Page 1
Hart Scientific Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com 9132 Portable IR Calibrator User’s Guide Rev. 5A2601...
1 Before You Start Symbols Used Before You Start Symbols Used Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual. Table 1 International Electrical Symbols Symbol Description AC (Alternating Current) AC-DC...
9132 Portable IR Calibrator User’s Guide Symbol Description Canadian Standards Association OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re- fers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.
1 Before You Start Safety Information • Overhead clearance is required. DO NOT place this instrument under a cabinet or other structure. • DO NOT use this unit for any application other than calibration work. • DO NOT use this unit in environments other than those listed in the user’s guide.
Page 10
9132 Portable IR Calibrator User’s Guide • Operate the instrument in room temperatures between 5 and 50°C. (41–122° F). Allow sufficient air circulation by leaving at least 6 inches of space between the instrument and nearby objects. • The instrument is a precision instrument. Although is has been designed for optimum durability and trouble free operation, it must be handled with care.
IR thermom- eters. The 9132 is small enough to use in the field, and accurate enough to use in the lab. Calibrations may be done over a range of 50°C to 500°C (122°F to 932°F).
3 Specifications and Environmental Conditions Specifications Specifications and Environmental Conditions Specifications Range 50°C to 500°C (122°F to 932°F) Accuracy ±0.5°C at 100°C (±0.9°F at 212°F) ±0.65°C at 300°C (±1.2°F at 572°F) ±0.8°C at 500°C (±1.4°F at 932°F) Stability ±0.1°C at 100°C (±0.18 at 212°F) ±0.15°C at 300°C (±0.27 at 572°F) ±0.3°C at 500°C (±0.54 at 932°F) Target Size...
9132 Portable IR Calibrator User’s Guide 9132 Gradient between probe and IR surface -0.1 -0.2 -0.3 -0.4 -0.5 -0.6 Reference Probe ⎠C Figure 1 Temperature Gradient Between Probe and IR Target Surface • temperature range: 5-35°C (41-95°F) • ambient relative humidity: 15-50% •...
Unpack the calibrator carefully and inspect it for any damage that may have oc- curred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: • 9132 Calibrator • Power Cord • User’s Guide • Serial Cable •...
9132 Portable IR Calibrator User’s Guide Setting the Temperature Section 8.2 explains in detail how to set the temperature set-point on the cali- brator using the front panel keys. The procedure is summarized here. 1. Press “SET” twice to access the set-point value.
5 Parts and Controls Back Panel Parts and Controls The user should become familiar with the calibrator and its parts. Back Panel The back panel (Figure 2) consists of the power cord inlet, power switch, heater voltage switch, serial port, and fan. 115V HEATER VOLTAGE...
9132 Portable IR Calibrator User’s Guide Fan – The fan inside the calibrator has two speeds and runs continuously when the unit is being operated to provide cooling for the instrument. The fan runs slow for heating and maintaining operation and runs fast for rapid cooling.
Page 20
5 Parts and Controls Front Panel DOWN – Used to decrement the displayed value of parameters. UP – Used to increment the displayed value. EXIT – Used to exit a function and to skip to the next function. Any changes made to the displayed value are ignored.
2. Press the “UP” or “DOWN” key to change units. Switching to 230 V Operation The 9132 is switchable from 115 VAC to 230 VAC 50/60 Hz. Switching the voltage can change the calibration. Therefore, we recommend that the instru- ment be calibrated after the voltage is switched.
7 Controller Operation Well Temperature Controller Operation This chapter discusses in detail how to operate the instrument temperature con- troller using the front control panel. Using the front panel key-switches and LED display the user may monitor the well temperature, set the temperature set-point in degrees C or F, monitor the heater output power, adjust the control- ler proportional band, and program the calibration parameters, operating pa- rameters, and serial interface configuration.
9132 Portable IR Calibrator User’s Guide Display Temperature DOWN Select Setpoint Displays Set-Point Resolution Adjust Setpoint Toggles °C / °F EXIT EXIT Units °C/°F Secondary Functions EXIT Scan On/Off Menu EXIT Display Power Scan Rate Set Proportional Band Adj. R0...
7 Controller Operation Temperature Scale Units 100.0 C Well temperature in degrees Celsius Access set-point memory 1. 100. Set-point memory 1, 100 C currently used. To change to another set-point memory press “UP” or “DOWN”. 4. 300. New set-point memory 4, 300 C Press “SET”...
9132 Portable IR Calibrator User’s Guide Scan The scan rate can be set and enabled so that when the set-point is changed the dry-well heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the dry-well heats or cools at the maximum possible rate.
7 Controller Operation Temperature Scale Units keys are released the temperature is again displayed. This value is used to cali- brate the unit and is not adjustable. Temperature Scale Units To toggle between °C and °F, press the “SET” and “UP” keys simultaneously when the temperature is displayed.
Page 27
9132 Portable IR Calibrator User’s Guide set-point. This range of temperature is called the proportional band. At the bot- tom of the proportional band the heater output is 100%. At the top of the pro- portional band the heater output is 0. Thus as the temperature rises the heater power is reduced, which consequently tends to lower the temperature back down.
7 Controller Operation Controller Configuration Accept the new proportional band setting 7.10 Controller Configuration The controller has a number of configuration and operating options and calibra- tion parameters, which are programmable via the front panel. These are ac- cessed from the secondary menu after the proportional band function by pressing “SET”.
9132 Portable IR Calibrator User’s Guide 7.12.1 Baud Rate The baud rate is the first parameter in the menu. The baud rate setting deter- mines the serial communications transmission rate. Flashes bAUd and then displays the setting bAUd 2400 b...
7 Controller Operation Calibration Parameters Adjust the setting using “UP” or “DOWN” and pressing “SET”. d=HALF New duplex mode setting 7.12.4 Linefeed The final parameter in the serial interface menu is the linefeed mode. This pa- rameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return.
9132 Portable IR Calibrator User’s Guide rameter to the new value. Pressing “EXIT” causes the parameter to be skipped ignoring any changes that may have been made. 7.13.1 This probe parameter refers to the resistance of the control probe at 0°C. The value of this parameter is set at the factory for best instrument accuracy.
8 Digital Communication Interface Serial Communications Digital Communication Interface The calibrator is capable of communicating with and being controlled by other equipment through the digital serial interface. With a digital interface the instrument may be connected to a computer or other equipment.
9132 Portable IR Calibrator User’s Guide simultaneously to enter the secondary menu. Press “SET” repeatedly until the display reads PAr. Press “UP” until the serial interface menu is indicated with SERIAL. Finally press “SET” to enter the serial parameter menu. In the serial interface parameters menu are the baud rate, the sample rate, the duplex mode, and the linefeed parameter.
8 Digital Communication Interface Interface Commands temperature set-point and view or program the various parameters. The inter- face commands are discussed in Section 8.3. All commands are ASCII charac- ter strings terminated with a carriage-return character (CR, ASCII 13). Interface Commands The various commands for accessing the calibrator functions via the digital in- terfaces are listed in this section (see Table 2).
Page 35
9132 Portable IR Calibrator User’s Guide Command Command Returned Acceptable Command Description Format Example Returned Example Values Set proportional band to n pr[opband]=n pr=8.83 Depends on Configuration Read heater power po[wer] po: 999.9 po: 1.0 (duty cycle) Configuration Menu Operating Parameters Menu...
9 Calibration Procedure Calibration Points Calibration Procedure Sometimes the user may want to calibrate the instrument to improve the tem- perature set-point accuracy. Calibration is done by adjusting the controller probe calibration constants R0, ALPHA, and DELTA so that the temperature of the dry-well as measured with a standard thermometer agrees more closely with the set-point.
9 Calibration Procedure Calibration Procedure 2. Press SET then use the UP or DOWN keys until the correct numerical setting is displayed. Press SET to accept the new value. Repeat step 2 for ALPHA and DELTA. 9.2.3 Accuracy and Repeatability Check the accuracy of the instrument at various points over the calibration range.
10 Maintenance Maintenance • The calibration instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in dirty or dusty environments.
Factory Reset Sequence. Hold the SET and EXIT buttons down at the same time while powering up the instrument. After the instrument displays -init-, release the buttons. The display shows -init-, then displays 9132, and then displays the firmware version. After performing the master reset sequence, all of the configuration parameters are reset to their default values.
9132 Portable IR Calibrator User’s Guide Problem Possible Causes and Solutions The display shows any Controller problem. The error messages signify the following problems with of the following: err 1, the controller. err 2, err 3, err Err 1 - a RAM error...
Need help?
Do you have a question about the 9132 and is the answer not in the manual?
Questions and answers