Download Print this page
HoldPeak 870N Operation Manual

HoldPeak 870N Operation Manual

6000 digits dual display ac/dc auto cal clamp multimeter
Hide thumbs Also See for 870N:

Advertisement

Quick Links

6000 DIGITS DUAL DISPLAY
AC/DC AUTO CAL CLAMP MULTIMETER
OPERATION MANUAL
This LCD Auto Range & Auto Power off Digital AC/DC AUTO
CAL clamp multimeter is a portable, 6000 digit multimeter. It is
ideally suited for field, laboratory, shop and home applications.
1. SAFETY INFORMATION
The following safety information must be observed to insure
maximum personal safety during the operation at this meter.
1) When measuring voltage ensure that instrument is not
switched to the current range, resistance range, diode and
continuity range, capacitance range or temperature range.
2) Use extreme care when measuring voltage above 50V.
especially from sources where high energy is existed.
3) Avoid making connections to "live" circuits whenever possible.
4) Before making resistance measurements, diode or continuity
test, capacitance test or temperature test, ensure that the
circuit under test is de-energized.
5) Always ensure that the correct function and range is selected.
6) Extreme care should be taken when using the instrument to
conjunction with a current transformer connected to the
terminals if an open circuit occurs.
7) Ensure that the test leads and probes are in good condition
with no damage to the insulation.
8) Take care not to exceed the over-load limits as given in the
specifications.
9) Before opening the cover of the battery cabinet to replace
batteries. disconnect the test leads from any external circuit,
set the selector switch to "OFF" position.
10) Keep the fingers after the protection ring when measuring
through the instrument lead.
1
1) Change the battery when the symbol "
incorrect data.
2. Panel Layout
1)
Rotary Switch: use this switch to select functions and ranges.
2) D.HOLD/ Back Light key:In any range, push the key, the
present display value will be locked and the " H " symbol will
appear, push it again to exit HOLD and the " H"symbol
disappear. Press "D.HOLD" button more than 2 seconds, the
back light will light, press it more than 2 seconds again, the
back light will light off or After 15 seconds Auto light out.
3) SELECT key:This key work on the "CAP
the key to choose resistance, diode ,continuity or capacitance
test, on the voltag or current range,change to DC/AC, and ℃/
℉ range , change to ℃/℉
4)
MAX/MIN key:Push the key to select MAX mode, push it
again to change MIN mode, push once again to change
max-min,press the key for more than 2 seconds to go back
auto range mode. And Push the key to change manual range
mode .But in Hz/Duty and Capacitance measurement, it can
not use.
5) Hz/DUTY key: In "
ACV/ACA
key, you can measure the Hz ,push again, can measure the
duty.
6) REL key: Push the key, the present display value will be stored
in memory, then the new display value is the difference
between input value and stored data. In Hz/Duty measurement,
it can not work. press the key for more than 2 seconds to
measure inrush current at ACA/DCA range
7) LCD Display:LCD Dual Display, facilitates reads the data.
8)T+V Ω Input Jack 、COM Input Jack、T- Input Jack.
870N
" appears to avoid
H P -870N
A U T O C A L
6 00A
F
IN R U S H
S EL E CT
M A X /M IN
H z/D u ty
R E L
T+
T-
C A T. I V
60 0V
TR U E R M S
Ω" range , Push
" or "Hz" range, push the
2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
Display: 6000 digit LCD with a max. reading of 6000.
Range control: Auto range control.
Polarity: Automatic negative polarity indication.
Zero adjustment: Automalic.
Overrange indication: Only the "OL" display.
Low battery: The " " is display when the battery voltage is below 6.2V.
Auto Power Off: 15 minutes after stopping the switch or no push
button, the meter automatically enter to power off
mode.Push SELECT button, auto power off disable.
Safety Standards: The meter is up to the standards of IEC1010
Double Insulation, Pollution Degree 2, Overvoltage
Category III 1000V.
Clamp opening size: 45mm.
Operating Environment: Temperature 32~104°F(0~40℃),
humidity<80%RH.
Storage Environment: Temperature -4~140°F(-20~60℃),
humidity<90%RH.
Power supply: 9V Zinc-carbon battery.
Dimension: 225(H)×97(W)×40(D)mm.
Weight: Approx. 350g (including batteries).
2.2 ELECTRICAL SPECIFICATIONS
Accuracies are ± (% of reading +numder in last digit) at 23±5℃,
≤70%RH.
2.2.1 DC Voltage
Range
600mV
6V
1.0% of rdg+5digits
60V
600V
1000V
1.5% of rdg+5 digits
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ,More than 100MΩ on 400mV scale
2.2.2 AC Voltage(RMS)
Range
Accuracy
600mV
3.0% of rdg+15 digits
6V
60V
1.5% of rdg+5 digits
600V
750V
2.5% of rdg+5 digits
Average sensing, calibrated to rms of sine wave
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ.
2.2.3 DC Current
Range
600A
3.0% of rdg+10 digits
3.5% of rdg+10 digits
1000A
Overload protection: 1000Arms within 60 seconds
2.2.4 AC Current
Range
Accuracy
600A
3.0% of rdg+10 digits
1000A
3.5% of rdg+10 digits
Average sensing, calibrated to rms of sine wave
Overload protection: 1000Arms within 60 seconds
2.2.5 Resistance
Range
600Ω
6kΩ
60kΩ
1.0% of rdg+8 digits
400kΩ
6MΩ
60MΩ
2.5% of rdg+15 digits
Overload protection: 250V DC/250Vrms AC
2.2.6 Capacitance
Range
9.999nF
2.5% of rdg+15 digits
99.99nF
999.9nF
9.999μF
2.5% of rdg+10 digits
99.99μF
999.9μF
3.0% of rdg+20 digits
9.999mF
5.0% of rdg+25digits
99.99mF
Overload protection: 250V DC/250Vrms AC
Accuracy
Resolution
0.1mV
100mV
Resolution
Frequency
0.1mV
40~100Hz
1mV
10mV
40~400Hz
100mV
1V
40~100Hz
Accuracy
Resolution
Resolution Frequency
0.1A
1A
Accuracy
Resolution
Accuracy
Resolution
100PF
100nF
1mV
10mV
1V
0.1A
1A
50~60Hz
0.1Ω
10Ω
100Ω
1kΩ
10kΩ
1PF
10PF
1nF
10nF
1uF
10uF

Advertisement

loading
Need help?

Need help?

Do you have a question about the 870N and is the answer not in the manual?

Questions and answers

Summary of Contents for HoldPeak 870N

  • Page 1 10mV 40~400Hz 600V 100mV 750V 2.5% of rdg+5 digits 40~100Hz H P -870N Average sensing, calibrated to rms of sine wave A U T O C A L Overload protection: 1000V DC/750Vrms AC Impedance: 10MΩ. 2.2.3 DC Current 6 00A...
  • Page 2 2.2.7 Frequency AND Duty socket and the red test lead to the " T+ " socket. 3) Put the sensor probe into the temperasure field under Range Accuracy Resolution Sensitivity measurement. 9.999Hz 0.001Hz 4) Read the result from the LCD panel. 99.99Hz 0.01Hz Max.input over-load: 250V rms<10sec...