What to do if...
40
Cause
• ORP value not stable
• Temperature not stable
10.2.2
Error message OFL, UFL
The measured value is outside the measuring range of the meter.
The measured value is obviously incorrect.
Cause
• Cable broken
• Gel electrolyte dried out
• Air bubble in front of the junction
• Air in the junction
10.3
General information
10.3.1
Symbol for calibration evaluation flashes
Cause
Calibration interval expired
10.3.2
[LoBat] display
Cause
Batteries almost empty
Remedy
• Measure with air excluded if
necessary
• Adjust temperature if necessary
Remedy
• Replace the electrode
• Replace the electrode
• Remove the air bubble
• Extract air or moisten the junction
Remedy
Recalibrate the measuring system
Remedy
Replace the batteries (see section
3.3.1 I
NSERTING THE BATTERIES
page 10)