Commands Listed in Alphabetical Order
SAVe:IMAGe:LAYout
Group
Syntax
Arguments
SAVe:SETUp (No Query Form)
Group
Syntax
Related Commands
Arguments
Examples
2-338
Sets or returns the layout to use for saved screen images.
Save and Recall
SAVe:IMAGe:LAYout {LANdscape|PORTRait}
SAVe:IMAGe:LAYout?
specifies that screen images are saved in landscape format.
LANdscape
specifies that screen images are saved in portrait format.
PORTRait
Stores the state of the oscilloscope to a specified memory location. You can later
use the *RCL command to restore the oscilloscope to this saved state.
Save and Recall
SAVe:SETUp {<file path>|<NR1>}
*RCL, RECAll:SETUp,
is the target location for storing the setup file. <file path> is a
<file path>
quoted string that defines the file name and path. Input the file path using the
form <drive>:<dir>/<filename>. <extension> and one or <dir>s are optional.
If you do not specify them, the oscilloscope will store the file in the current
working directory. <filename> stands for a filename. (Use of wildcard characters
in filenames is not supported.) Filename extensions are not required but are highly
recommended. For setups, use the extension ".SET".
specifies a location for saving the current front-panel setup. The
<NR1>
front-panel setup value ranges from 1 to 10. Using an out-of-range value causes
an execution error. Any settings that have been stored previously at this location
will be overwritten.
saves the current oscilloscope setup in memory location 5.
SAVE:SETUP 5
SAVE:SETUP "TEK00000.SET"
TEK00000.SET in the current working directory.
*SAV
saves the current oscilloscope setup in the file