Notice Congratulations on your purchase of the KODAK 9000 Extraoral Imaging System. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The User Guide for the KODAK 9000 3D Extraoral Imaging System includes information both on the Panoramic features as well as the 3D features.
CAUTION Alerts you to a condition that might cause serious damage. IMPORTANT Alerts you to a condition that might cause problems. NOTE Emphasizes important information. Provides extra information and hints. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 1–1...
To use and operate your panoramic unit and your digital imaging software you must follow the instructions contained in this guide. Warning and Safety Instructions When operating the Kodak 9000 3D unit, observe the following warning and safety instructions: DANGER OF ELECTRIC SHOCK This is an electrical Unit.
Page 7
Ask the patient to remain still until the Unit arm has stopped moving and the RESET movement has completed. • DO NOT use this Unit in conjunction with oxygen-rich environments. This Unit is not intended for use with flammable anesthetics or flammable agents. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 1–3...
Marking and Labeling Symbols Marking and Labeling Symbols Type B device symbol complying with the IEC 60601-1 standard In the EEC, this symbol indicates: DO NOT discard this product in a trash receptacle; use an appropriate recovery and recycling facility. Contact your local sales representative for additional information on the collection and recovery programs available for this product...
KODAK 9000 3D DIGITAL IMAGING UNIT OVERVIEW The KODAK 9000 3D digital imaging unit is compliant with the requirements of the EEC and international medical standards. The KODAK 9000 3D Unit has been designed to carry out the following radiological examinations: •...
Page 10
Control panel Generator Hand Grips Unit rotative arm Chin rest base X-Ray remote control Panoramic chin rest PC hosting the imaging and the acquisition software 3D chin rest and bite block 2–2 KODAK 9000 3D D IGITAL IMAGING UNIT OVERVIEW...
Page 11
Figure 2–3 KODAK 9000 3D Unit Positioning Laser Beam components 3D central positioning laser beam Mid-sagittal positioning laser beam Horizontal positioning laser beam 3D Field of View (FoV) positioning laser beam KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 2–3...
Display Screen: Displays the current acquisition parameters and the error messages. Ready Indicator LED: Green indicates the Unit is ready for acquisition. X-Ray Emission LED: Yellow indicates x-rays are being emitted. System Status LED: Red indicates error alerts. 2–4 KODAK 9000 3D D IGITAL IMAGING UNIT OVERVIEW...
You must press and hold the exposure button until the end of acquisition. Premature release of the exposure button interrupts the acquisition. Figure 2–5 X-Ray Remote Control Exposure button: launches image acquisition. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 2–5...
Panoramic standard bite block Bite block for edentulous patients Single use sheaths for bite blocks (500 pcs box) Table 2–2 3D Positioning Accessories and Replacement Parts Accessory Description 3D bite block 3D chin rest 2–6 KODAK 9000 3D D IGITAL IMAGING UNIT OVERVIEW...
Chapter 3 IMAGING SOFTWARE OVERVIEW PC System Requirements This section specifies the minimum PC system requirements for Kodak 9000 3D extraoral imaging system software. IMPORTANT It is MANDATORY to check that the PC system configuration is compatible with the PC system requirements for the Kodak 9000 3D software.
• Acquisition interface module Kodak Dental Imaging Software The Kodak dental imaging software is a user-friendly working interface that was designed and developed specifically for radiological diagnosis. It is the common imaging platform for all our digital systems for dentistry.
Kodak Dental Imaging Software Overview Patient Window Overview The Patient Window is the first window that appears when you launch the Kodak dental imaging software. This window is the main patient record management interface that provides you with all the patient record functions.
Page 18
Kodak Dental Imaging Software Overview Exit Exits the Patient window Picture Insert picture Inserts the patient picture Remove picture Removes the patient picture Utilities Reindex Regenerates all the index table Archival... Archives images before the given date Options Preferences... Configures the language and the...
New 3D image Accesses the 3D Acquisition Window for a 3D image History Open Opens the patient history record with image files Save Saves the acquired image Close Closes the patient image KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–5...
Page 20
Kodak Dental Imaging Software Overview Delete Deletes selected image Print Prints the patient image Print Setup Sets up the printing parameters Slide Show Displays a set of images in a particular sequence Report Creates customized letters Word Report Creates a report template...
Page 21
7 Contextual Toolbar: The toolbar type depends on the image type context. You can reposition the toolbar. NOTE For information and procedures on how to use these functions, from the menu bar click ? to access the online help. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–7...
Kodak Dental Imaging Software Overview 3D Imaging Windows Overview The 3D Imaging Windows are the main imaging interfaces that provide you with the 3D imaging functions. 3D Home Page Window Figure 3–3 Title Bar: Displays Software name and version. Window Tabs: Provide access to the indicated Windows.
Page 23
Click to display the cut plane as colored planes. 3D Filters Click to apply the smooth filter. Click to configure the smooth filter settings. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–9...
Page 24
Kodak Dental Imaging Software Overview Measurements Click to show or hide measurements. Click to measure the length. Click to measure the angle. Click to edit the measurements. Click to delete the measurements. Implant Click to show or hide all implants.
Page 25
Export Click to create a screen capture file. Click to access the screen capture file in the directory. Display options: Display the specific screen in different formats. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–11...
Page 26
Kodak Dental Imaging Software Overview To access the , from the Home Page click the Oblique Slicing Window Oblique Slicing Window tab. Figure 3–6 Oblique Slicing Window Title Bar: Displays software name and version. Window Tabs: Provides access to the indicated windows.
Page 27
Click to access the screen capture file in the directory. Display options: Display the specific screen in different formats. NOTE Click to use the default value. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–13...
Acquisition Interface Module Acquisition Interface Module Panoramic Acquisition Window Overview The Panoramic Acquisition Window is the main panoramic interface with the Kodak 9000 3D extraoral imaging system that provides you with imaging acquisition functions. Panoramic Acquisition Window Figure 3–7 Information button: identifies: •...
Page 29
The Selector button enables you to access the following 3 panes: • Program pane: Examination type options • Patient pane: Patient type parameter options • Parameters pane: Exposure parameter options KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–15...
Acquisition Interface Module Panoramic Program Pane The Panoramic Program pane enables you to choose different radiological exams. The diagram in the Panoramic Program pane represents a jaw, with the right side of the diagram corresponding to the right side of the patient. Click on a section of the jaw to select an anatomical zone for radiological exam.
Click for a normal dental arch. Click for a sharp dental arch. Incisors orientation: Click for a vestibular orientation. Click for an axial orientation. Click for a lingual orientation. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–17...
Acquisition Interface Module Panoramic Parameter Pane The Panoramic Parameter pane enables you to choose exposure parameters for the radiological image acquisition. You must adapt the parameter settings to the patient type. Figure 3–10 Panoramic Parameter Pan Fine-tuning buttons: Click to fine-tune the kV and click to fine-tune the mA.
Acquisition Interface Module 3D Acquisition Window Overview The 3D Acquisition Window is the main 3D interface with the Kodak 9000 3D extraoral imaging system that provides you with imaging acquisition functions. 3D Acquisition Window Figure 3–11 Information button: identifies: •...
Page 34
Acquisition Interface Module Exit button: Closes the Acquisition Window. X-Ray Emission Indicator: Yellow indicates the x-ray emission status. Selector button: Selects different acquisition setting options. • Click Program to select examination type options • Click Patient to select patient type parameters •...
Rotative Arm Position Indicator: Indicates the rotative arm position, using the 3D central positioning laser beam. Click to validate the region of interest selection. Click to cancel the region of interest selection. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–21...
Acquisition Interface Module 3D Patient Pane The 3D Patient pane enables you to choose different patient parameters. The selection of the patient parameters influences the quality of the image. The selected parameters must be based on the patient age and morphology. Figure 3–13 Panoramic Patient Pane Patient corpulence parameters:...
You must adapt the parameter settings to the patient type. Figure 3–14 3D Parameter Pane. Fine-tuning buttons: Click to fine-tune the kV and click to fine-tune the mA. Radiation dose options: Kilovolt milliampere KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 3–23...
To start your imaging software, follow these steps: 1. On your desktop, double-click From your PC, click Start > All Programs > Kodak > Kodak Dental Software. Blank Patient Window is displayed. 2. Create or open an existing patient record.
Accessing the Acquisition Windows 2. Enter the required patient information. The Last name, the First name and the Date of birth fields are required. 3. From the menu bar, select Picture > Insert Picture to add a *.tif or *.bmp picture of the patient to the record.
4. Position the appropriate chin rest on the chin rest support and cover the bite block with a hygienic barrier. If needed, use the edentulous bite block. Press and hold to raise the chin rest to the maximum height. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 5–1...
Acquiring a Panoramic, TMJ x2 or Sinus Image Preparing and Positioning the Patient To prepare and position the patient, follow these steps: 1. Ask the patient to remove all metal objects. 2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient shoulders.
Page 43
OFF automatically. IMPORTANT The spinal column and the nose of the patient must be aligned in a straight line (1). 6. Immobilize the patient head with the temple supports (3). KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 5–3...
Acquiring a Panoramic, TMJ x2 or Sinus Image 7. Ask the patient to close the eyes, to remain still, to swallow, to place the tongue in contact with the palate and to breath through the nose. Launching the X-ray To launch the x-ray, follow these steps: 1.
Page 45
4. Do the following when the acquisition is finished: • Open the temple supports and release the patient. • Remove the hygiene barrier of the bite block. • Reset the Unit rotative arm for the next acquisition. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 5–5...
Acquiring a TMJ x4 Image Acquiring a TMJ x4 Image Before acquiring a TMJ x4 image, check that you have: • Reset the Unit rotative arm to start position for patient entry • Selected the patient record • Accessed the Imaging Window •...
To prepare and position the patient, follow these steps: 1. Ask the patient to remove all metal objects. 2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient shoulders. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 5–7...
Page 48
Acquiring a 3D Image 3. Ask the patient to enter the Unit. On the Control Panel, press and hold to adjust the Unit to the patient height. 4. Ask the patient to do the following: • Stand up straight or sit on a stool. •...
Page 49
3D FoV (2) positioning laser beam. Tighten the temple supports (3). 8. Ask the patient to close the eyes, to remains still and to breath through the nose. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 5–9...
Acquiring a 3D Image Launching the x-ray To launch the x-ray, follow these steps: 1. Leave the x-ray room and close the door. You must keep eye contact with the patient during acquisition. IMPORTANT To stop the acquisition, if any problem, release the exposure button of the remote control or press the red emergency stop button.
Unit rotative arm. 4. Remove the plaster imprint and position the patient in the Unit and follow the appropriate steps to acquire the 3D image of the region of interest. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 5–11...
Page 52
Acquiring a 3D Image 5–12 ACQUIRING IMAGES...
Wipe the Unit with a dry cloth at the end of each day's operation. WARNING Do not use detergents or solvents to clean the outer covers of the Unit. Monthly Wipe the outer covers of the unit with a soft, dry cloth. KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 6–1...
Chapter 7 TECHNICAL SPECIFICATIONS Compliance with International Standards The KODAK 9000 3D digital imaging Unit is compliant with the International and EEC standards. Manufacturer TROPHY A subsidiary of Carestream Health, Inc. 4, Rue F. Pelloutier, Croissy-Beaubourg 77435 Marne la Vallée Cedex 2, France Model KODAK 9000 3D Unit, KODAK 9000 Unit.
Unit Technical Specifications Unit Technical Specifications Table 8-1 describes the technical specifications for the KODAK 9000 3D digital imaging unit and Kodak 9000 digital imaging unit. Table 7–1 Unit Technical Specification Kodak 9000 3D Kodak 9000 X-Ray Generator Tube voltage...
Page 57
1158mm (L) x 1595mm (D) x 2378mm (H)* * Unit with shorter column is available on special order Required space 1500 (L) x 2000 (D) x 2400 (H) mm Weight 160 kg (353 lb KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 7–3...
Unit Electronic Specifications Unit Electronic Specifications 230/240 V (± 10%) 50/60 Hz, 100/110/130V (± 10%) 50/60 Hz, Type of Electrical Power Supply Single-Phase Single-Phase Acceptable fluctuation ± 10% ± 10% Ω Ω Apparent resistance of the power (max.) 0.12 (max.) supply circuit Permanent absorbed current 0.45 A...
Radiation leakage after one hour's operation (maximum utilization rate of < 1 mGy 93W, i.e. 90 kV, 10 mA, 13.9 sec. every 2 minutes 15 sec.) Weight 8.2 kg Dimensions 235 x 245 x 120 mm KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 7–5...
Page 60
X-Ray Generator Technical Specifications Figure 7–1 Heating and Cooling Curves of the X-ray Generator. Heating and cooling curves 110, 0 100, 0 90, 0 80, 0 100 W 70, 0 66 W 60, 0 33 W 50, 0 15 W 40, 0 10 W 30, 0...
Page 61
X-Ray Generator Technical Specifications Figure 7–2 Heating and Cooling Curves of the X-ray Tube Figure 7–3 Single Load Chart of the X-ray Tube Figure 7–4 Filament Emissions of the X-ray Tube KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01 7–7...