Keep batteries away from children. Many types of batteries
are poisonous. If swallowed, immediately contact your
poison control centre.
50
WARNING
Error Detection Displays
The meter screen displays error codes (E plus a number) for test result
errors, strip errors, or system errors. When an error occurs, the meter
beeps 2 times and displays an error code. Press the OK button to
turn off the meter.
If you experience continued errors, contact Diabetes Care Customer
Service. See the back cover for contact information.
Error Code
What it Means
Strip Errors
E 1
Too Little Blood
E 2
Used Test Strip
E 3
Strip Upside Down
E 4
Wrong Strip Inserted
E 6
Moisture Damaged Strip
E 8
Strip or Test Errors
Testing Errors
E20
Testing Error
Too Cold to Test Control
E24
Solution
C
®
N
ONE blood glucose monitoring system
ONTOUR
EXT
TECHNICAL INFORMATION
What to Do
Remove the strip. Repeat
the test with a new strip.
Remove the strip. Repeat
the test with a new strip.
Remove the strip and insert
it correctly.
Remove the strip. Repeat the
test with a C
N
ONTOUR
EXT
test strip.
Remove the strip. Repeat
the test with a new strip.
Repeat the test with a new
strip. If the error persists,
contact Customer Service.
Repeat the test with a new
strip. If the error persists,
contact Customer Service.
Move the meter, strip, and
control solution to a warmer
area. Test in 20 minutes.
51
Need help?
Do you have a question about the NEXT ONE and is the answer not in the manual?
Questions and answers
My meter says 56 44 52 and can not see previous readings. How doing fix it?? Thanks
can the batteries be charged
No, the Contour NEXT ONE batteries cannot be charged. They are not rechargeable and must be replaced after about 1000 tests.
This answer is automatically generated