Quick Start Guide LIQ-QSG-499ADO, Rev K July 2017 ™ Rosemount 499ADO Dissolved Oxygen Sensor...
Page 2
NOTICE ROSEMOUNT (“SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL SELLER BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.
Page 3
CAUTION! EQUIPMENT DAMAGE Do not exceed pressure and temperature specifications. Pressure: 65 psig (549 kPa abs) max. Temperature: 32 to 122 °F (0 to 50 °C)
If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount immediately. Product description Figure 1-1: Rosemount 499ADO Sensor Parts Membrane retainer Membrane assembly O-ring...
Page 8
Process connections 1 inch MNPT Weight/shipping weight 1 lb/3 lb (0.5kg /1.5 kg) (1) Noryl is a registered trademark of General Electric. (2) Viton and Teflon are registered trademarks of E.I. duPoint de Nemours & Co. Rosemount 499ADO...
Low flow cell 2 to 5 gph (7.6 to 19 L/hr) The Rosemount 499ADO sensor can be installed in ponds or basins. Use a pipe screwed into the back-facing threads to protect the cable and keep liquid away from the back end of the sensor.
Page 10
Install Figure 2-2: Flow through 1-1/2 in. Tee Rosemount 499ADO...
Page 11
Install Figure 2-3: Flow through 2 in. Tee Quick Start Guide...
Wire Wire NOTICE For additional wiring information on this product, including sensor combinations not shown here, please refer to the Liquid Transmitter Wiring Diagrams. Figure 3-1: Rosemount 499ADO-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Quick Start Guide...
Page 16
Wire Figure 3-2: Rosemount 499ADO-54-60 and 499ADO-54-VP Sensor Wiring to Rosemount 1056 and 56 Transmitters Rosemount 499ADO...
Page 22
Wire Figure 3-8: Rosemount 499ADO-54-VP Sensor Wiring to Rosemount 1066 Transmitter Figure 3-9: Rosemount 499ADO Sensor Pin-out Diagram When making a connection through a junction box (PN 23550-00), wire point-to-point. NOTICE Use a wire nut and pigtail (included) when connecting several wires to the same terminal.
Calibrate Zero point calibration Even in the absence of oxygen, the Rosemount 499ADO sensor generates a small signal called the zero current. Failing to correct for the zero current can introduce a bias, particularly if the oxygen concentration is small (<1 ppm). Zero the sensor when it is first placed in service and every time the fill solution is changed.
Maintenance Maintenance Periodic maintenance and cleaning are required for best performance of the sensor. Generally, the membrane and fill solution should be replaced every four to six months. Sensors installed in harsh or dirty environments require more frequent maintenance. The optimum maintenance frequency is best determined by experience.
Page 26
Place a drop of electrolyte solution on the cathode; then place the membrane assembly over the cathode. Screw the membrane retainer in place. The sensor may require several hours operating at the polarizing voltage to equilibrate after the electrolyte solution has been replenished. Rosemount 499ADO...
Accessories Accessories Part number Description 23747-06 Interconnecting cable, VP 6, 2.5 ft (0.8 m) 23747-04 Interconnecting cable, VP 6, 4 ft (1.2 m) 23747-02 Interconnecting cable, VP 6, 10 ft (3.0 m) 23747-07 Interconnecting cable, VP 6, 15 ft (4.6 m) 23747-08 Interconnecting cable, VP 6, 20 ft (6.1 m) 23747-09...