Intended Use Range of Models Signum 1, 2 and 3 are precise and rugged complete scales that give you reliable weighing results. Three different types of weighing tech- nology are used in the Signum Series, The Signum Series of compact scales...
Contents Intended Use Configuring Printouts Warnings and Safety Precautions Interface Port Connecting a Second Weighing Getting Started Platform Unpacking the Scale Pin Assignment Chart Equipment Supplied Pin Assignments for COM1 Installation Pin Assignments for UniCOM Conditioning the Scale Wiring Diagram (Display and Checking the Geographical Data Control Unit <—>...
– The IP43 (or optional IP65) protection interference. This equipment generates, rating for the display and control unit – On request, Sartorius will provide infor- uses and can radiate radio frequency is ensured only if the rubber gasket is installed and all connections are...
Getting Started Unpacking the Scale Checking the Geographical Data § After unpacking the equipment, Entered for Use in Legal Metrology please check it immediately for any (SIWR Models Only) external damage. $ If you detect any damage, proceed Preparation as directed in the chapter entitled (See also the “Device Information”...
Getting Started Installing the Display and Control Unit The following options are available for installing the display and control unit: – Attached to the front of the weighing platform – On a column; part number YDH01P (optional) Fastening the display and control unit to the weighing platform: §...
Page 7
§ Loosen the two locking bolts at the top of the column to facilitate installation of the display and control unit. § Use the 6 hexagonal screws to attach the display and control unit to the top of the column. §...
If the voltage specified on the label or the plug design of the AC adapter do not match the rating or standard you use, please contact your Sartorius office or dealer. The power connection must be made in accordance with the regulations applicable in your country.
General View of the Equipment Signum 1 Display and Keypad Signum 1 Display (for details, see the chapter entitled “Operating Design”) On/off key Toggle key (toggle display between weighing platforms) Zero key Standby Tare key ISO- Zero Tare Test Function...
Operating Design Keys Operation of the Signum 1, Signum 2 Signum 3 or Signum 3 scale involves just a few keys. These keys have one function during measurement and another during configuration. Some of the keys have one function when pressed briefly,...
Input Through the Digital Input over the ASCII Port Signum 2 and 3 only: I Press to view either application data Control Port or manual tare values, depending See page 86, “Data Input Format.” on the key pressed subsequently You can connect a remote hand switch (e.g., )) or foot switch to the input control line, Input using a Bar Code...
Operating Design Display Modes Printing is active Calculated Scale weigh- Bar graph value ing capacity Tare assignment Battery symbol Measured value line Unit GMP printing is active Memory Application symbol Reference value for application Plus/minus sign; Counting: Data transfer stability symbol reference sample updating/ automatic totalizing There are two display modes:...
Appl. 1 Appl. 2 Appl. 3 Display in Weighing Mode Signum 2 and 3 only: Application 3: The illustration above depicts all of the 11. Numeric display; e.g., showing main display elements and symbols that reference value Totalizing can be shown during weighing. Signum 2 and 3 only: Net-total Formulation 1.
Operating Design Display of menu settings: Text menu (example) Display of menu settings: Numeric menu (example) Configuration Numeric input in Signum 3 operating Alphanumeric Input in the Menu menu: (Operating Menu) Enter values (date and time, etc.) using the 10-key numeric keypad The keys below the readout let you navigate the menu and define parameters for configuration.
You can connect two strip or label printers values) are saved when you change appli- (2 seconds) and an acoustic signal to Signum 1, 2 or 3, and have printouts cation programs or switch off the Signum. (double-beep) is emitted...
Configuration Setting the Language You can configure the Signum by selecting parameters in the operating menu. The parameters are combined Example: Selecting “U.S. Mode” for the language in the following groups (this is the first menu level): Switch on the scale –...
Configuring a Password Example: Assign a password (in this example, ) to protect the application program settings APPL and the device parameters SETUP from unauthorized changes 1 Switch on the Signum Enter the second character using the p and k keys (in this example: 2 While all segments are lit, 10 Save the character...
Menu levels are identified by texts, and numeric codes identify the individual settings. individual requirements by entering user data and setting selected parameters = Setting/function available in Signum 1 only in the operating menu. = Setting/function available in Signum 2 and 3 only...
Page 19
Operating Menu = Setting/function available in Signum 1 only = Setting/function available in Signum 2 and Signum 3 only * Factory setting Menu Application programs appl appl1** weigh. Basic Weighing Function 3.5. Minimum load for automatic taring and automatic printing 3.5.1*...
Page 20
Configuration appl appl 1** neutr.m Neutral Measurement 3.6. Minimum load for initialization Numeric menu as for Counting 3.9. Resolution for calculation of reference value 3.9.1* Display resolution 3.9.2 Display resolution + 1 decimal place 3.9.3 Display resolution + 2 decimal places 3.9.4 Internal resolution 3.10.
Page 21
appl appl 1** perc.wg Weighing in Percent 3.6. Minimum load for saving initialization Numeric menu as for Counting 3.9. Resolution for calculation of reference value 3.9.1* Display resolution 3.9.2 Display resolution + 1 decimal place 3.9.3 Display resolution + 2 decimal places 3.9.4 Internal resolution 3.10.
Page 23
No k key function off * gro net Signum 1 only: Gross/net toggling Show 2nd Weight unit 2. unit res 10 Signum 1 only: 10-fold increased resolution Display: max. 10 seconds Show the minimum permissible sample quantity SQmin * Factory setting...
Page 24
Configuration Device Parameters Setup Password prompt displayed if a password is configured wp-1 Weighing platform 1 (Display designation of this menu level: 1) 1.1. Adapt weighing instrument to ambient conditions (adapt filter) 1.1.1 Very stable conditions 1.1.2* Stable conditions 1.1.3 Unstable conditions 1.1.4 Very unstable conditions...
Page 25
Setup wp-1 Calibration and adjustment 1.9. 1.9.1* External calibration/adjustment; default weight 1.9.3 External calibration/adjustment; weight can be selected under menu item 1.18.1 1.9.4 Internal calibration/adjustment (models with built-in motorized calibration weight only) 1.9.8 Set preload 1.9.9 Clear preload No function when you press the J key 1.9.10 1.10.
Configuration Setup Com1 Interface port 1 (Display designation of this menu level: 2) Weighing platform 2 rs-232 RS-232 sbi-std SBI standard version sbi-app SBI trade version (for legal metrology) bpi-232 XBPI-232 1.1. through 1.8. Numeric menu as for WP1 1.9. Calibration and adjustment 1.9.1* External calibration/adjustment;...
Page 27
Setup Com1 datProt 6.3. Time-dependent automatic data output 6.3.1* 1 display update 6.3.2 2 display updates 6.3.4 10 display updates 6.3.7 100 display updates 7.2. Data output: line format for printout 7.2.1 For raw data: 16 characters 7.2.2* For other applications: 22 characters 9.1.
Page 28
Configuration Setup Com1 datProt SMA interface function 5.1. Baud rate 5.1.1 150 baud 5.1.2 300 baud 5.1.3 600 baud 5.1.4 1200 baud 5.1.5 2400 baud 5.1.6 4800 baud 5.1.7* 9600 baud 5.1.8 19,200 baud 5.2. through 5.6. Numeric menu as for SBI Printer Printer configuration YDP01IS...
Page 30
Numeric menu as for WP1 adc-232 ADC-232 1.1. through 9.1. Numeric menu as for WP1 rs-485 RS-485 is-485 Connection of Sartorius IS weighing platform 1.1. through 1.8. Numeric menu as for WP1 1.9. Calibration and adjustment 1.9.1* External calibration/adjustment; default weight 1.9.3 External calibration/adjustment;...
Setup unicom Interface port 2 (Optional) (Display designation of this menu level: 3) datprot Data protocol SBI: standard version 5.1. through 9.1. Numeric menu as for COM1 bpi-232 XBPI-232 bpi-485 XBPI-485 0 to 31 Network address: From 0 to 31 inclusive SMA interface function 5.1.
Page 32
Configuration Setup uniCom1 Printer Printer configuration YDP01IS YDP01IS line Strip printer label Label printer lab ff Label printer with manual feed YDP02 YDP02 Variants 5.1. through 5.4. Numeric menu as for COM1 YDP20 YDP20-0CE 5.1. through 5.4. Numeric menu as for COM1 YDP02IS YDP02IS line...
Page 33
Setup ctrl io Control input/output ports (Display designation of this menu level: 4) ctr inp Input ports 8.4. Function of control input ports (TTL) Trigger p key function 8.4.1* Trigger p (> 2 sec) function 8.4.2 Trigger ) key function 8.4.3 Trigger J function 8.4.4...
Page 34
Configuration Setup barcode Bar code (Display designation of this menu level: 5) Store value as reference tare Store value as tare Store value as ID code 1 Input Enter value on display (triggered when a key is pressed) Header Store value as tare or ID code, depending on bar code header ext.keyb External computer keypad prtprot...
Page 35
1 headr. 2 sn scal 7.13. GMP record (optional on Signum 1) 7.13.1* GMP off 7.13.3 GMP on for multiple results 7.14. Date with/without time 7.14.1* Date block includes time on printout 7.14.2 Time not printed 7.15.
Page 36
Configuration Setup utilit Operation (Display designation of this menu level: 7) 8.3. Keys 8.3.1* All available 8.3.2 All blocked 8.3.3 Numeric keypad Signum 3 only 8.3.4 Toggle weighing platform Signum 2 and 3 only 8.3.5 Zero 8.3.6 Tare 8.3.7 8.3.8 isoTEST 8.3.9 Print...
Page 37
Setup time Time (Option on Signum 1) Format for setting the time: 10.07.41 (hours.minutes.seconds) date Date (Option on Signum 1) Format for setting the date: 31.05.06 (day.month.year); U.S. mode: 05.31.06 (month.day.year) code Password Set, change and delete password here. Max. 8 characters; example: 12345678...
– Delete tare values by entering 0 1 digit (no minimum load) (press ) to save) 2 digits You can choose from the following 5 digits options: Signum 1 only: 10 digits – 8.3.1. All keys accessible – Press k to toggle the display 20 digits All keys blocked except e –...
Example with Signum 1: Switch on the Signum, zero the scale, tare the container weight, place sample in container, toggle display to gross weight or to second weight unit, print results. 1 Switch on the scale Display with tared scale and...
Operation Example with Signum 1: Example with Signum 3: Tare the scale by placing a container on the weighing platform Weigh with numerical input of the tare weight, Print the result 1 Switch on the scale. 1 Switch on the scale.
Example with Signum 3: Weigh with variable tare values, print the results and delete the tare values 1 Switch on the scale. Read the net weight The automatic self-test runs. When the weight readout is shown, the scale is ready to operate and automatically set to zero.
Operation Calibration and Adjustment Important Note: – Flashing symbol as adjustment On SIWABBP and SIWSBBP models, prompt. If more than one weighing Purpose the switch is located on the left-hand platform is connected, the platform side of the rear panel. number is also displayed: Perform calibration to determine the Setup...
Setting the Preload Clearing the Preload Internal Calibration SIWS models: Notes on Settings Notes on Settings Verified scale or scale equipped with !The preload can be set only when the !The preload can be cleared only when Option E7 menu access switch is open. the menu access switch is open.
Operation Preparation § Switch on the scale: Press e § While all segments are lit, press the ) key § Select the Setup menu: Press k repeatedly until Setup is displayed § Open the Setup menu: Press the ) key ”: Press the ) key, or $ Select weighing platform 1, “...
Page 45
Example: External calibration and manual adjustment with default weights (with factory settings for weighing parameters) The difference between 1 Zero the scale measured value and the true mass is shown, plus or minus sign. 2 Start calibration (e.g., after calibration prompt: flashing symbol).
Operation SQmin Function SQmin Display * Off Purpose To display the allowable minimum Printed in * Off sample quantity “SQmin” in accordance GMP header with the United States Pharmacopoeia (USP). According to USP guidelines, * = factory settings the uncertainty of measurement may not exceed 0.1 % of the sample See also “Operating Menu Overview”...
Page 47
Example Determining sample weights while monitoring the minimum sample quantity (in this example, SQmin = 100 g) Settings (changes in the factory settings required for this example): Setup: SQmin: Display Step Key (or action) Display/printout 1 Switch on the scale and configure settings as indicated above 2 Place the container for the sample on the scale and tare...
Operation Data ID Codes Factory settings for the ID code names: Signum 3 only ID1: You can assign codes (such as product ID2: name, batch number, etc.) for identifi- ID3: cation of measured values on printouts. ID4: Features Factory settings for the ID code values: No default values set.
Page 49
Example with Signum 3: 10 Save the character Enter ID code names. Enter “Batch no.” and “Cust.” as names for ID codes 1 and 2. 1 Switch on the scale 11 Proceed as described above to enter subsequent characters. After entering the last character, press ) to save the code.
Combining Application Programs Application Programs Signum 1 Applications 1 through 3: Overview Signum 1 Signum 2 Signum 3 Keypad 6 keys 14 keys 17 keys plus numeric keypad Standby Display 14-segment 14-segment 14-segment ISO- Test Zero Tare Function Print plus...
Application 1: Counting Z With the Counting program you can Signum 3 only: Before the quantity on the platform determine the number of parts that You can select the function of the can be calculated, the reference c key for clearing application data. each have approximately equal weight.
Page 52
Application 1: Counting Z Preparation Storage Parameter Minimum Load The minimum load required for § Switch on the scale: Press e The weight on the platform is saved as initialization of the weighing platform a reference when the scale has stabilized. is configured in the operating menu §...
Page 53
Reference Sample Updating Counting with Two Weighing In the operating menu, under: Platforms appl1 count 3.12. You can use two weighing instruments you can define whether or not the simultaneously with the Counting reference sample weight is automatically application. When using two platforms, updated during weighing.
Page 54
Example: Determining the number of uncounted parts. Settings (changes in the factory settings required for this example): Application parameters: Application: Counting Setup: Printout: PrtProt 7.6. ; then select menu line item of your choice (see “Configuration” for options) 5 Activate calculation of refer- ence sample weight 1 Place empty container on the platform...
Application 1: Neutral Measurement Z nM With this application you can use – Tare function: The initial application values are valid until deleted by pressing the c key, your weighing platform to measure the 1) If you store a tare (weight value) by pressing the ) key, you can later length, surface and volume of parts or until overwritten by new values.
Page 56
Application 1: Neutral Measurement Z nM Storage Parameter Minimum Load Neutral Measurement with The reference weight is saved when The minimum load required for Two Weighing Platforms the scale has stabilized. “Stability” is initialization of the weighing platform You can use two weighing platforms defined as the point at which fluctuation is configured in the operating menu, simultaneously with the Neutral...
Page 57
Example: Measuring 25 m electrical cable. Settings (changes in the factory settings required for this example): Application parameters: Application: Neutral Measurement Setup: Printout: PrtProt 7.6. ; then select menu line item of your choice (see “Configuration” for options) 1 Place empty container on the platform 5 Place the desired amount of cable in the container...
Application 1: Averaging (Animal Weighing) V With the Averaging application, you can – Closing application program; A number of subweighing operations use your scale for calculating weights as deleting parameters: are required to form the basis for the average of a number of individual The number of measurements remains calculation of an average weight.
Page 59
Preparation Minimum Load Starting the Measurements § Switch on the scale: Press e. The minimum load required for The averaging routine does not begin initialization of the averaging routine until the fluctuation in weight value § While all segments are lit, is configured in the operating menu, remains below a defined threshold over press the ) key...
Page 60
Application 1: Averaging (Animal Weighing) V Example: Measuring the weight of one mouse. Settings (changes in the factory settings required for this example): Application parameters: Application: Animal weighing Setup: Printout: PrtProt 7.6. ; then select menu line item of your choice (see “Configuration”...
Application 1: Weighing in Percent L With the Weighing in Percent applica- – Tare function: Preparation tion, you can have the value of the § Switch on the scale: Press e. 1) If you store a tare (weight value) by weight on the platform displayed as pressing the ) key, you can later a percentage calculated in relation...
Page 62
Application 1: Weighing in Percent L Parameter for saving weight Minimum Load Weighing in Percent with The reference weight is saved when The minimum load required for Two Weighing Platforms the platform has stabilized. “Stability” initialization of the weighing platform You can use two weighing platforms is defined as the point at which fluctua- is configured in the operating menu,...
Page 63
Example: Weighing in 100% of a sample material. Settings (changes in the factory settings required for this example): Application parameters: Application: Weighing in percent Setup: Printout: PrtProt 7.6. ; then select menu line item of your choice (see “Configuration” for options) Reduce the minimum load setting.
Application 2: Checkweighing O – Activate info-mode by pressing I With the Checkweighing application, – Entering absolute values using the you can check whether the sample on keypad or placing the desired amount of the weighing platform matches a target –...
Page 65
Minimum Load Digital Input/Output Interface Checking Range The minimum load required for auto- The Checkweighing application sup- 30 % Target weight 170% matic taring of the container weight on ports the digital input/output interface. Lower Upper the platform (“autotare first weight”), There are 4 control lines, or outputs, limit limit...
Page 66
Application 2: Checkweighing O Example: 1 Checkweighing samples with a target weight of 1250 g and a tolerance range from –10 g to +30 g Settings (changes in the factory settings required for this example): Application parameters: Application: Checkweighing Setup: Printout: PrtProt 7.6.
Application 2: Checkweighing toward Zero Example: 1 Checkweighing samples with a target weight of 1250 g and a tolerance range from –10 g to +30 g Settings (changes in the factory settings required for this example): Application: Checkweighing toward zero (menu item 4.7.2) Setup: Printout: PrtProt 7.6.
Application 2: Classification With the Classification application, – Tare function: Class 3: >115 g to 130 g you can determine whether the weight 1) If you store a tare (weight value) by Class 4: >130 g to 145 g pressing the ) key, you can later of a given sample lies within the limits Class 5: >...
Page 69
Minimum Load Digital Input/Output Interface Min. load Max. load The minimum load for the first class is The Classification application supports Class configured in the operating menu, the digital input/output-interface. under: There are 4 control lines, or outputs, Appl 2 which are activated as follows Class 1 class...
Page 70
Application 2: Classification Example: Defining three classes. Settings (changes in the factory settings required for this example): Applications: application 2: Classification Setup: Printout: PrtProt 7.6. ; then select the menu line item of your choice (see “Configuration” for options) 1 Begin input of class delimiters 6 Place the sample on the weighing platform Signum 3 only:...
Application 3: Totalizing With the Totalizing application, – Automatic taring of container weight. The Signum has a totalizing memory you can add weight values together Operating menu setting: for adding individual net and gross in the totalizing memory. In addition Appl a.tare 3.7.
Page 72
Application 3: Totalizing Preparation Minimum Load § Switch on the scale: Press e The minimum load required for automatic taring of the container weight on the § While all segments are lit, platform (“autotare first weight”) is config- press the ) key ured in the operating menu, under: Appl m.weigh...
Page 73
Example: Totalizing weight values. Settings (changes in the factory settings required for this example): Application parameters: Application: Totalizing Setup: Printout: PrtProt 7.6. Setup: “Print when value stored”; then select menu line items of your choice (see the chapter entitled “Configuration,” under “Printouts,” page 34) Setup: Printout: “Total printout: Print when CF pressed;”...
Application 3: Net-total Formulation R With the Net-total Formulation – Automatic taring of container weight. application, you can weigh in different Operating menu setting: components up to a defined total. Appl a.tare 3.7. Each component is saved in a net-total memory.
Page 75
Preparation Minimum Load Net-total Formulation with § Switch on the scale: Press e The minimum amount that a compo- Two Weighing Platforms nent must weigh before it can be saved This mode is used for weighing § While all segments are lit, in net-total memory is configured in large and small components at the press the ) key...
Page 76
Application 3: Net-total Formulation R Example: Weighing in 3 components of a formulation recipe. Settings (changes in the factory settings required for this example): Application parameters: Application: Net-total formulation Setup: Printout: PrtProt 7.6. “Component printout: Print when value stored;” then select the menu line item of your choice (see “Configuration” for options) Setup: Printout: PrtProt 7.6.
Page 77
The weighing platform is tared and the component counter value is increased by one. Prompt to fill and save the second component is shown. Add the third component to the container, bringing the total up to the desired target (in this example, 2000 g). 5 Add the second component to the container (in this example, 525 g)
Examples of Application Combinations in Signum 3 Z O Example 1: “Breaking bulk” (Counting, Checkweighing and Totalizing) Settings (changes in the factory settings required for this example): Setup: Application 1: Counting ( count Setup: Application 2: Checkweighing ( check.wg Setup: Application 3: Totalizing: Value saved: Net + calculated ( 3.23.3 Setup: Application 3: Totalizing: Autosave: On ( 3.16.2...
Page 79
If the weight is too light, Reduce the minimum load 11 Add desired number of pieces the error code Inf 29 setting or increase the shown in the main display. reference sample quantity and the number of parts in the container. Toggle to Checkweighing The number of pieces is saved automatically...
Configuring Printouts Purpose View information on a particular prod- Storing the average piece weight: uct or tare value: § Initialize the application. The product data memory stores initial- § Enter a memory cell number and press ization data and user data (product and the I key.
Page 81
Purpose Preparation Additional Functions § Switch on the scale: Press e You can specify which data items are Printing the “Selection” and “List” included printouts. When using the Settings § While all segments are lit, Totalizing or Net-total formulation – LIST : Print the currently selected list press the ) key...
Page 82
Example: Standard printout for data output from the Counting application Settings (changes in the factory settings required for this example): Application parameters: Application: Counting Then set the printing option: Printout: Interface 1: Print when p key pressed § Select the Setup menu: Press k repeatedly until Setup is displayed §...
Ethernet: Option B9, YDO01SW-ETH Office or production area network Profibus: Option B1, YDO01SW-DP Connect devices over Profibus DP field bus * For model SIWSDCS with a round connector, you will need a corresponding adapter cable. Contact your Sartorius dealer for more information.
Pin Assignment Chart Pin Assignment Chart Connecting a Second Weighing Platform Female connectors COM1 and UniCOM: Female connectors COM1 and UniCOM: 25-pin D-Submini round female con- 25-pin RS-232 with screw lock hardware On Signum 2 and 3 models, you can nector (DB25S) with screw lock hard- connect a second weighing platform to ware for:...
Configuring the Data Data Input Format Command Meaning F6: Trigger I key function kF6_ Interface as a COM Port (Commands) (Signum 2 and 3 only) ( datProt ) d key (Signum 3 only) KF7_ You can connect a computer to your scale to send commands controlling weighing D key (Signum 3 only) KF8_...
Interface Port Data Output Format Error Codes: Each line in a print job can contain up to 22 characters Pos. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 (up to 20 printable characters plus two control characters). The * * * E r r * * # # * * * * CR LF first 6 characters, called the “data header”, identify the subsequent * * * E r r * # # # * * * * CR LF...
Interface Port Automatic Data Output (SBI) You can have results of measurement printed automatically You can configure the autoprint function to print at certain intervals (measured in display updates and define whether printing is dependent on stability of the scale .
(Weighing platform model) -------------------- Dotted line GMP footer: -------------------- Dotted line 14.01.2007 09:45 Date and time Name: Field for operator signature Blank line -------------------- Dotted line Not applicable for Signum 1 display and control units Signum 2 and 3 only...
Solution A key is stuck Release the key, or Err 101 Key pressed when switching on the scale contact your local Sartorius Service Center Err 320 Operating program memory defective Contact your local Sartorius Service Center Verified platform not compatible...
In Germany and many other countries, outlet (mains supply). If you have an – the built-in power supply no longer Sartorius AG takes care of the return and cable connected to the interface port, functions properly. legally compliant disposal of its electrical unplug it.
Symbol “ – ~” on the label denotes AC or DC voltage Operation with rechargeable battery See Sartorius Installation Instructions for Option L9 Operation via built-in or external rechargeable battery (only available as an option that must be ordered with the scale) Emissions Acc.
Dimensions (Scale Drawings) SIWRDCP / SIWADCP / SIWSDCP Standard version: With built-in rechargeable battery: SIWSDCS 453.5 237.5 41.5 M x1 150.5 165.5 24.5 All dimensions are given in millimeters...
Page 100
Dimensions (Scale Drawings) SIWABBP / SIWSBBP Ø ±2 Ø ±2 ±2 Ø All dimensions are given in millimeters...
Overview Accessories/Options Accessories for the Display and Control Unit Printer and printer accessories Verifiable printer with functions for date, time and statistical evaluations YDP20-0CE Printer paper for data printer (5 rolls; length per roll: 50 m) 6906937 Replacement ink ribbon cartridge 6906918 YDP20-0CE Verifiable strip and label printer with thermal print head,...
Page 102
Stainless steel wall-mounting bracket, tiltable YDH01CIS Stainless steel wall-mounting bracket YDH02CIS Dust covers (set of 2) YDC01SW * Not available on model SIWSDCS ** Model SIWSDCS scales require a corresponding adapter cable for these options. Please contact your Sartorius dealer for more information.
Page 103
YCC01-091SM5 Round male connector for configuring own cable 69QC0010 Adapter cable, from round male connector to RS232-DB25 female connector, for connecting standard Sartorius accessories YCC01-0016M3 Extension cable, round male connector / round female connector (6m) YCC01-011SM6 Cable to connect YDP04/121S-0CEUV and YDP121S-0CEUVTH...
Quality Man- tion on verification and legal regulations The operator shall be responsible for agement System of Sartorius AG by the currently applicable in your country, and any modifications to Sartorius equipment Niedersächsische Mess- und Eichwesen,...
Appendix: General Password When you select the Setup menu item, the password prompt is displayed for 2 seconds, and then cursor flashes in the position of the first characters of the password. Repeatedly: Enter the password – Press k to enter letters and ) to save.
Page 116
Sartorius AG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 Fax +49.551.308.3289 www.sartorius.com Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG.