How To Change Methods With A Chipcard - Labitec CoaDATA 501 Operator's Manual

Coagulation analyze
Table of Contents

Advertisement

2.1.5

How to change methods with a ChipCARD

aPTT
Lot No.: 123456, Exper. Date 05/05
Insert ChipCARD
Reconstitute reagents according to
manufacturer information
TEST:
- pipette 50ul Plasma in cuvette
- add 50ul aPTT reagent
- 120s incubation time
- start with 50ul CaCl2
UTILITIES
Unless a measurement is currently running, a ChipCARD can be inserted
into the adapter at any time for additional methods.
The method parameters will be loaded in memory 7 Extr. Factor. Once
another method is introduced, the method Extr. Factor will be overwritten
and can only be reloaded again using a ChipCARD for Extr. Factor.
The ChipCARD Reader can be accessed through a side opening
underneath the right membrane keypad.
The ChipCARD will be inserted into the ChipCARD reader with the contact
ahead and method description readable.
To load a method:
• Insert the ChipCARD with the selected method into the ChipCARD
Reader the method name and the lot-no. will be displayed automatically.
<-reading card
aPTT
Lot
XXXXX
• Press Enter to confirm loading.
remove card
When remove card is displayed, remove the ChipCARD from the
ChipCARD Reader.
<-write param.
When writing parameter to internal is displayed, the method parameters
will be stored in method memory 7.
Incubate next samples and continue as described above.
cuv in 1
The menu Utilities is a group of menus in which instrument settings can be
performed after a "Secret no." (code number) has been entered.
The submenus are as follows: <printer>, <computer>, <beeper>, <clock>
<calibrate temp>, <secret number>, and <cuvette detect.>.
Refer to chapter 2.2.8 Utilities.
CoaDATA 501 – Operators Manual 2.1
..
eter
Operation
Page 15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents