Hanna Instruments HI 96735 Instruction Manual
Hanna Instruments HI 96735 Instruction Manual

Hanna Instruments HI 96735 Instruction Manual

Hardness ism

Advertisement

Quick Links

INSTRUCTION MANUAL
HI 96735
HI 96735
HI 96735
HI 96735
HI 96735
Hardness ISM
w w w . h a n n a i n s t . c o m
sale@ponpe.com
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the meter.
This manual will provide you with the necessary information for correct use
of the instrument. If you need additional technical information, do not
hesitate to e-mail us at tech@hannainst.com or view our worldwide
contact list at www.hannainst.com.
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 96735 Ion Selective Meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual
Note: Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and accessories
see "Accessories".
Technical specifications:
Range
LR
0 to 250 mg/L
MR
200 to 500 mg/L
HR
400 to 750 mg/L
Resolution
1 mg/L from 0 to 100 mg/L
5 mg/L from 100 to 750 mg/L
Accuracy
LR ±5mg/L ±4% of reading@ 25°C
MR ±7 mg/L ±3% of reading@ 25°C
HR ±10 mg/L ±2% of reading@ 25°C
Light Source
Light Emitting Diode
Light Detector
Silicon Photocell with narow band interference
filter @ 466 nm
Method
Adaptation of the EPA recommended method
130.1. The reaction between calcium,
magnezium and the reagents causes a
red-violet tint in the sample.
Environment
0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type
1 x 9 volt
Auto-Shut off
After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder.
Dimensions
192 x 104 x 69 mm (7.6 x 4.1 x 2.7")
320 g (11.3 oz.).
Weight
Functional description:
1. RANGE/GLP/
key: press to change the range or press and hold
for three seconds to enter GLP mode . In calibration mode press to edit
the date and time.
2. CAL CHECK key: press to perform the validation of the meter, or
press and hold for three seconds to enter calibration mode .
3. ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore. Press
and hold for three seconds to start a preprogramed countdown prior
to measurement.
4. READ/ /UNIT key: In measurement mode , press to make a
measurement, or press and hold for three seconds to change the
measurement unit. In GLP mode press to view the next screen.
5. ON/OFF key: to turn the meter on and off.
6. Liquid Crystal Display (LCD)
7. Cuvette alignment indicator
8. Cuvette holder
DISPLAY ELEMENTS DESCRIPTION:
1. The measuring scheme (lamp, cuvette, detector), appears during
different phases of zero or reading measurement
2. Error messages and warnings
3. The battery icon indicates the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display
Errors and warnings:
ON ZERO READING:
Light High: There is too much light to perform
a measurement. Please check the preparation of
the zero cuvette.
Light Low: There is not enough light to perform
a measurement. Please check the preparation of
the zero cuvette.
No Light: The instrument cannot adjust the
light level. Please check that the sample does
not contain any debris.
ON SAMPLE READING:
Inverted Cuvettes: The sample and the zero
cuvette are inverted.
Zero: A zero reading was not taken. Follow the
instructions of the measurement procedure for
zeroing the meter.
Under Range: A blinking "200" indicates that
the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.
Over Range: A flashing value of the maximum
concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and
re-run the test.
DURING CALIBRATION PROCEDURE:
Standard Low: The standard reading is less
than expected.
Standard High: The standard reading is higher
than expected.
OTHER ERRORS AND WARNINGS:
Cap Error: Appears when external light enters
in the analysis cell. Assure that the cuvette cap
is present.
Cooling lamp: The instrument waits for the
lamp to cool down.
Battery Low: The battery must be replaced
soon.
Dead Battery: This indicates that the battery
is dead and must be replaced. Once this
indication is displayed, normal operation of the
instrument will be interrupted. Change the
battery and restart the meter.
Measurement procedure:
Measurement
1• Turn the meter on by pressing ON/OFF.
2
2• When the beeper sounds briefly and the LCD
displays dashes and "P1" (Low range), "P2"
(Medium range) or "P3" (High range), the
meter is ready. The code that appears on
the secondary display is the one of the
last selected range. If necessary, press
3
RANGE/GLP/
to change range. The
blinking "ZERO" indicates that the
instrument needs to be zeroed first.
3• Add 0.5 mL of unreacted sample to the
cuvette. Add 0.5 mL of HI 93735IND-0
10 mL
indicator reagent. With the plastic dropper
fill the cuvette up to the 10 mL mark
4
adding HI 93735A-XR (X=L, M, H)
reagent appropriate to the selected range.
4• Add two drops of HI 93735B-0 buffer
reagent. Replace the cap and invert 5 times
to mix.
5
5• Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
6
positioned securely into the groove.
6• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
7• After a few seconds the display will show
8
"-0.0-". The meter is now zeroed and ready
for measurement. Remove the cuvette.
8• Add the content of 1 packet of HI 93735C-0
fixing reagent. Replace the cap and shake
gently to mix 20 seconds.
9
9• Replace the cuvette into the holder and
ensure that the notch on the cap is positioned
securely into the groove.
10
10•Press READ/ /UNIT. In all cases the lamp,
cuvette and detector icons will appear on the
display, depending on the measurement phase.
www.ponpe.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the HI 96735 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 96735

  • Page 1 No Light: The instrument cannot adjust the light level. Please check that the sample does 1• Turn the meter on by pressing ON/OFF. Each HI 96735 Ion Selective Meter is supplied complete with: not contain any debris. • Two Sample Cuvettes and Caps 2•...
  • Page 2 . CHECK or ON/OFF keys. HI 96735 is warranted for two years against defects in workmanship and increased. 1• Turn the meter on by pressing ON/OFF. materials when used for its intended purpose and maintained according Battery management 15•When the correct month has been set, press...