Calibration Procedure
4.3.4
Service Diagnostics
4.3.4.1
User Interface
All IABP controls become nonfunctional after entering Service Diagnostic mode. All Service
Diagnostic user input is provided through the keys indicated by the black arrows in
FIGURE 4-2.
FIGURE 4-2
To enter Service Diagnostics, press and hold the INFLATION INTERVAL and the FREEZE
DISPLAY keys on the Monitor keypad while powering on the CS300. Continue to hold the
keys until the Datascope logo is displayed. Following the logo is the warning screen:
WARNING - Do Not use Service Diagnostics
while the system is connected to a patient.
Use of Service Diagnostics while the system is
connected to a patient, may subject the patient
to injury. Verify patient is NOT connected to the
system and press the 'User Preferences'
key to proceed.
FIGURE 4-3
™
CS300
Service Manual
OPERATION MODE
Semi
AUTO
Auto
TRIGGER SOURCE
IAB FREQUENCY
ECG
1:1
Pressure
1:2
Pacer V / AV
Pacer A
1:3
Internal
Ref
Alarm
Help
Line
Mute
IAB
Aug
Fill
Alarm
Hold 2 Sec.
Clear Event Log
Scroll Left
Select
Scroll Up
User Interface
CS300 - Service Diagnostic
Service Diagnostic Warning
0070-10-0689
Toggle Transducer Reference
Zero
Manual
Pressure
Hold 2 Sec.
Pediatric Only
AUGMENTATION
IAB INFLATION
MAX
Earlier
OFF
ECG / AP
Sources
Pump
Open
Options
Menu
User
Preferences
Back
Scroll Down
Exit
Calibration
START
STANDBY
IAB DEFLATION
Later
Later
Earlier
Inflation
Interval
Freeze
Display
Print
Strip
Hold for continuous
Scroll Right
Toggle Recorder On/Off
4 - 5