Deepak kanujiya
January 18, 2025

I have face error 7 troubling problem please giving me the solution

1 comments:
Mr. Anderson
February 3, 2025

For error 7 (Turbine problem) on the SEFAM DreamStar Intro, the solution is as follows:

1. Unplug the device from all sources of electrical power.
2. Check that there is no foreign object or signs of water in the air outlet.
3. Plug the device back in.

If the error persists, contact Technical Support.

This answer is automatically generated

Summary of Contents for SEFAM DreamStar Intro

  • Page 1: User Manual

    Clinician and Home Care Provider User Manual DreamStar™ Product Range Upgradeable Humidifier Option DreamStar™ Intro DreamStar™ Intro Evolve DreamStar™ Info DreamStar™ Info Evolve DreamStar™ Auto DreamStar™ Auto Evolve DreamStar™ is a brand registered by SEFAM.
  • Page 2 Manufacturing plant and technical agency: SEFAM SEFAM 144 AV CHARLES DE GAULLE 10 ALLEE PELLETIER DOISY 92200 NEUILLY SUR SEINE 54600 VILLERS-LES-NANCY FRANCE FRANCE TEL: +33 (0) 3 83 44 85 00 www.sefam-medical.com en GB – 2 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 3: Table Of Contents

    Functional pneumatic block diagram ....49 Accessing the Settings menu ......23 Essential performance requirements for Settings menu ..........23 electromagnetic compatibility ......50 Service menu ...........26 Other settings and information ....29 DreamStar Intro, Info, Auto Clinician Manual en GB – 3...
  • Page 4: Before You Start

    The DreamStar™ product range devices are intended to treat patients suffering from Obstructive Sleep Apnea Syndrome (OSA) in patients over 30 kg (66 lbs.) and breathing spontaneously. They can be used at home or in a sleep center. en GB – 4 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 5: Contra-Indications

    Use only those accessories which can guarantee the patient’s treatment pressure and reduce re- breathing of CO . When a facial mask is necessary, always use a mask which is equipped with an anti- asphyxia valve. DreamStar Intro, Info, Auto Clinician Manual en GB – 5...
  • Page 6: Available Features

    The DreamStar™ device has a patient reminder feature to help the patient remember key events such as filter and mask changes, device inspections, and memory card compliance data downloads on the memory card. This feature is described under “Reminders about key events”, on page 31. en GB – 6 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 7: Comfort Calibration

    Ramp feature in AutoCPAP mode Ramp feature in AutoCPAP mode Ramp feature in CPAP mode Ramp feature in CPAP mode DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 7...
  • Page 8 CPAP or Auto or Auto-CPAP mode. End of ramp / detection of events detection of events End of ramp / / Max ramp time en GB – 8 DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual...
  • Page 9: Device Settings

    A session is started after the device has been on for at least 5 minutes without interruption. Water chamber: chamber containing the water necessary for humidification. DreamStar Intro, Info, Auto Clinician Manual en GB – 9...
  • Page 10: Settings Ranges

    (Evolve configuration) and the heat level is set from 1 to 10, the heated humidifier begins operating automatically. The heated humidifier indicator then displays on the screen. When the heat level is set at 0, the humidifier symbol is displayed without steam en GB – 10 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 11 Access to setting of language for the patient (locked) (locked) (autorised) Ramp type Default Time ramp Indexed ramp Indexed ramp Indexed ramp Ramp feature (T.Ramp) (I.Ramp) (I.Ramp) (I.Ramp) DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 11...
  • Page 12: Accessing The Device Menus

    (see table entitled “Description of symbols displayed on the screen“). en GB – 12 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 13 22. The "General Information" menu for the devices is described on page menu for the devices is described on page 29. DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 13...
  • Page 14: How To Set Dreamstar™ Intro / Intro Evolve Device

    Pressure Alter: feature that can be enabled to allow the patient to adjust the prescribed pressure ±1 Press at the bottom of the page or to return to the page header. Press to move to the next page. en GB – 14 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 15: Service Menu

    Software version: Version of the integrated device software. The version of the integrated device software can be viewed but not modified. Press at the bottom of the page or to return to the page header. Press to move to the next page. DreamStar Intro, Info, Auto Clinician Manual en GB – 15...
  • Page 16 The reminder expiration date is recalculated when you change the frequency. Press at the bottom of the page or to return to the page header. Press to move to the next page. en GB – 16 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 17 Press at the bottom of the page or to return to the page header. Press to exit the menu. DreamStar Intro, Info, Auto Clinician Manual en GB – 17...
  • Page 18: How To Set Dreamstar™ Info / Info Evolve Device

    Pressure Alter: Feature that can be enabled to allow the patient to adjust the prescribed pressure ± 1 Press at the bottom of the page or to return to the page header. Press to move to the next page. en GB – 18 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 19 See "Specific calibration of the pneumatic configuration" paragraph, on page 32. Press at the bottom of the page or to return to the page header. Press to exit the menu. DreamStar Intro, Info, Auto Clinician Manual en GB – 19...
  • Page 20: Service Menu

    The date can be set only when the device is in standby. Note: Press at the bottom of the page or to return to the page header. Press to move to the next page. en GB – 20 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 21 Press to move to the next page. to move to the next page. DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 21...
  • Page 22 Pressure unit: the unit of measurement used for the displayed pressure can be changed from cmH O to hPa, and vice versa. Press at the bottom of the page or to return to the page header. Press to exit the menu. en GB – 22 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 23: How To Set Dreamstar™ Auto / Auto Evolve Device

    Pressure Alter: feature that can be enabled to allow the patient to adjust the prescribed pressure ± 1 Press at the bottom of the page or to return to the page header. Press to move to the next page. DreamStar Intro, Info, Auto Clinician Manual en GB – 23...
  • Page 24 Then press Then press to access the next desired desired menu parameter. menu parameter. parameter parameter setting. setting. en GB – 24 DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual...
  • Page 25 If no calibration is PRESENT, the calibration goes back to the default calibration. See "Specific calibration of the pneumatic configuration" paragraph, on page 32. Press at the bottom of the page or to return to the page header. DreamStar Intro, Info, Auto Clinician Manual en GB – 25...
  • Page 26: Service Menu

    The date can be set only when the device is in standby. Note: Press at the bottom of the page or to return to the page header. Press to move to the next page. en GB – 26 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 27 Press to move to the next page. to move to the next page. DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 27...
  • Page 28 Pressure unit: the unit of measurement used for the displayed pressure can be changed from cmH O to hPa, and vice versa. Press at the bottom of the page or to return to the page header. Press to exit the menu. en GB – 28 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 29: Other Settings And Information

    Press the On/Off button to exit the menu. to exit the menu. DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 29...
  • Page 30: Setting The Heat Level (If Water Chamber Included)

    Press to move to the next page. to move to the next page. en GB – 30 DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual...
  • Page 31: Reminders About Key Events

    Press to exit the menu. DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 31...
  • Page 32: Specific Calibration Of The Pneumatic Configuration

    To do this, insert the memory card into the card reader supplied with the software, then connect the reader to a USB port on your computer. To retrieve the compliance data, consult the DreamStar™ Analyze software user manual. CAUTION Use approved memory cards only. en GB – 32 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 33 To configure a "service card", consult the DreamStar™ Analyze software user manual. To be able to use the service card with the device, hold down the ramp button pressed while inserting the card. DreamStar Intro, Info, Auto Clinician Manual en GB – 33...
  • Page 34: Cleaning And Maintenance

    Grip zone Inner tubing Upper part Gasket Base Center tab en GB – 34 DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual...
  • Page 35 • Reinstall the cover or the water chamber. Reinstall the cover or the water chamber. DreamStar Intro, Info, Auto Clinician Manual DreamStar Intro, Info, Auto Clinician Manual en GB – 35...
  • Page 36: Monthly

    Never use the device without making sure that an air intake filter is installed. • Check that the heating element is completely dry before reconnecting the device to the power supply. en GB – 36 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 37: Material Disinfection

    Keep the disinfected and dried water chamber in a single use clean bag, sealed until a new use. Discard the water chamber if showing any damage and replace it with a new one. Note : DreamStar Intro, Info, Auto Clinician Manual en GB – 37...
  • Page 38: What Should Not Be Used

    Do not use any pre-disinfecting or disinfecting product not active on the Mycobacterium tuberculosis (BK). • Never disinfect the internal circuit of air flow using disinfecting product or with a Airway Surface Disinfection System (DSVA). en GB – 38 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 39: Troubleshooting Tips

    If the error persists, contact Technical Support. Er 06 CheckSum code error. Unplug the device from all sources of electrical power. Plug it in again and turn on. If the error persists, contact Technical Support. DreamStar Intro, Info, Auto Clinician Manual en GB – 39...
  • Page 40 The device is operating without employing the heated humidification Er 102 high. feature. Unplug the device from all sources of electrical power. Plug it in again and turn on. If the error persists, contact Technical Support. en GB – 40 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 41 Unplug the device from all sources of electrical power. Plug it in again and device. turn on. If the error persists, contact Technical Support. DreamStar Intro, Info, Auto Clinician Manual en GB – 41...
  • Page 42: Technical Specifications

    +5° C to + 35 °C (41 ºF to 95 ºF) with humidifier Relative humidity: between 10% and 95% without condensation 0 − 2 700 m (0−9,000 ft) approximately Altitude range: en GB – 42 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 43: Transportation And Storage Conditions

    Maximum liquid fill level: Indicated on water chamber CE marking DreamStar™ Intro, DreamStar™ Intro Evolve : 2011 DreamStar™ Info, DreamStar™ Info Evolve : 2011 DreamStar™ Auto, DreamStar™ Auto Evolve : 2011. DreamStar Intro, Info, Auto Clinician Manual en GB – 43...
  • Page 44: Symbols On The Device

    European Parliament and of the Council on waste electrical and electronic of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE). en GB – 44 DreamStar Intro, Info, Auto DreamStar Intro, Info, Auto Clinician Manual...
  • Page 45: Disposing Of The Device At The End Of Its Life

    (see "Symbols on the device" page 44). In other countries, this device must be processed following local regulation. Unsuitable disposal of the device at the end of its life could harm the environment. DreamStar Intro, Info, Auto Clinician Manual en GB – 45...
  • Page 46: Nf En Iso 17510-1:2007-Compliant Features

    Sound pressure measured in accordance with ISO 17510-1:2009: 30 dBA with humidifier, 33 dBA without humidifier Sound power measured in accordance with ISO 17510-1:2009: 38 dBA with humidifier, 41 dBA without humidifier en GB – 46 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 47 0.05 0.04 0.26 -0.27 -0.13 0.01 -0.30 -0.22 -0.06 -0.31 -0.20 -0.08 -0.28 -0.25 -0.08 pressure (cmH Delta relative to dynamic 0.04 0.04 0.06 0.06 0.07 0.09 0.08 0.11 0.12 0.09 0.12 0.15 0.13 0.15 0.16 pressure (cmH DreamStar Intro, Info, Auto Clinician Manual en GB – 47...
  • Page 48 Sound pressure measured in accordance with ISO 17510-1:2009: 30 dBA with humidifier, 33 dBA without humidifier Sound power measured in accordance with ISO 17510-1:2009: 38 dBA with humidifier, 41 dBA without humidifier en GB – 48 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 49: Functional Pneumatic Block Diagram

    Functional pneumatic block diagram * except DreamStar™ Intro or DreamStar™ Intro Evolve DreamStar Intro, Info, Auto Clinician Manual en GB – 49...
  • Page 50: Essential Performance Requirements For Electromagnetic Compatibility

    25 cycles for 25 cycles < 5% U < 5% U (>95% dip in U (>95% dip in U for 5 s for 5 s en GB – 50 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 51 DreamStar™ product range. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V / m. DreamStar Intro, Info, Auto Clinician Manual en GB – 51...
  • Page 52 Ref.: M-214831-11 6.5 ft (2m) RS 232/mini USB 2 m cable (except with DreamStar™ Intro) Ref.: M-214831-12 50 ft (15m) RS 232/mini USB 15 m cable (except with DreamStar™ Intro) en GB – 52 DreamStar Intro, Info, Auto Clinician Manual...
  • Page 54 Manufacturer: Manufacturing plant: SEFAM SEFAM 144 AV CHARLES DE GAULLE 10 ALLEE PELLETIER DOISY 0459 92200 NEUILLY SUR SEINE 54600 VILLERS-LES-NANCY FRANCE FRANCE : M-155DFU00-60 Revision 8 2015-11...