G).
Portatherm (Optional Extra) ® Portatherm ® is an easy to use infrared thermometer which allows easy checking of a cylinders temperature prior to testing. This is useful when testing agents such as Carbon Dioxide or FE 13 in hot climates, which cannot be tested above their critical temperature.
Low Battery: This is represented by an icon. Turn the unit off and replace the batteries with 2 x AAA batteries. Ensure the unit is off before replacing batteries. Cleaning: Clean the thermometer lens with a soft cloth or cotton swab and water/ medical alcohol.
4. Maintenance This section contains instructions on how to maintain the equipment to extend its lifetime as long as possible. Maintenance of the equipment is simple and only requires the cleaning of the sensor. For any major works or re-calibrations the unit can be returned to the manufacturer for repairs.
Battery Replacement You will need a small flathead screwdriver and a 9V alkaline type 6LR61 Battery. Step 1: Ensure the unit is off and turn the unit so that the front fascia faces downwards. Step 2: Take a small, flat head screwdriver and insert it into the slit at the bottom of the back side of the unit.
5. Training This section is for new or untrained users, or for users who may need to remind themselves of how to operate a Portalevel ® The positions of the particular controls for the Portalevel Max are shown at the ®...
Page 134
The BENCH Test: Step 1: Connect the sensor to the Portalevel unit and turn the unit on. ® Step 2: Choose a vertical strip of your large container which is as continuously smooth as possible, free of chips, rust or any other kind of damage. Apply water or the ultrasonic gel along the length of this chosen vertical strip.
Page 135
Step 6: Move the sensor back up the vertical strip of the cylinder, again in steps of roughly 5cm (2 inches), observing the bar graph each time you place the sensor onto the container wall. Be sure the vertical strip of wall you chose of the container is still well covered by whatever couplant (water or gel) you have chosen.
Page 137
Contents Introduction and Key Features What does the Portalevel ® MINI do? Portalevel ® MINI How does a Portalevel ® MINI work? Operating Instructions Introduction Preparation 2.21 Assembly 2.22 Function Test 2.23 Dip Test 2.24 Cylinder Preparations 2.25 Temperature Calibration Procedure Identification (CPI) 2.31 Test Procedure 1 2.32 Test Procedure 2 Troubleshooting...
1. Introduction and key features What does the Portalevel® MINI do? The Portalevel® MINI ® is designed to be a portable way to non-invasively locate the liquid level inside any single skinned cylinder. Portalevel® MINIs are capable of detecting the presence of any liquid externally; from water and liquid pressurized gases to firefighting clean agents.
Portalevel® MINI The Portalevel® MINI is specifically designed for use in environments which are hard to access or in dark environments where hands-free use is beneficial. With a calibration feature, two power settings, LED liquid level display and connectors for a neck strap, the Portalevel®...
How does a Portalevel® MINI work? All sound is physical vibrations which move through a material, and in the case of the sound we are used to, the vibrations move through air. Ultrasound is the same as the sound we experience, but its frequency is beyond that which humans are capable of hearing.
2. Operating Instructions POINTS TO CONSIDER BEFORE PROCEEDING: Do not proceed before reading section 1. Do not proceed before familiarising yourself with the Training section on page 26. Introduction There are three basic procedures which must be carried out when using the Portalevel®...
DIP Test: The DIP Test checks whether the sensor you are using is working correctly and is communicating with the Portalevel® MINI ® main unit. Step 1: Connect extension rod to unit and turn on. Step 2: Lightly dip the tip of sensor into cup of water vertically. The amount of water in the cup is not important, as long as there is enough to dip the end of the sensor so the black central patch on the end of the senor is submerged, without it touching the bottom of the cup.
2. Calibration Procedure Identification (CPI) T he Purpose of this test is to identify whether method 1 or 2 should be used for the cylinder being tested. The details and differences between methods 1 and 2 are explained in the next section. To perform CPI: Step 1: Place a strip of gel or water down the vertical strip on the cylinder you chose during CYLINDER PREPARATIONS.
Page 144
3. Testing PROCEDURE 1: TO BE USED WHERE HIGHER READINGS ARE FOUND IN THE GAS PHASE COMPARED TO THE LIQUID PHASE. Step 1: Place a strip of gel or water down the vertical strip on the cylinder you chose during CYLINDER PREPARATIONS. Step 2: Place the sensor on the area covered with gel or water and on the UPPER portion of the cylinder, where you are certain the sensor will be above the liquid level.
Page 145
Below liquid level Note: Some cylinders are prone to giving ‘false levels’, that is the bar graph may disappear after moving the sensor even though the liquid level has not been passed. If you think you may have found a false liquid level, simply move the sensor slightly to the left or right to check if the bar graph returns, as false levels can usually be caused by irregularities in a specific part of the steel wall.
Page 146
PROCEDURE 2: TO BE USED WHERE HIGHER READINGS ARE FOUND IN THE LIQUID PHASE COMPARED TO THE GAS PHASE. Step 1: Place a strip of gel or water down the vertical strip on the cylinder you chose during CYLINDER PREPARATIONS. Step 2: Place the sensor on the area covered with gel or water and on the LOWER portion of the cylinder, where you are certain the sensor will be below the liquid level.
Page 147
Above liquid level Note: Some cylinders are prone to giving ‘false levels’, that is the bar graph may disappear after moving the sensor even though the liquid level has not been passed. If you think you may have found a false liquid level, simply move the sensor slightly to the left or right to check if the bar graph returns, as false levels can usually be caused by irregularities in a specific part of the steel wall.
Troubleshooting ISSUE 1: “The readings on the main unit did not change when the sensor was placed anywhere on the cylinder” SOLUTION: Firstly, ensure you are using an adequate quantity of water or gel to couple the sensor to the cylinder and also ensure that the rubber pad of the sensor is clean and undamaged.
Page 149
Step 7: Now move the sensor to the side of the level where the HIGHEST readings are found (full LED bar graph should appear) and work towards the opposing side in steps until the LED bar graph reduces and disappears. Step 8: Move sensor back along cylinder in smaller steps until the bar graph rises again.
Page 150
Step 4: Engage SPA. This will boost power to the sensor and is useful for dealing with dirty, rusty or damaged cylinders. SPA will not correct the problem however if the battery power is already low. Step 5: Trial ‘Reverse Calibration Procedure’ explained in the solution of ISSUE 3 above.
Page 151
ISSUE 5: “Debris/gel in sensor housing has caused the sensor mechanism to stick making it hard to use.” SOLUTION: The sensor can be removed from its housing and cleaned following the below steps. Step 1: With a damp cloth, clean the various parts, taking extra care on the sensor itself, the spring, and the inside of the housing/around the magnet.
Page 152
ISSUE 6: “The display values do not rest at or close to 0 (a value of 5 or more) after the unit it turned on, with the sensor connected but not placed on a cylinder.” SOLUTION: This is a technical fault with the sensor, please contact technical support service@sdifire.com...
Frequently Asked Questions Why do I have to use water or gel with the sensor? The use of water or ultrasonic couplant gel is essential to the operation of a Portalevel ® MINI unit. When the sensor both emits a high energy pulse and listens for the retuning echoes, excellent mechanical contact must be maintained between the senor and the container in order for the ultrasonic signals to pass into and out of the container efficiently and without interference.
Page 154
Why can’t I test CO and FE-13 in high temperatures? When testing some liquids, it is vital that testing is done under atmospheric temperatures lower than their critical temperature. At the critical temperature of a liquid, it transforms into a vaporous state in which a liquid level no longer exists inside the container to be measured.
3. Accessories Extension Rod Sensor Ultrasonic Gel Portalevel® MINI A Portalevel ® MINI shown in its carry case. Ultrasonic gel, sensor and extension rod extension rod sold separately Ultrasonic Gel The ultrasonic gel or couplant can be used instead of water if the user intends to test a cylinder for a reasonably long period of time, if the ambient temperature is very hot, or if the site does simply not allow open containers of water (data server rooms for example).
Portalevel® MINI Sensor Portalevel® MINI sensors are single crystal piezoelectric sensors which both emit and receive the ultrasound that the Portalevel® MINI uses to detect a liquid level. Whilst the sensors are designed to be rugged, they must be properly maintained in order to maximise their lifetime.
Page 157
Mode Button The instrument must first be turned on by pressing the ‘measure’ button before the mode can be changed. 1 Press - Minimum temperature: Pressing the mode button once, confirmed by then pressing the measure button will measure the minimum temperature when the measure button is pressed again.
Maintenance of the equipment is simple and only requires the cleaning of the sensor. For any major works or re-calibrations the unit can be returned to SDi for repairs. Sensor Care The ultrasonic sensor is the most delicate part of a Portalevel® MINI ®...
Battery Replacement You will need a Philips screwdriver and 4x AA 1.5V Batteries. Step 1: Turn unit onto its front, gently slide off battery hatch. Step 2: Replace batteries with 4x AA 1.5V batteries, taking care to match the correct orientation/polarities of the batteries into the slots.
5. Training This section is for new or untrained users, or for users who may need to remind themselves of how to operate a Portalevel® MINI ® The positions of the particular controls for the Portalevel® MINI ® are shown at the beginning of the manual in section 1.2, page 4, and the names and functions of Portalevel®...
Page 161
The BENCH Test: Step 1: Connect the extension rod to the Portalevel® MINI ® unit and turn the unit Step 2: Choose a vertical strip of your large container which is as continuously smooth as possible, free of chips, rust or any other kind of damage. Apply water or the ultrasonic gel along the length of this chosen vertical strip.
Page 162
Step 7: As you progress up the wall of the container, the numerical readout on the Portalevel® MINI ® unit will change, and the bar graph might fluctuate slightly. However, the bar graph will only disappear entirely once you place the sensor on the container wall above the liquid level.
Page 163
PORTAGAUGE®
3
OPERATING
I NSTRUCTIONS
Operation:
Plug
t he
s ensor
i nto
t he
P ortagauge®
3 ,
a ligning
t he
r ed
m arks.
Turn it on by pressing the ON switch Place a small amount of ultrasonic gel on to the sensor face Press the sensor face with medium pressure onto the 5mm calibration...
Page 164
Replacement necessary when “low bat” shows on display Battery:
Lever off both screw covers on bottom of unit, unscrew and replace with 9V battery Testing
P ipework:
F or
b est
r esults
k eep
c entral
d ivision
o f
p robe
a t
9 0
d eg
t o
axis
...
Page 166
Contents Introduction and Key Features 1.1 What does the Portagauge ® 1.2 The Portagauge ® 1.21 The Portagauge ® 4 – Unit and Measuring Screen 1.22 The Portagauge ® 4 – Definition of Menu Items How does a Portagauge ® 4 work? Operating Instructions Introduction...
1. Introduction and key features What does the Portagauge ® The Portagauge 4 is a portable, robust device used to accurately measure the ® thickness of a material using multiple echo, ultrasonic technology. The Portagauge 4 is ® capable of working across a large number of different applications and environments, whether this be different surface conditions, the presence of rust, small curvature surfaces or for any other maintenance and inspection requirement.
The Portagauge ® Lemo Connector: Allows for secure Display: Color LCD fastening of the cable for display for access to the the sensor probe. Measuring screen and Menu Right Option Key: screen. Selects the option at Left Option Key: the bottom right of the Selects the option at the LCD screen bottom left of the LCD...
Page 169
Definition of Menu Items iProbe Shows that an iProbe has been connected. The probes are — automatically recognised by the gauge and all the relevant probe data is transferred to the gauge to perfectly match the probe to the gauge for optimum performance. All materials have varying velocity of sound properties and —...
4 work? How does a Portagauge ® All sound is physical vibrations which move through a material, and in the case of the sound we are used to, the vibrations move through air. Ultrasound is the same as the sound we experience, but its frequency is beyond that which humans are capable of hearing.
2. Operating Instructions DO NOT PROCEED BEFORE READING SECTION 1. There are two basic procedures which must be carried out when using the Portagauge ® 1. Preparation: Calibration ensures that the Portagauge ® equipment functions properly for the specific metal being tested. 2.
PREPARATION Preparation for testing is simple, quick and ensures you get the most reliable and accurate results from your testing. It ensures the equipment is assembled correctly and that the appropriate probe is selected. Calibration ensures that the unit and probe work correctly for the specific metal being tested.
Probe Membranes All probes are fitted with protective membranes to help prolong the life of the probe. The membranes also provide a flexible interface which acts as an aid when working on rough surfaces. Membrane wear should be monitored and checked at regular intervals.
Calibration The gauge should be calibrated to the type of material that is being measured. This is because ultrasound travels at different speeds in different materials. There are two methods of changing the calibration. Either adjust the velocity of sound setting according to the metal being measured, or adjust the gauge to a known thickness of metal.
Speed Table The given values are in certain cases strongly dependent on the alloy, heat treatment, manufacturing and processing and are therefore only approximate standard values. They are given here as a guide only. Material Speed – m/s Aluminium 6200 – 6360 Rolled 6420 Duraluminium...
Page 176
STEP 4: Turn on the gauge using the ON/OFF button (see page 4). Note: If a compatible probe other than an SDi probe is attached to the unit, the gauge will automatically detect this and display the message “Unrecognised Probe”. If this occurs press the “OK”...
Page 177
Note: If the message “No Probe connected” appears, either connect a suitable probe or simply press the “OK” button to enable the navigation of the menus without a probe fitted. No measurements can be taken if a probe is not detected. The Coating Plus+ feature is automatically switched off when the unit is turned off.
Never try to repair a probe or try to remove scratches from the face of the probe. This may result in further damage. Always return it to SDi for investigation and possible repair.
STEP 5: Refit the retaining ring (3) and secure it in place using the membrane key (5). STEP 6: Place a few drops of membrane oil (6) onto the face of the probe (4). Do not use too much. STEP 7: Screw the knurled ring (1) back onto the probe (4) whilst applying pressure on the membrane (2) with the thumb to expel any air from behind the membrane.
Battery Replacement You will need a small flathead screwdriver and three 1.5V AA batteries. STEP 1: Ensure the unit is off and turn the unit so that the front fascia faces downwards. STEP 2: Take a small, flat head screwdriver and unscrew the two screws holding down the hatch of the battery compartment.
Page 182
Contents Introduction and Key Features What does the Portasteele ® The Portasteele ® How does a Portasteele work? ® Preparation Tablet Setup Testing Calculator Mode Calculating Mass Measurements Calculating Liquid Level Measurements Performing Calculations Database Mode Adding a New Cylinder Adding a New Folder Managing Saved Cylinders Editing Saved Cylinders...
1. Introduction and key features What does the Portasteele ® The Portasteele ® is an application which is readily installed on a rugged, secure tablet.The Portasteele uses information input by the user, including the dimensions ® of a cylinder and its liquid level, to calculate the mass of agent contained within a cylinder.
Page 184
Portasteele ® Micro USB port: cable may be connected to charge the device or to connect to a PC to export data from the Portasteele ® application. Power Button: Powers the device on and off when held down. Ruggedized Casing: Casing designed to be hard-wearing and shock resistant with a rating of...
How does the Portasteele work? ® The Portasteele uses measurements of the dimensions of a cylinder, such as height and circumference, to calculate the volume of the space inside a cylinder. From there it can use the value of the liquid level, also measured by the user, to find out how much of this volume is occupied by the agent inside the cylinder.
2. Preparation Before using the Portasteele it is important to ensure that the tablet is set up ® correctly. Tablet Setup The tablet should be fully charged on receipt. A charger will accompany the tablet on arrival. The tablet can be charged via PC or electrical outlet and the charger is plugged into the micro-USB socket as shown on page 4.
Testing After the initial setup of the tablet, it should be tested to ensure that the Portasteele ® tablet is functioning properly and that calculation results are correct. Begin by launching the Portasteele ® application, which is shown by its icon on the home screen of the tablet.
3. Calculator Mode Calculator Mode is used to quickly input measurements for a single cylinder, and receive the calculated result for either Actual Agent Mass or Expected Liquid Level, depending on what has been input. When the application boots it will open onto its home screen. To begin carrying out calculations, select the Calculator option by tapping on the upper half of the screen as highlighted...
Page 189
Selecting the Calculator button will open up Calculator mode. Start by using the drop down menus to select the agent type. To do this, tap on the drop down arrow next to agent, which can be seen here Then select the desired agent from the drop down menu by tapping on the correct option.
Page 190
Follow the same procedure to select the desired unit type (metric or imperial). There is a similar drop down menu which can be used to change between calculating Actual Agent Mass and calculating Expected Liquid Level, as shown here. Once these options have been selected, the measurements for the cylinder need to be entered into the calculator.
Part 1 – Calculating Mass Measurements To calculate mass, the calculate type must be set to Actual Agent Mass – ensure this is correct before proceeding. The measurements which need to be inputted into the calculator are slightly different depending on the agent in question. Please ensure you refer to the correct section for the cylinder you are measuring.
Page 192
The measurements which need to be taken are: Height, this needs to be measured from the base of the cylinder to the point at • which the cylinder begins to curve, shown as in the diagram overleaf. This must be measured in centimetres for metric units and inches for imperial units. This needs to be measured to the nearest 0.1 cm or nearest 0.1 ”.
Page 193
Cylinder The grey dashed line shows the point at which the cylinder begins to curve, it is important that measurements are correctly taken relative to this point. The red line shows the top of the cylinder, the dome height should not include any components of the valves or tubing at the top of the cylinder.
Page 194
NOVEC 1230 The measurements which need to be taken are: Fill Weight, this is the mass of agent with which the cylinder was originally filled, • and should be printed on the side of the cylinder. This must be measured in kilogrammes for metric units and pounds for imperial units.
Page 195
NOVEC 1230 Cylinder The grey dashed lines shows the points at which the cylinder begins to curve, it is important that measurements are taken correctly relative to these points. The red line shows the top of the cylinder, the dome height should not include any components of the valves or tubing at the top of the cylinder.
Page 196
FM-200 The measurements which need to be taken are: Height, this needs to be measured from the base of the cylinder to the point at • which the cylinder begins to curve, shown as in the diagram overleaf. This must be measured in centimetres for metric units and inches for imperial units. This needs to be measured to the nearest 0.1 cm or nearest 0.1 ”.
Page 197
FM-200 Cylinder The grey dashed lines shows the points at which the cylinder begins to curve, it is important that measurements are taken correctly relative to these points. The red line shows the top of the cylinder, the dome height should not include any components of the valves or tubing at the top of the cylinder.
Part 2 – Calculating Liquid Level Measurements To calculate liquid level, the calculate type must be set to Expected Liquid Level – ensure this is correct before proceeding. The measurements which need to be inputted into the calculator are slightly different depending on the agent in question. Please ensure you refer to the correct section for the cylinder you are measuring.
Page 199
The measurements which need to be taken are: Actual Agent Mass, this is the mass of agent with which the cylinder was • originally filled, and should be printed on the side of the cylinder. This must be measured in kilogrammes for metric units and pounds for imperial units. This needs to be measured to the nearest 0.1 kg or nearest 1 lb.
Page 200
Cylinder The grey dashed line shows the point at which the cylinder begins to curve, it is important that measurements are taken correctly relative to this point. The red line shows the top of the cylinder, the dome height should not include any components of the valves or tubing at the top of the cylinder.
Page 201
NOVEC 1230 The measurements which need to be taken are: Fill Weight, this is the mass of agent with which the cylinder was originally filled, • and should be printed on the side of the cylinder. This must be measured in kilogrammes for metric units and pounds for imperial units.
Page 202
NOVEC 1230 Cylinder The grey dashed lines shows the points at which the cylinder begins to curve, it is important that measurements are taken correctly relative to these points. The red line shows the top of the cylinder, the dome height should not include any components of the valves or tubing at the top of the cylinder.
Page 203
FM-200 The measurements which need to be taken are: Actual Agent Mass, this is the mass of agent with which the cylinder was • originally filled, and should be printed on the side of the cylinder. This must be measured in kilogrammes for metric units and pounds for imperial units. This needs to be measured to the nearest 0.1 kg or nearest 1 lb.
Page 204
FM-200 Cylinder The grey dashed lines shows the points at which the cylinder begins to curve, it is important that measurements are taken correctly relative to these points. The red line shows the top of the cylinder, the dome height should not include any components of the valves or tubing at the top of the cylinder.
Performing Calculations Once all the necessary measurements have been taken to calculate either mass or liquid level, they can inputted into the calculator. To input measurements, tap in the space above the line next to the corresponding measurement value to be entered.This will bring up the on-screen keyboard, which can then be used to input the measured value.
Page 206
Once all the measurements have been correctly inputted into the calculator, close the keyboard and tap the ‘Calculate’ below the input boxes. If something has been input incorrectly, an error message will appear at the bottom of the screen, detailing what it is that must be changed for the inputs to be valid.
4. Database Mode Cylinder Database The application also has the ability to create and organize records of cylinder dimensions and calculation results. This means that when cylinders are tested again, the record of their dimensions can be simply opened in the app, rather than having to measure all the cylinders each time they are tested.
Adding a New Cylinder to the Database The first time the database is loaded, it will contain only a single example cylinder. New saved cylinders for as many different cylinder types and dimensions as is needed can be added and stored in the database for later use. To add a new cylinder to the database, start by clicking the + button in the bottom right, as shown here.
Page 209
The New Cylinder creation screen is very similar to the Calculator Mode as detailed on page 8.The specifications and dimensions of the new cylinder should be input in exactly the same way as in Calculator Mode, the only differences being that the new cylinder being created must be given a unique name to identify it in the database, and that no measured liquid level or mass needs to be inputted, since no calculation is being performed at this stage.
Adding a New Folder to the Database In addition to adding cylinders to the database, new folders and sub-folders can also be created.Within every folder, new cylinders and sub-folders can be created, allowing cylinders to easily be organized. To add a new folder to the database, start by clicking the + button in the bottom right, as shown here.
Page 211
A dialog will now pop up to name this new folder. Simply click in the input box to bring up the on-screen keyboard, and use it to enter a unique name for this new folder. Selecting the Create button will then add this new folder to the database.Alternatively, select the Cancel button to abandon creating this new folder.
Managing Saved Cylinders The first time the database is loaded, it will contain only a single example cylinder. Saved cylinders can be managed from their drop down ribbon, located to the left of the cylinder image. The first time the database is loaded, it will contain only a single example cylinder.
Editing Saved Cylinders The first cylinder management option of the drop down ribbon allows the saved dimensions and information of a cylinder to be edited. Start by clicking on the first item in the drop down ribbon as shown here.This will bring up the cylinder editing screen.
Exporting Saved Cylinders The second cylinder management option of the drop down ribbon allows the saved dimensions and information of a cylinder to be exported via email. Start by clicking on the second item in the drop down ribbon as shown here.This will bring up the email client if an email account has already been setup, otherwise it will prompt you to do so.
Deleting Saved Cylinders The final cylinder management option of the drop down ribbon allows a cylinder to be completely removed from the database. Start by clicking on the third item in the drop down ribbon as shown here.This will then prompt the application to check that you really wish to delete the selected cylinder.
Navigating the Database After creating new folders, they can be opened and dealt with in the same way as the home folder. When the database first opens it will be in its home folder. Clicking on a folder will open the selected folder and navigate into this sub-folder.
Page 217
To navigate upwards in the file system, use the back button or the back arrow in the top left of the screen in order to navigate a single level upwards. In order words, this means you will be navigating into the folder containing the current folder you are within. Navigating upwards within the home folder will save and close the database.
Performing Calculations on Saved Cylinders After creating saved cylinders, calculations can be performed using this cylinder information. To perform a calculation on a cylinder, simply click on the selected cylinder.This will open up the saved cylinder calculator. This saved cylinder calculator is very similar to Calculator mode as detailed on page 8, except that the only 2 inputs which can be altered are Temperature and either Actual Liquid Level or Actual Agent Mass,...
Page 219
Once all the measurements have been correctly inputted into the calculator, close the keyboard and tap the ‘Calculate’ below the input boxes, in the same way as for Calculator mode, as detailed on page 26. Also in the same way as in Calculator mode, if something has been input incorrectly, an error message will appear.
Page 220
Managing Calculation Records After storing cylinder calculation results, they can be viewed, managed and exported from Database Records. To open up Database Records, simply click the Records icon as seen here. The Database Records organize calculation results by cylinder, as can be seen here.To view the results for a certain cylinder, simply click on the selected cylinder’s name.This will open up a drop down list of results.
Page 221
Opening up the records for a cylinder will show all of the stored records for that cylinder, organized by how new the record is, as seen here. A record can be removed by clicking the delete icon, as shown here.This will permanently remove the selected result from the records.
Page 222
It is also possible to remove every record for every cylinder in Database Records, by clicking the Clear All button, as shown here.
Page 223
Exporting Calculation Records It is possible to export all of the calculation results stored in Database Records via email with a single click. Start by clicking the Export All button, as shown here. This will bring up the email client if an email account has already been setup, otherwise it will prompt you to do so.
5. Maintenance Basic maintenance of the tablet involves simple procedures such as recharging the battery. Any major maintenance is carried out by the technical team. Please contact service@sdifire.com for any queries.
Page 227
The sensing face can be degreased as necessary using a mild detergent. Do not use aggressive chemicals. 3.2. Service and Calibration SDi recommends that the BEARING INDICATOR is serviced, checked and calibrated regularly through an authorized SDi service provider. The recommended recalibration period is 12 months.
Page 228
4. Safety Summary The sensing face of the BEARING INDICATOR contains a very strong magnet and should never be positioned in the vicinity of sensitive electrical/electronic equipment (for example; pacemakers or other sensitive medical equipment) or magnetic storage media such as hard disk drives. Do not carry out any instruction in this manual if it may cause risk of physical injury, damage to machinery, would be against site rules or could adversely affect...
Page 229
6. Operation 6.1. Switching BEARING INDICATOR ON and OFF Press the button to turn the unit on. The BEARING INDICATOR briefly displays its serial number and software version. The last measurement taken is then displayed. The BEARING INDICATOR switches off automatically after 30 seconds of inactivity.
Page 230
7. Taking Measurements 7.1. Mounting location The sensing face needs to be acoustically coupled to a static part of a machine’s external casing containing the bearing to be monitored. Positioning and orientation of the sensor is not critical although highest signal levels will be detected in the immediate vicinity of the bearing - avoid flimsy cowls, guards and nameplates.
Page 231
7.3. Displaying Readings After the countdown, the BEARING INDICATOR takes measurements for a further 10 seconds; the Measurement in progress light will flash. Measured values of Distress® and dB are then displayed. 7.4. How to get the best from the BEARING INDICATOR Do not move the BEARING INDICATOR during measurements and ensure nothing is touching the BEARING INDICATOR as this can affect the readings.
Page 232
9. Typical Interpretation of Readings from Bearings For progressive bearing deterioration due to pitting of race or rolling element surfaces in rotating machinery, the usual observation of readings is as follows: ® Distress Wear Stages dB Level Quality Numeric “OK” <10 1 Good condition Lowest for...
Page 233
Not all fault conditions follow the above progression. For example high frictional activity from the outset would give a high dB and possibly a low Distress®, whereas a bearing spinning in its housing usually gives a high Distress® and an abnormally low dB. In addition, improving bearing lubrication – especially in the early stages, often gives a reduction in Distress®...
Page 234
BEARING INDICATOR complements existing methods and procedures for dealing with service, maintenance and repair of machinery. Since SDi has no control over the machinery to which the BEARING INDICATOR will be applied, the care taken during its use, or the appropriateness of any actions taken, SDi cannot accept any responsibility or liability for any consequential loss howsoever arising from the use of the BEARING INDICATOR.
Page 235
SDi Fire Service Department service@sdifire.com (732)-751-9266 Instructions for Calculation of Mass or Liquid Level of Agent Before beginning any calculations, please note whether the bottom of the cylinder is similar to Cylinder 1 or Cylinder 2. This is important because the set of instructions for the calculations depend on the type of cylinder.
Page 236
SDi Fire Service Department service@sdifire.com (732)-751-9266 To convert a value in metres (m) to centimetres (cm), multiply the value by 100. (1m=100cm). To convert a value in millimetres (mm) to centimetres (cm), divide the value by 10. (1cm=10mm). To convert a value in grams (g) to kilograms (kg), multiply the value by 1000. (1kg=1000kg).
Page 237
SDi Fire Service Department service@sdifire.com (732)-751-9266 The measurements used in the example are displayed in the table below. 85 cm 0.7 cm a) Calculate the radius: circumference divided by 6.28 Write down: r = 13.5 b) Subtract the value of the thickness from the radius Write down: (b) = 12.8...
Page 238
SDi Fire Service Department service@sdifire.com (732)-751-9266 1. Mass of the full cylinder, w in kilograms, found by weighing the cylinder that has the agent in 2. Mass of the empty cylinder (i.e. when it has no agent in it), e, in kilograms. This is often printed on the side of the cylinder or is available from the manufacturer.
Page 239
SDi Fire Service Department service@sdifire.com (732)-751-9266 a) Calculate the radius: circumference divided by 6.28 Write down: r = 13.5 b) Subtract the value of the thickness from the radius Write down: (b) = 12.8 c) Multiply the number calculated in step (b) by itself Write down: (c) = 163.8...
Page 240
SDi Fire Service Department service@sdifire.com (732)-751-9266 1. Liquid level, l, this should be measured from the base of the cylinder in centimetres. (The liquid level can be measured using a Portalevel) 2. Circumference of the cylinder, c, in centimetres. This is the distance all the way around the cylinder.
Page 241
SDi Fire Service Department service@sdifire.com (732)-751-9266 85 cm 1400 61 cm a) Calculate the radius: circumference divided by 6.28 Write down: r = 13.5 b) Multiply the number calculated in step (a) by itself Write down: (b) = 182.3 c) Identify the correct value of density to use from table (page 10). For example, a cylinder of FM-200 at a temperature of 20 ⁰C has d = 1400...
Page 242
SDi Fire Service Department service@sdifire.com (732)-751-9266 2. Mass of the empty cylinder (i.e. when it has no agent in it), e, in kilograms. This is often printed on the side of the cylinder or is available from the manufacturer. 3. Circumference of the cylinder, c, in centimetres. This is the distance all the way around the cylinder.
Page 243
SDi Fire Service Department service@sdifire.com (732)-751-9266 Write down: r = 13.5 b) Multiply the number calculated in step (a) by itself Write down: (b) = 182.5 c) Identify the correct value of density to use from table (page 10). For example, a cylinder of at a temperature of 68⁰F has d = 1400...
Page 249
Contents Contact information and support 1. Introduction 2.General use and safety precautions Safety precautions when charging Safety precautions related to connectors 3. Getting started 4. Using the VOX-01 Quick STIPA STIPA pro SPL meter Real-Time Analyzer (RTA) Fast Fourier Transform Anazlyzer Reverberation Time Meter (RT60) LAEQ Logging Oscilloscope...
1. Introduction Congratulations on purchasing the VOX-01 STIPA meter, the most advanced instrument for measuring the Speech Transmission Index ever built, designed and manufactured by the inventors of STIPA. The VOX-01 is also a fully compliant Class 2 / Type 2 Sound Pressure Level meter, a 1/1 and 1/3 octave Real-Time Analyzer, a Fast Fourier Transform Analyzer, a Reverberation Time Meter, an Oscilloscope, and much more.
2. General use and safety precautions The VOX-01 is a sophisticated electronic measuring instrument that should be used, maintained and stored with care: • We recommend that you always use a lanyard to secure your device against dropping, by suspending it around your neck or wrist. If you use a tripod to mount the device during measurements, we recommend that you use a sturdy model with a sufficiently wide base.
The supplied charger includes adapters for use worldwide, and operates on a wide AC-voltage range of 100 – 240V . Do not use the charger if the casing appears cracked or broken, or if it has been exposed to fluids or moisture. Contact with components inside the charger while this is plugged in may result in serious injury or death.
3. Getting started You should receive your VOX-01 with sufficiently charged batteries to start the device and get acquainted with its features. We do recommend that you fully charge the device using the supplied charger before starting your first real measurement session. The device is powered on by pressing the red power button once.
The Embedded Acoustics reference STIPA test signal is included with the VOX-01 (on a USB flash drive) and can also be downloaded from the SDi website, free of charge. This is a fully IEC-60268-16 rev.4 compliant test signal, compatible with all STIPA meters that conform with the standard.
Page 255
• Tap “Start” on the VOX-01 • Wait for 25 seconds while the measurement takes place. Make sure that signal playback is not interrupted, and that the acoustic environment is stable and free from impulsive sounds (such as door slams). •...
Page 256
• Choose between presentation of the STI in large digits or as a bar plot • Choose whether or not you want to see the qualification bands displayed The “browse measurements” screen allows you to review earlier (saved) measurements without having to download your data to a PC first. •...
STIPA pro The STIPA pro module is also used to measure the Speech Transmission Index, but unlike the quick STIPA module, it also comprises features that allow you to analyze and manipulate your measurement data in greater detail. “STI and spectrum” and “measurement details” screen in the STIPA pro module The STIPA pro module has five screens: STI and spectrum : display measured STI, octave spectrum and dBA level Measurement details: display all measurement details including the MTF...
Page 258
The “measurement details” screen shows all the diagnostic details of an individual STI measurement: • OCT: The octave center frequency (Hz) • REL: The octave band levels relative to the STIPA signal spectrum (this is effectively the frequency transfer function of the tested channel in octaves). •...
Page 259
There are two ways to include the influence of background noise in your STI measure- ment: • Simply do your measurement in the actual noise environment. The STI method is designed to incorporate the effects of any noise present during the measurement representatively in the measured STI.
SPL meter The VOX-01 is also Sound Pressure Level meter (also known as SPL meter or SLM) that complies with IEC 61672 Class 2 and ANSI S1.4 Type 2. In fact, the VOX-01 electronics far exceed the requirements for a Class 2/Type 2 device. The layout and buttons of the SPL meter app are similar to the STIPA module, featuring three screens: SPL measurements, settings and browse measurements.
Page 261
The settings screen in the SPL module allows you to set the measurement time. Once you press start, the measurement will run for as many seconds as you choose here. You may also choose to let measurements run continuously; you then determine the measurement time manually, by choosing the moment you press “stop.”...
Real-Time Analyzer (RTA) The RTA module has two main modes of operation: as a 1/1 octave band analyzer (span- ning the octave bands from 31 Hz to 16 kHz) and as a 1/3 octave band analyzer (25 Hz through 20 kHz). The operation is generally similar to the SPL module, but instead of just the overall signal level, a frequency analysis (into 1/1 or 1/3 octave bands) is also shown.
Page 263
By pressing the “save” ( ) button in the main measurement screen, you store the current measurement on the devices internal storage, using the measurement number indicated above the battery indicator. Some remarks about the RTA module: • By definition, “max hold” and “EQ” are mutually exclusive settings. •...
Fast Fourier Transform Analyzer The FFT module allows you to perform spectral analyses with a much higher frequency resolution that the RTA module. This is very useful if you are looking to identify the exact frequency and level of certain signal components, or if you want to study the harmonic structure of sounds.
Page 265
The minimum and maximum frequency that can currently be displayed are indicated at the top of the screen, as well as the exact frequency at the cursor position and the level at the cursor position. The second screen of the FFT module allows you to set an integration time for the FFT analysis.
Reverberation Time Meter (RT60) The reverberation time (or RT60) is among the most commonly specified acoustic measures. Unfortunately, reverberation time measurements are also among the most complex to carry out accurately. RT60 measurements are prone to errors, which cannot always be detected by the measuring device. If you are new to measuring reverberation times, we recommend that you read this section of the manual carefully, and preferably also consult external literature and training materials on reverberation time measurements.
Page 267
• Start the RT60 module and press “arm” • To trigger the measurement, shut off the noise source. Once the device has been armed, it will start measuring the ambient noise field. At this stage the device is continuously sampling the audio, but the RT60 measurement itself has not yet been started.
Page 268
Once the measurement is completed, the energy decay curves are plotted. Use the arrows to cycle through the frequency bands (note that you can choose whether to measure in 1/1 or 1/3 octave bands through the “settings” screen). RT60 decay curves plotted once the measurement is completed Inspect the decay curves to visually verify that the measurement is correct.
Page 269
Cycle through EDT, T20 and T30 using the button on the right. Note that each of these three metrics is always calculated and saved for each measurement, even if only one metric at a time is displayed on the screen at a time. RT60 measurement results.
Page 270
The capability of averaging measurement data adds complexity to the user interface, but averaging is sometimes essential to arrive at valid end results. Note that averaging takes place on the level of the individualy measured data points along the decay curve (so not just across individually measured reverberation times).
Page 271
• A spherical loudspeaker configuration playing pink or white noise is the preferred source of measurement noise. • If you are using a single, non-spherical loudspeaker, avoid placement of the microphone on the loudspeaker axis. Even though the microphone is omnidirectional, avoid pointing the microphone straight at the loudspeaker.
LAEQ Logging The LAEQ logging module is intended for long-term monitoring and logging applica- tions. When running this module, the VOX-01 keeps track of how the equivalent- continuous A-weighted sound pressure level (LAEQ) develops over time. This is displayed on the device and also logged, for download to a PC later on. This module can be used for several purposes: •...
Page 273
The LAEQ logging module has a start/stop button similar to all other modules, but un- like other modules, it has no “save” button. Since this module is specifically intended for logging purposes, all data is always saved to a text file. Whenever a new measurement is started, the measurement number is automatically incremented and a new file is opened.
Page 274
Screen 2 of the LAEQ module, showing a graph of how the average LAEQ is developing. In cases where the operator has control over the measured sound level (i.e. through control of a PA installation), and is using the VOX-01 to make sure that the level is kept below the limit, the graph gives a useful insight into how much “room”...
Page 275
Browsing through measurement data Browsing through measurement data works a little differently than in other modules, since the amount of data associated with a measurement file can be very large. In many cases, it will not be practical to inspect the data from the device itself, and data will be downloaded to a PC first.
Oscilloscope The oscilloscope module is intended to quickly inspect acoustic signals, in particular peri- odic signals, in the time domain. Sine wave with a frequency of 1000Hz as seen through the oscilloscope module Directly from the first screen of the module (see above), you can control the horizontal scale (time base) and vertical scale of the graph, as well as the trigger mode.
USB audio device mode When starting this module, a special mode of the device is activated, in which it performs no other tasks than simply streaming audio to a PC, applying the VOX-01 level calibration. This is done in the form of a single-channel stream, at a fixed bandwidth of 48 kHz and a bit depth of 32 bits.
Settings The settings module will allow you to configure the generic settings of the device, in addition to the module-specific settings found in each module. The following attributes can be set: • Time and date • Time interval before the backlight automatically switches off to save power •...
Page 279
Microphone pre-amp settings (left) and power saving settings (right) The internal microphone amplifier of the VOX-01 features an adjustable gain setting. Adjust Adjustment of the pre-amp gain is hardly ever necessary, since the system has a very wide (linear) dynamic range of 130 dB. However, if you wish to connect an external signal source which operates at lower signal levels than the VOX-01 microphone, you can choose adjust the pre-amp gain by (approximately) +20 dB.
Calibration The VOX-01, the instrument itself as well as the microphone, is calibrated before leaving the factory. A calibration certificate is included with the device. However, it is recommended that you carry out your own “user calibrations” from time to time, especially under the following conditions: •...
Page 281
Calibration module showing calibration results. The new calibration factor (sensitivity adjust) of the device is shown, as well as the original calibration factor as currently used in the device. The differences between the old and new value are typically small (less than one dB), unless you have connected a different signal source or microphone than the device is currently calibrated for.
Status System status information can be read using the status module If you contact SDi for technical support, you may be asked to provide some of the information displayed here (such as the firmware version and hardware revision). The status app is not used to make changes to the system.
Page 283
Remote control from a PC In certain situations you may wish to control your VOX-01 remotely from a PC. To achieve this, keep the VOX-01 connected via USB, and run the “Remote Display” application under Windows. This application works on any 64-bit version of Windows. A self-extracting installer of the application is included on the flashdrive bundled with your device, and can also be downloaded from our website.
5. Power and charging Charging instructions We recommend that you charge your VOX-01 only with the USB charger supplied with the instrument. Alternatively, you can charge the device through any external USB wall charger (“wall wart”) with a charge current of 1500 mA or higher. Some PCs allow increased charge currents to be drawn from their USB ports (sometimes through special ports marked “Boost”...
• It is normal for the total battery capacity to slowly decrease over time. After 500 charge/recharge cycles, the battery capacity is normally at 80% to 90% of its original capacity, depending on the way the device is used and recharged. •...
6. Cleaning, maintenance and calibration With normal use, no further maintenance is needed beyond normal cleaning periodic re-calibration of the device. We recommend that you have the device checked and calibrated at our factory or at an external calibration specialist at least once every 2 years. We recommend having the battery pack replaced after approx.
If you did not purchase directly from SDi, also include dated proof of your purchase. In general, our warranty programme covers all defects except for those resulting from acci- dents, misuse (including improper electrical connections) and improper maintenance.
Page 288
Please do not attempt to open your VOX-01 yourself, as there are no user-serviceable parts inside. If you do see a legitimate need to open the device yourself, please contact SDi first, so we can discuss preserving your warranty rights. Please be aware that the product is protected by tamper-proof seals.
8. Firmware updates Firmware updates are released through the website www.sdifire.com. Before updating your firmware, always check: www.sdifire.com/support/#productsoftwareupdates • If the update you downloaded is intended for your device type (the VOX-01) • If you have sufficient battery charge to complete the update cycle without interruption Interrupting the power supply during the update process may render your device unusable.
10. Troubleshooting Problem Possible cause Solution System does not power on after Battery empty Connect wall charger for at pressing power button least 30 minutes System does not recharge after Charge current too low Use bundled wall charger; connecting through USB standard PC USB ports do not supply enough current to charge...
Page 292
New in v2.0 6.3 Full STI The Speech Transmission Index (STI) can be measured by means of various test methods. The STIPA method, as implemented by the Quick STIPA and STIPA pro modules, is a fast and efficient version of the STI originally developed for public address systems. Although fast and robust, STIPA has its limitations: only 2 modulation frequencies are measured per octave band, whereas the complete Speech Transmission Index measurement scheme calls for 14 modulations per octave band.
Page 293
• A novel test signal is used, which repeats 7 different patterns (segments) in a circular fashion. • Each pattern roughly follows the approach of STIPA: each octave band is simultane- ously modulated with 2 modulation frequencies. • Each pattern takes approx. 9 seconds to complete. The whole measurement is com- pleted in 65 seconds.
Page 294
ended Full STI measurement can be thought of as a compromise in between STIPA and Full STI. Please refer to the section on the STIPA pro module (page 11) for a description of the various features and controls of the module (additive noise, qualifications bands, disabling level-dependent masking, etc).
Page 295
6.4 Speech Level Meter In the context of speech intelligibility measurements, special care is needed when deter- mining the sound pressure level of running speech. As human talkers use their articulators to form speech utterances, they alternately produce high-intensity bursts of acoustic energy and naturally occurring pauses.
Page 296
8.2. Noise curves The “noise curves” module implements a series of measurement methods to evaluate noise levels, based on sets of standardized spectral curves. The objective of these methods is to set criteria in terms of a single-number rating, while taking the spectral characteristics of noise into account.
Page 297
Each method uses its own procedure for rating noise spectra based on the series of curves associated with the method. The so-called “tangency method” plays a central role for NC and related measures; the rating is effectively determined by the highest NC curve “touched”...
Page 298
The second screen of the module is used to set the duration of the measument. Selecting curves (and the associated rating method) is done in the third screen of the module. Choose from the available noise curve metrics VOX-01 Operating Manual...
Page 299
9.2. AC Volt meter The AC Volt meter module is a simple tool to quickly assess the output level of line-level audio equipment. The line output of a device is connected directly to the XLR input of the VOX-01 (after removing the microphone). The VOX-01 then measures the signal level and expresses it in the following three units: •...
Page 302
Please direct your warranty claims to: • Now use a STIPA measuring instrument (such as the VOX-01) to obtain STI readings at all SDi, relevant listener locations. You can choose to have the STIPA signal play continuously, or pause/ 3535 Route 66, Building 6, Neptune, NJ 07753 mute inbetween measurements.
Page 305
Contact information and support The VOX product line, including the VOXBOX TalkBox, is designed and manufactured in the Netherlands and sold by SDi in the United States. SDi can be contacted at the following address: SDi Fire 3535 Route 66...
1. Introduction Congratulations on purchasing your VOXBOX talkbox. The VOXBOX is a versatile and reliable signal source for acoustic measurements, including STIPA and full STI speech intelligibility measurements. Using the VOXBOX is quite easy, thanks to its intuitive touch-screen interface. Most users will get the hang of doing measurements with the VOXBOX without spending too much time with this manual.
Page 310
The VOXBOX features a class II laser pointer to assist in placement of microphones. Never look directly into the laser, and make sure that others cannot accidentally look directly into the laser either. Do not use around children or animals. Do not disassemble or remove the laser from its enclosure.
Page 312
Select any of the tabs on the top-left bar to access any of the four categories of test signals: Speech intelligibility (STIPA), Noise, Sine or Speech. Then choose the test signal which you need for your measurements. Consult chapter 4 of this manual for further explanations on the different types of test signals.
3.5. External power supply The SDi VOXBOX operates off its external 12 V DC power supply. It does not have an internal battery; the device shuts off immediately when the power supply is disconnected. In most use scenarios, the VOXBOX is used near a lectern or paging microphone, in which case a mains outlet (100-250 V AC) will be available nearby.
NOTE: IEC-60268-16 recommends a 3 dB adjustment when matching the level of a STI- PA signal to the level of recorded speech. This adjustment is needed to account for the pauses that occur naturally in speech, while the STIPA signal is fully continuous. This correction is already applied by your VOXBOX, according to the procedure described in Appendix J of IEC-60268-16 rev.
5. Noise, sine waves and sweeps 5.1 Noise The VOXBOX offers two kinds of noise: pink noise and white noise. Pink noise has a flat spectrum on a logarithmic scale (equal energy per octave band), whereas white noise has a flat spectrum on a linear scale (equal energy per FFT bin).
7. Calibration and accuracy The frequency transfer function and sensitivity is measured individually for each VOX- BOX at the factory. Before the start of the calibration process, a 15-hour “burn in” cycle is carried out to improve the stability of the loudspeaker. Based on these measurements, the signals played back by the VOXBOX are amplified and equalized to accurately match the intended level and spectrum.
If you did not purchase directly from SDi, also include dated proof of your purchase. In general, our warranty program covers all defects except for those resulting from acci- dents, misuse (including improper electrical connections) and improper maintenance.
11. Technical specifications Functionality Playback of acoustic test signals via balanced XLR line-out and loudspeaker. Supported signals: STIPA, Full STI, sine waves (125 Hz - 8 kHz), log sweep, pink noise, white noise, speech (male/female, 6 languages: US English, UK English, French, German, Spanish, Dutch).
Page 326
testifire ® M U L T I - S T I M U L U S D E T E C T O R T E S T E R M U T I - S T I M U L U S D E T E C T O R T E S T E R User Manual detectortesters testing technology from No Climb...
Page 327
testifire ® M U L T I - S T I M U L U S D E T E C T O R T E S T E R M U T I - S T I M U L U S D E T E C T O R T E S T E R User Manual This manual provides information for the correct testifire®...
Page 328
Important Information Read this User Manual completely before using your Testifire. Save this User Manual - Save all safety and operational instructions for future reference. Take note of the Warnings - Read carefully and follow all warning labels on the product and those described in this User Manual Water and Moisture - This product is designed for indoor use only and should not be subject to, or used in,...
Page 329
Meanings of Symbols and Terms The following symbols are used throughout this User Manual. This symbol on the product indicates that there is a safety hazard. You must read the appropriate sections of the User Manual to understand the nature and severity of all the potential hazards present and the action you must take.
Safety General Safety Information CAUTION This product is intended to be used at height. Exercise great care and always wear appropriate PPE (personal protective equipment) when operating above head height in order to avoid the risk of injury. DO NOT OVER REACH .
Page 331
Table of Contents Page No. General Instructions Warranty Acknowledgement Recycling Introduction Parts Identification Preparation For Use Charging the Battery Using the Battery Removal and Replacement of the Smoke and Carbon Monoxide (CO) Capsules 9 Turning the Unit On Using the Menus Adjusting the Head Unit Angle Testing High Profile Detectors Using Your...
1. General Instructions 1.1 Warranty In addition to any other express warranty given in writing by the Company in relation to the Goods, the Company warrants that the Goods supplied under these terms and conditions will be in accordance with the specification (if any) contained in the Purchase Order, and will be free from defects in workmanship and material for a period of 18 months from the date of delivery to the Buyer or for a period of 12 months after the date of sale by the Buyer to the final customer whichever period is the shorter.
2. Introduction Thank you for purchasing Testifire Multi-Stimulus Detector Tester. This manual is designed to assist you to get the best and most efficient use of the Testifire 1000 and 2000 model range, and provides all the information required to perform routine service and maintenance tasks with ease.
3. Parts Identification Bellows Replacement Capsules Testifire Smoke Capsule TS3 RFID Antenna** Inner Clear Infrared Beam Main Duct for Heat, Platform Smoke and CO* Testifire CO Clear Capsule TC3* Testifire 100 RFID ® Bluetooth Module** Smoke Capsule ‘Test Type’ LED CO Capsule TC3* ‘Status’...
4. Preparation For Use 4.1 Charging the Battery WARNING Do not attempt to use or charge the battery if either the unit or the battery connection point are damaged. Never connect AC and DC power at the same time. Solo NiMH Battery Batons are used to power Testifire and enable it to be fitted to Solo Access Poles. The Solo 760 Battery Baton must be charged before using Testifire.
CAUTION Store charger in a dry place (indoor use only when connected to AC mains). Danger of fire and electric shock! Do not fast-charge a hot battery, allow the Battery Baton to cool down naturally before starting a charge cycle. Allow the charger to cool down for at least 15 minutes after one fast charge.
4.3 Removal and Replacement of the Smoke and Carbon Monoxide (CO) Capsules CAUTION This symbol on the product indicates that this part of the device is susceptible to static damage. DO NOT TOUCH the contacts on the PCB on the capsule. Static electricity may damage them and contamination of the contacts must be avoided.
CO Capsule Removal Instructions (Testifire 2000 series only) Only replace the CO capsule when it is empty. Testifire should be turned off for five minutes to allow the ‘CO’ capsule to cool down before removing it. Release the lower access cover (Fig.
4.5 Using the Menus Before turning Testifire on for the first time, it is recommended that you become familiar with system menus. Testifire’s menu system is simple to navigate. Using the ‘UP’ and ‘DOWN’ menu navigation keys , you can move the cursor through the ‘MAIN’...
4.7 Testing High Profile Detectors When testing high profile detectors, it may be necessary to remove the inner clear cup to enable the detector to sit in the correct position for a successful test. The inner clear cup has a semi-circular cut-out on the side nearest to the user interface.
5. Using Your Testifire CAUTION This product is intended to be used at height. Exercise great care and always wear appropriate PPE (personal protective equipment) when operating above head height in order to avoid the risk of injury. DO NOT OVER REACH .
Once the stimulus has been selected, the ‘Test Type’ LED will change colour to indicate the following: Red = Heat Test Blue = Smoke Test Green = Carbon Monoxide Test Clear = Clearing The ‘Status’ LED provides feedback on the modes of operation and alerts you to errors or when a smoke or CO capsule is empty or incorrectly fitted.
NOTE: Lowering Testifire during a test will do the following: For less than 2 seconds: Nothing. The test will continue as soon as the detector re-enters the cup. For more than 2 seconds: During a Sequential Test, Testifire will advance to the next test that is programmed.
Page 344
A ‘Hi Heat’ setting for fixed temperature detectors which operate at more than 90ºC (194ºF) is available. It is recommended that for most tests the ‘Heat’ (normal heat) setting is used as this will conserve battery power. To select a ‘Hi Heat’ test, use the ‘UP’ and ‘DOWN’ keys on the keypad to highlight ‘Heat’.
5.5 Carbon Monoxide (CO) Testing WARNING This product emits small amounts of carbon monoxide (CO) gas which is a harmful, odourless gas. DO NOT INHALE directly from the duct during a CO test or within 5 minutes of conducting a CO test Note: Under normal operating conditions, this emission will present no harm to the user.
5.7 Capsule Fuel Level The remaining level of ‘Smoke’ and ‘CO’ fuel available in the capsules can be displayed on the screen by highlighting ‘Smoke’ or ‘CO’ in the ‘MAIN’ menu by using the ‘UP’ and ‘DOWN’ keys on the keypad (Fig.
5.8 Simultaneous Testing Using Testifire to carry out a Simultaneous Test means that a number of operations (Smoke, Heat and CO) can be pre-programmed into the unit before it is raised up to the detector. This saves time, reduces handling and enables the testing of certain multi-sensor detectors if they are so enabled.
Another identical test sequence will automatically be ready to begin at this point, and there is no need to re- select the test on the keypad. Fig. 49 illustrates the procedure for a Sequential Test (in this example for ‘Smoke’ followed by ‘Heat’ and then ‘CO’). Programme Raise Testifire Lower Testifire...
5.11 Infrared Remote Control (Testifire 25) Infrared beam LED Operation indicator LED Some types of detector may not automatically initiate a test as they enter the clear cup of Testifire. The Infrared Remote Control can be used to manually start a test, and/or move on to the next test in the sequence.
5.13 Indicating LED Reference Chart ‘Test Type’ LED ‘Status’ LED Blue Solid Green Slow Flashing Smoke Heat Smoke Test in progress ‘STANDBY’ Mode Menu Status Red Solid Green Fast Flashing Smoke Heat Heat Test in progress ‘OPERATIONAL’ Mode Menu Status (test in progress) Green Solid Red Flashing...
6. Troubleshooting If you experience any type of problem with your Testifire, turn the power off, wait for a few seconds and then turn the power back on. This may clear the fault. If the fault continues, the following information may help you solve the problem.
6.3 Error Messages and Indicating LED Reference Chart Note: If a problem occurs with your Testifire unit, the ‘Status’ LED will flash or remain on in red and the ‘Test Type’ LED will not be illuminated. The error message will be displayed on the screen. If the ‘Status’...
7. Support 7.1 Technical Support For support, service and further information, please contact: No Climb Products Ltd Edison House, 163 Dixons Hill Road, Welham Green, Herts, AL9 7JE United Kingdom Tel: +44 (0) 1707 282760 Fax: +44 (0) 1707 282777 Email: support@testifire.com Online help available at...
7.5 Servicing To ensure that your Testifire unit continues to perform as intended, it is recommended Service Due that the unit is returned annually for a service. Refer to Manual Press any key In some countries, Testifire will require an annual, traceable service in accordance with national or international standards for testing equipment.
7.6 Service Reminders By default Testifire will notify the user when a Service is due or over due. If required the Service reminders can be turned off (Firmware version 5.4 or later). System Info To disable service reminders, press the 'MENU' key on the keypad and use the 'DOWN' Language key to navigate to 'Service' (Fig 61).
7.7 Testifire Flash Loader Software Instructions For Testifire 1000 & 2000 Models Welcome to the Testifire Flash Loader Software Before connecting Testifire to the computer, first install the Flash Loader software. Introduction Testifire Flash Loader is a free Windows based application that allows you to update your Testifire firmware on your Testifire without the need to return it for service.
Page 357
Step 3 Step 6 Select "I don't have the disc. Show me other options." This driver may not yet be digitally signed, however it Windows will now install the Testifire USB drivers. was obtained from either www.testifire.com or the disc Click "Close" to exit the wizard which was supplied with your Testifire then it is safe to use on your computer.
Page 358
Windows XP -Testifire Device Driver Installation Step 1 Step 5 Run the Flash Loader software and when prompted Select "Install from a This driver may not yet be tested by Windows Logo testing, however if it list or specified location" and connect Testifire to the computer using the was obtained from either www.testifire.com or the disc which was supplied then click "Next"...
Page 359
Windows 2000 - Testifire Device Driver Installation Step 1 Step 5 Run the Flash Loader software and when prompted connect Testifire to the computer using the supplied USB cable. The USB drivers are located in the install directory of the Testifire Flash The "Found New Hardware"...
Page 360
Updating Your Testifire Firmware Before connecting Testifire to the computer, first install the Flash Loader software. Step 4 Step 1 The application will now scan the computers USB ports to detect the After installing the Testifire Flash Loader software, start the application devices connected.
Page 361
Step 5 When programming is completed successfully the following screen is shown. Click on the Next button. The next stage is to load the Testifire firmware file into the application. Use the Open button to select a file stored on the computers hard disk or use the Download button to automatically download the latest firmware version from the Testifire website.
8. Technical Specifications Power source Battery Baton: nominal 7.2V 2.2Ah NiMH rechargeable battery pack with internal overcurrent protection connects directly to Testifire with no leads or wires. Must be charged by Solo 726 Battery Charger (using 100-230VAC or 12VDC input). Battery Charge Time Approx.
Weights Testifire 1000: 990g (including Battery Baton) Testifire 2000: 1.23kg (including Battery Baton) Solo 760 Battery Baton: 0.5kg Dimensions Testifire 1000: Bellows width: 153mm max. Head Unit height: 224mm max. (397mm including handle) Testifire 2000: Bellows width: 153mm max. Head Unit height: 273mm max. (397mm including handle) Note: Product specifications may be subject to change without prior notice or obligation on the part of the manufacturer.
10. Agreement for Supply of Products and Associated Software IMPORTANT NOTICE: PLEASE READ CAREFULLY BEFORE EITHER: (A) DOWNLOADING ANY SOFTWARE FROM THIS WEBSITE, OR (B)OPENING THE PACKAGING, OR (C) INSTALLING THE SOFTWARE: This Agreement for Supply of Products and Software (Agreement) is a legal agreement between you (Licensee or you) and No Climb Products Limited (Licensor or we) for (i) the Licensor's Product and (ii) all related software and all future upgrades thereto (Software).
Page 366
3. INTELLECTUAL PROPERTY RIGHTS You acknowledge that all intellectual property rights in the Software and the Licensor's Products and the Documentation anywhere in the world belong to the Licensor, that rights in the Software are licensed (not sold) to you, and that you have no rights in, or to, the Software or the Documentation other than the right to use them in accordance with the terms of this Agreement.
Page 367
7. TRANSFER OF RIGHTS AND OBLIGATIONS This Agreement is binding on you and us, and on our respective successors and assigns. SRH/SM/SM/573.4/502256.8 You may transfer, assign, charge or otherwise dispose of this Agreement, or any of your rights or obligations arising under it, without our prior written consent provided that the beneficiary of any such any transfer, assignment, charge or disposal must have agreed to abide by the terms of this Agreement prior to or at the time of any such transfer, assignment, charge or disposal.
Page 368
As our policy is one of continuous improvement, details of products described within this publication are subject to change without notice. All information provided here is believed to be correct at the time of going to press. Every effort has been made to ensure the accuracy of information which is provided in good faith but nothing contained herein is intended to incorporate any representation or warranty, either express or implied or to form the basis of any legal...
Page 395
3535 Route 66, Building 6, N , NJ 07753...
Need help?
Do you have a question about the Solo 365 and is the answer not in the manual?
Questions and answers