VPD 3in1 User Manual

Multiparameter for self testing use

Advertisement

Quick Links

Multiparameter for
self testing use
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3in1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for VPD 3in1

  • Page 1 Multiparameter for self testing use USER MANUAL...
  • Page 2: Table Of Contents

    Contents: •.3in1..meter •.3in1. glucose.strips.10.pcs •.Lancing.device.1.pc •.Sterile.lancets.10.pcs. •.User.manual •.Rapid.guide Introduction..................1 System.................. 1 Meter..................2 Display................... 3 Mode..................3 The.measurement................4 Coding.the.meter..............4 How.to.obtain.a.drop.of.blood............ 6 Application.of.blood..............7 Procedure.to.test.glucose............7 Procedure.to.test.cholesterol.and.triglycerides......8 Warnings....................9 Strips....................10 Control.solutions................11...
  • Page 3: 3In1

    System Thank.you.for.choosing.the.3in1..system. 3in1..is.a.new.system,.useful.for.the.monitoring.of.glucose,. cholesterol.and.triglycerides.in.the.blood. 3in1..has.two.different.technologies: •.amperometric,.with.3in1..glucose.electrodes.strips. •.reflectometric,.with.3in1..cholesterol.and.triglycerides.strips. Self–monitoring.of.glucose,.cholesterol.and.triglycerides.could.be. helpful,.but.the.treatment.to.be.applied.should.be.agreed.with.the. physician..The.3in1..system.is.plasma-calibrated. Information about system line 3in1. The.3in1..system.line.includes.everything.necessary.for.self.testing: •.3in1..meter •.3in1..strips.10.pcs •.Lancing.device.1.pc •.Sterile.lancets.10.pcs •.3in1..control.solutions.-.available.by.extra.order •.3in1..data.manager.software.-.by.special.request. Read.carefully.the.instructions.for.use.inside.the.box.of.3in1..strips. that.you.want.to.use.and.read.carefully.the.instructions.for.the.use.of. the.lancing.device.that.are.included.in.the.3in1..meter.box.
  • Page 4: Meter

    Meter Display.–.shows.test.results,.messages,.and.results. stored.in.memory. Power.Button.–.press.this.button.to: Turn.on.the.meter.or.turn.off.the.meter. Enter.the.SET.mode.in.the.STANDBY.mode. Go.to.the.next.step.in.the.SET.mode. Exit.the.SET.mode.and.enter.the.STANDBY. mode. “+”.Button.--.press.this.button.to: Enter.the.MEMORY.mode. Recall.previous.stored.results.in.the.MEMORY. mode. Change.parameters.displayed.and.increase. values.in.the.SET.mode. “-”.Button.–.press.this.button.to: Enter.the.MEMORY.mode. . R ecall.next.stored.results.in.the.MEMORY. mode. . C hange.parameters.displayed.and.decrease. values.in.the.SET.mode. . R eset.the.test.results. Test.Strip.Slot.–.insert.test.strip.here..The.meter. automatically.enters.the.TEST.mode.when.a.test. strip.is.inserted. PC.Connector.–.transfers.data.from.the.meter.to.a. computer. Cover.–.open.this.for.cleaning.. Battery.Cover. Put.your.finger.on.this.to.let.the.battery.cover.slide. Data-Chip.Slot.–.insert.the.data-chip. Ejector.–.press.to.discharge.the.used.strip.
  • Page 5: Display

    Display Data-chip.inserting Meter.mode Low.battery.warning Test.results.and.messages Unit.of.measurement Test.analytes Abnormal.temperature.warning Time Day.and.month Alarm Mode The.3in1..meter.has.three.modes. You.can.check.the.present.mode.of.the.meter.by.seeing.the.top.of display. 1. Standby Mode Press.and.release.the.power.button. A.strip-shaped.symbol.will.twinkle.and.the.meter.will.display.the.data. chip.number,.time.and.date. 2. Memory Mode Press.and.release.the.“+”.or.“-”.button.in.the.standby.mode. The.meter.will.display.a.“MEM”.symbol.on.the.top.of.the.display. 3. Set Mode Press.and.release.the.power.button.in.the.standby.mode. Press.again.the.power.button.for.3.seconds. The.meter.will.display.a.“SET”.symbol.on.the.top.of.the.display.
  • Page 6: The.measurement

    Coding the meter ....Attention!..The.code.chip.is.not.necessary.for.glucose.measuring. The.3in1..meter.uses.a.data-chip.that.is.inside.the.strips box..It.will.be.necessary.to.change.the.data-chip.in.the.following cases: before.using.a.new.strips.box, when.you.want.to.change.the.analyte. Remove.the.data-chip.from.the.strips.box. Insert.the.data-chip.in.the.proper.space.in.the.back.part.of.the. meter. Gently.press.the.data-chip.up.to.the.complete.insertion.
  • Page 7 Press.power.button.and.check.the.three.digit.number.is.the.same. of.the.one.printed.on.the.strips.vial. If.the.data-chip.is.not.inserted,.it.is.impossible.to.perform.the. measurement. Analytes:. CH = CHOLESTEROL (GREEN) TGL=TRIGLYCERIDES (YELLOW)
  • Page 8: How.to.obtain.a.drop.of.blood

    When switching on the meter, check that the date and the hour are correct. How to obtain a drop of blood To.perform.a.measurement.with.the.3in1..system,.it.is.necessary.to withdraw.a.drop.of.capillary.blood.by.following.the.procedure: •. Wash.the.hands.with.warm.water.to.increase.blood.circulation.in. the.fingers.to.simplify.the.withdrawal.of.blood. •. Dry.your.hands. •. Check.that.the.finger.to.be.pricked.is.completely.dried. •. Prepare.the.lancing.device.that.is.inside.the.3in1..kit.by.inserting. the.sterile.lancet. Read.carefully.the.instruction.for.use.of.the.lancing.device. contained.in.the.meter.kit. Use.a.new.lancet.for.each.prick.and.use.the.lancing.device.only.on. yourself. Before.insertion,.verify.that.the.lancets.and.the.protection.of.the. lancets.are.intact. Only.an.intact.lancet.protection.can.guarantee.that.the.lancet.remains.
  • Page 9: Application.of.blood

    Application of blood Read.carefully.the.instructions.for.the.use.strips.that.you.want.to.use. to.perform.the.measurement.. Press.the.Power.button..Insert.data-chip. only.for.cholesterol.and.triglycerides. measurements..The.symbol.that.appears. on.the.display.(3in1..for.glucose.or.a. three.digit.number.for.cholesterol.and. triglycerides).has.to.match.the.symbol. printed.on.the.strips.vial..If.they.do.not. match,.contact.Customer.Service. Glucose.measurements.do.no.require. coding.. Take.a.strip.from.the.vial.and.insert.it.in.the.meter.with.the.printed. or.coloured.part.on.the.upper.side..You.will.hear.a.click.and.the. meter.will.emit.an.acoustic.signal.. 3. Procedure.to.test.glucose: 3.a.Perform.the.prick.on.the.finger.with.the. lancing.device.. Apply.a.drop.of.blood.on.the.part.of.the. strip.that.comes.outside.the.meter.. Let.the.strip.aspirate.the.blood.up.to.see. the.moving.segments.on.the.display..If the volume is not sufficient the measurement does not start.
  • Page 10: Procedure.to.test.cholesterol.and.triglycerides

    Procedure.to.test.cholesterol.and.triglycerides 3.b.Sit.down.for.one.minute.with.the.arm.along.the.body.. Massage.the.finger.to.be.pricked.in.order.to.increase.the.blood. flow.. Prick.the.finger,.remove.the.first.drop.with.sterile.gauze.and. gently.press.the.finger.to.obtain.a.second.drop..Without.touching. and.pressing.the.finger.on.the.strip,.apply.the.drop.on.the.white. circular.area.of.the.strip.that.comes.out.of.the.meter.. Wait.until.moving.segments.will.appear.on.the.display.. The.meter.will.emit.an.acoustic.signal. and.the.result.will.be.visualised.on.the. display.and.automatically.stored.in.the. memory.of.the.meter.(refers.to.the. instruction.for.use,.that.you.are.using. to.know.the.reading.time). Press.the.ejector.key.to.remove.the.strip.that.you.used,.pointing. the.meter.down.for.the.low.
  • Page 11: Warnings

    Warnings If.the.results.obtained.do.not.correspond.to.the.well-being.of. the.user,.proceed.with.a.check.up.of.the.3in1. instrument.and. contact.your.physician..If.the.results.are.too.high.(Hi).or.too.low. (Lo),.contact.your.physician.immediately. When.the.measuring.is.finished,.it.is.possible.to.turn.OFF.the.3in1. meter..The.instrument.will.shut.down.automatically.after.60.seconds.if. not.used.for.other.operations. The.measured.value,.along.with.date.and.time,.will.be.saved.in. position.1.in.the.memory.of.the.instrument.. The.values.previously.measured.will.be.saved.in.positions.2,.3.etc.. starting.from.the.most.recent..Descriptions.on.how.to.recall.data.from. the.memory.and.how.to.erase.memorised.values.are.presented.on. page.12. Strips.and.used.materials.have.to.be.properly. disposed,.avoiding.contamination..The.meter. has.to.be.disposed.following.the.rules.regarding. electric.and.electronic.waste.disposal. In.a.laboratory.and.in.a.hospital,.the.procedures. of.biological.safety.must.be.followed.when. handling.potentially.infected.material,.in. particular.when.measurements.are.performed. on.different.patients. To.avoid.infections.clean.the.instrument.regularly. and.dispose.of.the.used.strips.in.a.way.that. there.is.no.risk.of.infections.for.other.people.
  • Page 12: Strips

    Strips The.correct.use.of.3in1..strips.is.very. important.for.obtaining.accurate.results.. Therefore.it.is.important.to.keep.in.mind.the. following:. •. The.strips.are.sensitive.to.light.and. humidity:. keep.the.strips.only.in.their.original.vial. •. Protect.the.vial.from.direct.sunlight.and. keep.at.a.temperature.between.5.and. 30 °C.(41.-.86°F). •. Use.the.strips.immediately.after.removing. them.from.the.vial.and.close.the.vial. immediately •. The.strips.must.not.be.used.after.the.expiry. date. •. Write.the.date.of.the.first.opening.on.the. vial:. the.strips.must.be.used.within.90.days.after. the.vial.has.been.opened. •. Do.not.use.strips.that.are.damaged. •. Carefully.read.the.instructions.contained.in. the.package.of.strips. •. Take.measurements.only.with.original. 3in1..strips.
  • Page 13: Control.solutions

    Control Solutions Read.carefully.the.instructions.for.use.of.the.control.solution.that.you. want.to.use. It.is.possible.to.check.the.working.of.the.3in1..meter.by.using.the. 3in1..control.solutions. The.control.solutions.can.be.obtained.as.a.separate.accessory. through.specialised.retailers. The.control.solution.is.applied.on.the.strip.just.like.blood. When.testing.is.done.with.the.control.solution,.the.obtained.result. must.be.within.the.interval.as.shown.on.the.label.of.the.strip’s.vial. In.this.way,.it.is.possible.to.learn.how.to.use.the.3in1..properly and.at.the.same.time.to.check.if.it.is.functioning.properly. Write.the.date.of.opening.on.the.vial.when.the.solution.is.opened.for the.first.time. The.control.solution.must.be.used.within.90.days.from.first.opening,. if.the.expiry.date.is.valid. When.is.it.necessary.to.perform.a.check-up.with.control.solutions: •. When.suspecting.a.possible.defect.of.the.3in1..meter. •. When.suspecting.that.the.possibility.of.a.malfunction.or.wrong. measurement. The.result.obtained.with.the.control.solutions.has.to.be.deleted.by. the.memory.of.the.meter.in.order.to.not.interfere.with.the.average. calculation.of.the.glycaemia.
  • Page 14: Memory.functions

    Memory functions: recalling and deleting results from the memory The.3in1..meter.stores.up.to.500.test.results.for.all. analytes,.letting.you.to.review.them.in.the.order.from. the.most.recent.(“1”).to.the.oldest.(“500”). If.you.insert.the.glucose.data-chip.in.the.meter,.you. allow.to.recall.only.previous.glucose.test.results..The. same.applies.for.other.analytes.. Press.and.release.the.“+”.or.“-”.button.in.the. STANDBY.mode.. The.meter.will.enter.MEMORY.mode.and.the. average.of.test.results.is.displayed.. While.pressing.the.“+”.or.“-”.button,.memory. number.will.be.displayed..If.you.hold.down.the. “+”.or.“-”.button,.the.meter.scrolls.quickly.and. only.memory.positions.are.displayed..Result.is. not.shown.until.you.release.the.button..When. releasing.the.“+”.or.“-”.button,.the.corresponding. result.will.be.displayed.. Repeat.step.2.to.see.the.remaining.stored.results.. Keep.the.power.button.pressed.and.then.press. the.“-”.button.for.1.second.to.delete.one.test. result.stored.in.the.meter.. Press.“-”.button.for.5.seconds.together.with.the. power.button.to.delete.all.test.results.stored.in.the. meter.
  • Page 15: General.settings

    Modification of settings Press.and.release.the.power.button.to.select.the. month.. Press.and.release.the.“+”.or.“-”.button.up.to.get. the.correct.month.. Press.and.release.the.power.button.to.select.the. day.. Press.and.release.the.power.button.to.change.. Press.and.release.the.power.button.to.choose.the. date.format.. Press.and.release.the.“+”.or.“-”.button.to.choose. between.month-day.(mm-dd).or.day-month.(dd- mm).. Press.and.release.the.power.button.to.change.the. year..Press.and.release.the.“+”.or.“-”.button.up.to. reach.the.current.year.. The.meter.enables.to.change.the.year.from.2011. to.2025.. Press.continuously.the.power.button.for.3. seconds..The.meter.will.enter.in.the.SET.mode. and.it.will.be.possible.to.change.the.hour.. Press.and.release.the.“+”.or.“-”.button.up.to.reach. the.correct.hour.. Press.and.release.the.power.button.to.change.the. minutes..Press.and.release.the.“+”.or.“-”.button. up.to.get.the.correct.minutes..
  • Page 16: Setting.the.acoustic.signal

    Press.and.release.the.power.button.to.change. the.hour.format..Press.and.release.the.“+”.or.“-”. button.to.choose.between.12.or.24.hours.. Press.and.release.the.power.button. to.change.the.unit.of.measurement.. Press.and.release.the.“+”.or.“-”. button.to.choose.between.mg/dL. and.mmol/L.. Press.and.release.the.power.button.to.select. how.many.days.will.be.included.in.the.average. result..Press.and.release.the.“+”.or.“-”.button.up. to.get.the.desired.number.of.days.(0,7,14,21,28. days.-“0”.means.that.the.meter.will.not.show.any. average.of.the.measurements). Setting the acoustic signal Switch.on.the.meter.and. continuously.press.the.power. button.to.enter.in.the.SET.mode.. Press.and.release.the.power.button. up.to.see.“bp”.fixed.and.“On”. blinking.on.the.display. Press.the.“+”.or.“-”.button.to. activate.or.deactivate.the.acoustic. signal. Setting the alarms Follow.these.instructions.to.set.the.alarms: Press.and.release.the.power.button.. Press.the.power.button.continuously.to.enter.in. the.SET.mode..Press.and.release.up.to.see.the. icon.of.the.alarm.fixed.and.“0”.blinking.
  • Page 17: Enabling.data.transfer.to.pc

    Press.the.“+”.or.“-”.button.to.select.the.number.of. the.alarm.to.be.set.(0,.1,.2,.or.3). Press.the.power.button.again.to.select.the.hour. and.the.minutes.related.to.the.alarms.that.you.are. setting. You.can.exit.by.using.the.SET.mode.at.any. moment.by.pressing.continuously.the.power. button.for.three.seconds.. The.alarms.are.not.activated.if.they.are.set.to. 12.00.or.0.00. Enabling data transfer to PC All.the.test.results.that.are.stored.in.the.3in1..meter.can. be.downloaded.to.a.computer. Install.the.software.according.to.the.manufacturer’s.instructions. Insert.the.cable.into.the.USB.port.of.the.computer.and.the.jack.of. the.cable.into.the.meter. Press.and.release.the.power.button..“PC”.will.appear.on.the. meter’s.display. Set.up.your.computer.software.to.receive.information.from.the. meter.. After.the.data.has.been.transferred,.you.can.turn.off.the.meter.by. pressing.the.power.button.and.removing.the.jack.cable.from.the. port.of.the.meter.
  • Page 18: Maintenance.and.preservation.conditions

    For. taking. measurements. on. different. people,. like. in. a. hospital. for. example,. clean. the. instrument. after. every. measurement. with. a. 70%. alcohol. solution. before. proceeding.with.the.next.measurement. Preservation and measuring conditions To.assure.the.accuracy.of.each.measurement,.it.is.important.to.meet. the.following.conditions. The.measurement.with.3in1..strips.has.to.be.performed.between. 10.and.40°C.(50-104°F)..The.3in1..kit.(meter.and.strips).may.not. be.stored.at.temperature.lower.than.5°C.(41°F).or.higher.than.30°C. (86°F). Note. If.the.instrument.is.too.cold.or.too.hot.(e.g..after.being.in.a.car.at. very.low.or.high.temperatures),.allow.it.to.cool.down.or.to.heat. up.slowly.at.room.temperature.(about.20.minutes).
  • Page 19: Error.messages

    Error messages Error. Problem Solution message The.temperature.is.out. Place.the.instrument.in.an.area. of.the.allowed.range. with.temperature.between. 10°C.(50°F).and.40°C.(104°F). and.wait.for.the.instrument.to. equilibrate.with.the.external. temperature,.before.performing. the.measurement. The.strip.is.used.or. Use.a.new.strip.to.perform.the. the.reactive.area.is. measurement. coloured. The.blood.was.applied. Take.a.new.strip,.insert.it.in. on.the.strip.before. the.instrument.and,.only.after. waiting.for.the.drop. visualised.blinking.drop.on.the. blinking.on.the.display. display,.aspirate.the.blood.on. the.strip. The.reading.window. Open.the.plastic.cover.where.you. is.dirty. insert.the.strip.and.clean.the.glass. window.using.only.water..Dry.the. glass.without.scratching.it..Do.not. use.solvents.or.soap..Close.the. plastic.counter.and.repeat.the.test. The.strip.is.removed. Switch.off.the.meter.by.using. from.the.meter.without. the.power.button..Insert.the.strip.
  • Page 20: Error Messages

    Error messages Error. Problem Solution message Data-chip.error Remove.the.data-chip.and.insert.it. again..If.E05.still.remains,.contact. 3in1. .Customer.Service. There.is.not.enough. Repeat.the.test.and.apply.more. blood blood. The.data-chip.is.not. Insert.the.data-chip.corresponding. inserted to.the.strips.that.you.want.to.use. The.result.is.lower. If.the.result.does.not.correspond. than.the.limit.of.that. to.the.personal.sensations,.repeat. parameter.(see.also. the.test,.applying.a.big.drop.of. instructions.for.use.of. blood... the.strips) If.LO.is.obtained.again,.contact. your.physician.for.additional. information.
  • Page 21 Error messages Error. Problem Solution message Value.is.higher.than.the. Immediately.contact.your. limit.of.reading.of.that. physician.and.repeat.the.test.in.a. parameter. hospital.laboratory. Get.new.batteries. Replace.the.batteries.before. taking.a.measurement. Check.expiry.date.of.the.strips. and.use.new.strips.to.perform.the. measurements..Blinking.result.is. not.saved.in.the.memory.
  • Page 22: Change.of.batteries

    Changing the batteries The.3in1..meter.houses.two.3V.lithium.batteries.(type.CR.2032).. If.the.battery.icon.is.blinking,.the.charge.is.low.and.the.battery.must. be.replaced.before.taking.a.measurement. Always.insert.new.batteries.and.not.a.used.one.. Shut.down.the.instrument.before.removing.the.batteries.. Check.the.new.batteries.to.see.if.they.apply. How.to.change.the.batteries: •. Open.the.sliding.rear.panel.of.the.instrument.(1). •. Remove.the.batteries.from.the.plastic.holder.(2). •. Place.the.new.batteries.so.that.the.type.is.readable..The.negative. side.of.the.batteries.must.be.at.the.top.(3). •. Close.the.back.cover.until.a.click.is.heard. Flat.batteries.can.be.disposed.of.in.recycling.centers. NOTE. If.the.batteries.are.replaced.or.simply.removed.for.more.than.60. seconds,.it.is.necessary.to.set.time.and.date.again.
  • Page 23: Warranty

    Information on the conditions of warranty VPD,.Bled,.d.o.o..company.grants.to.the.customer.that.the. instrument.will.be.free.of.material.defects.and.manufacturer.handling. for.3.years.starting.from.the.purchasing.date..During.such.period,. VPD,.Bled,.d.o.o..should.replace.the.defected.instrument.with.an. equivalent.version.or.the.same.model.after.verifying.the.defect. Such.warranty.is.subject.to.exception.and.limitation.as.described. below..The.warranty.is.limited.to.the.replacement.due.to.defects. in.some.parts.and.handling..Instruments.with.defects.caused.by.. improper.use,.alteration,.negligence,.technical.assistance.performed. by.unauthorised.personnel,.or.due.to.an.accident,.abuse.or.using. strips.that.differ.from.the.original.3in1..strips.will.not.be.replaced. VPD,.Bled,.d.o.o..reserves.the.right.to.apply.technical.modifications. to.3in1..meter.without.the.obligation.to.insert.such.specifications.to. already.marketed.products. Other.warranty.will.be.granted.for.this.product.if.the.replacement. option.described.above.shall.not.deem.possible. The.responsibility.for.special,.indirect.or.consequential.damage.is. excluded.
  • Page 24 Name Surname Street City.&.Postal.code Telephone E-mail Serial.number Date.of.purchase Date Signature Send.to: VPD,.Bled,.d.o.o. Pot.na.Lisice.4,.4260.Bled,.Slovenia,.EU...
  • Page 25: Symbols

    Symbols used on the packaging Follow.instructions In-Vitro-Diagnosticum Use.until... Batch.number... Store.at.temperatures 3V.direct.current.and.battery.type CE.mark Do.not.dispose.the.instrument.in.urban.waste..Collect. separately,.specifically.for.electric.and.electronic. waste.or.give.the.instrument.to.the.distributor.when. you.buy.a.new.one.or.when.you.want.to.substitute.it.. Please.refer.to.local.updated.regulations.
  • Page 26: Performance

    10-600.mg/dL.(0.6-33.3.mmol/L).glucose.electrodes 130-400.mg/dL.(3.3..-.10.2.mmol/L).cholesterol 50-500.mg/dL.(0.56-5.6.mmol/L).triglycerides Power.source: 2x3V.lithium.button.cell,.type.CR.2032.. The.battery.can.be.used.for.approx..1.year.with.3. measurements.per.day.(approx..1000.measur.) Dimensions: 97.mm.–.length 49.mm.–.height 20.5.mm.-.width Weight: 65.g.(incl..battery) Display: Temperature.limits: 10.to.40°C.(50°F.to.104°F) Humidly.limits: 0%.to.90%.relative.humidity.(non-condensing) Memory.capacity: 500.measured.values.with.time.and.date Interface.for.data. Connecting.cable processing: Special.requirements. The.CE.mark.on.the.device.indicates.that.it.complies. for.electromagnetic. with.the.requirements.of.the.EC.directives.89/336/EEC. radiation: relating.to.electro-magnetic.compatibility.73/23/EEC. for.electrical.equipment.intended.for.the.use.in.areas.of. specific.voltage.levels Hint.for.Healthcare. The.appropriate.legal.provisions.apply.for.the.use.of. 3in1. Professionals: ..The.accuracy.of.the.instrument.should.be. checked.after.3.years. Waste.management. Please.refer.to.local.updated.regulation! disposal:...
  • Page 28 VPD, Bled, d.o.o. Pot na Lisice 4, 4260 Bled Slovenia, EU tel.: +386 4 574 50 70 fax: +386 4 574 50 71 email: info@3in1.si web: www.3in1.si EDitioN DAtE AND LASt REViSioN DAtE: 2014-02-01...