Therapeutic Wearable System™
TENS + EMS Wireless Muscle Stimulator
M O D E L N O . E T - 5 0 5 0
Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ET-5050 and is the answer not in the manual?

Questions and answers

Summary of Contents for iReliev ET-5050

  • Page 1 Therapeutic Wearable System™ TENS + EMS Wireless Muscle Stimulator M O D E L N O . E T - 5 0 5 0 Instruction Manual...
  • Page 2 3. Self-Adhesive Electrode Pads x 2 4. Dual USB Charging Cable x 1 5. AC Adaptor x 1 6. Carry Case x 1 Visit www.iReliev.com or your authorized reseller to purchase additional CH3 and CH4 wireless receiver pods, accessories, and replacement electrode pads. www.iReliev.com...
  • Page 3: Table Of Contents

    CARE AND MAINTENANCE…………………………………………………............................................26 TROUBLESHOOTING.………..……………………………………………….…..............................................28 TECHNICAL SPECIFICATIONS……………………………………………………..........................................29 RADIO FREQUENCY WIRELESS INFORMATION ………………………………..................................32 CONFORMITY OF STANDARDS …………………………………………………..........................................33 INFORMATION ABOUT ELECTROMAGNETIC COMPATIBILITY ………..............................34 FCC INFORMATION ………………………………………………………………..............................................38 WARRANTY…………………………………………........................................................39 REGISTER YOUR DEVICE …………………....................................................40 REGISTRATION..............................................................41 www.iReliev.com...
  • Page 4: Indications And Contraindications

    -for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. (Choose TENS Mode P8) -for use by healthy adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance. (Choose EMS Modes P1 through P6) www.iReliev.com...
  • Page 5: Warnings And Precautions

    Stimulation over the neck could also have an adverse effect on hearing or blood pressure. • Do not apply stimulation across the chest because the introduction of electrical current into the chest may cause rhythm disturbances to the heart. www.iReliev.com...
  • Page 6 • Application of electrodes pads near the thorax may increase the risk of cardiac fibrillation. • The system is internally powered (power from battery pack), IP 22, continuous operation. (Please refer to www.iReliev.com...
  • Page 7 User experiences skin irritation due to electrical stimulation or the electrical conductive medium used to remove the electrode pads. Discontinue stimulation and consult a clinician. Irritation may be reduced by an alternative conductive medium or an alternative electrode pad www.iReliev.com...
  • Page 8 ® • This unit should only be used with iReliev brand electrode pads and accessories. • The device is not intended for medical use, for the treatment of any medical condition or for any per- manent physical changes.
  • Page 9 Turn off the stimulator if the stimulation feels unpleasant or does not provide pain relief. • Keep your system out of the reach of children. ® • Use your stimulator only with iReliev brand electrode pads and accessories. • Do not use this system when driving, operating machinery, or when swimming. •...
  • Page 10 Do not use this system if you are wearing a belly button ring. Remove ring before session. ® • Use the device with only the leads and electrode pads provided for use by iReliev with your device. Any others may not be compatible with your unit and could degrade the minimum safety levels. Use only the electrode pad placements and stimulation settings prescribed by your practitioner.
  • Page 11 ® • If there is any other problem, please consult ExcelHealth or return the device to an authorized iReliev reseller . Do not try to repair a defective device. • For plug-in equipment, the power outlet socket must be located near the equipment and must be easily accessible.
  • Page 12: General Information

    Power ON key SET key Power OFF key CH key ON/+ key: increase OFF/- key: decrease USB Charger Socket About Receiver Pods CH1 Receiver Pod: Power ON/OFF key CH2 Receiver Pod: Power ON/OFF key Snaps for Electrode Pad USB Charger Socket www.iReliev.com...
  • Page 13: About Rechargeable Battery

    ▲Note: You may not use the handheld remote while it's charging. When charging, the handheld remote will turn off automatically. ▲Note: Do not charge receiver pods with electrode pads snapped on to the receiver pods as it will reduce lon- gevity of the electrode pads due to excessive heat from charging. www.iReliev.com...
  • Page 14 Remote Battery Status Remote ON Remote Low Battery Remote Battery Drained Battery Symbol Showing Remote Charging Remote Done Charging Battery Symbol Showing NO Battery Symbol Showing www.iReliev.com...
  • Page 15 Receiver Pod Battery Status Receiver Pod ON Receiver Pod Low Battery Receiver Pod Battery Drained Light Blinking Receiver Pod Charging Receiver Pod Done Charging Orange Light On Light Off www.iReliev.com...
  • Page 16 • If the battery fluid leaks and comes into contact with the skin, wash with clean water. • Do not throw into a fire. • When discarding or recycling batteries, make sure not to short circuit (+)(-) the terminals. www.iReliev.com...
  • Page 17: Step By Step Operation Guide For Training

    When removing electrode pads, always remove by pulling in the direction of hair growth. ® It may be helpful to apply iReliev after use skin lotion on electrode pad placement area when system is not in use.
  • Page 18 www.iReliev.com...
  • Page 19 To turn off the receiver pods, press and hold the OFF/- button for three (3) seconds. ▲Note: In an emergency, you may take off the receiver pod(s) and electrode pads directly from treatment site(s). ▲Note: To prevent unpleasant electric shocks, never remove the device while it is still turned on. www.iReliev.com...
  • Page 20: Communication Between Remote And Receiver Pod(S)

    LCD symbol on the remote control and LED indicator guide on wireless unit(s): The ET-5050 system includes one (1) remote and two (2) receiver pods. Each receiver pod is labeled with "CH1" or "CH2" indicating the respective channel of the receiver pod. Two (2) additional expandable re- ceiver pods (Model # ET-5555) may be purchased with your ET-5050 system.
  • Page 21 2. LED in blue blinks on with the channel in use when it receives signal transmitting from the remote. 3. If LED in orange is flashing, battery power is low. Receiver pod will turn on when charging the battery. Orange light will remain until fully charged, then turns off. www.iReliev.com...
  • Page 22: Selecting A Program

    3. Press the ON/+ button (to increase) or the OFF/– button (to decrease) for the choice of program of the selected modality. ▲Note: Make sure the identical receiver pod is turned on and properly placed. www.iReliev.com...
  • Page 23 -for symptomatic relief and man- Variable comfortable tingling and pulsing sensation (sensation agement of chronic, intractable pain should appear to come in waves). Pain should ease and there and relief of pain associated with should be relief after treatment. arthritis. www.iReliev.com...
  • Page 24 Strengthening It improves their anaerobic capacity and is used Increase the stimulation intensity until you get a for improving maximum muscle strength. strong and deep contraction. Do not exceed your comfort level; 20 min/duration. www.iReliev.com...
  • Page 25: Selecting Therapy Time

    ▲Note: If you change programs during the course of a therapy session, the treatment time will not set unless you manually reset it by performing the steps de- scribed above. www.iReliev.com...
  • Page 26: Selecting The Therapy Intensity Level

    If the user's area has any pets or pests, the system device(s) should be cleaned each time before use and kept safe away in a drawer or somewhere pets and children won’t be able to reach. • Never use aggressive cleaning products or stiff brushes to clean the device. www.iReliev.com...
  • Page 27 Do not dispose of the device(s) in a fire. The battery could explode causing injury or death.
  • Page 28: Troubleshooting

    • Electrode pads are worn down and must be replaced. • Change the positioning of the electrode pads slightly. • Choose a program that is appropriate to your needs or gives you the greatest pleasure. www.iReliev.com...
  • Page 29: Technical Specifications

    TECHNICAL SPECIFICATIONS Receiver Pod Capability: Up to 4 receiver pods may be operated independently using the ET-5050 hand control remote including program, intensity level, timer, & TENS or EMS modes. Pulse Rate: As pre-programmed, in operation mode. Pulse Width: As pre-programmed, in operation mode.
  • Page 30 Transport and Storage Conditions: +14°F (-10°C) to +140° (60°C), 30-95% max. Relative humidity Operation Altitude: 3000m. Operating Atmospheric Pressure Range: 700~1013 hPa Transport and Storage Atmospheric Pressure Range: 500~1060 hPa (i) There are a number of technical symbols on your system, explained as follows: Remote Label Receiver Pod Label www.iReliev.com...
  • Page 31 Interference may occur in the vicinity of equipment marked with the following symbol. Ingress Protecting Rating IP22 The symbol on the device means it is protected against hazards from entry of objects larger than a finger and vertically falling water drops when tilted at 15 degrees. Stand-by. www.iReliev.com...
  • Page 32: Radio Frequency Wireless Information

    9.8 feet range between remote and receiver pod. Wireless Security Measures ® Device Identification 8 bits, shared only with other PlayMakar devices. Data Integrity Checks Each data frame is protected by a 16 (b) length CRC. Acknowledgment Each point-to-point communication is acknowledged. www.iReliev.com...
  • Page 33: Conformity Of Standards

    -If the PlayMakar Therapeutic Wearable System™'s RF channel is interfered by emission from other devices, fail-safe outcome is that iReliev's remote control will not operate, and/or reset the stimulation Interference from other devices level to “0”; the receiver pods will stop stimulation.
  • Page 34: Information About Electromagnetic Compatibility

    RF disturbances are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF com- ® munications equipment (transmitters) and the PlayMakar Therapeutic Wearable System™ as recommend- ed below, according to the maximum output power of the communications equipment. www.iReliev.com...
  • Page 35 Radiated RF 3 V/m 10V/m mended separation distance calculated from the IEC 61000-4-3 80 MHz to 2.5 GHz equation applicable to the frequency of the trans- mitter. Interference may occur in the vicinity of equipment marked with the following symbol: www.iReliev.com...
  • Page 36 70 % U ; 25/30 cycle recommended that the EQUIPMENT or IEC 61000-4-11 Single phase: at 0o 70 % U ; 25/30 cycle SYSTEM be powered from an uninter- Single phase: at 0o ruptible power supply or a battery. www.iReliev.com...
  • Page 37 Harmonic emissions domestic establishments and those directly con- IEC 61000-3-2 Class A nected to the public low-voltage power supply network that supplies buildings used for domes- Voltage fluctuations/Flicker Complies tic purposes. emissions IEC 61000-3-3 www.iReliev.com...
  • Page 38: Fcc Information

    To comply with the FCC RF exposure compliance requirements, no change to the antenna or the de- vice is permitted. Any change to the antenna or the device could result in the device exceeding the RF exposure requirements and void user’s authority to operate the device. www.iReliev.com...
  • Page 39: Warranty

    The warranty does not apply to damage resulting from failure to follow the operating instructions, accidents, abuse, alterations or disassembly by unauthorized individuals. ExcelHealth reserves the right to replace or repair the unit at their discretion. US Tel no. 855-723-2582 ExcelHealth Inc. 1603 Hart Street Southlake, TX 76092 www.iReliev.com www.iReliev.com...
  • Page 40: Register Your Device

    Your new iReliev product is one of the best in the industry, and in many ways, leads the industry, particular- ® ly in the warranty coverage and customer satisfaction. Customer satisfaction is a key factor in every iReliev transaction. ®...
  • Page 41: Registration

    www.iReliev.com...
  • Page 44: Www.ireliev.com

    ExcelHealth Inc. www.iReliev.com 1603 Hart Street Southlake, TX 76092 If you have any questions whatsoever regarding your ® PlayMakar Therapeutic Wearable System™ Model # ET-5050, contact your reseller or ExcelHealth Inc. at: 855-723-2582 or visit www.iReliev.com...