CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
•
Hang hand control from side
rail (or end rail) of table when
not in use to avoid possible
damage to the control.
•
Route the hand control cord
(and optional HERMES
Ready
1
or ACT Enabled™ in-
terface cord and/or optional
foot control cord, if appli-
cable) clear of any pinch
points where the cord(s)
could be damaged.
1
HERMES-Ready is a registered trade-
mark of Computer Motion.
3.3 Hand Control
Interchangeability
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: The
®
HERMES
-Ready
Enabled hand controls are in-
terchangeable for control of
table functions; however,
ORCS voice-activation control
will NOT operate properly with
improper hand control.
Installation Instructions
2. Press FLOOR LOCK Function button in center row of control buttons, and
within five seconds press LOCK button (to the left of FLOOR LOCK button,
see Figure 3-4). Table is locked in position as floor locks are lowered and
casters are raised. Table will remain locked (immobile) until UNLOCK*
function is actuated.
NOTE: The tripodal floor locks are self-compensating for floor irregularities of up
to 1/4" (6 mm), and should not require adjustment. Floor locks should engage
simultaneously and the table base should rise evenly. Casters should swing
freely when the table is in the LOCKED position.
3. Hang hand control on table side rail or end rail (see Figure 3-2) and route
control cord away from possible pinch points.
4. Check floor locks to ensure each is snug against floor (see Figure 3-5).
IMPORTANT: If table was in storage for longer than four weeks, operate table
through all articulations prior to usage.
* To unlock table, press the FLOOR LOCK button in the center row of control buttons,
®
-
and within five seconds press the UNLOCK button adjacent to it on the right (see
Figure 3-4). Floor locks will retract and table will rest on casters.
Casters Swing Freely
The 3080 RL, 3080 SP, and 3080 RL/SP/3085 SP hand control will plug into and
operate the 3080 RL, 3080 SP, and 3085 SP tables. However, note the following
exceptions:
• The 3080 RC hand control will plug into and operate the 3080 RL, 3080 SP, and
3085 SP in a normal fashion EXCEPT there is no Return-to-Level button.
• The 3080 RL, 3080 SP, and 3080 RL/SP/3085 SP hand control will plug into and
1
and ACT
operate the 3080 RC EXCEPT the 3080 RC does not have the Return-to-Level
capability.
• The standard hand control (for the 3080 RC, 3080 RL, 3080 SP or 3085 SP
tables) has a six pin plug. The Operating Room Control System (ORCS) hand
control (for either 3085 SP HERMES
different plug (18 pins) and cannot be plugged into or used on any 3080 or
3085 table except HERMES
Figure 3-5. Check Floor Locks
®
-Ready
®
-Ready
1
Operator Manual
Check Floor Locks
1
or ACT Enabled™ table) has a
or ACT Enabled™ 3085 SP tables.
3-5
150830-353