SciCan STATIM 2000 Operator's Manual

SciCan STATIM 2000 Operator's Manual

Cassette autoclave
Hide thumbs Also See for STATIM 2000:

Advertisement

STATIM 2000/5000
CASSETTE AUTOCLAVE™
• Operator's Manual
• Manual del Operador
• Manual do Operação
Latin America

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the STATIM 2000 and is the answer not in the manual?

Questions and answers

Summary of Contents for SciCan STATIM 2000

  • Page 1 STATIM 2000/5000 CASSETTE AUTOCLAVE™ • Operator’s Manual • Manual del Operador • Manual do Operação Latin America...
  • Page 2: Table Of Contents

    3.10 Preparing Unit for Use .............8 3.11 Shipping the Unit .............8 4. Instructions for Use ............9 4.1 STATIM 2000 — Using the Cassette .........9 4.2 STATIM 5000 — Using the Cassette .......10 4.3 STATIM 5000—STAT-DRI Plates ........11 4.4 Preparing and Loading Instruments ......... 11 4.5 Using Biological/Chemical Indicators ......13...
  • Page 3: Introduction

    SciCan Dri-Tec Drying Technology. To dry instruments, the STATIM 2000 uses a combination of forced filtered air and convection heat. The convection heat is derived from utilizing the remaining heat in the system after the sterilization phase has been completed.
  • Page 4: How To Use This Book

    Cassette Tray and Lid Waste Bottle and Bottle Lid STAT-DRI PLUS Spray Fitting Unwrapped Instrument Rack Power Cord Tube Mounting Hardware and/or Wrapped Instrument Rack Operator’s Manual Exhaust Tube (STATIM 5000) SciCan Ltd. Introduction STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 5: Statim Unit Overview

    12. Cassette Power indicator 13. Biological air filter Active indicator 14. Optional internal printer (STATIM 5000) Power switch 15. RS232 port (not on all models) Power cable port 16. Serial/Model/Number Label SciCan Ltd. Unit Overview STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 6: Installation

    • The unit should be stable and all four feet should be securely in contact with the counter surface. This will prevent the unit from moving freely. Figure 3 SciCan Ltd. Installation STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 7: Connecting The Waste Bottle

    10 µS/cm). The impurities and additives in other water sources will cause an error reading on the LCD. If you have a water conductivity meter (available from SciCan, order number 01-103139S) check each new water container before filling the reservoir. If your unit is equipped with a Rev 7 PCB or higher (software SxxxR6xx), you can read the water conductivity by refering to Section 5.9 Reading Water Quality.
  • Page 8: Setting The Time And Date

    WRAPPED cycle button will decrease it. Use the RUBBER/PLASTIC to move to the next digit. To save changes and return to the regular operating mode, press the STOP button. SciCan Ltd. Installation STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 9: Adjusting The Cycle Drying Time

    Screw-in the three leveler feet found underneath the unit. Repack the unit in the original packing materials and include all accessories originally included with the unit. Specify heated and insured shipping. Figure 7 SciCan Ltd. Installation STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 10: Instructions For Use

    STAT-DRI Treatment of the interior surfaces of the cassette with the STAT- DRI drying agent, provided with your unit, will enhance the drying process. (Replacement bottles are available from SciCan, order number 2OZPLUS, 8OZPLUS, 32OZPLUS). SciCan Ltd. Instructions for Use...
  • Page 11: Statim 5000 - Using The Cassette

    STAT-DRI Treatment of the interior surfaces of the cassette with the STAT- DRI drying agent, provided with your unit, will enhance the drying process. (Replacement bottles are available from SciCan, order number 2OZPLUS, 8OZPLUST, 32OZPLUS). SciCan Ltd. Instructions for Use...
  • Page 12: Statim 5000-Stat-Dri Plates

    Up to ten plates can be arranged along the length of the rack. Additional plates may be ordered from SciCan (SciCan part number 01-103935). To install and adjust a STAT-DRI plate, follow these steps: Hold a plate over the desired position, tilted slightly forward with the tabs down.
  • Page 13 SciCan recommends the use of plastic/paper autoclave bags such as SPS™, Medi-Plus™ and Chex All II™. Steri-Stik™ paper/paper autoclave bags may also be used. The use of cloth wraps in the STATIM is not recommended. Loosely pack instruments in the bags to allow steam penetration to all instrument surfaces.
  • Page 14: Using Biological/Chemical Indicators

    Ensure the cassette was not overloaded. Consult the earlier portion of this section for proper loading instructions. If the results do not change, do not process any more instruments within the STATIM and contact your SciCan dealer for further assistance.
  • Page 15 Wrapped Cycle Sterilization Temperature: 134°C (273°F) 10 minutes (STATIM 2000), 6 minutes (STATIM 5000) Sterilization Hold Time: The Wrapped Cycle is used to sterilize solid and hollow metal instruments which have been sealed in paper/ paper, or paper/plastic autoclave bags. Dental handpieces may be sterilized in this cycle.
  • Page 16: Running A Cycle

    During sterilization, the pressure of the chamber and the time remaining for that stage is also displayed. During the air drying phase, the time remaining is displayed. Message will vary depending on the cycle choosen. SciCan Ltd. Instructions for Use STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 17: Stopping A Cycle

    If the air drying stage of the cycle is interrupted, the display reads the following messages: STOP BUTTON PRESSED STOP BUTTON PRESSED STOP BUTTON PRESSED DRYING INTERRUPTED LOAD NOT STERILE PRESS STOP TO RESET SciCan Ltd. Instructions for Use STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 18: Maintenance

    5. Maintenance 5.1 Cleaning the Cassette Keeping the STATIM cassette clean is good clinical practice and assists in the function of the unit. SciCan recommends that the interior surface be cleaned at least once a week. • Use dishwashing soap or a mild detergent that does not contain chlorine.
  • Page 19: Changing The Statim 2000 Air Filter

    The filters should be replaced every six months or after 500 cycles to maintain an adequate supply of clean air during the air drying cycle. STATIM 5000 To change the bacteria retentive air filter on the STATIM 2000 and 5000, follow these steps (see Figures 11 & 12): Power the STATIM OFF.
  • Page 20: Replacing The Cassette Seal

    To ensure optimum performance of your STATIM cassette autoclave, change the cassette seal every 500 cycles or every six months, whichever comes first. Replacement seals are available from SciCan (order number 01-  100028S for STATIM 2000 and 01-101649S for STATIM 5000).
  • Page 21: Preventative Maintenance Schedule

    Make sure no steam is leaking from the valve. Replace if there are any leaks. Check the reservoir for dirt. Clean and rinse with Water Reservoir • steam process distilled water if necessary. Calibrate the unit. Calibration • SciCan Ltd. Maintenance STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 22: The Communication Port

    For printing, you will need to purchase a recommended printer (see list below) from your local computer or electronics store. For data storage, you can purchase SciCan’s USB Data Logger to record and store cycle information onto a mass storage device (MSD) such as a USB Flash Drive or SD memory card.
  • Page 23: Installing The Scican Data Logger

    Drying Unwrapped Please ensure Time/Date and Unit ID are set as per Section 3.6 and Drying Wrapped Section 3.8. To install the SciCan Data logger, access the user setup Save and Exit menu by powering up the unit while pressing the button.
  • Page 24 (see images after STEP 5 below): STEP 5: Connecting the SciCan Data Logger Ensure that both the STATIM unit and the SciCan Data Logger are off. Connect the SciCan Data Logger to the STATIM unit by using the serial cable.
  • Page 25: Cycle Printout Overview

    Time Venting started: VENTING 5:57 5:57 (start of ‘E’ phase) AIR DRYING 6:42 Time Air Drying started: 6:42 (start of ‘F’ phase) CYCLE COMPLETE 16:42 Cycle completion time: 16:42. SciCan Ltd. The Communication Port STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 26: Internal Printer

    Use only paper approved for use with the optional internal printer on the STATIM 5000. The use of any other paper will damage the printer and will void the warranty. Thermal paper is available from your dealer (SciCan order no. 01-101657S).
  • Page 27: Removing Internal Printer Paper Jams

    Press the printer power button to the position. Load paper into the printer following the procedures in Section 7.1 Installing Paper on the Figure 16 Internal Printer. SciCan Ltd. Internal Printer STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 28: Troubleshooting

    SciCan dealer for a replacement. For the STATIM 2000, make sure the compressor is working. To check, remove the exhaust tube from the waste bottle. Start the Air Drying Only Cycle, and place the free end into a glass of water. If there is not a strong, steady flow of bubbles, the compressor is not functioning properly.
  • Page 29 Check the power connection. Ensure the lower red LED is lit. Check that the Mass Storage Device is properly inserted. Repeat the instructions for Installing the SciCan Data Logger on your STATIM. Message: MSD/FLASH FULL <> The MSD is full. Export the data.
  • Page 30: Spare Parts List

    STAT-DRI Plus 32 oz. Bottle w/ sprayer 01-101757S Cassette Lid w/ Box (5000) 8OZPLUS STAT-DRI Plus 8 oz. Bottle w/ cap 01-101763S Power Cord Japan (2000 & 5000) 99-108332 Chemical Emulator 134˚C, 3.5 min. SciCan Ltd. Spare Parts STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 31: Warranty

    10. Warranty Limited Warranty For a period of one year, SciCan guarantees that the STATIM 2000/5000, when manufactured by SciCan in new and unused condition, will not fail during normal service due to defects in material and workmanship that are not due to apparent abuse, misuse, or accident.
  • Page 32: Specifications

    15°C to 25°C (59°F to 77°F) and 25% to 70% 15°C to 25°C (59°F to 77°F) and 25% to 70% and Humidity: Altitude: Up to 2000 meters (6600 ft) Up to 2000 meters (6600 ft) Installation Category: SciCan Ltd. Specifications STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 33: Appendix: Cycle Profile And Capacity

    Load Capacity: 0.4 kg (0.9 lbs) Load Capacity: 0.4 kg (0.9 lbs) LENGEND A) Warming up D) Sterilizing B) Conditioning E) Venting C) Pressurizing F) Air Drying Time Heavy Duty Load Capacity: 1.5 kg (4.4 lbs) SciCan Ltd. Appendix STATIM 2000/5000 Cassette Autoclave Operator Manual...
  • Page 34 STATIM Cassette Autoclave and STATIM are registered trademarks and STAT-DRI, Your Infection Control Specialist, DriTec R and DriTec C logos are trademarks of SciCan Ltd. All other trademarks referred to in this manual are the property of their respective owners.