Horizontal Amplifier Response Adj Ustmen Ts; Vertical Amplifier Response Adjustments; Phase Shift Adjustment; Ments - HP 1317A Operating And Service Manual

Table of Contents

Advertisement

Adjustments
j.
Set GAIN A3R12 for 10 inches deflection on
CRT.
5-65.
HORIZONTAL AMPLIFIER RESPONSE AD-
JUSTMENTS.
5-66. References. Schematic
1,
schematic
2,
schematic 3, table 5-1, figure 5-13, and figure 5-16.
5-67.
Procedure.
a. Perform preoperational procedure (Section
III).
b. Set
500/10K
switches (on boards
Al
and A3)
to 500.
c. Connect equipment as shown in figure 5-13.
d. Set square wave generator controls for 10
kHz, maximum amplitude output signal.
e. Center display with POSITION X and POSI-
TION Y.
f. Set monitor oscilloscope controls for stable,
full-screen display of two square-wave periods.
g. Set LF ADJ NO. 1 A2R24, and LF ADJ NO.
2 A2C17 for flattest pulse response on ground side of
display.
h.
Disconnect
square wave generator from X
INPUT and connect constant amplitude signal gen-
erator.
i. Set frequency of constant amplitude signal
generator to 50 kHz and amplitude to cause 5 inches
deflection on
CRT.
j.
Change frequency of constant amplitude
signal
generator to 5 MHz.
k.
Set HF ADJ NO. 1 A2C8 and HF ADJ NO.2
A2C 12 for vertical deflection of 3.7 inches.
5-68.
VERTICAL AMPLIFIER RESPONSE ADJUST-
MENTs.
5-69. References. Schematic
4,
schematic
5,
schematic 6, table
5-1,
figure
5-14,
and figure
5-16.
5-70.
Procedure.
a
.
Perform preoperational procedure (Section
III).
b.
Set
500/
10K
switches (on boards Al and A3)
to 500.
c. Connect equipment as shown in figure 5-14.
5-16
Model 1317A
d. Set square wave generator controls for 10
kHz, maximum amplitude output signal.
e. Center display with POSITION X and POSI-
TION Y.
f. Set monitor oscilloscope for stable, full-
screen display of two square-wave periods.
g. Set LF ADJ NO.1 A4R24 and LF ADJ NO.2
A4C17 for flattest pulse response on ground side of
display.
h.
Disconnect square wave generator from Y
INPUT and connect constant amplitude signal gener-
ator.
i.
Set frequency of constant amplitude signal
generator to 50 kHz and amplitude to cause 5 inches
of deflection on CRT.
j.
Change frequency of constant amplitude
signal generator to 5 MHz.
k. Set HF ADJ NO.1 A4C8 and HF ADJ. No.2
A4C12 for vertical deflection of 3.7
inches.
5-71.
PHASE SHIFT ADJUSTMENT.
5-72.
References. Schematic
3,
schematic
6,
table
5-1,
figure 5-15 and figure 5-16.
5-73.
Procedure.
a. Perform preoperational procedure (Section
III).
b. Set
500/10K
switches (on boards Al and A3)
to 500.
c. Connect equipment as shown in figure 5-15.
d. Set frequency of constant amplitude signal
generator to 250 kHz and amplitude for full-screen
diagonal line (or ellipse) on CRT.
e.
If separation of ellipse at center of display
is greater than 0.1 inch, make small adjustments to
A2C8,
A2C12, A4C8, and A4C12 until ellipse separa-
tion is less than 0.1 inch.
Note
The procedures in step d above, in
para-
graph 5-70 step n
,
and in paragraph 5-67
step n are interactive. Repeat these pro-
cedures until the requirements of all
three are satisfied.