Page 5
Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
Do Not Immerse in Liquids. Clean the probe using only a damp cloth. The cloth must be dampened with either isopropyl alcohol or a mild detergent and water solution. Do not use any other chemicals or abrasives to clean the probe. P6243 Service Manual...
CAUTION indicates a hazard to property including the product. Symbols on the Product. These symbols may appear on the product: DANGER Protective Ground ATTENTION Double High Voltage (Earth) Terminal Refer to Manual Insulated P6243 Service Manual...
Specifications The warranted specifications in Table 1 apply to a P6243 Active Probe when the probe and the host instrument are allowed to warm up for 20 minutes before measurements are taken. The warranted specifications that appear in boldface type are specifications checked in the Performance Verification section of the service manual.
Non-Destructive Input Voltage Range (See Figure 4 on page 6) DC Offset Drift 100 V/°C or less at output of probe 1 mV/°C or less displayed on screen with TEKPROBE interface 5.3 ns ±0.2 ns Delay Time P6243 Service Manual...
Specifications 25 V 20 V 15 V 10V at 1 GHz 10 V 6V at 3 GHz 1 MHz 10 MHz 100 MHz 1 GHz Figure 4: Typical Non-Destructive Peak Voltage Derating vs. Frequency P6243 Service Manual...
Cycle 1 Assurance Level II for packaged products 0 -- 20 lbs. Test 2 for Warehouse and Vehicle Stacking (Compression) is omitted Tektronix standard 062--2858--00, Rev. B, Class 5 Altitude Operating: 4,572 m (15,000 ft) Non--Operating: 15,240 m (50,000 ft)
Input Linear Dynamic Range The probe head amplifier used by the P6243 Active Probe has a limited linear operating range. To keep the input linearity error less than 4% you must limit the signal input voltage to ±8 V (DC + peak AC).
Probe TEKPROBE Cable Interface Figure 5: P6243 Simplified Schematic Diagram Probe Head and Cable Assembly The probe head assembly contains an active amplifier circuit that buffers and amplifies the input signal. The amplifier receives power from the compensation box assembly via the cable assembly. All signal amplification and buffering is performed in the probe head assembly.
EEPROM. Figure 6 shows the TEKPROBE interface pin functions. Refer to the service documentation of your oscilloscope for more detailed specifications. - - 5 V Ground Signal Data Clock +5 V Figure 6: TEKPROBE Interface P6243 Service Manual...
Performance Verification This section contains the procedures for verifying the performance of the P6243 Active Probe. Equipment Required Table 5: Equipment Required for Performance Verification Performance Requirement Item Description Recommended Example Oscilloscope TEKPROBE interface Tektronix TDS 744A TEKPROBE Power Supply Tektronix 1103 10.00 ±.01 VDC at 1 mA...
1. Connect the test equipment as shown in Figure 7. 2. Attach the probe to a TEKPROBE interface on the 1103 power supply. 3. Set the 1103 offset VAR/0V to 0V. 4. Set the multimeter to read DC volts. P6243 Service Manual...
Terminator BNC to Dual Banana Adapter Figure 7: P6243 DC Attenuation Accuracy Setup 6. Set the power supply to +2.000 V. (This setting will be referred to later on as Vmax). 7. Record the value measured by the multimeter as M1. (With the 10X attenuation of the probe, the voltage measured will be about +0.200 V.)
Page 25
10. Determine the percent error by using the formula below: %Error = [ M1 − M2 − 1] × 100% The calculated error should be ≤2%. NOTE. An unacceptable error value may result if a low tolerance termination is substituted for the recommended termination. P6243 Service Manual...
Maintenance This section details the maintenance, repair, and troubleshooting procedures for the P6243. Replacing TEKPROBE™ Interface Pins TEKPROBE interface pins can stick and fail to make contact after time. Periodically check to see that each of the interface pins move freely and fully extends out of the interface.
2. Hold the open edge apart, and use the tool to open the other side of the compensation box. 3. With both sides of the box open, gently separate the two halves of the compensation box. Cover Catches Figure 9: Removing the Compensation Box Covers P6243 Service Manual...
2. Press the cover catches in so that the cover can be lowered. 3. Slide the tab into the notch. 4. Firmly press the pieces together until the cover catches snap into place. Figure 10: Replacing the Compensation Box Cover P6243 Service Manual...
See Figure 11. Align the tab to the slot and gently press the two pieces together. See Figure 11. Once installed, the TEKPROBE collar should rotate freely to lock and unlock. Slot Figure 11: Replacing the TEKPROBE Collar P6243 Service Manual...
Replacement Parts Refer to the Replaceable Parts section for a list of customer replacement parts. Due to the sophisticated design of the P6243, there are no user replaceable parts within the probe. Preparation for Shipment...
Replaceable Parts This section contains a list of the replaceable parts for the P6243 Active Probe. Use this list to identify and order replacement parts. Parts Ordering Information Replacement parts are available from or through your local Tektronix, Inc. service center or representative.