CALIBRATION PROCEDURE

This scale has been factory calibrated and does not require calibration prior to
use. If required, the scale can be calibrated. The following procedure assumes the
scale is set to LB mode. The minimum calibration weight for this scale is 300 lb. If
calibrating in KG mode, the minimum calibration weight is 130 kg.
Note: Only weights certified and traceable to national standards are to be used for
calibration procedures.
Scale Calibration
1. Make sure the display is powered off, then press and hold the
HOLD/RELEASE button while pressing the ON/OFF button. (Two
alphanumeric characters will appear next to U followed by 4-5
characters.)
2. Press the ENTER button.
3. Make sure there is nothing on the scale, then press the ENTER
button.
4. Use the ► and ▲ buttons to set the calibration weight to a
value other than the default weight. The calibration weight should
be as high as possible, with 300 lb as a minimum calibration
weight.
5. Press the ENTER button.
6. Load the scale with the required weight.
7. Without touching or shaking the scale, allow the display to
stabilize, then press the ENTER button.
8. If the calibration is successful, the display will show SUCC and
power off. If the calibration is not successful, the display will show
Err1 and power off. Contact customer service for an Err1 display
(+1 800-638-3722).
Operator Action
13
Display
U"XX" followed by a 4-5
digit 'idle scale count'
number.
"CAL".
"Clear" followed by
"0300.0LB"
Shows "0300.0LB" or
entered calibration weight.
Displays "Load" followed by
"0300.0LB" or entered
calibration weight
Display increments showing
raw weight data.
"SUCC" or "Err1".
-------