Do you have a question about the CMC 156 and is the answer not in the manual?
Questions and answers
Sushil Hibare
February 6, 2025
Hi
We want to purchase
VEHZ2007 Cable
Qty 1 nos
Haochi Tang
March 27, 2025
Why the overload current light is red when we switch it on
1 comments:
Mr. Anderson
March 27, 2025
The overload current light on the Omicron CMC 156 turns red when an overload occurs at one of the current or voltage outputs. This triggers a corresponding display in the CMC software, indicating the overload condition. Additionally, if the device switches off due to overtemperature caused by a low ohmic burden with a high ON period, messages such as "Overtemperature at voltage output!" or "Overtemperature at current output!" appear, and all voltage (V) and current (I) outputs are switched off.
Need help?
Do you have a question about the CMC 156 and is the answer not in the manual?
Questions and answers
Hi We want to purchase VEHZ2007 Cable Qty 1 nos
Why the overload current light is red when we switch it on
The overload current light on the Omicron CMC 156 turns red when an overload occurs at one of the current or voltage outputs. This triggers a corresponding display in the CMC software, indicating the overload condition. Additionally, if the device switches off due to overtemperature caused by a low ohmic burden with a high ON period, messages such as "Overtemperature at voltage output!" or "Overtemperature at current output!" appear, and all voltage (V) and current (I) outputs are switched off.
This answer is automatically generated