Advertisement

Making Measurements with the HP E-Series Power Sensors

Making Measurements with the HP E-Series Power
Sensors
This section describes how to make continuous wave measurements using
the HP E-series power
sensor calibration tables stored in EEPROM. This allows the frequency
and calibration data to be downloaded to the power meter automatically.
To make measurements, perform the following steps:
1. Zero and calibrate the power meter.
2. Set the frequency for the signal you want to measure.
3. Take a reading.

Procedure

The following procedure details how to make a measurement on channel A
of the power meter.
1. Ensure that no power is applied to the power sensor.
Zero
2. Press
.
Cal
3. Press
. During zeroing, which takes approximately 10
Zero
seconds, the wait symbol is displayed.
4. Connect the power sensor to the POWER REF output.
5. Press
to calibrate the power meter. During calibration the
Cal
wait symbol is displayed. (The power meter automatically turns
on the POWER REF output.)
Frequency
6. Press
Cal Fac
under the
Freq
7. To change this setting press
frequency in a pop up window. Modify this frequency (see below)
as desired.
Use
currently positioned.
Use
8. To confirm your choice press the appropriate frequency units.
9. Connect the power sensor to the signal to be measured.
10. The measurement result is now displayed.
HP E4418B User's Guide
sensors.
The HP E-series power sensors have their
. The current setting of the frequency is displayed
softkey.
Freq
or
to modify the digit on which the cursor is
or
to move to other digits.
Power Meter Operation
. The power meter displays the
2-15