Extech Instruments TKG100 User Manual

Extech Instruments TKG100 User Manual

Digital ultrasonic thickness gauge and gauge with datalogger
Hide thumbs Also See for TKG100:

Advertisement

USER GUIDE
TKG100
Digital Ultrasonic Thickness
Gauge
TKG150
Digital Ultrasonic Thickness
Gauge with Datalogger

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TKG100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Extech Instruments TKG100

  • Page 1 USER GUIDE TKG100 Digital Ultrasonic Thickness Gauge TKG150 Digital Ultrasonic Thickness Gauge with Datalogger...
  • Page 2 The information in this document is subject to change without notice and describes only the product defined in the introduction of this documentation. This document is intended for the use of Extech Instruments customers only (a Flir Systems, Inc. brand) for the purposes of the agreement under which the document is submitted, and no part of it may be reproduced or transmitted in any form or means without the prior written permission.
  • Page 3 LIABILITY Ultrasonic testing is a function of using the proper equipment (electronics, transducer, cable and couplant combination) for the inspection and a qualified operator who knows how to use this manual, the instruments and all calibration procedures. The improper use of this equipment, along with the improper calibration can cause serious damage to components, factories, facilities, personal injury and even death.
  • Page 4: Table Of Contents

    Replacing the Batteries ..............13 2 Basic Gauge Operations ..............14 Power on the Gauge ................14 Performing a Reset (TKG100) ............. 16 Performing a Reset (TKG150) ............. 17 About Screen ..................19 3 Calibrating the Gauge/Making Measurements ........ 20 Velocity Calibration Only ...............
  • Page 5 Automatic Zero ..................24 Measurement Mode with a Datalogger (TKG150) ....... 26 Measurement Mode without a Datalogger (TKG100) ....27 4 Measurement Mode Setup Options ..........29 Using the Clock ..................31 Setting LCD Contrast ................33 Understanding a LOS Reading ............33 Using the Hold Option ................
  • Page 6 Extech Instruments Copying a File ..................71 Deleting a File ..................74 7 Technical Specifications ..............76 8 Software Options ................... 79 B-Scan (TKG150 only) ................80 9 Technical Assistance ................81 TKG100_TKG150 Ultrasonic Thickness Gauge User Guide v2.8 11/13...
  • Page 7: Getting Started

    1 Getting Started About the TKG Series The TKG100 and TKG150 are portable, digital, handheld thickness gauges. The TKG series of ultrasonic thickness gauges are specifically designed to measure the remaining wall thickness of primarily steel structures.  Vibralarm (vibrates gauge below/above minimum/maximum thickness value) ...
  • Page 8: Probe Zero

    Extech Instruments The TKG100 is a base gauge offering a simple user interface packaged in the same custom molded high- density plastic case with rubber keypad as the other more sophisticated models. This gauge offers reliable, accurate thickness readings on mostly steel structures with access to only one side.
  • Page 9 working in accordance with electronic zeroing procedures. This feature is particularly important on high temperature materials and when the transducer becomes worn. Important Notice: Please make sure the transducer is not coupled to the test piece when the gauge is first turned on and that there is no couplant on the end of the transducer.
  • Page 10: Keypad Functions

    Extech Instruments Keypad Functions The figure below shows the full keypad of the TKG Series thickness gauges. Refer to the following table that lists the function of each key on the keypad. Function Up arrow key Down arrow key Left arrow key...
  • Page 11: Function Keys

    MENU/OK key F1 key F2 key F3 key Function Keys Function keys, or F keys, such as F1, F2, and F3 have various gauge functions and may change depending on the display screen. View the bottom of the display screen for the function that corresponds with the appropriate F key.
  • Page 12 Extech Instruments The TKG series is designed to show the parameters selected in the center of the display screen. The top part of the display screen for TKG150 model shows the File name, ID number, AA, and 0001 for grid column and row.
  • Page 13: Battery Power

    Battery Power Monitoring the Battery from the Display The TKG series constantly displays the percent battery life on the bottom right hand corner of the display screen. When the gauge gets below 20 %, the indicator will flash. Replacing the Batteries To replace the batteries, unscrew the battery door on the bottom of the gauge and slide out the two “AA”...
  • Page 14: Basic Gauge Operations

    Extech Instruments 2 Basic Gauge Operations Power on the Gauge To power on any of the TKG series thickness gauges follow these steps: Press and hold the F1 key for more than 3 seconds. The power symbol is under the F1 key printed on the keypad as shown below.
  • Page 15 4. Use the up and down arrow keys to select a transducer and press OK key. (Please note that the transducer selection will automatically display the transducer last used prior to shut off. Press F1 to select the Setup option. Press F2 to select the Reset option.
  • Page 16: Performing A Reset (Tkg100)

    Extech Instruments Performing a Reset (TKG100) You can reset the TKG100 gauge settings back to the default settings by performing a gauge reset. Note: Performing a reset permanently deletes all of the saved parameters from the gauge and replaces the settings with default values.
  • Page 17: Performing A Reset (Tkg150)

    Performing a Reset (TKG150) You can reset both the TKG150 DATABASE and PARAMETER settings back to their default settings in one gauge reset or you can reset the PARAMETER and DATABASE data individually. Note: Performing a reset permanently deletes all of the saved parameters from the gauge and replaces the settings with default values.
  • Page 18 Extech Instruments Note that performing a Database Reset clears all the files in the database and keeps the first three: Linear, Row increment and Col increment database files with cleared readings. All the user created files in the database will be cleared and deleted.
  • Page 19: About Screen

    About Screen You can find information about any of the TKG series thickness gauges such as the model number, version number and contact information from the gauge‟s About screen. To go to the About Screen, follow these steps: From the Home screen (transducer select screen), press the F2 –...
  • Page 20: Calibrating The Gauge/Making Measurements

    Extech Instruments 3 Calibrating the Gauge Calibrating is the process of adjusting the gauge for a specific material and transducer before testing material to make sure that all measurements are accurate. You must always calibrate before measuring material for standard accuracy.
  • Page 21: Velocity And Zero Calibration

    OK. To proceed to the Velocity calibration, follow the following instructions. Cal screen for TKG100 non-datalogger version While measuring the thicker step, select VEL by pressing F3. After selecting VEL, you can take the transducer off the test block.
  • Page 22: Zero Calibration Only

    Extech Instruments To calibrate both: velocity and zero at the same time first go to the Calibration mode as shown in Velocity Calibration Only section. While measuring the thicker step, select VEL by pressing F3. After selecting VEL you can take the transducer off the test block. If the displayed...
  • Page 23: Delay Line Calibration

    and electronics. This is called delay line and zero calibrated value for the transducer. To perform a zero calibration, go to the Cal mode as shown in Velocity Calibration Only section. Then while measuring the thinner step, select ZERO by pressing F2. After selecting ZERO, you can take the transducer off the test block.
  • Page 24: Automatic Zero

    Extech Instruments the sensitivity of the transducer is too low and whether the transducer should be replaced. Automatic Zero To perform an Automatic Zero or Auto Zero, you must first select a transducer option from the Transducer selection screen and press the Menu/OK key.
  • Page 25 After three seconds the gauge automatically begins to zero the transducer. During the delay line calibration a screen appears displaying the message “Performing Auto Zero”. A warning message will appear on the display screen if the delay line is below the acceptable limit for accurate thickness measurements.
  • Page 26: Measurement Mode With A Datalogger (Tkg150)

    Extech Instruments 4 Measurement Mode Measurement Mode with a Datalogger (TKG150) For the TKG150 series datalogger the display screen in Measurement mode will look similar to the screen below: Measurement To continue in mode, follow these steps: Press the F1 key to select the Save T option.
  • Page 27: Measurement Mode Without A Datalogger (Tkg100)

    Press the F3 key to control the display backlight. (See: Display Backlight) Press the Menu / OK key to select the Menu option. (See: Menu section.) Note: The available choices for F1, F2, F3 are different for the TKG100 series thickness gauge.
  • Page 28 Extech Instruments Also, the file symbol, filename, ID: xx,xxxx are not displayed in the TKG100 series thickness gauge. TKG100_TKG150 Ultrasonic Thickness Gauge User Guide v2.8 11/13...
  • Page 29: Measurement Mode Setup Options

    Measurement Mode Setup Options The following modes are available (please note that in the TKG100 model the Datalogger mode is not available): To select the Measurements setup mode, first access the main MENU (using the MENU/OK button from the Home screen), and then follow these steps: Use the arrow buttons to highlight MEASUREMENTS and then press Menu OK.
  • Page 30 (indicated in the top right hand of the display screen) to change the setting of the highlighted parameter. The entire list of available MEAUREMENT MODE parameters for the TKG100 and TKG150 (red items TKG150 only) are as follows: FAST...
  • Page 31: Using The Clock

    Setting the Clock The TKG series thickness gauges have an internal real- time clock that includes the year, month, day and time. This option appears under the Initial Settings option. The entire list of available clock setup options are as follows: TIME FORMAT 12 HR...
  • Page 32 Extech Instruments previous setup. To change the Time Format, select TIME FORMAT using up/down arrow keys. Use the left/right arrow keys to select between 12 HR and 24 HR options. To change the Date Format, select DATE FORMAT using up/down arrow keys. Use the left/right arrow keys to select between MM/DD and DD/MM options.
  • Page 33: Setting Lcd Contrast

    To set Day, select the DAY option using up/down arrow keys. Use left/right arrow keys to decrease/increase the day value. You can set the day value from 01 to 31. To set Month, select the MONTH option using up/down arrow keys. Use left/right arrow keys to select a Month from JAN to DEC.
  • Page 34: Using The Hold Option

    Extech Instruments  Not enough couplant  Surface is too rough or rusty  Low Transducer sensitivity  Temperature is too high  Dis-bond exists between the coating and the steel In the event you try to save a reading in LOS with the TKG150, the Notes screen will automatically appear so that a reason is given for saving an LOS.
  • Page 35: Using The Fast Option

    Using the Fast Option The Fast option is available by accessing the main MENU, selecting MEASUREMENTS, and then selecting FAST from the list. To select the Fast option, follow these steps: Use the up and down arrow keys to highlight the Fast option and press the OK key.
  • Page 36 Extech Instruments In the Measurement the selected Fast option setting is displayed similar to the screen below: Helpful Hint: When using the FASTMIN mode, if you press the F2 key “Freeze” prior to LOS, the gauge will display the last minimum prior to the LOS so that the actual couplant on the end of the transducer is not registered as a legitimate minimum reading.
  • Page 37 value. If you select the On parameter from the Fast option, the LOS will work normally and not show the measured reading. The screens below reflect different parameters selected from the Fast option. For example, the top screen shows the FASTMAX with LAST MAX measurement on LOS, and the bottom screen shows the FASTMIN with LAST MIN measurement on LOS.
  • Page 38: Using The Gain Option (Tkg150 Only)

    Extech Instruments To reset the tracking of LAST MAX and LAST MIN measured value press the OK (Menu) key. Using the Gain Option (TKG150 only) Gain refers to an increase in signal power (echo height) and is typically measured in decibels (dBs). The Gain function is useful for setting a reference level where by making it easier to add or subtract gain.
  • Page 39: Using The Diff Option

    In the Measure mode screen the Gain option that you select is displayed on the center/right side of the display. Using the Diff Option There are 3 Diff options:  ABSOLUTE: Absolute, where the gauge will display a thickness value as an absolute number. For example, diff abs = 0.500”...
  • Page 40: Using The Alarm Option

    Extech Instruments 5 Alarm Options Using the Alarm Option The TKG series thickness gauges offers various alarm settings to alert you of low or high thresholds via audible sounds, display flashes, vibrations (if on) and illuminating keypad. There are several alarms options, which include: ...
  • Page 41 Quick Access mode by using the right or left keys once Alarm is highlighted). For the TKG100 model you can also press the F1 (Alarm) key to go directly to the Alarm option without having to go through the Main MENU/Measurement Mode list.
  • Page 42 Extech Instruments the previous page. Use the up and down arrow keys to select the desired alarm type and then press the OK key. The following screen below appears (screens will differ slightly based on alarm type selection). Use the up and down arrow keys to...
  • Page 43 Visual and Audible Alarm The Visual Alarm conditions can be viewed from either a red, yellow or green LED that light up under the F1, F2, or F3 keys on the keypad. As an example, when using the 5 step English test block with thicknesses of 0.100, 0.200, 0.300, 0.400 and 0.500 inches, if you enter alarm high-low you will first need to enter the high value.
  • Page 44: Vibra' Alarm (Tkg150 Only)

    Extech Instruments The Audible Alarm turns the beeper on the gauge either on or off based on the same alarm conditions used for visual alarm stated above. ‘Vibra’ Alarm (TKG150 only) The gauge will physically vibrate when the Vibrating Alarm is active based on the same alarm condition used for visual alarm.
  • Page 45 Visual, Audible, and Vibrating alarm enabled .)) symbol appears next to the alarm icon. The letter H (High) appears in the alarm icon Alarm condition has occurred meaning the measured reading is greater than the High Alarm limit. The letter H (High) appears in the alarm icon.
  • Page 46 The screen above has the Vibrating alarm enabled as the .)) symbol appears next to the alarm icon. Note: For the TKG100, the F1, F2, and F3 choices are different and the file symbol or the filename along with ID# are not shown.
  • Page 47: Using The Echo-To-Echo Option (Tkg150 Only)

    6 Using the Echo-to-Echo Option (TKG150 only) The Echo-to-Echo option allows you to make measurements between two consecutive back-wall echoes. Therefore, a good usage of the Echo-to-Echo option is for measuring through coatings to measure only the true metal thickness. There are two Echo-to-Echo options you can select, which are as follows: ...
  • Page 48: Special Gauge Functions

    Extech Instruments 7 Special Gauge Functions The TKG series thickness gauges have many special functions that go beyond the basics, mainly in the TKG150 model. This section will discuss these special gauge functions in detail. Using the Save Option (TKG150 only) You are able to save your data via the Save option.
  • Page 49 If the Notes option is set to Always then the thickness reading is stored at the ID location [Linear], [Row], [Col] with notes. To change the Notes settings, select the Notes option from the Datalogger screen in the MAIN Menu. Now when you press Save T (T for Thickness) to store a reading in measurement mode, the following Notes options will be presented:...
  • Page 50: Using The Freeze Option

    Extech Instruments Using the Freeze Option You can freeze your data via the Freeze option. To use the Freeze option, follow these steps: From the Measurement mode screen, press [F2] (Freeze). The following screen comes up after pressing [F2] (Freeze) whereby showing the flag “Freeze”...
  • Page 51 TKG150 Datalogger version TKG100 Non-datalogger version Press [F1] (Save) to save the reading. (See Using the Save Option for more information.) Press [F2] (Unfreeze) to disable the Freeze mode Press [F3] (Dir) to perform file operations. (See Using the Directory Option for more information.)
  • Page 52 Extech Instruments TKG150 Datalogger version TKG100 Non-datalogger version TKG100_TKG150 Ultrasonic Thickness Gauge User Guide v2.8 11/13...
  • Page 53 Using the Menu Option For the TKG100 and TKG150 models you can operate in Measure mode. To use the Measure mode, follow these steps: Press the Menu (OK) key to select the Menu options from Measure mode. Measure mode of the TKG150 Datalogger version...
  • Page 54 Extech Instruments The following Menu options are available: TKG150 Datalogger version TKG100 Non-datalogger version TKG100_TKG150 Ultrasonic Thickness Gauge User Guide v2.8 11/13...
  • Page 55: Using Datalogger Directory Mode (Tkg150)

    8 Using Datalogger Directory Mode (TKG150 only) Note: This section applies only to the TKG150 Thickness Gauge with a datalogger. If you do not have a datalogger, you may skip this section. To use the datalogger Directory mode, follow these steps: From the Measurement mode screen press [F3] (Dir) to select the Directory mode.
  • Page 56 Extech Instruments Press the OK key to select a file in the directory view. To create a custom file see Creating Custom Files section. To review thickness readings, see Reviewing a File section. To rename custom file, see Renaming a File section.
  • Page 57 To create copy of the existing file structures, see Copying a File section. 10. To delete a custom file, see Deleting a File section. The first three files cannot be deleted or renamed. These files are factory default files available for you to start storing thickness reading or make a quick copy of the file structure (Linear or Grid [Col, Row]) in order to start saving data in the...
  • Page 58 Extech Instruments Managing the Datalogger File System The TKG150 supports the following two files types:  Linear file : Consists of ID# from 0001 to 5000. The extended memory option consists of ID# from 1 to 500000.  Grid file : Consists of ID# as COLUMN and ROW.
  • Page 59 A,1 to A,156 then B,1 to B,156 then Z,1 to Z,156 then AA,1 to AA,156 then AB,1 to AB,156 then AF,1 to AF,156 If a new GRID file with END COLUMN = ZZ, then the maximum ROW you can slew up to is limited to 5000/(26 for A to Z + 26*26 for AA to ZZ) = 7.
  • Page 60 Extech Instruments Note that there are two different types of Grid files: ROWINC and COLINC. The above two examples show how the ID locations are incremented in ROWINC type file only. Here the Row value is incremented first, while updating ID location. Whereas, in a COLINC type file, Column value is incremented first.
  • Page 61 Creating Custom Files To create a custom file in the datalogger, follow these steps: Press [F3] (Dir) from the Measurement mode screen. The following screen appears: Use the up and down arrow keys to highlight the next Empty file in the list.
  • Page 62 Extech Instruments Press the OK key and select RENAME, the following screen appears: Use the up, down, left, and right arrow keys to select the characters of the filename. Press the OK key to accept the selected characters. If you make an error, press [F2] (Del) to delete the last character of the entered filename.
  • Page 63: Selecting A Custom Linear File

    Selecting a Custom Linear File When you select a LINEAR file format you are prompted to select START and END file points. With the extended memory, you can store up to 100,000 readings allowing you more total files. Pressing the F3 (Done) key again will take you to the Measure mode displaying the new filename in the top left corner and the next ID location in the top right corner.
  • Page 64 Extech Instruments Use the up and down arrow keys to select the END ROW option. Now press left or right arrow key to change the END ROW value. Then press down arrow key to select END COL option. Press left or right arrow key to change the END COL value.
  • Page 65: Reviewing A File

    Reviewing a File To review a file from the Measurement mode screen in datalogger version only, follow these steps: Press [F3] (Dir) from the Measurement mode screen. The following screen appears: Select the file you would like to review using the up and down arrow keys and press the OK key.
  • Page 66: Reviewing A Linear File

    Extech Instruments ABCD CLEAR ALL READI NGS COPY DELETE FI LE RENAM E REVI EW DETAI LS REVI EW GRI D EXI T The three factory default files have only 3 options: Clear All Readings, Copy, and Review Details. Custom files have 5 options: Clear all readings, Copy, Delete file, Rename, Review details, and Review Grid.
  • Page 67: Reviewing A Grid File

    F2 (Clear). Note that once a reading is cleared, it cannot be restored. To store another reading in the file location that has been emptied by the Clear operation, highlight that location by pressing up or down arrow key. Then press OK to go to the measure mode.
  • Page 68 Extech Instruments The following screen appears: Use the up and down arrow keys to select the file you would like to rename and press the OK key. Note: The Rename option is not available for the first three factory default files.
  • Page 69 Use the up, down, left, and right arrow keys to select the characters of the filename. Press the OK key to accept the selected characters. If you make an error, press [F2] (Del) to delete the last character of the entered filename. Press [F3] (Done) when you are finished entering the filename.
  • Page 70 Extech Instruments Clearing a File To clear (or delete) a single thickness reading at a time from a file, see Reviewing a File section. To clear all the readings from a file, use the clear file option. To do this, from measurement mode, press [F3] (Dir) for the Directory mode.
  • Page 71: Copying A File

    YES/NO/EXIT confirmation: Press [F1] (Exit) to exit without clearing the thickness readings. Press [F2] (Yes) to clear the stored thickness from the entire file. Press [F3] (No) to exit without clearing the thickness readings. If the file is already clear, the unit will show „NO READINGS TO CLEAR.‟...
  • Page 72 Extech Instruments The following screen appears: Use the up and down arrow keys to select a file and press the OK key. Any file structure (factory default or custom) can be copied. Note: When you copy a file only the structure of the file ID# is copied and not the associated file readings.
  • Page 73 Press up or down to select an Empty file and press OK to assign it a new filename. The following screen appears. Use the up, down, left, right arrow keys to select the characters of the filename and press the OK key to accept the selected characters. If you make an error, press [F2] (Del) to erase last character before the cursor.
  • Page 74: Deleting A File

    Extech Instruments Deleting a File To delete a file, from the Measurement mode screen in the datalogger version only, follow these steps: Press [F3] (Dir) for the Directory mode. The following screen appears: Use the up and down arrow keys to select the file to be deleted and press the OK key.
  • Page 75 confirmation prompt for YES/NO/EXIT appears. Press [F1] (Exit) to exit the screen without deleting the file. Press [F3] (No) to exit the screen without deleting the file. Press [F2] (Yes) to delete the file and remove the file from the directory.
  • Page 76: Technical Specifications

    Extech Instruments 7 Technical Specifications Specifications for the TKG100/TKG150 Ultrasonic Thickness Gauges Size: 5" (127 mm) (L) x 3" (76.2 mm) (W) x 1.25" (31.75 mm) (H) Weight: 8 OZ (.23 kg) Thickness range: 0.008 - 20 inches (0.20 mm - 508 mm), depending on material, temperature and transducer selection Material velocity calibration range: 0.200-0.7362 in/μS (0.508-18.699...
  • Page 77 Differential Mode: Displays the difference from the actual thickness measurement and a user entered reference value Alarms: Minimum/Maximum depth, vibralarm, audible/visual indications as well as keypad illumination Illuminating keypad: F1 = Red, F2 = Yellow and F3 = Green for easy, go/no-go testing (Patent Pending) Automatic probe wear indicator ( Transducer attendant): Automatically informs the operator to replace the transducer (Patent Pending)
  • Page 78 Extech Instruments Specification differences for TKG100 and TKG150 Models Item Specification Thickness range 0.008-0.20 inches (20mm-508mm) Auto at power up with listed numeric value. Delay line zero measurement Ideal for correcting delay line Displays minimum or maximum thickness Scan mode...
  • Page 79: Software Options

    8 Software Options Echo-to-Echo (TKG150 only) The Echo-to-Echo feature should and can be used within the thickness range of the chart listed below. To calibrate in Echo-to-Echo, first press MENU OK, scroll to set up, press MENU OK and find where E to E is displayed. Use the right arrow to turn ON or press MENU OK and scroll to ON.
  • Page 80: B-Scan (Tkg150 Only)

    Extech Instruments B-Scan (TKG150 only) The B-Scan represents a cross sectional view on the test piece. The simplest example is to show a 5 step test block. To turn on the B-Scan on the TKG150, press MENU/OK, scroll to MEASUREMENTS and press OK, select B-scan and press MENU/OK.
  • Page 81: Technical Assistance

    Calibration, Repair, and Customer Care Services FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we sell. NIST certification for most products is also provided. Call the Customer Service Department for information on calibration services available for this product. Annual calibrations should be performed to verify meter performance and accuracy.
  • Page 82 Calibrage, réparation et services après-vente FLIR Systems, Inc. offre des services de calibrage et de réparation pour les produits Extech Instruments que nous commercialisons. Nous fournissons également une certification NIST pour la plupart des produits. Contactez notre service client pour toute information sur les services de calibrage disponibles pour ce produit.
  • Page 83 FLIR Systems, Inc., ofrece servicios de reparación y calibración para los productos que vendemos de Extech Instruments. Además ofrecemos certificación NIST para la mayoría de los productos. Llame al Departamento de Servicio al Cliente para solicitar información de calibración para este producto.
  • Page 84 Extech Instruments TKG100_TKG150 Ultrasonic Thickness Gauge User Guide v2.8 11/13...

This manual is also suitable for:

Tkg150

Table of Contents