3.20 ESC K ........................18 3.21 A ....................... 18 RROW 3.22 ENTER ......................... 18 3.23 MULTI MODE K ..................... 18 3.24 ON/OFF K ......................19 3.25 DFX-7 O ......................19 VERVIEW 3.26 N ..................... 21 AVIGATING THE...
Page 4
3.27 T & B ....................22 OTTOM ....................23 RANSDUCER ONNECTORS & B ..................23 ROBE ATTERY OVER USB-C C ........................ 23 ONNECTOR CHAPTER FOUR SETTING UP FOR MEASUREMENT ......... 24 4.1 S ....................24 ELECTING ISPLAY 4.2 A ....................25 DJUSTING THE DISPLAY ........................
Page 5
EFINITIONS 10.5 A ..................... 76 TTENUATION ARAMETERS 10.6 C DGS C ....................77 REATING A URVE CHAPTER ELEVEN ADDITIONAL FEATURES OF THE DFX-7 ....84 11.1 P ........................84 ULSE IDTH 11.2 P (PRF) ................85 ULSE EPETITION REQUENCY 11.3 D .........................
Page 6
14.1 C ................151 OMPUTER YSTEM EQUIREMENTS 14.2 I ..................... 151 NSTALLING 14.3 C DFX-7 ................. 152 OMMUNICATING WITH THE 14.4 U DFX-7 ....................152 PGRADING THE APPENDIX A - VELOCITY TABLE ..............154 APPENDIX B - SETUP LIBRARY ..............156 ...
A/B Scan thickness scope, in a single unit. It’s a combination of both our CMX DL+, as well as our new DFX-7 flaw detector series. Why is this advantageous? Thickness gauge are specifically setup to very accurately measure thickness, locate pits, flaws and blind surface corrosion.
1.2 Electrical Warning The DFX-7 contains a high voltage pulser. It’s recommended that the gauge be powered off prior to connecting or disconnecting your transducer, to avoid damaging the DFX-7.
DFX-7, use the UP/DOWN arrow keys to select “flaw detector”, followed by pressing the OK key start the DFX-7 in flaw mode. If this is the first time the gauge has been used, a general startup configuration will automatically be loaded and the DFX-7 is ready for general inspections.
Note: The following features are set to off: GT2, TRIG, AWS, TCG, DAC, DGS. 2.3 Top & Submenu Reference The following table is a quick menu reference guide. The DFX-7 has 15 top level menu titles, and multiple submenu items as illustrated below. Refer to Chapter Three for additional definitions and information on the keypad and menu items.
REF TYPE REF SIZE 2.4 Loading the Default Setup If the DFX-7 has previously been used, the basic default settings can be reloaded at any time, as outlined below: Load Default Setup The following section assumes the gauge is powered up, the flaw detector option selected, and the main measurement screen is being displayed.
Page 12
Dakota Ultrasonics 3) Press the ENTER key to display the confirmation screen. 4) Press the OK key to load the default setup and return to the measurement screen.
CHAPTER THREE KEYBOARD, MENU, & CONNECTOR REFERENCE 3.1 Menu Key (Operation & Sub Menus) The Menu key activates the primary menu structure containing 15 menu tab groups. These tab groups then contain sub menu items, or functions. The sub menu items have been organized in the tab groups according to how closely they are related to the individual tab group names.
Page 14
Dakota Ultrasonics Activating and Navigation of the Menu Items 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left until the desired tab group is highlighted and displaying the submenu items.
ZERO: Used to zero the transducer and calibrate the screen. The DFX-7 is zeroed in much the same way that a mechanical micrometer is zeroed. If the DFX-7 is not zeroed correctly, all of the measurements made using the DFX-7 may be in error by...
Refer to page 99 for further info. 3.5 TUNE – Menu FILTER: The DFX-7 is a wide band flaw detector with a frequency range of 1.8 to 19 MHz (-3dB). The DFX-7+ is additionally equipped with selectable narrow band filters...
DFX-7 Ultrasonic Flaw Detector GAIN: The DFX-7 has a 0-110dB gain range with 2dB resolution. This feature is used to increase/decrease the strength or amplitude of the signal. This might easily be considered as similar to turning the volume up or down on a stereo receiver.
Refer to page 51 for further info. THICKNESS: Used to set the thickness of the test material. The thickness will be used to determine which skip/leg the DFX-7 is detecting on. Refer to page 51 for further info.
DFX-7 Ultrasonic Flaw Detector ANGLE: Used to set the angle of the transducer, and used in the calculations above. Refer to page 51 for further info. X-OFFSET: This feature is used to calculate the surface distance in the trigonometry function, as is a physical measurement from the index point of the probe to the front of the wedge/can/shoe/case.
Dakota Ultrasonics 3.11 DAC – Menu DAC: Toggles the DAC feature on/off/audible. A Distance Amplitude Correction (DAC) curve is used to configure a variable threshold for the gate, using the same size reflector at different distances. Refer to page 61 for further info.
DFX-7 Ultrasonic Flaw Detector REFERENCE GAIN: Toggles the DRAW mode on/off, and used to set the reference gain from a given reference standard. Refer to page 75 for further info. TRIGGER CURVE: Used to set the alarm to detect on either the DGS curve, or independent GATE.
Refer to page 146 for further info. DEFAULT SETUP: Loads a basic default setup. Use only as a last resort when the setups in the DFX-7 have been corrupted and a computer is not accessible. Refer to page 148 for further info.
DFX-7. Refer to page 108 for further info. SHOW DATE: Gives the user the ability to display the date and time in the waveform area of the DFX-7. The options are OFF, DATE, TIME, BOTH. Refer to page 110 for further info.
Dakota Ultrasonics The MEAS key puts the DFX-7 into its primary mode of operation. In this mode, the user has a complete view of the LCD. When pressed multiple times, navigates the hot menu items. 3.19 OK Key The primary function of the OK key is confirmation of a change or selection.
3.25 DFX-7 Overview In order to understand how to operate the DFX-7, it’s best to start off with an understanding of what we’re looking at. The DFX-7 has a lot of great features and tools that will prove to be a huge benefit for the variety of applications you’re constantly facing on a continual basis.
Page 26
Dakota Ultrasonics B. Battery Icon – Indicates the amount of battery life the DFX-7 has remaining. C. Height – Shows the signal amplitude height as a percent of 100% full screen height. D. Feature Status Bar – Indicates the features currently enabled and in use in the following order: ...
3.26 Navigating the Hot Menu The diagram above is a screenshot of the “Hot Menu” feature of the DFX-7. The Hot Menu’s contain all the most regularly adjusted features of the gauge. The primary purpose of the design, was to provide the user with an efficient way to make adjustments on the fly during inspections.
Dakota Ultrasonics Note: This is a coarse adjustment. It’s the fastest and easiest way to increment, decrement, or toggle the status, however it does so in coarse steps. Set Detect 1) Press the UP and DOWN arrow keys until DETECT is highlighted.
C female connector. It is designed to connect directly from the DFX-7 to a standard USB type A port on a PC. The cable supplied with the DFX-7 is a USB type C to a USB type A (pt# N-003-0330). Note: This connector is also used to upgrade...
CHAPTER FOUR SETTING UP FOR MEASUREMENT 4.1 Selecting Display View The DFX-7 includes a variety of viewing options for the waveform display. There are three primary options waveform options with selectable polarity. These are all outline below as follows: Radio Frequency (RF): Displays the entire sine wave signal, much like an oscilloscope, from a reflector or return echo.
The Delay represents the left side of the display, and can be adjusted to start at any thickness value within the overall range of the DFX-7. Note: Once the delay is set, it will remain the same for the views: RF, +RECT, - RECT, RECTFW.
Page 32
Dakota Ultrasonics Setting Delay 1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to move left, until the DELAY cell is highlighted. 2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value.
Page 33
DFX-7 Ultrasonic Flaw Detector The Range is the distance from the delay that will be viewable on the screen. It can be adjusted to any value greater than the Delay. The procedures to adjust the range are outlined below: Setting Range 1) Press the MEAS key once to activate measure menu items.
If it’s turned down too much, you can’t hear it at all. The DFX-7 commands a 110dB gain range using 3 stages of amplification. The overall range makes the DFX-7 extremely versatile regarding a variety of applications and material types.
Page 35
DFX-7 Ultrasonic Flaw Detector 4) Press the UP and DOWN arrow keys to scroll the highlighted value. 5) Press the LEFT and RIGHT arrow keys to scroll the digit locations. 6) Repeat steps 4 & 5 until the GAIN value is correctly displayed.
Page 36
Dakota Ultrasonics The user can also access and adjust the gain from the tabbed menus. However, this method is more tedious than making the adjustments using the Hot Menus. The procedure using the tabbed menus is outlined below: Adjusting the Gain using the Tabbed Menus 1) Press the MENU key once to activate the menu items tab.
70%. The width and range are adjusted to cover the range tested, and the inspection is underway. The DFX-7 is equipped with 2 fully independent gates. Each gate has a start and stop or width, a threshold, a polarity switch, and a detect option. A gate can only detect on reflections that occur inside the boundaries of the gate.
Page 38
Dakota Ultrasonics No Gate Gate On The screenshots above show a rectified waveform with and without a gate enabled. The following steps outline how to enable a gate as follows: Activating a Gate Gate1 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the GT1 menu is highlighted and displaying the submenu items.
Page 39
DFX-7 Ultrasonic Flaw Detector Note: Above will trigger the alarm when a signal meets or exceeds the height of the gate, and below will detect on anything that meets or is below the height of the gate. Set Gate1 Start 1) Press the UP and DOWN arrow keys until GATE1 is highlighted, followed by pressing the LEFT and RIGHT arrow keys to scroll the value.
Page 40
Dakota Ultrasonics 6) Press the OK key to set the GATE1 WIDTH and return to the menu screen, or ESC to cancel entering the GATE1 WIDTH. Note: This sets the overall width of the gate in terms of the horizontal axis or depth.
Page 41
DFX-7 Ultrasonic Flaw Detector Note: Sets where the detection will occur, on the negative or positive going echoes.
CHAPTER FIVE CALIBRATION In this section we’ll focus on calibrating the DFX-7. There are a number of ways to go about performing a calibration, but we’ll be focusing on the simplest and most accurate method, which is performing a two point calibration. This method is very convenient, as it will automatically calculate the material velocity and zero using a minimum and maximum calibration standard of know thickness and material type.
Page 43
However, the user should invest the necessary time getting to know the features and options included in the DFX-7, as well as gaining a good understanding of their function: Loading Default Setup 1) Press the MENU key once to activate the menu items tab.
Page 44
Dakota Ultrasonics 2) Press the ENTER key to display the Digits Edit Box. 3) Press the UP and DOWN arrow keys to scroll the highlighted value. 4) Press the LEFT and RIGHT arrow keys to scroll the digit locations. 5) Repeat steps 3 & 4 until the desired range value is displayed.
Page 45
DFX-7 Ultrasonic Flaw Detector Activating a Gate 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the GT1 menu is highlighted and displaying the submenu items.
Page 46
Dakota Ultrasonics 5) Repeat steps 3 & 4 until the desired GATE1 value is displayed. Note: In our example, a value less than 1 inch should be selected. 6) Press the OK key to set the value and return to the GT1 menu.
5.2 Straight Beam Calibration Now that the DFX-7 is setup with a valid range and a gate enabled, we’re ready to perform a two point, calibration using a single element contact style transducer, as follows: Two Point Calibration –...
Page 48
Dakota Ultrasonics Coupling to the 1” Cal Standard (1 Point) Apply a drop of couplant on the transducer and place the transducer in steady contact with the 1” calibration standard. Be sure that the reading is stable and the repeatability indicator, in the top left corner of the display, is fully lit and stable.
Page 49
DFX-7 Ultrasonic Flaw Detector 3) Press the ENTER key to display the Digits Edit Box. 4) Press the UP and DOWN arrow keys to scroll the highlighted value. 5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.
Page 50
Dakota Ultrasonics stable. Also, be sure there is sufficient signal amplitude in order to avoid ‘peak jumping’ or missing the first cycle. Adjusting the Gate 1) Press the MEAS key once to activate the measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to move left, until the GATE1 cell is highlighted.
Page 51
DFX-7 Ultrasonic Flaw Detector 4) Press the UP and DOWN arrow keys to scroll the highlighted value. 5) Press the LEFT and RIGHT arrow keys to scroll the digit locations. 6) Repeat steps 4 & 5 until the known thickness value is correctly displayed.
Page 52
Dakota Ultrasonics Adjusting the Gate to Calibrated Range 1) Press the MEAS key once to activate the measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to move left, until the GATE1 cell is highlighted.
Refer to section 5.1 and 5.2, regarding the procedures to setup and calibrate the DFX-7. The following are a list of additional items to consider when using an angle beam...
Page 54
Selecting the Material Type: The user can also select a material type directly from a table built into the DFX-7 with shear wave values. This would only get the user close, in terms of velocity, to the...
Page 55
DFX-7 Ultrasonic Flaw Detector wave values, the TRIG mode must be enabled. The following outlines the steps to enable the shear wave velocity table and select a material type from the chart: Selecting the Material Type 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the TRIG menu is highlighted and displaying the submenu items.
Page 56
Dakota Ultrasonics 7) Press the ENTER key to select the material type and initiate the overwrite confirmation screen. 8) Press the OK key to confirm the material selection, or ESC key to cancel changing the material type.
CHAPTER SIX TRIGONOMETRY MODE 6.1 Introduction to TRIG Trig mode is most commonly used for inspecting welded joints using an angle beam transducer. Based on the location, orientation, and geometry of the weld, a straight beam transducer is typically not suitable for inspection. As a result, a transducer emitting sound at a given angle will allow the operator to position the transducer away from the actual weld, and “skip”...
(B). Curved Surface Correction: The DFX-7 will also compensate for convex curved surfaces of longitudinally welded pipes. A curved surface will produce a greater sound path than a flat surface will, using a transducer with a given angle. The user should consider the angle used on curved surfaces, with respect to the material thickness, in order to be sure a reflection will occur from the opposite surface of the test material.
Page 59
DFX-7 Ultrasonic Flaw Detector Activating & Setting the Parameters Activate Trig Mode 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the TRIG menu is highlighted and displaying the submenu items.
Page 60
Dakota Ultrasonics 4) Press the LEFT and RIGHT arrow keys to scroll the digit locations. 5) Repeat steps 3 & 4 until the thickness is correctly displayed. 6) Press the OK key to set the THICKNESS and return to the menu screen, or ESC to cancel entering the THICKNESS.
Page 61
DFX-7 Ultrasonic Flaw Detector 6) Press the OK key to set the X-OFFSET and return to the menu screen, or ESC to cancel entering the X-OFFSET. Note: The X-Offset is used for purposes of marking, and is physically measured with a ruler from the index point to the front edge of the transducer.
(SP), surface distance (SD), and depth (D). The following sections provide an outline for using the AWS feature. 7.2 Additional Comments Since the DFX-7 is a gain controlled instrument, the following formula applies: A – B – C = D Where:...
Page 63
DFX-7 Ultrasonic Flaw Detector inspection, and should additionally be setup prior to enabling the AWS feature. Refer to Chapter Six for instructions on setting up TRIG mode. Activating & Setting AWS Activating AWS Mode 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the AWS menu is highlighted and displaying the submenu items.
Page 64
GATE1 cell is highlighted. Note: A gray cursor should be visible in the waveform area of the display. Also, since the DFX-7 has been previous calibrated and the TRIG feature has been setup, a gate should be enabled and active.
Page 65
DFX-7 Ultrasonic Flaw Detector 2) Press the LEFT and RIGHT arrow keys to position the start of GATE1 to trigger on the waveform. 3) Optionally, press the MEAS key multiple times to move right and the ESC key multiple times to move left, until the WIDTH1 cell is highlighted.
Page 66
Dakota Ultrasonics Note: The AWS screen can be toggle on by pressing the OK key while in the main measurement screen. Equally, the Hot Menus can be toggled by pressing the MEAS at any time.
Therefore, if the -6/-12dB curve option has been selected, the user can toggle the TRIGGER options: GATE, DAC, -6dB, or -12dB. Note: Based on the screen resolution of the DFX-7, the dynamic range of the DAC is approximately 18dB. As a result, inspection of any materials that attenuate more than 18dB over the inspection range, the DAC feature should not be used, and TCG (Time Corrected Gain) should optionally be considered.
Dakota Ultrasonics %FSH – The top of the waveform display represents 100% full screen height. Any waveform amplitudes generated greater than 100% full screen height will not be displayed. 100% is the greatest value that can be displayed. Note: The MEAS option selected is overridden with %FSH while the DAC curve is in DRAW mode.
Page 69
DFX-7 Ultrasonic Flaw Detector 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until DAC is highlighted. 3) Press the LEFT and RIGHT arrow keys to toggle the DAC feature ON or AUDIBLE. The menu items should now become active and adjustable, as illustrated above.
Page 70
Dakota Ultrasonics 1) Press the UP and DOWN arrow keys multiple times until FIT is highlighted, followed by pressing the LEFT and RIGHT arrow keys to select SPLINE or LINEAR. 2) Proceed to the next section – Drawing the DAC.
Page 71
DFX-7 Ultrasonic Flaw Detector ex. 1 ex. 2 2) Press the MEAS key multiple times to move right, and the ESC key multiple times to move left until the DAC BOX: cell is highlighted. Now that the DAC BOX: cell is highlighted, the cursor can be controlled by moving the LEFT and RIGHT arrow keys.
Page 72
Dakota Ultrasonics 6) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the DAC menu is highlighted and displaying the submenu items.
CHAPTER NINE TIME CORRECTED GAIN (TCG) 9.1 Introduction to TCG The TCG feature increases gain as distance increases, in order to achieve an overall level of sensitivity for the same reflector at different distances. It is very similar to the DAC feature noted in a previous chapter.
Page 74
Dakota Ultrasonics Activating TCG 4) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the TCG menu is highlighted and displaying the submenu items.
Page 75
DFX-7 Ultrasonic Flaw Detector 11) Press the OK key to set the % SCRN HEIGHT and return to the menu screen, or ESC to cancel entering the % SCRN HEIGHT. Activating Draw Mode 4) Use the UP and DOWN arrow keys to scroll through the sub menu items until DRAW is highlighted.
Page 76
Dakota Ultrasonics ex. 1 ex. 2 2) Press the MEAS key multiple times to move right, and the ESC key multiple times to move left until the TCG BOX: cell is highlighted. Now that the TCG BOX: cell is highlighted, the cursor can be controlled by moving the LEFT and RIGHT arrow keys.
Page 77
DFX-7 Ultrasonic Flaw Detector 7) Press the UP and DOWN arrow keys until DRAW is highlighted, followed by pressing the LEFT and RIGHT arrow keys until OFF is displayed. The TCG curve is now set. 8) Optionally, the TCG curve can be displayed by pressing the UP and DOWN arrow keys until CURVE is highlighted, followed by pressing the LEFT and RIGHT arrow keys toggle ON.
CHAPTER TEN DISTANCE, GAIN, SIZE (DGS) 10.1 Introduction to DGS (AVG) The DGS (Distance, Gain, Size) method was introduced in 1958 as a way to standardize the process of evaluating and sizing defects, also referred to by the German acronym AVG. This method uses a combined empirical and mathematical model to eliminate the need for a large range of reference standards, generally required when using the DAC or TCG methods.
If DGS/AVG transducers are being used, the values for effective diameter and frequency can simply be entered into the DFX-7. However, if a standard transducer will be used and has a standard transducer certificate, the effective diameter must be estimated in order to determine the near field length (NFL).
Page 80
Dakota Ultrasonics Where: N = Near Field Length = Effective Diameter = Wave Length (velocity/frequency) In order to approximate the effective diameter, the shape of the crystal must be taken into account. Transducers are either round or square/rectangular in shape, and can...
DFX-7 Ultrasonic Flaw Detector = Effective Diameter K = Aspect Ratio A = Largest Dimension 0.97 = Fixed Ratio Calculated results can also be compared to the following table: Ratio of the sides 1.37 1.25 1.15 1.09 1.04 1.01 1.00 0.99...
DGS(AVG) chart in section 10.1. If the value entered is not an exact curve value represented in the diagram, the DFX-7 will compare and interpolate between the two closest curves, and mathematically derive a curve for that value.
DFX-7 Ultrasonic Flaw Detector DVK (Curvature Correction): If a V1/K1 or V2/K2 (A2/A4, IIW/Kidney) calibration standard is used as a reference reflector, the curved surface offers additional focus normally resulting in greater signal output. DVK is a correction to the gain, in order to simulate the amplitude generated from a flat back wall.
Page 84
Dakota Ultrasonics Activating DGS 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the DGS menu is highlighted and displaying the submenu items.
Page 85
DFX-7 Ultrasonic Flaw Detector 2) Alternatively, press the ENTER key to display the Digits Edit Box. 3) Press the UP and DOWN arrow keys multiple times to scroll the highlighted value. 4) Press the LEFT and RIGHT arrow keys multiple times to scroll the digit locations.
Page 86
Dakota Ultrasonics 2) Press the UP and DOWN arrow keys multiple times until ERS is highlighted, followed by pressing the LEFT and RIGHT arrow keys to scroll the value. When the correct ERS is displayed, proceed to the next setting.
Page 87
DFX-7 Ultrasonic Flaw Detector 6) Press the LEFT and RIGHT arrow keys multiple times to scroll the digit locations. 7) Repeat steps 5 & 6 until the parameter value is correctly displayed. 8) Press the OK key to set the value and return to the ATTENUATION sub menu screen, or ESC to cancel entering the parameter value.
Page 88
Dakota Ultrasonics 2) Couple the transducer to the first calibration standard, and peak up on the reflector, adjusting the gain to keep the signal height within the boundaries of the waveform display area, and at a suitable height to allow for attenuation at greater depths in the overall range to be tested.
Page 89
DFX-7 Ultrasonic Flaw Detector 5) Press the UP and DOWN arrow keys multiple times until REFERENCE GAIN is highlighted, followed by pressing the LEFT and RIGHT arrow keys to toggle OFF. The DGS curve is now set and ready to be used for...
ADDITIONAL FEATURES OF THE DFX-7 11.1 Pulse Width The DFX-7 has a square wave pulser with adjustable pulse width option. Pulse width, refers to the duration of time the pulser is left on. This time results in increased energy sent into the test material. There are three width options (SPIKE, THIN, and WIDE).
DFX-7 Ultrasonic Flaw Detector 11.2 Pulse Repetition Frequency (PRF) The DFX-7 is equipped an adjustable pulse repetition rate, with adjustable settings of 8, 16, 32, 66, 125, 250, and 333Hz. The primary purpose is to control how fast the pulser is firing. The repetition rate is tied into the alarm LEDS. Therefore, when scanning at higher speeds, a higher repetition rate will make the alarm trigger more responsive.
11.3 Damping The DFX-7 has an adjustable damping feature, enabling the user to select from a variety of settings according the transducer frequency and overall output. The range of the damping feature is 50-1500 ohms. The available steps are 50, 75, 100, 300, 600 and 1500 ohms.
The DFX-7 has a 200 volt square wave pulser that can be adjusted for specific applications and transducers. The Pulser Volt feature offers a 50 volt cut/boost option to the user. The standard setting is 200 volts. This enables the DFX-7 to offer greater penetration for difficult material types, or increased resolution on noisy materials.
Dakota Ultrasonics 11.5 Material Velocity Charts The DFX-7 has two material velocity charts built into the gauge with both longitudinal and shear material velocities. It should be noted that in order to display the chart containing the shear wave velocities the TRIG mode must be activated, as we’ve assumed an angle beam transducer will be used for the inspection process.
11.6 Material Velocity The DFX-7 has a built in feature to manually adjust the material velocity to a know value in the calibration section of the DFX-7. It should be noted that this does not calculate the zero value like the two point calibration option does. Therefore, if the material velocity is changed you should check the calibration.
9) Finally, press the MEAS key to return to the measurement screen. 11.7 Zero The DFX-7 has a feature to manually adjust the electronic zero. This feature used in conjunction with manually adjusting the velocity, are the primary features necessary...
Page 97
DFX-7 Ultrasonic Flaw Detector to perform a ‘crude’ manual screen calibration. The procedure for adjusting the zero is listed below: Adjusting Zero 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the CAL menu is highlighted and displaying the submenu items.
9) Finally, press the MEAS key to return to the measurement screen. 11.8 Brightness The DFX-7 is equipped with a brightness feature to adjust the display visibility and optimize battery life. It has an arbitrary scale with a values from 1-20, with 20 representing the brightest setting.
11.9 Color Scheme The DFX-7 has a variety of display color options to select from. These will change the look and feel of the according to the users preference. The procedures for changing the VIEW, are outlined below:...
Page 100
Dakota Ultrasonics 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the DISP menu is highlighted and displaying the submenu items.
11.10 Landscape The DFX-7 is a very portable and compact gauge with a 2.7” AMOLED display that is used in a portrait mode as standard. In order to maximize the viewable A-Scan area of the display, the orientation of the display can be rotated to a landscape view for either right or left hand use.
However, if the gauge is turned on, set for a DIM of 120sec, and then left idle for 5 minutes, the DFX-7 will dim in 2 minutes, then power off in 5 minutes. The procedures for adjusting the DIM time are outlined below:...
11.12 Graphics Options (look & feel) The DFX-7 includes a couple of cosmetic features for user preference when viewing the waveform. These features only serve as cosmetic items, and do not change the functionality of the DFX-7 in any way.
Page 104
Dakota Ultrasonics Filled – Draws a filled version of the waveform. The procedure to select the look and feel option is outlined below: Setting the Look & Feel...
11.13 Detect Mark Another look and feel option of the DFX-7 is the DETECT MARK feature. The detect mark give the user the option to select: none, line, dots, or box. Depending on the select made, an indication will be displayed on the waveform, showing where the detection is occurring.
Dots 11.14 Filters (Wide & Narrow Bands) There are two versions of the DFX-7 series: DFX-7 and DFX-7+. The DFX-7 is a wide band only gauge, which includes a single wide band filter with a range of 1.8MHz to 19MHz. The DFX-7+ includes the same wide band filter, but also includes a few narrow band filter options at 2MHz, 5MHz and 10MHz.
DFX-7 Ultrasonic Flaw Detector 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the TUNE menu is highlighted and displaying the submenu items.
Page 108
Dakota Ultrasonics 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the TUNE menu is highlighted and displaying the submenu items.
DFX-7 Ultrasonic Flaw Detector 9) Finally, press the MEAS key to return to the measurement screen. 11.16 Detect Modes The DFX-7 has three selectable modes of detection: Z-Cross, Flank, and Peak. The following diagram illustrates the difference between the options: Z-Cross...
11.17 Polarity The DFX-7 has a number of features and views that offer selectable polarity. The important thing to note is the polarity view is entirely different than the polarity selection of a gate. If a polarity view of +RECT is selected, only the positive portion of the waveform will be shown on the screen.
Page 111
Toggle Polarity (+/-) Note: The screenshots above show where all the possible locations for polarity selection are found in the DFX-7. For more information on the above, refer to the sections on VIEW, DAC, GT1, & GT2. 1) Press the MENU key once to activate the menu items tab. Press the MENU...
11.18 Auto Find The DFX-7 is equipped with a very handy feature for locating the detection point of a return echo that is not currently in the viewable region of the display. The AUTO FIND feature is a convenient way to automatically bring the detected echo into view on the display.
DFX-7 Ultrasonic Flaw Detector 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the UTIL menu is highlighted and displaying the submenu items.
Note: Pressing MULTI MODE followed by ENTER again will clear the display. 11.20 Set Date & Time The DFX-7 is equipped with an internal clock to time and date stamp the log, setup and screen capture files for reporting/documentation purposes. The procedures for setting the time and date are outlined below: Setting Date &...
Page 115
DFX-7 Ultrasonic Flaw Detector 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the UTIL menu is highlighted and displaying the submenu items.
11.21 Show Date & Time The DFX-7 can be configured to show the date & time in the active A-Scan window as needed or preferred (off/date/time/both). The procedure for activating and displaying the above options are outlined below: Live Time &...
4) Press the MEAS key to return to the measurement screen. 11.23 Freeze & Capture The DFX-7 freeze feature enables a user to immediately freeze what’s currently being displayed on the screen for further review or analysis. The freeze feature used...
Page 118
.tif (tagged image) file format that can be opened using any graphics viewer. With the storage capacity of the DFX-7 the user can store as many screens shots as needed. It should also be mentioned that this feature is immediately activated on boot up.
4) Press the MEAS key to return to the measurement screen. 11.24 Peak Hold The DFX-7 includes a peak hold feature that draws and saves the maximum waveform amplitude on the screen. This allows an inspector to move the transducer in all directions, in an effort to “peak up”...
11.25 Freeze Waveform The DFX-7 also includes a FREEZE feature that allows an operator the ability to freeze the current waveform being display on the screen at anytime. Once the operator freezes the waveform, the transducer can be removed from the test material.
Page 121
DFX-7 Ultrasonic Flaw Detector Note: Assuming the operator currently has a waveform that they’d like to freeze on the display. 1) Press the MULTI MODE key located in the bottom left corner of the keypad to display the mode options.
Multiple grids can be created and stored until the DFX-7’s memory is full. If the user attempts to store a new file in the DFX-7 and the size of the file exceeds the capacity of memory, the DFX-7 will respond with an error message indicating that the memory is unable to store the new file.
DFX-7 Ultrasonic Flaw Detector Grid File Formats Sequential Log Formats Important Note: For the duration of this chapter, all references to GRIDS and SEQ LOGS should be considered synonymous with references to FILES. 12.2 Creating a new Grid or Sequential Log (File) Important Note: This entire section is a step by step guide to successfully create a grid or sequential log.
Page 124
Dakota Ultrasonics Creating a Name Grid/Seq Log Name : Can contain a combination of up to 20 numeric, alpha, or special characters listed in the first section of this chapter. Grid Log Sequential Log 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right, and the ESC key multiple times to tab left, until the DATA menu is highlighted and displaying the submenu items.
Page 125
DFX-7 Ultrasonic Flaw Detector Grid Log Sequential Log 4) Press the ENTER key to display the new Grid or Seq Edit Box. 5) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq List Items until NAME is highlighted.
Page 126
Dakota Ultrasonics 9) Use the CLR key to backspace if necessary. 10) Repeat steps 6 - 9 until the Grid or Seq Name (File Name) is completed. 11) Press the OK key to save the Grid or Seq Name and return to the Grid or Seq List Items menu, or ESC to cancel entering the Grid or Seq Name (File Name).
Page 127
DFX-7 Ultrasonic Flaw Detector 3) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate alpha characters. 4) Press the ENTER key to select a character and advance to the next field of the Grid or Seq Note.
Page 128
20 of the same character set. Note: The identifier cannot start or end with a special character. Once a start and end ID are entered into the DFX-7 and the log created, the DFX-7 will automatically generate all the identifiers within that range.
Page 129
DFX-7 Ultrasonic Flaw Detector Grid Log Sequential Log 3) GRID LOG: Use the LEFT, & RIGHT arrow keys to scroll the Columns, and the UP, DOWN arrow keys to scroll the Rows. SEQ LOG: Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate alpha characters.
Page 130
Dakota Ultrasonics Grid Log Sequential Log 1) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq List Items until LOWER RIGHT or END ID is highlighted. 2) Press the ENTER key to activate the Coordinate or End ID Edit Box.
Page 131
OK or ESC key to return to the Grid or Seq List Items menu. It may be necessary to free some memory in the DFX-7 at this time. Refer to page 132 for more information on Deleting a File.
Page 132
Dakota Ultrasonics Selecting the Auto Increment Direction The Auto Increment feature gives the user the ability to specify which direction to advance the cursor after storing a reading. Grid Log Sequential Log 5) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq List Items until INCR.
Page 133
DFX-7 Ultrasonic Flaw Detector Saving Graphics The DFX-7 provides the user with the ability to save a snapshot of the display screen and all the current settings of the DFX-7 with every reading, or just save the reading only. Saving the graphics might be advantageous to the user when the A-Scan views will be used to graphically save a picture of the scanned areas for reporting purposes.
Page 134
Dakota Ultrasonics Saving a Grid Once all the parameters are set, the user has the option of saving or canceling the new grid. Grid Log Sequential Log 4) Use the UP and DOWN arrow keys to scroll through the new Grid or Seq List Items until CREATE GRID or LOG? is highlighted.
DFX-7 Ultrasonic Flaw Detector Grid Log Sequential Log 6) Press the OK key to save the New Grid or Seq Log, or the ESC key to cancel the New Grid or Seq Log setup and return to the DATA menu.
Dakota Ultrasonics Grid Log Sequential Log 1) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the target cell cursor to the desired storage location. 2) Press the ENTER key to save the current reading in the highlighted cell location.
Page 137
DFX-7 Ultrasonic Flaw Detector Viewing Stored Waveforms Grid Log Sequential Log 1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to move right and the ESC key multiple times to move left until the LOG cell is highlighted.
Dakota Ultrasonics 5) Abort the Grid/Seq Log View Box by pressing the MEAS key at any time. 12.5 Deleting Grids (Files) Deleting One or All Files from Memory Deleting One File 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the DATA menu is highlighted and displaying the submenu items.
Page 139
DFX-7 Ultrasonic Flaw Detector 3) Press the ENTER key to display the File List Box. 4) Use the UP and DOWN arrow keys to scroll through the stored Files until the target File to delete is highlighted. 5) Press the OK key to delete the File.
Page 140
Dakota Ultrasonics 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until DELETE ALL DATA is highlighted. 3) Press the ENTER key to activate the confirmation screen. 4) Press the OK key to delete All Files from memory, or the ESC key to abort.
DFX-7 Ultrasonic Flaw Detector 12.6 Editing a Grid (File) Once a grid has been created and saved to memory, the user can edit the Comments or Increment Direction at a later time. The following procedures outline this process: Editing a Grid 1) Press the MENU key once to activate the menu items tab.
Page 142
Dakota Ultrasonics 5) Press the ENTER key to activate the Alpha Edit box – Only used when editing the NOTE. 6) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate alpha characters. 7) Press the ENTER key to select a character and advance to the next field of the Comments.
The user may have transferred grid/seq templates from a PC to the DFX-7, or setup grids/seq using the DFX-7 at an earlier time. The name of the currently active file is always displayed at the top of the Grid/Seq Box in measurement mode (refer to photo below).
Page 144
Dakota Ultrasonics Grid Log Sequential Log 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the DATA menu is highlighted and displaying the submenu items.
DFX-7 Ultrasonic Flaw Detector 5) Press the ENTER key to activate the confirmation screen. 6) Press the OK key to load the file from memory. 7) Press the MEAS key to return to the measure screen. 12.8 Closing an active File - Close A user might not have a current requirement to store measurements, but a file is currently open or active and needs to be closed.
Page 146
Dakota Ultrasonics 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until CLOSE is highlighted. 3) Press the ENTER key to close the active file. Note: Following the key press, the CLOSE text will be grayed out indicating the file has been close...
13.2 Opening a Setup The DFX-7 is loaded with a number of setups from the factory. These setups can be opened, edited, and saved to any one of 64 setup locations. If a factory setup is written over, the user can simply reload the default factory setups at anytime using the utility software included with the DFX-7.
Page 148
Dakota Ultrasonics Opening a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the SETUP menu is highlighted and displaying the submenu items.
7) Press the MEAS key to return to the measure screen. 13.3 Saving a Setup Once the DFX-7 parameters and features have be adjusted for an application, the user may elect to save these setting to a specific setup location for future use. This can potentially save time and reduce error between users.
Page 150
Dakota Ultrasonics Saving a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the SETUP menu is highlighted and displaying the submenu items.
Page 151
DFX-7 Ultrasonic Flaw Detector 5) When the parameter to edit is highlighted, press the ENTER key to activate the Alpha Edit Box. 6) Use the UP, DOWN, LEFT, and RIGHT arrow keys to scroll through the characters, the ENTER key to select characters, and the CLR key to backspace through the characters, until the Name or Note fields have been edited.
Dakota Ultrasonics 11) Use the UP and DOWN arrow keys to scroll through the setups until the target location to save the Setup is highlighted. 12) Press the OK key to activate the confirmation screen. 13) Press the OK key to save the Setup, or ESC to cancel saving the Setup.
Page 153
DFX-7 Ultrasonic Flaw Detector Deleting a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the SETUP menu is highlighted and displaying the submenu items.
7) Finally, press the MEAS key to return to the measurement screen. 13.5 Using the Default Setup The default setup feature was added to the DFX-7 to use, as a last resort, if there are no setups stored in the gauge –factory or otherwise. The only time this might...
Page 155
DFX-7 Ultrasonic Flaw Detector Using the Default Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the SETUP menu is highlighted and displaying the submenu items.
Dakota Ultrasonics 13.6 Selecting a Language The DFX-7 is equipped with a language option. Currently, the only languages supported are English, Spanish, and German. The steps to select one of these languages are outlined as follows: Selecting a Language 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the SETUP menu is highlighted and displaying the submenu items.
CHAPTER FOURTEEN USING THE UTILITY SOFTWARE 14.1 Computer System Requirements DakView will run on many different operating systems: Windows XP, Windows 2000 Professional, Windows 2000 Server, Windows 2000, Vista, Windows 7, 8, 10, or advanced Server operating systems running on Intel or AMD hardware. Additionally there is a version for the Mac OSX platform.
DFX-7 series gauges. 14.4 Upgrading the DFX-7 The DFX-7 can be upgraded to the latest revision of firmware at any time. Simply download the latest version posted on the Dakota Ultrasonics website, copy the upgrade file to the main drive directory DFX-7, and use the upgrade utility located in the XFER menu.
Page 159
DFX-7 Ultrasonic Flaw Detector 2) Press the ENTER key to display the upgrade confirmation screen, followed by pressing OK key to start the upgrade process. Note: The DFX-7 will power down during the upgrade process and restart when completed.
APPENDIX B - SETUP LIBRARY Name Comment 1 Gn/AGC Velocity Enter Custom Name … … … … … … …...
Page 163
Additionally, Dakota Ultrasonics warrants transducers and accessories against such defects for a period of 90 days from receipt by the end user. If Dakota Ultrasonics receives notice of such defects during the warranty period, Dakota Ultrasonics will either, at its option, repair or replace products that prove to be defective.
Need help?
Do you have a question about the DFX-7 and is the answer not in the manual?
Questions and answers