> At the end of its useful life, send the product to the separate collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal. WEEE Reg. Nr. DE 75334352...
2 Specifications Specifications 2.1. The testo 184 data loggers are used to save and read out individual readings and measurement series. They have been specifically designed for monitoring the transportation of products subject to cold chain requirements. Temperature and humidity readings are saved throughout the entire duration of the measurement program.
Page 5
T2, T3, T4: 45 g Directives, 2014/30/EC, EN 12830, HACCP-certified, standards, temperature calibration certificate traceable certificates according to ISO 17025 testo 184 H1, G1 Feature Values Display Probe type Internal NTC temperature sensor H1: Internal digital humidity sensor G1: Internal digital humidity sensor & internal 3-axis...
Page 6
For continuous application in high humidity (> 80 % RH at ≤ 30 °C for > 12 h, > 60 % RH at > 30 °C for > 12 h), please contact us via www.testo.com/service- contact Regenerate the sensor for 12 hours at < 20% RH and 50 °C, or for 24 hours at...
Warranty www.testo.com/warranty EU conformity The EU Declaration of Conformity can be found on the testo homepage www.testo.com under the product specific downloads. EU countries: Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland (IE),...
3 Product description Alarm Feature LED colour No alarm green Alarm Battery Feature LED colour Battery life > 10 days green Battery life < 10 days Mode Feature LED colour WAIT mode (wait for program start) green and red mode (measurement program running) green mode (measurement program finished) 3.2.
3 Product description 3 Wait for start of measurement program 4 Highest saved reading 5 Lowest saved reading 6 Reading 7 Status information: start criterion Date/time programmed/time mark, measurement axes for acceleration measurement, Alarm set limit value(s) exceeded, set limit value(s) not exceeded 8 Units 9 Battery capacity:...
Page 10
3 Product description Display T2 T3 H1 G1 Current temperature reading (°C / °F) Current average value (Mean Kinetic Temperature) Current relative humidity reading (%) Current acceleration reading, X axis (x, g) Current acceleration reading, Y axis (y, g) Current acceleration reading, Z axis (z, g) Maximum temperature reading (Max, °C / °F) X...
Page 11
3 Product description Hibernation mode is activated. 3.4. Important information and glossary of terms • Single-use data logger (versions T1 and T2): the data logger has a time-limited service life, which begins from the first time the program is started. •...
The status LEDs are disabled, is displayed (instruments with display). The device drivers are installed automatically. The window Automatic playback is displayed. 2. Click on Open folder to view files. The file explorer opens. 3. Open the file testo 184 configuration.pdf.
Page 13
Configuring several data loggers with the same settings > Configure your measurement protocol using the configuration PDF or import an existing XML file. > Connect the testo 184 data logger to a USB connection. > Click to save the configuration on the connected testo 184 data logger.
Page 14
Alternatively, the data logger can also be configured using testo Comfort Software Professional (version 4.3 service pack 2 or later) or testo Comfort Software CFR (version 4.3 service pack 2 or later). Please consult the relevant software instruction manual. In order to ensure conformity of the testo Comfort Software 21...
To carry out detailed analysis or to further process readings, the software testo Comfort Software Professional (version 4.3 service pack 2 or later) or testo Comfort Software CFR (version 4.3 service pack 2 of later) is required (accessory). Please consult the relevant software instruction manual.
However, stored measurement data are preserved. An interruption of the power supply leads to the testo 184 data logger's time settings being reset. In order to restore the correct time setting, configuration via the PDF file or the Comfort Software must be carried out.
21 CFR Part 11 for the configuration, configuration via the PDF file is disabled. The size of the PDF Copy the PDF file of another testo 184 configuration file is 0 kB or data logger or download the the file is damaged.
Page 18
6 Tips and assistance Question Possible causes/solution > The testo 184 data logger has not Different time or time zone been configured after a battery in the report replacement. Repeat the configuration to restore the correct time settings. > Check whether the PC which was used for the configuration has the correct time settings.
Page 19
> Store the device for 12 hours at 20°C to 30°C and a relative humidity of around 75%. If you have any questions please contact your local dealer or the Testo Customer Service. Contact details can be found on the internet: www.testo.com/service-contact...