Preventive Maintenance Tasks
M3015A
NOTE
1. Allow 5 seconds between individual service procedures in order to ensure stable
equipment conditions.
2. When certain Monitor procedures are running (for example, AutoZero or purging),
service procedures are not possible and trying to start them will result in a message
"Service Operation Failed" in the status line on the Monitor. Wait until the Monitor has
completed the current operation, then restart the service procedure.
Preventive maintenance comprises the following activities:
• Perform visual inspection (see "Inspecting the Instrument" on page 122)
• Check the operating time for the pump and CO
• If required, (operating time >15,000 hours), replace the pump and the CO
with serial number prefix DE020xxxxx only)
• Perform barometric pressure check and calibrate, if necessary, see "Barometric Pressure
Check and Calibration" on page 149.
• Perform leakage check, see "Leakage Check" on page 150.
• Perform Pump Check
• Perform flow check and calibrate, if necessary, see "Flow Rate Check and Calibration" on
page 151.
• Perform Noise Check
• Perform CO
sary, see "CO
• If calibration done, perform CO
• If M3015A has been opened, perform a system safety test, see "Patient Safety Checks" on
page 159.
Tools Required
• Standard tools, e.g. screwdriver, tweezers
• ESD protection equipment
• Electronic flowmeter, M1026-60144
• Gas calibration equipment
•
•
•
•
Required Replacement Part
• Pump assembly and CO
124
calibration check; check the date of the last calibration and calibrate, if neces-
2
Gas Measurement Calibration Check" on page 152.
2
Cal 1 gas 15210-64010 (5% CO
Cal 2 gas 15210-64020 (10% CO
Cal gas flow regulator M2267A
Cal tube 13907A
scrubber (M3015-29303)
2
scrubbers (Pump OpTime).
2
, and reset the operating time counter.
calibration verification using 2nd calibration gas
2
)
2
)
2
(units
scrubber
2
Maintaining the Instrument
Need help?
Do you have a question about the M3001A and is the answer not in the manual?
Questions and answers
Is there any data stored in the philips physiological monitor module for brick model 3001A or can this be disposed as is?
The Philips M3001A physiological monitor module stores data in two separate databases: a short-term and a long-term database. These databases are battery-buffered to prevent data loss during a power failure. Therefore, it should not be disposed of without considering the stored data.
This answer is automatically generated