Error Codes - Inscale TFC 150 User Manual

Tfc series
Table of Contents

Advertisement

ERROR CODES

During the initial power-on testing or during operation it is possible the scale may show
an error message. The meaning of the error messages is described below.
If an error message is shown repeat the procedure that caused the message, turning
the balance on, calibration or other functions. If the error message still is shown then
contact your dealer for further support.
ERROR CODE
Err 4
Err 5
Err 6
Err 8
DESCRIPTION
Initial Zero is greater than
allowed (typically 4% of
maximum capacity) when
power is turned on or when
the ZERO key is pressed,
Keyboard Error.
A/D count is not correct
when turning the scale on.
Calibration Error
POSSIBLE CAUSES
Weight on the pan when
turning the scale on.
Excessive weight on the
pan when zeroing the
scale.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Improper operation of the
scale.
Platform not installed.
Load cell damaged.
Electronics damaged.
Attempt calibration again,
make sure the scale is
stable.