The following points are included in the monthly maintenance:
• Monthly Maintenance Procedure
• Lubrication of o-rings
• Cleaning of the Sample Carousel
• Outside cleaning
• Deletion of instrument log files
To perform Monthly Maintenance you need ImmunoCAP Maintenance Solution
Kit (10-9476-01). For preparing the Maintenance Solution add one ampule (23
ml) of Maintenance Concentrate to 1500 ml purified water. Mix gently in a glass
beaker. Please read and follow the warnings and instructions carefully. Monthly
Maintenance is also used as a corrective action, see Action List.
Monthly Maintenance Procedure
A cleaning of Phadia 100 has to be done once a month. When handling
Maintenance Solution protective gloves should be used.
1.
In Maintenance mode
press the number 2 key.
2.
Press the yes key.
3.
Press lid open and open
the Processing Lid. Place four
CheckCAP in position 1 to 4 in
the ImmunoCAP Holder Disc
and close the Processing Lid.
Press enter to confirm this step.
NOTE! ImmunoCAP Holder Disc is fragile and should be handled as glass.
4.
Fill a clean and empty
conjugate vial (diameter 22
mm) with 5 ml Maintenance
Solution and place the vial in
the Sample Carousel in posi-
tion G. Press the enter key to
confirm this step.
12-3501-20/12
Monthly Maintenance
6.9.Maintenance
1.Daily Maintenance?
6.9.Maintenance
2.Monthly Maintenance?
6.9.2.Monthly Maintenance
LOAD 4 CheckCAP IN POS 1-4
6.9.2.Monthly Maintenance
LOAD MAINT SOLUTION IN POS.G
Maintenance
2
5
6
11