Description of Device Functions Proline Promass 80 Coriolis Mass Flow Measuring System MEASURING MASS FLOW VOLUME FLOW DENSITY TEMPERATURE VALUES UNIT UNIT MASS FLOW UNIT MASS UNIT VOLUME SYSTEM UNITS MEASURING VOLUME FLOW MASS FLOW VOLUME FLOW DENSITY TEMPERATURE VALUES...
Device Functions Proline Promass 80 1 Function matrix Promass 80 Function matrix Promass 80 The function matrix: layout and use The function matrix is a two-level construct: the groups form one level, functions the other. The groups are the highest-level grouping of the control options for the measuring device. Each group comprises a number of functions.
Page 6
1 Function matrix Promass 80 Device Functions Proline Promass 80 Graphical illustration of the function matrix Endress+Hauser...
Device Functions Proline Promass 80 2 Group MEASURING VALUES Group MEASURING VALUES Function description MEASURING VALUES Note! • The engineering unit of the measured variable shown here can be set in the “SYSTEM UNITS” group (see Page 8). • If the fluid in the pipe flows backwards, a negative sign prefixes the flow reading on the display.
3 Group SYSTEM UNITS Device Functions Proline Promass 80 Group SYSTEM UNITS Function description SYSTEM UNITS You can select the unit for the measured variable in this function group. UNIT MASS FLOW Use this function to select the unit for displaying the mass flow (mass/time).
Page 9
Device Functions Proline Promass 80 3 Group SYSTEM UNITS Function description SYSTEM UNITS UNIT VOLUME FLOW Use this function to select the unit for the volume flow. The unit you select here is also valid for: • Current output • Frequency output •...
Page 10
3 Group SYSTEM UNITS Device Functions Proline Promass 80 Function description SYSTEM UNITS UNIT CORRECTED Use this function to select the unit for displaying the corrected volume flow (corrected VOLUME FLOW volume/time). The unit you select here is also valid for: •...
Page 11
Device Functions Proline Promass 80 3 Group SYSTEM UNITS Function description SYSTEM UNITS UNIT DENSITY Use this function to select the unit for displaying the fluid density. The unit you select here is also valid for: • Current output • Frequency output •...
Page 12
3 Group SYSTEM UNITS Device Functions Proline Promass 80 Function description SYSTEM UNITS UNIT LENGTH Use this function to select the unit for the unit of length for nominal diameter. The unit you select here is also valid for: • nominal diameter of the sensor (see the NOMINAL DIAMETER function on Page 57).
Device Functions Proline Promass 80 4 Group QUICK SETUP Group QUICK SETUP Function description QUICK SETUP QUICK SETUP Use this function to start the Setup menu for commissioning. COMMISSIONING Options: NO YES Factory setting: Note! • The display returns to the QUICK SETUP COMMISSIONING cell if you press the X key combination during parameter interrogation.
Page 14
4 Group QUICK SETUP Device Functions Proline Promass 80 XXX.XXX.XX Quick Setup Quick Setup Commission HOME-POSITION Language Pre-setting Selection pre-settings Deliver Settings Actual Settings Selection System Units Mass flow Volume flow Corr. Vol. flow Density Temperature Quit Unit Unit Unit...
Device Functions Proline Promass 80 5 Group OPERATION Group OPERATION Function description OPERATION LANGUAGE Use this function to select the language for all texts, parameters and messages shown on the local display. Note! The displayed options depend on the available language group shown in the LANGUAGE GROUP function.
Page 16
HOME position. • You can also disable programming in this function by entering any number (other than the defined private code). • The Endress+Hauser service organization can be of assistance if you lose your personal code. DEFINE PRIVATE CODE Use this function to define a personal code number for enabling programming.
Device Functions Proline Promass 80 6 Group USER INTERFACE Group USER INTERFACE Function description USER INTERFACE ASSIGN LINE 1 Use this function to define the display value assigned to the main line (the upper line of the local display) for display during normal measuring operation.
Page 18
6 Group USER INTERFACE Device Functions Proline Promass 80 Function description USER INTERFACE 100% VALUE Note! (Line 1) This function is not available unless one of the following was selected in the ASSIGN LINE 1 function: • MASS FLOW IN % •...
Page 19
Device Functions Proline Promass 80 6 Group USER INTERFACE Function description USER INTERFACE DISPLAY DAMPING Use this function to enter a time constant defining how the display reacts to severely fluctuating flow variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).
7 Group TOTALIZER 1/2 Device Functions Proline Promass 80 Group TOTALIZER 1/2 Function description TOTALIZER 1/2 ASSIGN TOTALIZER Use this function to assign a measured variable to the totalizer. Options: MASS FLOW VOLUME FLOW CORRECTED VOLUME FLOW Factory setting: MASS FLOW Note! The totalizer is reset to “0”...
Page 21
Device Functions Proline Promass 80 7 Group TOTALIZER 1/2 Function description TOTALIZER 1/2 UNIT Use this function to define the unit for the totalizer measured variable, as selected TOTALIZER beforehand. Options (for the MASS FLOW assignment): → Metric g; kg; t →...
8 Group HANDLING TOTALIZER Device Functions Proline Promass 80 Group HANDLING TOTALIZER Function description HANDLING TOTALIZER RESET ALL TOTALIZERS Use this function to reset the totals (including all overflows) of the totalizers (1...2) to “zero” (= RESET). Options: Factory setting:...
Device Functions Proline Promass 80 9 Group CURRENT OUTPUT 1/2 Group CURRENT OUTPUT 1/2 Function description CURRENT OUTPUT 1/2 ASSIGN CURRENT Use this function to assign a measured variable to the current output. OUTPUT Options: MASS FLOW VOLUME FLOW CORRECTED VOLUME FLOW...
Page 24
9 Group CURRENT OUTPUT 1/2 Device Functions Proline Promass 80 Function description CURRENT OUTPUT 1/2 CURRENT SPAN Current span, operational range and signal on alarm level (contd) I[mA] 0-20 mA 0 - 20.5 mA 4-20 mA 4 - 20.5 mA 4-20 mA HART 4 - 20.5 mA...
Page 25
Device Functions Proline Promass 80 9 Group CURRENT OUTPUT 1/2 Function description CURRENT OUTPUT 1/2 Setting the span by means of The span for the measured variable selected in the ASSIGN CURRENT OUTPUT function the 0_4 mA and 20 mA is specified via the VALUE 0_4 mA and VALUE 20 mA functions.
Page 26
9 Group CURRENT OUTPUT 1/2 Device Functions Proline Promass 80 Function description CURRENT OUTPUT 1/2 Setting the span by means of DENSITY and REFERENCE DENSITY the 0_4 mA and 20 mA • The density value for the 0/4 mA current is defined in the VALUE 0_4 mA function, value (input range 0.0000 to +99999).
Page 27
Device Functions Proline Promass 80 9 Group CURRENT OUTPUT 1/2 Function description CURRENT OUTPUT 1/2 TIME CONSTANT Use this function to enter a time constant defining how the current output signal reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).
Page 28
9 Group CURRENT OUTPUT 1/2 Device Functions Proline Promass 80 Function description CURRENT OUTPUT 1/2 VALUE SIMULATION Note! CURRENT This function is not available unless the SIMULATION CURRENT function is active (= ON). Use this function to define a selectable value (e.g. 12 mA) to be output at the current output.
Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT Group PULSE/FREQUENCY OUTPUT Function description PULSE/FREQUENCY OUTPUT OPERATION MODE Use this function to configure the output as a pulse or frequency output. The functions available in this function group vary, depending on which option you select here.
Page 30
10 Group PULSE/FREQUENCY OUTPUT Device Functions Proline Promass 80 Function description PULSE/FREQUENCY OUTPUT VALUE F LOW Note! This function is not available unless the DENSITY, REFERENCE DENSITY or TEMPERATURE option was selected in the ASSIGN FREQUENCY function. Use this function to assign a value to the start value frequency (0 Hz), (see “Setting the span by means of the f-min.
Page 31
Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT Function description PULSE/FREQUENCY OUTPUT Setting the span means of the DENSITY and REFERENCE DENSITY f-min. and • The density value for the start value frequency is defined in the VALUE F LOW f-max.
Page 32
10 Group PULSE/FREQUENCY OUTPUT Device Functions Proline Promass 80 Function description PULSE/FREQUENCY OUTPUT OUTPUT SIGNAL Note! This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function. In this function the output can be configured in such a way that it suits an external totalizer.
Page 33
Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT Function description PULSE/FREQUENCY OUTPUT OUTPUT SIGNAL Example for output configuration PASSIVE-POSITIVE: (contd) Output configuration with an external pull-down resistance. In the quiescent state (at zero flow), a positive voltage level is measured via the pull- down resistance.
Page 34
10 Group PULSE/FREQUENCY OUTPUT Device Functions Proline Promass 80 Function description PULSE/FREQUENCY OUTPUT TIME CONSTANT Note! This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function. Use this function to enter a time constant defining how the frequency output signal reacts to severely fluctuating measured variables, either very quickly (enter a low time constant) or with damping (enter a high time constant).
Page 35
Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT Function description PULSE/FREQUENCY OUTPUT SIMULATION Note! FREQUENCY This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function. Use this function to activate simulation of the frequency output.
Page 36
10 Group PULSE/FREQUENCY OUTPUT Device Functions Proline Promass 80 Function description PULSE/FREQUENCY OUTPUT PULSE VALUE Note! This function is not available unless the PULSE setting was selected in the OPERATION MODE function. Use this function to define the flow at which a pulse is triggered. These pulses can be totaled by an external totalizer, and the total flow quantity since measuring started can be registered in this way.
Page 37
Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT Function description PULSE/FREQUENCY OUTPUT OUTPUT SIGNAL Note! This function is not available unless the PULSE setting was selected in the OPERATION MODE function. In this function the output can be configured in such a way that it suits an external totalizer.
Page 38
10 Group PULSE/FREQUENCY OUTPUT Device Functions Proline Promass 80 Function description PULSE/FREQUENCY OUTPUT OUTPUT SIGNAL Example for output configuration PASSIVE-POSITIVE: (contd) Output configuration with an external pull-down resistance. In the quiescent state (at zero flow), a positive voltage level is measured via the pull- down resistance.
Page 39
Device Functions Proline Promass 80 10 Group PULSE/FREQUENCY OUTPUT Function description PULSE/FREQUENCY OUTPUT FAILSAFE MODE Note! This function is not available unless the PULSE setting was selected in the OPERATION MODE function. For reasons of safety it is advisable to ensure that the pulse output assumes a predefined state in the event of a fault.
Page 40
10 Group PULSE/FREQUENCY OUTPUT Device Functions Proline Promass 80 Function description PULSE/FREQUENCY OUTPUT VALUE SIMULATION Note! PULSE This function is not available unless the COUNTDOWN option was selected in the SIMULATION PULSE function. Use this function to specify the number of pulses (e.g. 50) which are output during the simulation.
Device Functions Proline Promass 80 11 Group STATUS OUTPUT Group STATUS OUTPUT Function description STATUS OUTPUT This group is not available unless the measuring device is fitted with a status output. ASSIGN STATUS Use this function to assign a switching function to the status output.
Page 42
11 Group STATUS OUTPUT Device Functions Proline Promass 80 Function description STATUS OUTPUT TIME CONSTANT Note! This function is not available unless one of the following was selected in the ASSIGN STATUS function: • FLOW DIRECTION • LIMIT MASS FLOW •...
Device Functions Proline Promass 80 11 Group STATUS OUTPUT 11.1 Information on the response of the status output General If you have configured the status output for “LIMIT” or “FLOW DIRECTION”, you can define the requisite switching points in the ON-VALUE and OFF-VALUE functions. When the measured variable in question reaches one of these predefined values, the status output signal switches as shown in the illustrations below.
11 Group STATUS OUTPUT Device Functions Proline Promass 80 11.2 Switching action of the status output Function State Open Collector (Transistor) ON (operation) System in measuring mode XXX.XXX.XX conductive A0001052 A0001237 System not in measuring mode (power supply failure) XXX.XXX.XX...
Page 45
Device Functions Proline Promass 80 11 Group STATUS OUTPUT Function State Open Collector (Transistor) Flow direction forward conductive A0001241 A0001237 reverse conductive A0001242 A0001238 Limit value Limit value not overshot or • Mass flow undershot • Volume flow conductive • corrected volume...
12 Group STATUS INPUT Device Functions Proline Promass 80 Group STATUS INPUT Function description STATUS INPUT This group is not available unless the measuring device is fitted with a status input I/O module. ASSIGN STATUS INPUT Use this function to assign a switching function to the status input.
Page 47
Device Functions Proline Promass 80 12 Group STATUS INPUT Function description STATUS INPUT VALUE SIMULATION Note! STATUS INPUT This function is not available unless the ON setting was selected in the SIMULATION STATUS INPUT function. Use this function to select the level to be assumed at the status input during the simulation.
13 Group COMMUNICATION Device Functions Proline Promass 80 Group COMMUNICATION Function description COMMUNICATION TAG NAME Use this function to enter a tag name for the measuring device. You can edit and read this tag name at the local display or via the HART protocol.
Device Functions Proline Promass 80 14 Group PROCESS PARAMETER Group PROCESS PARAMETER Function description PROCESS PARAMETER ASSIGN LF CUT OFF Use this function to assign the switching point for the low flow cut off. Options: MASS FLOW VOLUME FLOW CORRECTED VOLUME FLOW...
Page 50
14 Group PROCESS PARAMETER Device Functions Proline Promass 80 Function description PROCESS PARAMETER EMPTY PIPE DETECTION Use this function to activate the empty pipe detection (EPD). With empty measuring (EPD) tubes the density of the fluid falls below a specified value (see EPD LOW VALUE function).
Page 51
Device Functions Proline Promass 80 14 Group PROCESS PARAMETER Function description PROCESS PARAMETER EPD EXC. CURRENT Use this function to activate the empty pipe detection (EPD). In the event of air pockets or if fluids are not homogeneous, the excitation current of the measuring pipes increases.
Page 52
Use this function to configure an automatic pressure correction. In this way, the effect of a pressure deviation between the calibration and process pressures on the measured error for mass flow is compensated for, (see also Operating Instructions "Proline Promass 80", BA 057D/06/en/, Accuracy Chapter).
Page 53
Device Functions Proline Promass 80 14 Group PROCESS PARAMETER Function description PROCESS PARAMETER PRESSURE Note! This function is not available unless FIX was selected in the PRESSURE MODE function. Use this function to enter the value for the process pressure which should be used during pressure correction.
15 Group SYSTEM PARAMETER Device Functions Proline Promass 80 Group SYSTEM PARAMETER Function description SYSTEM PARAMETER INSTALLATION Use this function to reverse the sign of the measured variable, if necessary. DIRECTION SENSOR Note! Ascertain the actual direction of fluid flow with reference to the direction indicated by the arrow on the sensor (nameplate).
Page 55
Device Functions Proline Promass 80 15 Group SYSTEM PARAMETER Function description SYSTEM PARAMETER MEASURING MODE Pulse output (contd) STANDARD Only positive flow components are totaled. Negative components are not taken into account. SYMMETRY Positive and negative flow components are taken into account.
Page 56
15 Group SYSTEM PARAMETER Device Functions Proline Promass 80 Function description SYSTEM PARAMETER DENSITY DAMPING The density filter allows the sensitivity of the density measuring signal to be lowered with respect to variations in the density of the fluid, e.g. with inhomogeneous liquids.
Device Functions Proline Promass 80 16 Group SENSOR DATA Group SENSOR DATA Function description SENSOR DATA All sensor data, including calibration factor, zero point, nominal diameter, etc. are set at the factory. All the sensor's parameter settings are saved on the S-DAT memory chip.
Page 58
16 Group SENSOR DATA Device Functions Proline Promass 80 Function description SENSOR DATA DENSITY COEFFICIENT This function shows the actual density coefficient C 0. " Caution! A density adjustment can alter the calibration value of this coefficient. DENSITY COEFFICIENT This function shows the actual density coefficient C 1.
Device Functions Proline Promass 80 17 Group SUPERVISION Group SUPERVISION Function description SUPERVISION CURRENT SYSTEM Use this function to check the current system status. CONDITION Display: “SYSTEM OK” or the fault / notice message with the highest priority. PREVIOUS SYSTEM...
Page 60
17 Group SUPERVISION Device Functions Proline Promass 80 Function description SUPERVISION ERROR CATEGORY Use this function to define whether a process error triggers a notice message or a fault message. If you select “FAULT MESSAGES”, all outputs respond to a fault in accordance with their defined error response patterns.
Device Functions Proline Promass 80 18 Group SIMULATION SYSTEM Group SIMULATION SYSTEM Function description SIMULATION SYSTEM SIMULATION Use this function to set all inputs, outputs and totalizer to their defined fault response FAILSAFE MODE modes, in order to check whether they respond correctly. During this time, the words “SIMULATION FAILSAFE MODE”...
19 Group SENSOR VERSION Device Functions Proline Promass 80 Group SENSOR VERSION Function description SENSOR VERSION SERIAL NUMBER Use this function to view the serial number of the sensor. SENSOR TYPE Use this function to view the sensor type (e.g. Promass F).
Device Functions Proline Promass 80 21 Factory settings Factory settings 21.1 SI units (not for USA and Canada) Low flow cut off, full scale value, pulse value Nom. diameter Low flow cut off Full scale value Pulse value [mm] (approx. v = 0.04 m/s) (approx.
21 Factory settings Device Functions Proline Promass 80 Unit Density kg/l Length Temperature °C Reference density kg/Nl 21.2 US units (only for USA and Canada) Low flow cut off, full scale value, pulse value Nominal Low flow cut off Full scale value...
Page 65
Device Functions Proline Promass 80 Index of key words Index of key words Numerics Temperature KM ........57 100% Value KM 2 .
Page 66
Index of key words Device Functions Proline Promass 80 Failsafe mode Language ........15 Current output .
Page 67
Device Functions Proline Promass 80 Index of key words Name ........48 Temperature .