Feature
Accumulated Air
Air-in-Line Settings
(single bolus)
Auto-Restart Attempts
KVO (Keep Vein Open)
Max Rate
Max VTBI
Pressure Mode
•
Mode Selection
•
Lock Status
•
Max Occlusion Pressure
•
Default Starting Occlusion
Pressure
SEC to PRI Alert
Secondary (Dual Rate
Sequential Piggybacking)
General Information
2-150
Configurable Settings (Continued)
Pump Module
Default Setting
Enabled
75 mcL
0
1 mL/h
999 mL/h
9999 mL
Pump
Unlocked
525 mmHg
525 mmHg
Enabled
Disabled
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Options
Enabled - Disabled
50, 75, or 250 mcL
Anesthesia Mode only: 500 mcL
0 - 9 attempts
Anesthesia Mode only: 9 attempts
-
mL/h
0.1
20
-
mL/h in
mL/h increments;
0.1
99.9
0.1
-
mL/h in
mL/h increments
100
999
1
-
mL
0.1
9999
Pump, Selectable
Locked, Unlocked
50 - 525 mmHg in 25 mmHg increments (adjustable
only in Selectable Pressure Mode)
50 - 525 mmHg in 25 mmHg increments (configured
by Profile and adjustable only in Selectable Pressure
Mode)
Enabled - Disabled
Enabled - Disabled
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART