Just received 525DS . Machine has red light on and beeps occasionally.
Can you help
1 comments:
Mr. Anderson
March 7, 2025
The red "Service Required" light and beeping on the DeVilbiss 525DS machine indicate a problem that needs attention. Possible causes include:
1. The flow meter is not properly set. Ensure it is set to the prescribed number. 2. The air filter is blocked. Check and clean it if needed. 3. The exhaust is blocked. Ensure nothing is restricting the exhaust area.
If these steps do not resolve the issue, contact your DeVilbiss provider. Additionally, if the oxygen concentration drops below 60%, the red light and beeping alert will activate.
Page 1
® ilbiss iter xygen OnCentratOr erViCe anual DANGER – NO SMOKING MODEL 525DS MODEL 525DS-Q MODEL 525KS MODEL 525KS-LT MODEL 525PS CAUTION-Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
525DS Unit - Serial #F & 525DS / 525DS-Q Units - Serial #J (115V Units) ........................23 525DS Unit - Serial #N & 525DS-Q Unit - Serial #R (115V Units with Auxiliary Oxygen Port) ................... 25 525KS Unit - Serial #F & 525KS / 525KS-LT Units - Serial #J (220/230/240V Units) ......................28 525KS-LT Unit - Serial #N &...
Use only water-based lotions or salves that are oxygen-compatible before and during oxygen therapy. Never use petroleum or oil-based lotions or NOTE–DeVilbiss reserves the right to alter or change the design of the DeVilbiss salves to avoid the risk of fire and burns.
Page 4
The DeVilbiss oxygen concentrators manufactured since July 1, 2012 are equipped with a fire mitigating outlet fitting that prevents propagation of fire into the unit. DeVilbiss oxygen concentrators manufactured prior to July 1, 2012 must be retrofitted by your provider with the fire mitigating outlet fitting.
Locate the unit a minimum of 5 feet (1.6 meters) from fireplaces, radiators, heaters, and hot-air registers. WARNING The plug on the DeVilbiss 525DS concentrators has one blade wider than WARNING the other. To reduce the risk of electric shock, this plug is intended to fit in Oxygen causes rapid burning.
Page 6
U N PA C K I N G A N D S E T U P “Power” and “Normal Oxygen” lights will remain on. NOTE– DeVilbiss recommends for optimal service life that the DeVilbiss Oxygen Concentrator to be operated for at least 30 minutes after it is powered on.
M A I N T E N A N C E PATIENT ALERT SYSTEM ROUTINE PATIENT MAINTENANCE DeVilbiss recommends using only original DeVilbiss parts and filters in order to The DeVilbiss Oxygen Concentrator patient alert system will detect unit guarantee a reliable operation of the product.
NOTE–If the following described complete processing of the concentrator by a the front panel indicator lights are operating. qualified DeVilbiss provider/technician is not possible, the device must not be Inspect intake filter (part #MC44D-605) during every PM Check. Replace used by another patient! as necessary.
From the flow meter the oxygen passes through the final bacteria filter and finally the oxygen outlet port to the patient. The DeVilbiss Oxygen Concentrator operates on a timed cycle (3.6 sec. @ 5LPM) that is controlled by the PC board. The PC board will send voltage to the valve causing it to shift and alternately pressurize the sieve beds.
T R O U B L E S H O O T I N G SIMPLIFIED TROUBLESHOOTING Type II—Operation Issues (Service Indicator Light and/or an audible alert is activated or the The key to simple troubleshooting is to recognize which type of problem exists Pressure Relief Valve releases pressure with a and select the most effective approach to solving the problem.
T R O U B L E S H O O T I N G TROUBLESHOOTING CHART A Visible Alarm Audible Alarm Compressor Power Light OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY Pulsating air noise Intake filter not in place or defective Check filter and replace if necessary Compressor intake hose disconnected Reconnect hose...
T R O U B L E S H O O T I N G TROUBLESHOOTING CHART B Visible Alarm Audible Alarm Compressor Power Light Blinking Pulsing OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY Fan off Line cord not properly installed or defective Insert plug in receptacle or replace line cord.
T R O U B L E S H O O T I N G TROUBLESHOOTING CHART E Visible Alarm Audible Alarm Compressor Power Light OTHER SYMPTOMS POSSIBLE CAUSE POSSIBLE REMEDY Fluctuating or no flow System pressure below 9 psi (62.1 kPa) due to: 1.
Remove the two screws (located directly above the hour meter) that hold The DeVilbiss Oxygen Concentrator is designed for ease of service. To aid the front cabinet to the unit’s internal structure.
C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T To loosen the bib: See Figures below showing pressures being checked at accumulator tank "T"...
Remove the two screws from the back of the oxygen outlet fitting assembly CAUTION–The 525DS concentrators use a GSE compressor with a 22 mfd and remove the assembly. capacitor. If replacement is necessary, be sure the correct capacitor is installed.
Page 17
C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T A compressor that is worn or defective may: To test the compressor operating voltage: •...
Replace if badly worn or damaged. FLOW METER To replace cup seal: There are two flow meters available for the DeVilbiss 525 series oxygen Remove rod screw from top of piston. concentrators: a standard 5 lpm flow meter (505DZ-607) and a low output flow Remove the cup retainer plate.
C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T Replace the flow meter. To replace the flow meter: Remove all tubing from the current flow meter.
C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T Ensure any contamination problem has been corrected. MOLECULAR SIEVE BEDS Ensure the unit is unplugged from the wall outlet.
The printed circuit (PC) board is responsible for monitoring and controlling the the pressure building until the pressure relief valve, on the compressor, releases DeVilbiss Oxygen Concentrator. the excess with a sputtering sound. The PC board has preset alerts for low flow and power failure. Should any of the alert values be exceeded, the patient alert system will activate.
Page 22
C O M P O N E N T T E S T I N G , R E PA I R A N D R E P L A C E M E N T If it is determined that the valve did not shift properly or the high pressures or low pressures are uneven, continue testing with step 3.
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "F" (115V Units) 525DS & 525DS-Q Units - Serial # "J" (115V Units) Front View...
Page 24
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "F" (115V Units) 525DS & 525DS-Q Units - Serial # "J" (115V Units) 525DS...
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "N" (115V Units with Auxiliary Oxygen Port) 525DS-Q Unit - Serial # "R" (115V Units with Auxiliary Oxygen Port)
Page 26
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "N" (115V Units with Auxiliary Oxygen Port) 525DS-Q Unit - Serial # "R" (115V Units with Auxiliary Oxygen Port) Rear Inside - Serial # "N"...
Page 27
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525DS Unit - Serial # "N" (115V Units with Auxiliary Oxygen Port) 525DS-Q Unit - Serial # "R" (115V Units with Auxiliary Oxygen Port)
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS Unit - Serial # "F" (220/230/240V Units) 525KS & 525KS-LT Units - Serial # "J" (220/230/240V Units) Front View Rear View 10-11...
Page 29
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS Unit - Serial # "F" (220/230/240V Units) 525KS & 525KS-LT Units - Serial # "J" (220/230/240V Units) 525KS 525KS 525KS-LT...
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS-LT Unit - Serial # "N" (220/230/240V Units with Auxiliary Oxygen Port) 525KS Unit - Serial # "R" (220/230/240V Units with Auxiliary Oxygen Port) Front View Rear View 29-30...
Page 31
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS-LT Unit - Serial # "N" (220/230/240V Units with Auxiliary Oxygen Port) 525KS Unit - Serial # "R" (220/230/240V Units with Auxiliary Oxygen Port) Rear Inside - Serial # "N"...
Page 32
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525KS-LT Unit - Serial # "N" (220/230/240V Units with Auxiliary Oxygen Port) 525KS Unit - Serial # "R" (220/230/240V Units with Auxiliary Oxygen Port) 525KS-LT 525KS PART DESCRIPTION...
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525PS Unit - Serial # "F" (220/230/240V Units) 525PS Unit - Serial # "J" (220/230/240V Units) Front View Rear View 10-11...
Page 34
F I G U R E S , D I A G R A M S A N D PA R T S L I S T 525PS Unit - Serial # "F" (220/230/240V Units) 525PS Unit - Serial # "J" (220/230/240V Units) 525PS 525PS PART DESCRIPTION...
F I G U R E S , D I A G R A M S A N D PA R T S L I S T ACCESSORIES Carton with Shipping Inserts 525D-606 Oxygen Outlet Connector - Plastic XM-1 Casters, Locking 525DS-603 Low Output Flow Meter 515LF-607 Transfill Caddy 525DD-650 Transfill Hose PF1100TUB...
PARTS RETURN Before returning parts or units to the factory, call the DeVilbiss Healthcare Customer Service Department (800-338-1988) or (814-443-4881) to obtain a return authorization number. Include in the package a note indicating the return authorization number along with your company name, address, phone number, and account number.
914-1524 M (3,000-5,000 Ft) *NOTE: The OSD performance at 5°C to 40°C, 95% R.H. through voltage range on the 525DS verified at 670m. **NOTE: This oxygen concentrator has been tested to and meets the EMC requirements of EN 60601-1-2. Do not place the concentrator near other equipment or devices that create or attract electromagnetic fields.
D E V I L B I S S G U I D A N C E A N D M A N U F A C T U R E R ’ S D E C L A R AT I O N WARNING Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the Electromagnetic Compatibility [EMC] information provided in the accompanying documents.
Need help?
Do you have a question about the 525DS and is the answer not in the manual?
Questions and answers
Just received 525DS . Machine has red light on and beeps occasionally. Can you help
The red "Service Required" light and beeping on the DeVilbiss 525DS machine indicate a problem that needs attention. Possible causes include:
1. The flow meter is not properly set. Ensure it is set to the prescribed number.
2. The air filter is blocked. Check and clean it if needed.
3. The exhaust is blocked. Ensure nothing is restricting the exhaust area.
If these steps do not resolve the issue, contact your DeVilbiss provider. Additionally, if the oxygen concentration drops below 60%, the red light and beeping alert will activate.
This answer is automatically generated