7.1 Malfunction during start-up
7 Troubleshooting
7.1
Malfunction during start-up
Problem
The instrument does
not start.
7.2
Error messages from the software
Problem
Error 4209 "Lamp is
not on"
Error 4210 "Sample
holder in wrong
position"
Error 4211 "Sample
holder moved dur-
ing scan"
Error 4234 "Sample
cup jammed"
Error 4236 "Door
open"
■■■■■■■■
32
Cause
The power supply cable is
not or incorrectly con-
nected.
The fuse has melted.
Cause
The lamp is no longer burn-
ing.
The sample cup is inserted
incorrectly.
The sample compartment
window lid is open.
The sample cup is inserted
incorrectly.
The sample cup is jammed.
The sample compartment
window lid is open.
■■■■■■■■■■■■■■■■■■■■■■
Remedy
Correctly connect the power supply cable (see
"Connecting the power supply cable", page
8).
Replace the fuse (see Chapter 6.2.3, page 30).
Remedy
Replace the lamp (see Chapter 6.2.1, page
20).
Insert the sample cup correctly (see Chapter
3.7, page 15).
Close the lid of the sample compartment.
Insert the sample cup correctly (see Chapter
3.7, page 15).
Insert the sample cup correctly (see Chapter
3.7, page 15).
Close the lid of the sample compartment.
Need help?
Do you have a question about the NIRS DS2500 and is the answer not in the manual?
Questions and answers