Warnings  And  Maintenance; Cleaning; Troubleshooting; Spares - Covita iCO Smokerlyzer User Manual

Table of Contents

Advertisement

 
Warnings  and  Maintenance  
 
1. Hands  should  be  washed  regularly  in  accordance  with  infection  control  practice.    
2. Please  do  not  attempt  to  modify  the  equipment  in  any  way  or  use  accessories  not  specified  
by   the   manufacturer.   Any   attempt   to   do   so   will   invalidate   the   warranty   and   may  
compromise  the  safety  of  the  device.    
 
Cleaning  
 
1. The  iCO  Smokerlyzer®  product  is  moulded  with  PC/ABS  plastic,  which  is  easy  to  keep  clean.  
2. Use  non-­‐alcohol  based  anti-­‐bacterial  cleaning  wipes  only.  
3. NEVER   use   alcohol   or   cleaning   agents   containing   alcohol   or   other   organic   solvents   as   long  
term  exposure  to  these  vapours  will  damage  the  CO  sensor  inside.    
4. Under   no   circumstances   should   the   instrument   be   immersed   in   liquid   or   splashed   with  
liquid.  
 
Troubleshooting  
 
"An  iCO  device  is  not  properly  connected,  please  reconnect  the  device  or  choose  the  manual  option"  
Make  sure  the  Audio  Cable  is  connected  at  both  ends,  between  Smartphone  and  iCO.  
 
"An  iCO  device  is  not  properly  connected  or  the  volume  is  not  high  enough...."  
Make  sure  the  Audio  Cable  is  connected  at  both  ends,  between  Smartphone  and  iCO,    
Also  check  that  the  volume  control  on  your  Smartphone  is  on  maximum  (100%).  
Repeated   test   done   too   soon   after   previous   test;   holding   the   cable   end   of   the   device,   wave   iCO  
Smokerlyzer   about   for   15   –   20   seconds   to   remove   residual   contents   from   previous   test   and   then  
repeat  test.  
 
Repeated  zero  readings  after  breath  test  
Even  a  non-­‐smoker  should  give  a  low  reading  of  less  than  6ppm.      
Check  that  you  are  completely  emptying  your  lungs  when  exhaling  in  to  the  device.    
Possible  Faulty  Sensor.  
 
Unexpected  high  reading  after  breath  test  
High  levels  of  Carbon  Monoxide  in  vicinity  –  (i.e.  faulty  Boiler),  re-­‐take  test  outdoors  if  possible.  
Interference   from   a   Radio   or   TV   transmitter   if   you   live   close   to   one,   re-­‐take   test   away   from  
transmitter.  
Sensor   has   been   exposed   to   Alcohol,   do   not   use   Alcohol   based   wipes.   Leave   iCO   Smokerlyzer   for  
24Hrs  in  fresh  air  before  use.    
Incorrect  Pin  Code,  re-­‐enter  Pin  Code  to  value  on  the  rear  label  of  the  iCO  Smokerlyzer.  
 
 
 
Spares  
 
iCO  Cable  
Audio  Lead  
 
 
 
 
 
 
10  

Advertisement