Adjustment Of Conductivity And Concentration; Adjusting And Calibrating Process - Nipro SURDIAL Service Manual

Single patient dialysis machine
Hide thumbs Also See for SURDIAL:
Table of Contents

Advertisement

10.16 Adjustment of conductivity and concentration

Adjust dilution ratio of dialysate concentrate intended to be used to input the adjusted ratio and
calibrate indicated dialysate conductivity and concentration by manipulation of operation panel
based on the following procedures.
Bicarbonate dialysis 1 or 2
Input of concentrate data
Adjustment of dilution ratio
of A concentrate
Adjustment of dilution ratio
of B concentrate
Calibration of indicated
conductivity and concentration
Start of adjustment
Selection of dialysis mode to be adjusted
Completion of adjustment and calibration
Fig. 10.16-1 Adjusting and calibrating process
Acetate dialysis
Input of concentrate data
Adjustment of dilution ratio
of acetate concentrate
Calibration of indicated
conductivity and concentration
10.16-1