Do you have a question about the JPD-FR401 and is the answer not in the manual?
Questions and answers
Summary of Contents for 1byone JPD-FR401
Page 1
Infrared Thermometer Model: JPD-FR401 1byone Products Inc. Instruction Manual 2313 E Philadelphia Street, Unit M, Ontario, CA 91761 www.1byone.com...
Page 2
Introduction Thank you for choosing 1byone infrared thermometer. This infrared thermometer tests body temperature by measuring the ear and forehead, making it suitable for both adults and children (noting that ear test mode is only suitable for children over 3 months). Please read the instruction manual carefully and keep it for future reference.
Page 3
Notes: Operating Conditions: 1. Battery polarities should be correctly installed. Otherwise damage may be caused to the device. Temperature: 10°C to 40°C (50°F to 104°F) Humidity: ≤80% RH, non-condensing 2. Please remove the batteries if the thermometer will not be used for an extended period of time. Atmospheric pressure: 860hPa to 1060hPa Warnings Storage and Shipping:...
2. Product Description Probe cover (put on when using HEAD mode) Probe (take off cover when using EAR mode) Head mode LCD display Ear mode Battery cover Display Description 1. Ambient temperature 2. Ready for measurement 3. Ear mode 4. Head mode 5.
3. Measuring Forehead Temperature 4. Measuring Ear Temperature 1. With the sensor/probe cover attached, position the thermometer at the center of the forehead, just 1. Gently remove the probe cover to reveal the ear probe. above the eyebrow. Ensure that the thermometer is touching the forehead. 2.
Page 6
To avoid inaccuracy: 5. Operation and Display Instructions 1. Please make sure there is no dirt in and on the temperature sensor. LCD display Operation and Display Instructions Sound and backlight 2. Do not use the thermometer in a windy environment. 3.
Page 7
3. Press the HEAD or EAR button again to make another will be 7 short beeps 3 short beeps with a red temperature measurement. and the reading will The measured temperature exceeds 100°C/212°F. backlight for 3 seconds. blink with a red backlight for 3 seconds.
Page 8
When the thermometer is off, press and hold the EAR Repeat the previous steps to view the remaining button for 3 seconds and the LCD will display “- - -” None None saved measurements (up to 20 total). with a blinking "M." The LCD will display “- - -”...
Page 9
The voltage of the batteries has fallen below Release the EAR button when "°C" or "°F" appears on None 2.51V±0.15V and should be replaced. the LCD. Press the EAR button again within 5 seconds to change the temperature unit. Power-off Error messages The thermometer will power off automatically if no activity is detected for 10 seconds.
6. Cleaning and Disinfecting Disinfecting Cleaning 1. Disinfect the thermometer body and the area around the temperature probe with a cloth slightly moistened with 75% medical alcohol. 1. Clean the temperature probe with a soft cloth. Clean the lens of the temperature probe with a cotton swab.
9. Troubleshooting If the warranty has not yet The thermometer is faulty. expired, contact Customer Symptom Possible Cause Solution Service. The battery level is extremely low. Replace with new batteries. Only the battery symbol is displayed after the The battery level is low. Replace with new batteries.
Page 12
Items Standards The distance between the The temperature reading is Model 1byone Infrared Thermometer Move the thermometer temperature probe and the target lower than the typical closer to the target. is too great. Applicable laws and regulations ASTM 1965/EN12470-5/GB/T 19146-2010 body temperature range.
11. Security Type ±0.2 ° C / ±0.4 ° F Precision The signal indicates that the thermometer is a BF-type device with internal power supply. Displayed measurement 0.1 ° C / 0.1 ° F resolution 12. Authorized European Representative Latency Time 1 second Wellkang Ltd The thermometer shall display the low-voltage signal if the...
13. Symbols Personal injury or thermometer damage may occur if the thermome- Warning ter is not correctly used. Symbol Description Inaccurate readings or thermometer damage may occur if the Attention BF-type applied part. thermometer is not correctly used. Attention must be paid. 14.Declaration Manufacturer information, such as name and address.
Page 15
CAUTION: Emissions test Compliance Electromagnetic environment - guidance This infrared Thermometer needs special precautions regarding EMC and needs to be installed The Infrared Thermometer uses RF energy only for its internal and put into service according to the EMC information provided for in the ACCOMPANYING function.
Page 16
Power frequency Guidance and manufacturer´s declaration – electromagnetic immunity Power frequency magnetic fields should be (50/60 Hz) at levels characteristic of a typical location in magnetic 3 A/m 3 A/m The Infrared Thermometer is intended for use in the electromagnetic environment specified a typical commercial or hospital environment.
Page 17
IEC 60601 Compliance Immunity test Electromagnetic environment guidance where “p” is the maximum output power rating of test level level the transmitter in watts (W) according to the Portable and mobile RF communications equipment transmitter manufacturer and “d” is the recommend- should be used no closer to any part of the Infrared ed separation distance in metres (m).
Page 18
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. SYSTEM -for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING Note 2: These guidelines may not apply in all situations. Its electromagnetic ability is affected by absorption and reflection from structures, objects, and people.
Page 19
Separation distance according to frequency of transmitter m For transmitters rated at a maximum output power not listed above, the recommended Rated maximum separation distance d in metres (m) can be estimated using the equation applicable to the 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz output power...
Page 20
Disposal The appliance’s packaging is made from environmentally friendly material and can be disposed Disposal of the appliance of at your local recycling plant. Under no circumstances should you dispose of the appliance in normal domestic waste. This product is subject to the provisions of European Directive 2012/19/EU. Dispose of the appliance via an approved disposal company or your municipal waste facility.