ECG Monitoring and Pacemakers

When the unit performs ECG monitoring on a patient with an implantable pacemaker, the unit's
Pacer indicator feature can indicate the occurrence of pacemaker signals.
If the Pacer Indicator setting is
detects the implantable pacemaker pulses
blanks the pacemaker pulses from the waveform—preventing them from disturbing the
ECG waveform and allowing for an accurate QRS detection
displays and prints vertical dashed lines to indicate the detected pacemaker signals
If the Pacer Indicator setting is
detect the implantable pacemaker pulses
blank the pacemaker pulses from the waveform
display or print the vertical dashed line pacer markers
You can turn the Pacer Indicator
To turn the Pacer Indicator
1. Press the More quick access key (
Press the Setup quick access key (
2.
Use the navigation keys to select ECG then press
3.
Use the navigation keys to highlight and select Patient Pace Marker then press
4.
Use the navigation keys to select On or Off as appropriate. When you select Off, the Pacer
5.
Off Indicator(
pacer indicator is off. If the patient has a cardiac pacemaker, the Paced Indicator should be
set to On indicating that pace pulse detection is On.
There are situations where ECG artifact could simulate pacemaker signals which could cause
false pacemaker detection and blanking. These situations may cause inaccurate QRS detection
and it may be desirable to turn the Pacer Indicator
setting is
be desirable to turn the Pacer Indicator on.
9650-001355-01 Rev. M
ON
OFF
ON/OFF
ON/OFF
) icon appears in the top ECG trace on the Display Screen to indicate the
, implantable pacemaker signals may cause inaccurate QRS detection and it may
OFF
X Series Operator's Guide
, the X Series performs the following actions:
, the X Series does not perform the following actions:
from the Setup>ECG menu.
:
).
).
.
. Inversely, when the Pacer Indicator
off
ECG Monitoring and Pacemakers
.
6-9

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the X Series and is the answer not in the manual?

Questions and answers

Thomas Loughlin
March 31, 2025

What does the letter “P” with a red slash on the main screen of the X Series monitor indicate?

1 comments:
Jonathan
April 7, 2025

@Thomas Loughlin It means the pacer indicator setting is turned off. See page 99 of the manual.