ECHELON FLEX™ Powered & Powered Plus Staplers
Optimized Device Performance Guide
Loading the device
Proper reload selection: Tissue thickness should be carefully evaluated prior to using the instrument. Refer to the
for proper reload selection. Selection of the appropriate staple reload should be based upon the combined thickness of both the tissue and the staple line reinforcement material.
1. Verify that the reload size matches the instrument size being used (e.g. use 60mm
reload with 60mm stapler).
3. Insert the reload by sliding it against the bottom of the reload jaw until the
stapler alignment tab stops in the reload alignment slot of the jaw. Positioning
the reload at the bottom of the reload jaw could prevent inadvertent sled
movement and lockout.
Device closure
1. Position the instrument around the tissue to be stapled. When closing the device, ensure that the tissue lies flat and is positioned properly between the jaws. Any "bunching" of
tissue along the reload, particularly in the crotch of the jaws, may result in an incomplete staple line or inadvertent lockout. Holding the jaws in place for 15 seconds after closing
and prior to firing may result in better compression and staple formation.
Reload Product Codes
2. Ensure the staple retaining cap is present during loading to minimize the potential
for premature sled movement, which could lead to inadvertent lockout.
4. Remove the staple retaining cap by lifting it straight up and discard.
5. Observe the surface of the new reload to ensure no drivers are visible and that the
reload lies flat in the jaw channel.
Chart, located in the Instructions for Use,
For complete product details, see Instructions for Use.

Advertisement

loading
Need help?

Need help?

Do you have a question about the ECHELON FLEX and is the answer not in the manual?

Questions and answers

Summary of Contents for Ethicon ECHELON FLEX

  • Page 1 ECHELON FLEX™ Powered & Powered Plus Staplers Optimized Device Performance Guide Loading the device Proper reload selection: Tissue thickness should be carefully evaluated prior to using the instrument. Refer to the Reload Product Codes Chart, located in the Instructions for Use, for proper reload selection.
  • Page 2 Maintaining pressure on the anvil release button, slowly release the closing trigger. Note: After the manual override is used, the device becomes nonfunctional. For complete product details, see Instructions for Use. ©2017 Ethicon US, LLC. All rights reserved. 019926-170815...