Yaohua Weighing System XK3190-A12+SS User Manual

Digital weighing indicator

Advertisement

Digital Weighing Indicator
User Manual
V 1.10
Ltd
Shanghai Yaohua Weighing System Co.,

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the XK3190-A12+SS and is the answer not in the manual?

Questions and answers

Summary of Contents for Yaohua Weighing System XK3190-A12+SS

  • Page 1 Digital Weighing Indicator User Manual V 1.10 Shanghai Yaohua Weighing System Co.,...
  • Page 2: Table Of Contents

    Content CHAPTER 1 MAIN SPECIFICATIONS…………………………….…………………….. 2 CHAPTER 2 APPEARANCE AND INSTALLATION…………………….…………….. 3 2.1 KEY FUNCTIONS…………………………………………………………………..….. 3 2.2 LOAD CELL CONNECTION…………………………………………………………... 3 2.3 SCOREBOARD CONNECTION TO THE INDICATOR…………………………..…4 2.4 SERIAL COMMUNICATION AND INDICATOR CONNECTION ……………….….4 CHAPTER 3 CALIBRATION……………………………………………………………..5 3.1 GENERAL CALIBRATION……………………………………………………………...5 3.2 FAST CALIBRATION…………………………………………………………………...
  • Page 3: Chapter 1 Main Specifications

    Chapter 1 Main Specifications 1. Model: XK3190-A12+SS weighing indicator 2. Accuracy: Grade Ⅲ, n=3000 3. Sample Rate: 10 times / second 4. Load cell sensitivity: 1.5~3mV / V 5. Scale interval: 1/2/5/10/20/50 for option 6. Display: 6 bits LCD,6 state indicating signals 7.
  • Page 4: Chapter 2 Appearance And Installation

    CHAPTER 2 APPEARANCE AND INSTALLATION 2.1 KEY FUNCTIONS Function 【 # 】 When turning on the indicator, keep pressing this key to enter the calibration mode 【Func 】 In weighing mode, keep pressing this key for more than 5 seconds to enter the user parameters setting mode, less than 5 seconds will enter into counting mode.
  • Page 5: Scoreboard Connection To The Indicator

    Non-professional staff is supposed not to be involved in this regard. XK3190-A12+SS can use optional RS232 serial communication interface to communicate with computer. The connection method, please refer to picture 2-4.
  • Page 6: Chapter 3 Calibration

    The format of N.W: wn000.000kg or wn000.000lb The format of Tare: wt000.000kg or wt000.000lb Note:The position of above decimal is decided by the decimal set on the indicator. (2) In command mode (all of them are ASCII): The indicator performs the corresponding operation according to the command transmitted from the indicator.
  • Page 7: Fast Calibration

    [TARE] Press to select the Full capacity setting digit bit; For example: Press when displayed【025000】, [ZERO] Press to change then the full capacity is set to be “25000”, and the [FULL ] the value; indicator enters zero point calibration Press for confirm the input of full capacity Make sure there is no load...
  • Page 8: Chapter 4 Operation

    Chapter 4 Operation 4.1 POWER ON AND AUTO ZERO-SETTING 4.1.1 The indicator will perform “000000-999999” to self-checking when turning on. Then it will enter weighing mode. 4.1.2 When power on, if loading weight on the scale deviates from the zero point, but still within zero set range, the indicator will set zero automatically;...
  • Page 9 In weighing mode, keep pressing [ Func ] for 5 seconds more, it will enter operator setting mode (mode P), there are 12 modes from P1 to P12 for option, press [ * ] to choose the mode and press [Tare] to change the parameter. The description of parameter as follows: 1、P1 kg Lb change x=1:...
  • Page 10: Chapter 5 Maintenance

    x=6: 3.0e x=7: 5.0e x=8: Tracking forbidden 8、P8 Zero key scope x=1: 2%FS x=2: 4%FS x=3: 10%FS x=4: 20%FS x=5: 100%FS 9、P9 Zero scope upon starting x=1: 2%FS x=2: 4%FS x=3: 10%FS x=4: 20%FS x=5: 100%FS x=6: Start ZERO forbidden 10、P10 x Digital filtering intensity x=1:...
  • Page 11: Chapter 6 Error Indication

    explosive materials. ▲! Do not use it where inflammable gases or steams exist. Don’t use it for canning system of compressive container. ▲! In the area where lightning and thunder happen frequently, reliable lightning arrester should be installed to ensure the personal safety and to prevent any damage to the indicator and relative equipment caused by lightning stroke.
  • Page 12 Please make sure no load is on the scale when turning on ERR 4 The input sample quantity is zero when sampling in counting mode Please input the right sample quantity ERR 5 When full capacity calibrating, the weight input is zero Please input the weight in accordance with the load on scale ERR 6 The unit weight is less than 0.25e when sampling in counting mode...