After completing the initial steps in
Bioanalyzer"
assay, as described in the following procedures.
Preparing the Gel-Dye Mix
Handling DMSO
WA R N I N G
Kit components contain DMSO. Because the dye binds to nucleic acids, it should be
treated as a potential mutagen and used with appropriate care.
➔
➔
1 Allow the blue-capped DNA dye concentrate (blue ) and red-capped DNA
It is important that all the reagents have room temperature before starting the next step.
N O T E
Protect the dye concentrate from light.
Agilent DNA 7500 and
DNA 12000 Assay Protocol
on page 7, you can prepare the assay, load the chip, and run the
Wear hand and eye protection and follow good laboratory practices when preparing
and handling reagents and samples.
Handle the DMSO stock solutions with particular caution as DMSO is known to
facilitate the entry of organic molecules into tissues.
gel matrix (red ) to equilibrate to room temperature for 30 minutes.
"Setting up the Assay Equipment and
Agilent Technologies
5
12