Configuration
Saving the
configuration
Loading
configuration
Resetting the
configuration
Exporting the
configuration
Importing the
configuration
Changing the
access codes
Note
Resetting
access codes
Formatting the
CF card
162
All the settings configured, except date and time, must be saved if they are to
be available the next time the corpuls
The "Load" function allows a reset to the last version saved while device is
running. So the device does not have to be switched off to undo temporary
changes to the settings.
With the "Set to default" function, the corpuls
(see also Appendix C Factory Settings, p. 278).
The function "Export" allows the export of the configuration settings that were
saved last onto the CF card. So, configurations once defined can be transferred
3
to another corpuls
. Settings for GSM/GPRS, the internal phonebook, IP
address, access codes and some master data (see also chapter 8.4 Master
Data, page 188) are not exported. In case of a software update all saved
configurations remain unchanged.
The configuration settings exported and saved on a CF card can be imported
onto another corpuls
configuration settings, the corpuls
configuration to take effect.
4.
To store, load or reset the changes in the configuration, select them with
the jog dial and confirm.
The person responsible for the device (Operator) can manually change the
access codes for the
- default user (Default),
- restricted access use of the of the manual defibrillator (Manual defibrillation)
and
- for the person responsible for the device (Operator).
For this purpose, a 4-digit access code has to be selected and repeated.
It is recommended to change the access codes for the person responsible for
the device (Operator) and – if appropriate – for restricted access use of manual
defibrillation (Manual defibrillation) on commissioning the device.
The person responsible for the device can reset the access codes for the
default user (Default), the "manual defibrillation" user (Manual defibrillation) and
for the person responsible for the device (Operator) to the factory settings. For
this purpose, a confirmation prompt must be confirmed with the softkey [OK].
If the operator access code is no longer known, the authorised corpuls sales
and service partner can reset the access codes to factory settings.(see also
Appendix C Factory Settings, p. 278).
The function „Format" allows the user to format the CF inserted in the device.
After formatting, a reboot is recommended.
3
is switched on.
3
with the function "Import". After importing the
3
has to be restarted in order for the new
User Manual corpuls
3
is reset to the factory settings
ENG - Version 2.1 – P/N 04130.2