6.3.1 Calculate the correction factor ................7 6.3.2 Calculate the calibration factor ................7 6.3.3 Change the calibration factor ................8 6.4 Change the displayed units ..................9 6.5 Change auto shut off time .................... 9 6.6 Change backlight ......................10 Original instructions SKF LAGM 1000E...
EMC DIRECTIVE 2004/108/EC as outlined in the harmonized norm for EN 61000-6-2:2005 EN 61000-6-4:2007 EN 61000-6-3:2007 EUROPEAN ROHS DIRECTIVE 2011/65/EU Nieuwegein, The Netherlands, January 2013 Sébastien David Manager Product Development and Quality SKF LAGM 1000E...
1. Application The SKF Grease Meter LAGM 1000E accurately measures the amount of grease applied to a bearing or other lubrication point. The grease meter can be connected to grease pumps that are air driven, electrically driven or hand operated.
The Grease Meter measures the amount passing through it. Use the button to turn the display on or off (>2 seconds to turn it off). When this mode is active there is a “STBY” in the lower right corner on the display. SKF LAGM 1000E...
NEXT (right button) for the first option in the menu. It is always possible to exit the menus and return to normal measuring state by pressing buttons at the same time. SKF LAGM 1000E...
The new calibration factor = the old calibration factor x the correction factor. Example: The old calibration factor is 700. The correction factor is calculated as 1,03. The new calibration factor = 700 x 1,03 = 721 SKF LAGM 1000E...
When the last digit has been accepted by pressing ENTER, the display shows: Press ENTER (left button) to accept and save the new calibration or press CANCEL (right button) to cancel the changes and return to the old calibration factor. SKF LAGM 1000E...
1 minute, 4 minutes or 8 minutes or work in STANDBY mode. Press ENTER (left button): the display will show: Press STEP (right button) to move forward to the next time or STANDBY mode. Press ENTER (left button) to accept and save the new time. SKF LAGM 1000E...
Press ENTER (left button) to choose low power setting or press STEP (right button) to change to: Press ENTER (left button) to choose high power setting or press STEP (right button) to go back to previous choice. SKF LAGM 1000E...
Need help?
Do you have a question about the LAGM 1000E and is the answer not in the manual?
Questions and answers