To disconnect the charger from the main voltage, Factory Settings remove the power plug from the electrical outlet. Initial Steps 4. Use supplied 3M Unitek charger. The use of any Charger other charger can result in damage to the battery. Light Guide/Handpiece Removing and Inserting the Light Guide 5.
8. Ortholux Luminous Curing Light may be operated Ortholux Luminous Curing Light unless adequate only with the supplied light guide or 3M Unitek safety measures, such as the use of protective replacement light guide. The use of other light goggles to filter blue light, are undertaken.
Refer to the table under Recommended Cure Techniques The light source is a high-performance light-emitting with 3M Unitek Adhesives. diode (LED). The emitted light covers the light wavelengths between 430 and 480 nm. - Though the majority of light-curing orthodontic...
Classification: Protection class II, Installation of the Unit Factory Settings Handpiece The factory settings of the unit are as follows: Power supply: lithium-ion battery, nominal voltage 3.7 V • When first turned on the unit is in Extended Mode, i.e. only the Power level LED is lit on the hand piece. Utilizable wavelength range: 430-480 nm...
Battery Charging Insert the battery: • Remove the protective cap from the battery • The unit contains a powerful lithium-ion and keep with the unit packaging. rechargeable battery. The battery does not have any memory effect and can therefore be recharged at • ...
If the Power Level LED is off, briefly press the START Recommended Cure Techniques with button to activate the hand piece. The Power Level LED 3M Unitek Adhesives will come on and a short audible signal is emitted. The hand piece is now ready for operation; the handpiece • ...
If the measures described Sleep Mode above do not result in any improvement, call 3M Unitek Customer Service. Once the handpiece is placed in the charger, all internal functions and LEDs are automatically turned Acoustical Signals - Handpiece off as the handpiece switches to sleep mode.
Troubleshooting Error Cause Error Cause • Solution • Solution The charge level display The remaining battery During light emission in The handpiece has become extended mode, an error overheated in the course on the handpiece glows charge is adequate only red steadily.
- Solvents or abrasive cleaners may not be used in battery inserted in the handpiece! any case, as they damage the plastic components! Use 3M Unitek batteries only! The use of other manufacturer batteries or non-rechargeable batteries/ - Cleaning agents must not enter the unit!
Warranty 3M Unitek warrants this product will be free from defects in material and manufacture for a period of two years from date of purchase. 3M UNITEK MAKES NO OTHER WARRANTIES INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
3M Unitek Service Centers Tedensa Plandent oyi Denta n.v. Kari Sami Eddy Dirk Lorenzo Raul Corrales Marques Hämäläinen Oksanen Van den Eynde Backsteen Asentajankatu 6 Heiveldekens 2 C/López Recuero, 3 - local B B – 2550 E - 28038 Madrid...
Page 13
Guidance and manufacturers declaration – electromagnetic emission The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance...
Page 14
Guidance and manufacturer´s declaration – electromagnetic immunity The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that it is used in such an environment.
Page 15
Guidance and manufacturer´s declaration – electromagnetic immunity The Ortholux™ Luminous Curing Light unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Ortholux Luminous Curing Light unit should assure that it is used in such an environment.
Page 16
Recommended separation distances between portable and mobile RF communications equipment and the Ortholux™ Luminous Curing Light unit The [EQUIPMENT or SYSTEM] is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Ortholux™ Luminous Curing Light unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Ortholux Luminous Curing Light unit as recommended below, according to the maximum output power of the communications equipment...