LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is two years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling.
Introduction The 1732 and 1734 Energy Loggers (the Logger or Product) are compact devices for energy and power quality surveys. With a built-in touch screen and USB flash drive support, it is easy to configure, verify, and download measurement sessions without the need of a computer at the measurement location.
1732/1734 Users Manual Safety Information • Replace the mains power cord if the insulation is damaged or if the insulation shows signs of wear. identifies hazardous conditions and procedures that are Warning • dangerous to the user. A identifies conditions and procedures...
Page 9
Energy Logger Safety Information • • Do not exceed the Measurement Category (CAT) rating Do not use USB accessories when the Product is of the lowest rated individual component of a Product, installed in environment with wires or exposed metal probe, or accessory.
Page 10
This product contains a Lithium-ion battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler per local regulations. Contact your authorized Fluke Service Center for recycling information.
• Energy Logger • Set of 4 Magnet Probes for 4 mm Banana Plugs • Power Supply These items are available for the 1732 Energy Logger as optional • Voltage Test Lead, 3-phase + N accessories. Note • 4x Dolphin Clips, Black The WiFi/BLE adapter is included only when the radio •...
Remote Control on page 52 for more information about VNC. • Display and store data of up to 2 modules of the Fluke FC 3000 series together with the instrument data into logging sessions (requires WiFi/BLE adapter feature, available with firmware version 2.0).
Energy Logger Before You Start Magnet Hanger Kit The accessory shown in Figure 2 is used to: • Hang the Logger with power supply attached (use two magnets) • Hang the Logger separately (use two magnets) • Hang the power supply separately (use one magnet) Voltage Test Leads Voltage test leads are four-core, flat, test leads that do not tangle and can be installed in tight spaces.
1732/1734 Users Manual Thin-Flexi Current Probe The R-coil has advantages over other types of current transformers: • It is not a closed loop. The second terminal is passed back through The Thin-Flexi Current Probe works on the Rogowski coil (R-coil)
Energy Logger Before You Start Use the color clips for easy identification of the four current probes. Kensington Lock Apply the clips that are appropriate for your local wiring codes on both A Kensington Security Slot (also called a K-Slot or Kensington lock) is ends of the current probe cable.
Auxiliary Input Adapter for up to 2 DC voltages (0 V to 10 V and 0 V to 1000 V) MP1-MAGNET PROBE 1 Set of 4 Magnet Probes for 4 mm banana plugs 1732/UPGRADE Upgrade Kit for 1732 to 1734 (includes: Hanger, Magnet Probes and WiFi BLE Dongle)
Energy Logger Storage Storage The power cord/measurement line cover slides to select the input source. When not in use, keep the Logger in the protective storage bag/case. Warning The bag/case has sufficient space for the Logger and all the To prevent possible electrical shock, fire, or personal accessories.
1732/1734 Users Manual Battery Operation To charge the battery: Connect the mains cord to the ac input socket on the power Caution supply. To prevent damage to the Product: Fit the power supply to the Logger or use the dc power cord to •...
Energy Logger Navigation and User Interface Navigation and User Interface See Table 4 for a list of the front panel controls and their functions. See Table 5 for a list of the connectors and their functions. Table 4. Front Panel Item Control Description...
Page 20
1732/1734 Users Manual Table 5. Connector Panel Item Description Current measurement inputs (3 phases) Voltage measurement inputs (3 phases + N) Power Cord/Measurement Line Slide-Cover Power Cord AC Input 100-240 V 50/60 Hz 15 VA Measurement Line AC Input ...
Energy Logger Connector Panel Decal Connector Panel Decal Table 6. Decal for Connector Panel A self-adhesive decal is supplied with the Logger. The decal corresponds to the wiring color codes used in your local area of operation. Apply the decal around the current and voltage inputs on the connector panel as shown in Table 6.
1732/1734 Users Manual Power Measurement Line Power Source Attach the Power Supply to the Logger or use the dc power cord to The Logger has options for power source: connect the Power Supply to the Logger. • mains Move the slide-cover on the power supply to access the safety •...
Energy Logger Power Battery Power Source Table 7. Power/Battery Status The Logger can operate on battery power without a connection to the power supply or dc power cord. Logger On Push . The Logger turns on and is ready to use in <30 seconds. Power Source Battery Symbol Power LED Color...
1732/1734 Users Manual Touch Screen Calibration The touch screen is pre-calibrated in the factory. If you notice that the The touch screen lets you interact directly with what is on the display. targets do not align with your touch on the display, you can calibrate To change parameters, touch a target on the display.
Energy Logger First-time Use/Setup Wizard First-time Use/Setup Wizard The Logger is now ready for the first measurements or energy study. Note To start the Logger: Be aware that for power measurements in 3-phase systems: Install WiFi/BLE or WiFi only adapter (see WiFi and WiFi/BLE-to- •...
1732/1734 Users Manual First Measurements Connect the voltage test leads to neutral, phase A/L1, phase B/L2, and phase C/L3. At the energy study site, look at the information in the panel and the With all of the connections done, check that the voltages for rating plates on the machines.
Energy Logger Function Selection Buttons Function Selection Buttons 17. Review the logged data using the softkeys V, A, Hz, +, Power, and Energy. The Logger has three buttons that switch between the Meter, Power, 18. To prevent unwanted operation, touch the Lock Screen target. and Logger function modes.
1732/1734 Users Manual THD Voltage includes a traffic light indicator: Study Type Depending on the application, select either Load Study or Energy • green: <2 % Study. • yellow: 2 % to 8 % • Energy Study: Select this study type when power and energy •...
Page 29
Energy Logger Function Selection Buttons Single Phase Single Phase IT Example: Branch circuit at an outlet. The logger has a galvanic isolation between the voltage inputs and ground based signals like USB and mains input. Example: Used in Norway and in some hospitals. This would be the A/L1 connection at a branch circuit.
Page 30
1732/1734 Users Manual Split Phase 3-Ф Wye Example: A North American residential installation at the service Example: Also called “Star” or four-wire connection. Typical entrance. commercial building power. A/L1 A/L1 B/L2 B/L2 C/L3 Energy Study Energy Study A/L1 A/L1 B/L2...
Page 31
Energy Logger Function Selection Buttons 3-Ф Wye IT 3-Ф Wye Balanced The logger has a galvanic isolation between the voltage inputs and Example: For symmetrical loads like motors the connection can be ground based signals like USB and mains input. simplified by measuring only one phase and assuming the same voltages/currents on the other phases.
Page 32
1732/1734 Users Manual 3-Ф Delta 2 Element Delta (Aron/Blondel) Example: Often found in industrial settings where electric motors are Example: Blondel or Aron connection, simplifies the connection by the used. use of only two current sensors. A/L1 A/L1 A/L1 B/L2...
Page 33
Energy Logger Function Selection Buttons 3-Ф Delta Open Leg 3-Ф High Leg Delta Example: A variant of power transformer winding type. Example: This topology is used to provide an additional voltage that is half the phase to phase voltage. A/L1 A/L1 B/L2 C/L3...
Page 34
1732/1734 Users Manual 3-Ф Delta Balanced Nominal Voltage Example: For symmetrical loads like motors, the connection is In load studies only, select a nominal voltage from the list. If a voltage simplified with only one phase measurement and assuming the same is not shown in the list, enter a custom voltage.
On the Logger, go to the Meter screen to view the measurements The Logger supports up to two additional measurement channels with from the connected module. either the wired AUX input or wireless radio signals from Fluke Note Connect sensors.
Page 37
Energy Logger Function Selection Buttons • Wired Connection Use up to 8 characters to configure the measurement Unit of the parameter. Configure the Auxiliary input to show the readings for the attached sensor. In addition to the default setting of ±10 V, up to five custom •...
Example 2: Connection Verification and Correction Once the measurement is configured and the voltage and current Fluke 80TK Thermocouple Module inputs are connected to the system under test, go back to the Meter Output: 0.1 V/°C, 0.1 V/°F mode and use the Verify Connection touch button to confirm the connection.
Energy Logger Function Selection Buttons Auto Correct is not available if the algorithm is not able to detect a Power better phase map or when no errors are detected. – In the Power mode you can get the values and a live trend Note chart for each phase (A, B, C or L1, L2, L3) and total as: It is not possible to detect all incorrect hook-ups automatically.
1732/1734 Users Manual The menu also provides the access to live values of Energy as: Logger • Active Energy (Ep) Wh – In Logger mode, you can: • Reactive Energy (E r) in varh • Configure a new logging session •...
Page 41
Energy Logger Function Selection Buttons To select between Load Study and Energy Study: Duration and Recording Start/Stop Date and Time Go to Meter > Change Configuration. This Configuration screen You can set the duration of the measurement from a list. No end contains the measurement configuration parameters such as configures the maximum possible duration based on the available Topology, current range, voltage, and current ratios.
Page 42
1732/1734 Users Manual Options to configure the logging session: The Logger recommends an interval based on the duration to get the best balance between resolution and data size. • Duration and manual start A memory gauge shows the memory used by recorded sessions and •...
Page 43
Energy Logger Function Selection Buttons Energy costs From the Logger home screen you have access to: • V, A, Hz, + (A, Hz, + for load studies) Enter the costs/kWh for demand energy. The energy costs are applied to forward energy (positive energy) using the demand interval and can •...
Page 44
1732/1734 Users Manual The algorithm to calculate Voltage min/max values is in accordance to For all power values, except PF and DPF, the three highest values established power quality standards to detect dips, swells, and during the logging session are available. Use (Reverse interruptions.
The screen shots are stored on the USB flash drive in the Sessions for more details. \Fluke173x\<serialnumber>\screenshots folder. Note Fluke 173x = your model number, such as Fluke 1734. It is not possible to review a completed logging session when Instrument Settings another session is active.
Page 46
1732/1734 Users Manual Phase Color/Phase Labels Name for Instrument The phase colors are configurable to match with the connector panel You can assign a name to the Logger. This name is attached to the decal: measurement files when you review these files in Energy Analyze Plus software.
Page 47
Energy Logger Function Selection Buttons Date/Time Zone To set the date format: The logger stores the measurement data in universal time coordinate Push (UTC) to ensure continuity in time and accounts for time changes due Push (Instrument Settings). to daylight saving time (DST).
PIN is also supported. Note Contact your local Fluke office for a master PIN in the case of a lost or unknown PIN. You will need the Product serial number and date shown on the screen to request a master...
Energy Logger Function Selection Buttons Touch Screen Calibration Copy Service Data to USB The touch screen has been calibrated at the factory before shipment. If requested for customer support, use this function to copy all In case you do experience misalignment with the touch targets, use measurement files in raw format and system information to a USB the touch screen calibration feature.
1732/1734 Users Manual Reset to Factory Defaults Copy the firmware file (*.bin) into this folder. The reset function deletes all user data, such as logging sessions and Make sure the Logger is powered from mains and turned on. screen captures. It also deletes the WiFi access point connection Plug the flash drive into the Logger.
Select Product Family > Power Quality Tools. 10. Copy the license file (*.txt) to this folder. Select Model Name > Fluke 1732 or Fluke 1734. 11. Make sure the Logger is powered from mains and turned on. Enter the serial number of the Logger.
1732/1734 Users Manual Maintenance How to Clean Caution If the Logger is used appropriately it does not require special To avoid damage, do not use abrasives or solvents on maintenance. Maintenance should only be done at a company related this instrument.
Logger. The recommended calibration cycle is order parts and accessories, see How to Contact Fluke on page 1. 2 years. For more information see How to Contact Fluke on page 1. Table 8. Replacement Parts Fluke Part or Ref.
Energy Analyze Plus Software System Requirements The computer hardware requirements for the software are: Purchase of the Logger includes Fluke Energy Analyze Plus software. • With the software, you can do many tasks from a computer: Free Hard Disk space 50 MB, >10 GB (for measurement data) recommended •...
WiFi Support PC Connections To connect the PC to the Logger: With the USB WiFi Adapter you can use the Fluke Connect App for Turn on the computer and the Logger. asset management, trending and sharing measurement data, wirelessly control the Logger with a PC/smartphone/tablet, and Install the Energy Analyze Plus Software.
Energy Logger WiFi Support WiFi Direct Connection WiFi-Infrastructure The WiFi direct connection uses WPA2-PSK (pre-shared key) with The WiFi connection requires the WiFi-Infrastructure license and AES encryption. The passphrase shown on the screen is required to supports WPA2-PSK. This connection requires a DHCP service establish a connection from a client to the device.
1732/1734 Users Manual Remote Control Configuration IP Address You can remotely control the instrument with a free third-party VNC Direct Connection ......10.10.10.1 client available for Windows, Android, Apple iOS, and Windows Phone after the WiFi connection is setup. VNC (Virtual Network WiFi Infrastructure......
Fluke Connect Wireless System Select the WiFi network that begins with "Fluke173x<serial-no>". Go to the Fluke Connect App and select the Logger from the list. The Logger supports the Fluke Connect Wireless System (may not be available in all regions). Fluke Connect is a system that wirelessly...
Page 61
Energy Logger Wiring Configurations Power A fund B fund C fund Total Total fund A fund B fund ...
1732/1734 Users Manual General Specifications Color LCD Display..........4.3-inch active matrix color TFT, 480 pixels x 272 pixels, resistive touch panel Power/Charging/LED Indicator Warranty Logger and Power Supply ........2 years (battery not included) Accessories ............1 year Calibration Cycle ...........2 years Dimensions Logger ..............19.8 cm x 16.7 cm x 5.5 cm (7.8 in x 6.6 in x 2.2 in)
Page 63
Energy Logger Environmental Specifications Safety IEC 61010-1 IEC Mains Input..........Overvoltage Category II, Pollution Degree 2 Voltage Terminals ...........Overvoltage Category IV, Pollution Degree 2 IEC 61010-2-033..........CAT IV 600 V / CAT III 1000 V Electromagnetic Compatibility (EMC) International ............IEC 61326-1: Industrial CISPR 11: Group 1, Class A Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
1732/1734 Users Manual Electrical Specifications Power Supply Voltage Range ............nominal 100 V to 500 V (85 V min to 550 V max) using safety plug input Mains Power ............nominal 100 V to 240 V (85 V min to 265 V max) using IEC 60320 C7 input (figure 8 power cord) Power consumption..........Maximum 50 VA (max.
Page 65
Input Range.............0 V dc to ± 10 V dc Wireless Connection (requires WiFi/BLE adapter USB1 FC) Number of Inputs..........2 Supported Modules .........Fluke Connect 3000 series Acquisition............1 reading/s Scale factor ............Format: mx + b (Gain and offset) user configurable Displayed units............User configurable (up to 8 characters, for example °C, psi, or m/s) Data Acquisition Resolution ............16-bit synchronous sampling...
Page 66
USB-A ..............File transfer via USB Flash Drive, Firmware updates, max. supply current: 120 mA WiFi Supported modes..........Direct connection and connection to infrastructure (requires WiFi-Infrastructure license) Security ............WPA2-AES with pre-shared key Bluetooth .............read auxiliary measurement data from Fluke Connect 3000 series modules (requires WiFi/BLE adapter USB1 FC) USB-mini .............Data download device to PC...
Page 67
Energy Logger Electrical Specifications Accuracy at Reference Conditions Maximum Intrinsic Accuracy at Reference Conditions Parameter Range Resolution (% of Reading + % of Range) ±(0.2 % + 0.01 %) Voltage 1000 V 0.1 V ±(0.3 % + 0.02 %) 15 mV 0.01 mV Rogowski Mode ±(0.3 % + 0.02 %)
Page 68
1732/1734 Users Manual Power/Energy Parameter Direct Input iFlex1500-12 iFlex3000-24 iFlex6000-36 i40S-EL Clamp: 50 mV/500 mV 150 A / 1500 A 300 A / 3000 A 600 A / 6000 A 4 A / 40 A Rogowski: 15 mV/150 mV Power Range W, VA, var Clamp: 50 W/500 W 150 kW / 1.5 MW...
Page 69
Energy Logger Electrical Specifications Intrinsic Uncertainty ±(% of measurement value + % of power range) Direct Input iFlex1500-12 iFlex3000-24 iFlex6000-36 i40S-EL Influence Parameter Quantity Clamp: 50 mV/500 mV 150 A / 1500 A 300 A / 3000 A 600 A / 6000 A 4 A / 40 A Rogowski: 15 mV/150 mV PF ≥0.99...
Energy Logger Electrical Specifications iFlex Probe Specifications Measuring range iFlex 1500-12 ............1 to 150 A ac / 10 to 1500 A ac iFlex 3000-24 ............3 to 300 A ac / 30 to 3000 A ac iFlex 6000-36 ............6 to 600 A ac / 60 to 6000 A ac Nondestructive current ........100 kA (50/60 Hz) ....±0.7 % of reading Intrinsic Error at reference condition...
1732/1734 Users Manual Phase shift...............< ± 0.5° iFlex1500-12: 88mm iFlex3000-24: 185mm iFlex6000-36: 282mm Figure 9. iFlex Probe Window Bandwidth..............10 Hz to 23.5 kHz Frequency derating ..........I x f ≤ 385 kA Hz Working Voltage ............1000 V CAT III, 600 V CAT IV [1] Reference Condition: •...
Page 73
Energy Logger Electrical Specifications Minimum bending radius .........38 mm (1.5 in) Output cable length iFlex 1500-12 ............2 m (6.6 ft) iFlex 3000-24 & iFlex 6000-36 ......3 m (9.8 ft) Weight iFlex 1500-12 ............115 g iFlex 3000-24 ............170 g iFlex 6000-36 ............190 g Material Transducer cable ..........TPR Coupling ..............POM + ABS/PC...
Page 74
1732/1734 Users Manual i40s-EL Current Clamp Specifications See Table 10 for setup instructions. Table 10. i40s-EL Setup Item Description Single Insulated current carrying conductor Release button Load direction arrow Tactile barrier Measuring range .............40 mA to 4 Aac / 0.4 Aac to 40 Aac .............≤...
Energy Logger Electrical Specifications Influence of conductor position ............± in jaw opening 0.5 % of reading (@ 50/60 Hz) Bandwidth ...............10 Hz to 2.5 kHz Working Voltage............600 V CAT III, 300 V CAT IV [1] Reference Condition: • Environmental: 23 °C ±5 °C, no external electrical/magnetic field, RH 65 % •...