Gpib Error Messages; Pulse Generator And Scope Commands; Keithley Pulse Card Kxci Commands - Keithley 4200-SCS Reference Manual

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Model 4200-SCS Reference Manual

GPIB error messages

KXCI error messages and numbers are shown in
Table 9-9
KXCI error messages

Pulse generator and scope commands

The KXCI commands to control pulse generator and scope are documented in the following topics:

Keithley pulse card KXCI commands

The KXCI commands to control the pulse generator are summarized in
these commands
Table 9-10
Keithley pulse card commands
Command Function
PD
Set pulse load (output
impedance)
PG
Set pulse trigger mode
PH
Halt pulse output
4200-901-01 Rev. S / May 2017
Error No.
Error Message
-999
"IEEE32.DLL GPIB driver is not loaded."
-998
"Unable to initialize shared memory."
-997
"Could not establish communication with console."
-996
"GPIB address not sent as argv[1]."
-995
"GPIB address not in 0<= addr <= 31."
-994
"Could not find configuration file."
-993
"GPIB argument error."
-992
"GPIB command error."
-991
"Illegal setup error."
-990
"Trigger Master card not found."
-989
"Command not valid on this page."
-988
"Instrument not mapped."
-987
"Skipping instrument."
-986
"Unsupported command received."
-985
"Unsupported file format error."
-984
"Could not open specified file."
Keithley pulse card KXCI commands
KXCI commands to control scope card
(Pulse card command
Command String
PD A, BBB.BBBB
A = 1, 2, ... or n pulse card channel number. The largest permissible value of n
equals the number of channels in the system (8 maximum).
BBB.BBBB = Output (load) impedance in ohms: 1.0 to 10 e6
PG A, B, C
A = 1, 2, ... or n pulse card channel number. The largest permissible value of n
equals the number of channels in the system (8 maximum).
B = Pulse trigger mode: 0 (Burst), 1 (Continuous) or 2 (Trig Burst)
C = Number of pulses to output (Burst or Trig Burst): 1 to 2
PH A
A = 1, 2, ... or n pulse card channel number. The largest permissible value of n
equals the number of channels in the system (8 maximum).
Return to
Section 9: Keithley External Control Interface (KXCI)
Table
details) follow the table.
= Default: 50.0 ohms
Section Topics
9-9.
Table
9-10. Details on
32
-1
9-57

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4200-SCS and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF