718 Calibration (V2.0 And Later) - Fluke 712 Calibration Manual

71x series process calibrators
Hide thumbs Also See for 712:
Table of Contents

Advertisement

CALU
8. Release pressure by venting the system.
9. Press the
m
calibration. If calibrating vacuum, the display should show:
CAL
10. Make sure the pressure/vacuum switch is in the vacuum position. Forward
(clockwise) is for pressure and backward (counter-clockwise) is for vacuum.
11. Set up the deadweight tester to apply -12.000 PSI.
12. Wait for the pressure standard to stabilize, then press any key.
13. The display should show:
---------------------------
14. The unit will then reset power.
15. The 718 Calibrator is now out of the CAL mode and into the normal mode of
operation. Carefully vent all pressure, press the O key to turn the calibrator off, and
disconnect all pressure fittings.

718 Calibration (V2.0 and Later)

mA Measure
1. Hold down both U and D on the 718 Pressure Calibrator. Press and release O.
For versions 3.00 and later, c has changed to a.
2. The display shows CAL xxx, as long as Uand D are pressed. Where xxx is a
number, indicating the number of times the 718Ex has been through the calibration
adjustment. When Uand D are released,
required before you can enter the calibration mode. The password is 817. Enter the
most significant digit first. The password appears below
the displayed number, h to decrease it. When the required digit appears, press D
to proceed to the next digit, or the password check. If the password entry was
correct, the 718 will go to the CAL mode. If the password is incorrect,
shows on the display and then the 718Ex returns to the normal operation mode. In the
CAL mode, the 718 should display:
0.000 mA
in the upper display and
CAL 0 mA
on the lower line of the display
The upper display shows an uncalibrated reading. This reading is approximately
+/- 10 % of full scale of the applied reading. When D is pressed, the calibration
value is calculated relative to this reading. The lower display indicates what function
and point is being calibrated.
3. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 718 Calibrator (black to black and red to red).
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. When 0.0 mA is sourced from the 5520A the upper display indicates about
"0.000 mA". When D is pushed, the unit establishes the calibration constant for
that point and the lower display changes to "CAL 24 mA". The upper display still
key to go to vacuum calibration. Otherwise, press any key to finish
-12.000 PSI
Note
.
is displayed. The password is
PASS
. Use c to increase
PASS
Process Calibrators
Calibration
BAD PASS
69

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 712 and is the answer not in the manual?

Questions and answers

Table of Contents